K Number
K212748

Validate with FDA (Live)

Date Cleared
2022-01-20

(143 days)

Product Code
Regulation Number
876.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Dual Action Tissue Clip is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only.

The Disposable Dual Action Tissue Clip is indicated for clip placement within the gastrointestinal tract for the purpose of:

Endoscopic marking,

Hemostasis for

·Mucosal/sub-mucosal defects < 3 cm,

·Bleeding ulcers,

·Polyps < 1.5 cm in diameter,

·Diverticula in the colon,

·Arteries < 2 mm,

As a supplementary method, closure of GI tract luminal perforations < 20 mm that can be treated conservatively.

Device Description

The proposed device Disposable Dual Action Tissue Clip is a sterile, single-use endoscopic clipping device, intended to be used in flexible endoscopy for the compression of tissue in the gastrointestinal tract. It consists of two main components, delivery system and clip assembly. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10 ° and placed in a sterility maintenance package to ensure a shelf life of 1 year.

AI/ML Overview

I am unable to provide a description of the acceptance criteria and the study that proves the device meets the acceptance criteria, along with the requested details, because the provided text is a 510(k) summary for a medical device (Disposable Dual Action Tissue Clip) and does not contain detailed information about specific acceptance criteria or a dedicated study report with performance data.

The document states that "Performance testing was conducted to demonstrate the essential performance of the proposed device... and confirmed that the proposed device works as intended with the compatible devices." It then lists various bench tests performed (Dimension, Release Force, Clamping Strength, Tensile Strength, etc.) and mentions that these were "tested and evaluated as substantially equivalent to the predicate device."

However, it does not provide:

  • A table of specific acceptance criteria (e.g., "Clamping Strength must be > X N") and corresponding reported device performance values.
  • Sample sizes for test sets (beyond implying "bench tests").
  • Data provenance for any performance data.
  • Information about expert involvement for ground truth, adjudication methods, or MRMC studies, as it explicitly states "No animal study is included in this submission" and "No clinical study is included in this submission."
  • Details about the training set or its ground truth, as this is a mechanical device subject to bench testing, not an AI/ML algorithm.

Therefore, I cannot fulfill your request for the specific comparative performance data, sample sizes, and expert details as they are not present in the provided FDA 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.

January 20, 2022

Micro-Tech (Nanjing) Co., Ltd. Cecilia Sun RA Engineer No.10 Gaoke Third Road, Nanjing National Hi-Tech Industrial Developement Zone Nanjing, Jiangsu Province 210032 CHINA

Re: K212748

Trade/Device Name: Disposable Dual Action Tissue Clip Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: Class II Product Code: PKL, Dated: December 3, 2021 Received: December 20, 2021

Dear Cecilia Sun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

K212748

Device Name Disposable Dual Action Tissue Clip

Indications for Use (Describe)

The Disposable Dual Action Tissue Clip is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only.

The Disposable Dual Action Tissue Clip is indicated for clip placement within the gastrointestinal tract for the purpose of:

Endoscopic marking,

Hemostasis for

·Mucosal/sub-mucosal defects < 3 cm,

·Bleeding ulcers,

·Polyps < 1.5 cm in diameter,

·Diverticula in the colon,

·Arteries < 2 mm,

As a supplementary method, closure of GI tract luminal perforations < 20 mm that can be treated conservatively.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212748 Page 1 of 6

Image /page/3/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, blocky font. The letters are connected and appear to be three-dimensional. There is a small circle with an "R" inside it in the upper right corner of the image, indicating a registered trademark.

Section 5 510(k) Summary

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________

1. Date of Preparation: 2021-12-03

2. Sponsor Identification

Micro-Tech (Nanjing) Co., Ltd.

No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing,

Jiangsu Province, PRC

Establishment Registration Number: 3004837686

Contact Person: Cecilia Sun

Position: RA Engineer

Tel: +86-25-58646395

Fax: +86-25-58350006

Email: RA.Micro-Tech@outlook.com

3. Identification of Proposed Device

Trade Name: Disposable Dual Action Tissue Clip

Common Name: Hemostasis Clip

Regulatory Information

Classification Name: Hemostatic Metal Clip For The GI Tract

Classification: II

Product Code: PKL

Regulation Number: 876.4400

Review Panel: Gastroenterology/Urology

{4}------------------------------------------------

K212748 Page 2 of 6

Image /page/4/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional font. The letters are connected and appear to be leaning slightly to the right, giving the logo a sense of depth. A registered trademark symbol is visible in the upper right corner of the image.

Section 5 510(k) Summary

4. Identification of Predicate Device

Predicate Device

510(k) Number: K202333

Product Name: Lockado™ Repositionable Hemostasis Clip

Manufacturer: Micro-Tech (Nanjing) Co., Ltd.

5. Indications for Use

The Disposable Dual Action Tissue Clip is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only.

The Disposable Dual Action Tissue Clip is indicated for clip placement within the gastrointestinal tract for the purpose of:

  • Endoscopic marking,
  • Hemostasis for
    • O Mucosal/sub-mucosal defects < 3 cm,
    • O Bleeding ulcers,
    • O Polyps < 1.5 cm in diameter,
    • O Diverticula in the colon,
    • O Arteries < 2 mm,
  • As a supplementary method, closure of GI tract luminal perforations < 20 mm that can be ● treated conservatively.

6. Device Description

The proposed device Disposable Dual Action Tissue Clip is a sterile, single-use endoscopic clipping device, intended to be used in flexible endoscopy for the compression of tissue in the gastrointestinal tract. It consists of two main components, delivery system and clip assembly. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10 ° and placed in a sterility maintenance package to ensure a shelf life of 1 year.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional font. The letters are connected and appear to be leaning slightly to the right, giving the logo a dynamic feel. A small circle with an 'R' inside, indicating a registered trademark symbol, is positioned in the upper right corner of the logo.

7. Comparison of Technological Characteristics

The Disposable Dual Action Tissue Clip incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device Lockado™ Repositionable Hemostasis Clip cleared under K202333.

ItemProposed DeviceDisposable Dual Action Tissue ClipPredicate DeviceLockado™ RepositionableHemostasis Clip (K202333)Remark
Product CodePKLPKLSame
Regulation No.876.4400876.4400Same
Class22Same
Indications for UseThe Disposable Dual Action TissueClip is intended for use in flexibleendoscopy for the compression oftissue in the gastrointestinal tract foradult patient only.The Disposable Dual Action TissueClip is indicated for clip placementwithin the gastrointestinal tract forthe purpose of:• Endoscopic marking,• Hemostasis for○Mucosal/sub-mucosal defects < 3cm,○Bleeding ulcers,○Polyps < 1.5 cm in diameter,○Diverticula in the colon,○Arteries < 2 mm,• As a supplementary method,closure of GI tract luminalperforations < 20 mm that can betreated conservatively.The Lockado™ RepositionableHemostasis Clip is indicated forEndoscopic clip placement within theGastrointestinal tract in adultpopulations only via a straight or sideviewing flexible endoscope for thepurpose of :(1) Endoscopic marking;(2) Hemostasis for(a) Mucosal / sub-mucosal defects <3cm,(b) Bleeding ulcers,(c) polyps < 1.5cm in diameter,(d) diverticula in the colon,(e) Arteries < 2 mm,(f) Prophylactic clipping to reducethe risk of delayed bleeding postlesion resection.(3) as a supplementary method,closure of GI tract luminalperforations <20mm that can betreated conservatively.(4) Anchoring to affix jejunalCoveredby thepredicateddevice
ItemProposed DevicePredicate DeviceRemark
Disposable Dual Action Tissue ClipLockadoTM RepositionableHemostasis Clip (K202333)feeding tubes to the wall of the smallbowel;
Configuration2 delivery units of the deliverysystem and 2 separate clips of theclip assemblyDelivery system and clip assemblySimilar
Principles ofOperationPull the steel wire through theoperating handle to drive theopening and closing of the clip toclamp the tissue. It is a metallicdevice with a clasping mechanism.Pull the steel wire through theoperating handle to drive the openingand closing of the clip to clamp thetissue. It is a metallic device with aclasping mechanism.Same
Open width (mm)158,11,16,22Similar
Working Length(mm)1650,1950,23501650,1950,2350,2700Similar
Minimal workingchannel ofendoscopy (mm)3.22.8Different
Supplied in SterileYesYesSame
Single UseYesYesSame
PackagingSingle-use EO sterilized pouch withone device per pouchSingle-use EO sterilized pouch withone device per pouchSame
Shelf LifeOne yearTwo yearsDifferent
BiocompatibilityConform to ISO 10993-1Conform to ISO 10993-1Same
SterilizationEO Sterilized, SAL:10-6EO Sterilized, SAL:10-6Same
LabelingConform to 21 CFR part 801Conform to 21 CFR part 801Same
MRI informationComply with ASTM F 2503, ASTMF 2052, ASTM F2119, ASTMF2182, ASTM F2213Comply with ASTM F 2503, ASTMF 2052, ASTM F2119, ASTMF2182, ASTM F2213Same

Table 7.1 Comparison to predicate Devices

{6}------------------------------------------------

K212748
Page 4 of 6

Image /page/6/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, blocky font. The letters are interconnected and appear to be three-dimensional. A small circle with the letter "R" inside is located in the upper right corner of the logo, indicating a registered trademark.

Section 5 510(k) Summary

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows a blue logo with the letters 'M' and 'T' stacked on top of each other. The logo has a three-dimensional effect, with the letters appearing to be slightly raised and angled. A registered trademark symbol is located in the upper right corner of the logo. The logo is simple and modern, and the blue color gives it a professional look.

8. Performance Data

Performance testing was conducted to demonstrate the essential performance of the proposed device Disposable Dual Action Tissue Clip and confirmed that the proposed device works as intended with the compatible devices.

The bench tests below were tested and evaluated as substantially equivalent to the predicate device.

  • Dimension;

  • ア Release Force;
  • Clamping Strength;

  • Tensile Strength:

  • Clip Assembly Repeated Open/Close:

  • Clip Open And Close Force

  • Scope Compatibility/Usability;

  • Endoscope Damage;

  • Biopsy Valve Compatibility;

  • Clip Approach;

  • Coil to Handle Tensile

  • Mechanical integrity of Clip Assembly Test

Shelf-life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Two-year aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test.

Sterilization validation was carried out in accordance with ISO 11135:2014 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices".

Biocompatibility testing was performed in accordance with the FDA Guidance, Use of

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a logo with the letters 'MT' in a bold, sans-serif font. The letters are rendered in a deep blue color and have a three-dimensional effect, achieved through shading and perspective. A registered trademark symbol is visible in the upper right corner of the logo.

Section 5 510(k) Summary

International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

MR compatibility was evaluated in accordance with ASTM F 2503, ASTM F 2052, ASTM F2119, ASTM F2182, ASTM F2213 and FDA guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment," issued on May 20, 2021.

9. Animal Study

No animal study is included in this submission.

10. Clinical Study

No clinical study is included in this submission.

11. Substantially Equivalent (SE) Conclusion

Based on the indications for use, technological characteristics, and safety and performance testing, the Disposable Dual Action Tissue Clip has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Lockado™ Repositionable Hemostasis Clip cleared under K202333.

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.