K Number
K212776
Date Cleared
2021-12-20

(110 days)

Product Code
Regulation Number
870.1220
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

Device Description

The Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled, is an irrigated steerable, flexible, insulated electrophysiology catheter constructed of thermoplastic elastomer material and noble metal electrodes. The shaft curvature is manipulated by the control mechanism located on the handle at the catheter's proximal end. To adjust the curve on the catheter, use the actuator to deflect the catheter in either direction. The catheter is compatible with St. Jude Medical (Abbott) visualization and 3D navigation systems.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA for a reprocessed medical device, specifically a Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled. It details the device, its indications for use, and a summary of testing conducted to demonstrate substantial equivalence to predicate devices.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance (Summary)
Functional & Safety Testing
BiocompatibilityMet
Cleaning ValidationMet
Sterilization ValidationMet
Visual InspectionMet (part of functional testing)
Dimensional VerificationMet (part of functional testing)
Electrical ContinuityMet (part of functional testing)
Simulated UseMet (part of functional testing)
Mechanical CharacteristicsMet (part of functional testing)
Electrical Safety TestingMet (Dielectric and Current Leakage)
Packaging ValidationMet
Reprocessing Cycle LimitReprocessed no more than 3 times; device rejected from further reprocessing after reaching maximum cycles.
Reprocessing ExclusivityReprocessing performed only by Innovative Health; excludes devices previously reprocessed by other reprocessors.

Note: The document states that the functional and safety testing was conducted "to demonstrate performance (safety and effectiveness)" and the conclusion is that the device is "substantially equivalent," implying that all these tests met their respective acceptance criteria. Specific quantitative acceptance criteria (e.g., "electrical continuity must be X ohms") are not detailed in this summary.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide specific sample sizes for the functional and safety testing mentioned.
Therefore, the sample size for the test set is not specified.
The data provenance is from bench and laboratory testing conducted by Innovative Health, LLC. The country of origin of the data is implicitly the United States, as Innovative Health, LLC. is based in Scottsdale, Arizona, and is interacting with the U.S. FDA. The testing conducted is likely prospective in nature, as it was performed to support the 510(k) submission for the reprocessed device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document does not describe the establishment of a "ground truth" in the sense of expert consensus on diagnostic imaging or clinical findings, as it pertains to a physical medical device (a catheter) and its reprocessing. The "testing" involved demonstrating physical, electrical, and sterilization performance. Therefore, this section is not applicable in the context of this device and submission.

4. Adjudication Method for the Test Set

This section is not applicable as the tests performed are objective measurements (e.g., electrical continuity, dimensional verification) or validations (e.g., sterilization, cleaning), not subjective assessments requiring expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This section is not applicable. The device is a physical medical instrument (a mapping catheter), not an AI-powered diagnostic tool. Therefore, an MRMC study related to human reader performance with or without AI assistance is irrelevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This section is not applicable. The device is a physical medical instrument (a mapping catheter), not a standalone algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance characteristics of this device is established by:

  • Engineering specifications and standards: These define acceptable dimensional tolerances, electrical properties, and mechanical characteristics for a functional catheter.
  • Biocompatibility standards: Defines acceptable levels of biological response to materials.
  • Sterilization and cleaning validation protocols: These define validated processes to ensure sterility and cleanliness, often with established microbial reduction or residue limits.
  • Predicate device performance: The reprocessed device is compared to the performance characteristics of the original cleared device and other predicate reprocessed devices to demonstrate substantial equivalence.

8. The Sample Size for the Training Set

This section is not applicable. There is no "training set" in the context of this device. The device is a physical product, not an algorithm that requires training data.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, and on the right is the FDA logo. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

December 20, 2021

Innovative Health, LLC. Amanda Babcock Regulatory Affairs Manager 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257

Re: K212776

Trade/Device Name: Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: NLG Dated: August 31, 2021 Received: September 1, 2021

Dear Amanda Babcock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

The item number included in the scope of this submission is as follows:

The item number included in the scope of this submission is as follows:

Item NumberDescriptionUsableLength(cm)FrenchSizeCurveSpacing(mm)ElectrodesSystemCompatibility
D-AVHD-DF16ReprocessedAdvisor HD GridMapping Catheter,Sensor Enabled1108FDF316EnSiteVelocity andEnSitePrecisionCardiacMappingSystems

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K212776

Device Name

Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled

Indications for Use (Describe)

The Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

As required by 21 CFR 807.92

Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

Contact Name and Information:

Amanda Babcock Requlatory Affairs Manager Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com

Date prepared:

August 31, 2021

Device Information:

Trade/Proprietary Name:Reprocessed Advisor HD Grid Mapping Catheter,Sensor Enabled
Common or Usual Name:Diagnostic Electrophysiology Mapping Catheter
Classification Name:Electrode Recording Catheter or Electrode RecordingProbe
Classification Number:Class II, 21 CFR 870.1220
Product Code:NLG

Predicate Device:

510(k) NumberDeviceManufacturer
K200212Reprocessed Advisor HD Grid MappingCatheter, Sensor EnabledInnovative Health, LLC.
K172393Advisor HD Grid Mapping Catheter, SensorEnabledSt. Jude Medical

Reference Device:

510(k) NumberDeviceManufacturer
K181458Reprocessed Advisor FL, Sensor Enabled,Circular Mapping CatheterInnovative Health, LLC.
K190785Reprocessed PentaRay Nav eco High-Density Mapping CatheterInnovative Health, LLC.

Device Description:

The Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled, is an irrigated steerable, flexible, insulated electrophysiology catheter constructed of thermoplastic elastomer material and noble metal electrodes. The shaft curvature is manipulated by the control mechanism located on the handle at the catheter's proximal end. To adjust the curve on the catheter, use the actuator to deflect the catheter in either direction. The catheter is compatible with St. Jude Medical (Abbott) visualization and 3D navigation systems.

The item number in scope of this submission is as follows:

{5}------------------------------------------------

ltem NumberDescriptionUsableLength(cm)FrenchSizeCurveSpacing(mm)ElectrodesSystem Compatibility
D-AVHD-DF16ReprocessedAdvisor HD GridMapping Catheter,Sensor Enabled1108FDF316EnSite Velocity andEnSite PrecisionCardiac MappingSystems

Table 5.1: Device Scope

Indications for Use:

The Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

Technological Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed Advisor HD Grid Mapping Catheter (Sensor Enabled) are identical to the predicate device. There are no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the Catheter includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled. This included the following:

  • Biocompatibility .
  • . Cleaning Validation
  • Sterilization Validation ●
  • Functional testing ●
    • . Visual Inspection
    • . Dimensional Verification
    • Electrical Continuity .
    • . Simulated Use
    • = Mechanical Characteristics
  • . Electrical Safety Testing
    • · Dielectric and Current Leakage
  • . Packaging Validation

The Reprocessed Advisor HD Grid Mapping Catheter, Sensor Enabled, is reprocessed no more than three (3) times. Each device is marked, serialized and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

Conclusion:

Innovative Health concludes that the Reprocessed Advisor HD Grid Mapping Catheter. Sensor Enabled, is substantially equivalent to the predicate devices described herein.

Page 2 of 2 Traditional 510(k) K212776

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).