K Number
K212668
Date Cleared
2022-02-23

(184 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is used for endoscopic injection into gastrointestinal mucosa, or to endoscopically introduce a sclerosing agent into selected sites to control actual or potential bleeding lesions in the digestive system.

Device Description

The Sclerotherapy Needle device consists of Luer connector, handle, sheath, infusion tube, fixed button connection tube and needle. EO sterilization and use for single use only. The Sclerotherapy Needle device is designed to pass through the channel of endoscope to mark the lesions of the digestive tract and use it for injection. Compatibility with endoscopes, working lengths are 1200mm,1800mm, 2000mm, 2300mm, and the minimum working channel is ф 2.8 mm.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Sclerotherapy Needle by Zhejiang Chuangxiang Medical Technology Co., LTD., which has been determined substantially equivalent to the predicate device. However, this document does not contain the specific acceptance criteria or a detailed study outlining device performance against such criteria for the Sclerotherapy Needle.

Instead, the document primarily focuses on establishing substantial equivalence to a predicate device (Hangzhou AGS's Disposable Sclerotherapy Needle, K190032) based on similarities in design, configuration, intended use, and fundamental technology.

Here's an analysis of the information provided and what is missing in relation to your request:

1. A table of acceptance criteria and the reported device performance:

  • Missing. The document explicitly states: "The following bench tests were performed on Sclerotherapy Needle: Appearance, Physical properties. The results of all testing were passing." However, it does not provide a table detailing specific acceptance criteria (e.g., maximum force for needle deployment, flow rate through the infusion tube, specific dimensions, etc.) or the measured performance values for these tests. It only states that the tests "were passing," which indicates compliance with internal or regulatory requirements but lacks the specific data requested.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

  • Missing. The document mentions "bench tests," but it does not specify the sample size for these tests or the provenance of the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

  • Not Applicable / Missing. This device is a physical medical instrument (sclerotherapy needle), not an AI/imaging device requiring expert interpretation for ground truth establishment. The "bench tests" would typically be evaluated against engineering specifications, not expert consensus on diagnostic images.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not Applicable / Missing. As above, this concept typically applies to diagnostic studies with multiple readers, not bench testing of a physical medical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is a physical medical device, not an AI diagnostic tool. Therefore, an MRMC study related to AI assistance for human readers is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Not Applicable. For the bench tests performed ("Appearance, Physical properties"), the "ground truth" would be the device's adherence to engineering drawings, material specifications, and functional requirements, as validated by standard measurement techniques and test methods. It's not a diagnostic "ground truth" in the clinical sense.

8. The sample size for the training set:

  • Not Applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established:

  • Not Applicable. As above, no training set.

Summary based on the provided text:

The document confirms that the Sclerotherapy Needle underwent "bench tests" for "Appearance, Physical properties" and that the results "were passing." It also notes compliance with biocompatibility (ISO 10993), sterilization (ISO 11135), and ethylene oxide residuals standards (ISO 10993-7). However, the detailed acceptance criteria (what constitutes "passing" for each specific test of appearance and physical properties) and the actual performance data are not disclosed in this summary. The entire premise of this 510(k) summary is to demonstrate substantial equivalence to an existing predicate device rather than to provide a comprehensive performance study with explicit acceptance criteria and results.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

February 23, 2022

Zhejiang Chuangxiang Medical Technology Co., LTD. Lucius Long, RA Manager Building 50, No. 650 Hongfeng Road Donghu Street Yuhang District, Hangzhou, 311100 CHINA

Re: K212668

Trade/Device Name: Sclerotherapy Needle Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FBK Dated: August 24, 2021 Received: January 27, 2022

Dear Lucius Long:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212668

Device Name Sclerotherapy Needle

Indications for Use (Describe)

The device is used for endoscopic into gastrointestinal mucosa, or to endoscopically introduce a sclerosing agent into selected sites to control actual or potential bleeding lesions in the digestive system.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212668 Page 1 of 5

Image /page/3/Picture/1 description: The image contains a logo with a geometric shape on the left, resembling a stylized hexagon with a gradient from green to blue. To the right of the shape, there are Chinese characters in blue, followed by the word "MEDNOVA" in a combination of blue and green. The overall design appears to represent a company or organization, possibly in the medical or technological field, given the name and the clean, modern aesthetic.

g Medical Technology Co., LTD.

Page: 1 / 5

510(k) Summary( 21CFR 807.92)

1. Submitter's information

Name: Zhejiang Chuangxiang Medical Technology Co., LTD.

Address: Room 101,201,301,401,501, Building 50, No.650 Hongfeng Road Donghu Street, Yuhang District, Hangzhou City, Zhejiang Province, 311100, P.R. China

Contact person: Lucius.Long Email: lucius.long@med-nova.com Telephone: 86-571-89167088 Fax: 86-571-89167086

2. Date of Submission

09-Aug- 2021

3. Device

Trade/Device Name: Sclerotherapy Needle Regulation name: Endoscope and accessories Regulation class: II Regulation number:876.1500 Panel: Gastroenterology/Urology Product code: FBK

4. Predicative device

4.1) 510(k) Number: K190032 Device Name: Disposable Sclerotherapy Needle

5. Device description

The Sclerotherapy Needle device consists of Luer connector, handle, sheath, infusion tube, fixed button connection tube and needle. EO sterilization and use for single use only.

The Sclerotherapy Needle device is designed to pass through the channel of endoscope to mark the lesions of the digestive tract and use it for injection. Compatibility with endoscopes, working lengths are 1200mm,1800mm, 2000mm, 2300mm, and the minimum working channel is ф 2.8 mm.

6. Indications for use

The device is used for endoscopic injection into gastrointestinal mucosa, or to endoscopically introduce a sclerosing agent into selected sites to control actual or potential bleeding lesions in the digestive system.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains a logo with a geometric shape on the left, transitioning from green to blue. To the right of the logo, there is text in blue, followed by the word "MEDNOVA" in a smaller font size, with the letters transitioning from green to blue. The text appears to be a company or organization name, with the logo serving as its visual identifier.

Zhejiang Chuangxiang Medical Technology Co., LTD.

Page:

7. Comparison of Technological Characteristics:

The Sclerotherapy Needle has substantially equivalent device design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Hangzhou AGS's Disposable Sclerotherapy Needle ,K190032.The differences between the proposed device and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below.

ItemProposed devicePredicate deviceComparison toPredicate Devices
Device nameSclerotherapy NeedleDisposable Sclerotherapy Needle/
Indicationsfor UseThe device is used forendoscopic injection intogastrointestinal mucosa, or toendoscopically introduce asclerosing agent into selectedsites to control actual orpotential bleeding lesions inthe digestive system.The Disposable SclerotherapyNeedle is intended for endoscopicinjection into thegastrointestinal mucosa.SubstantialEquivalence
ConfigurationLuer connector,Handle,Infusion tube,Sheath,Fixedbutton,Connection tube,NeedleLuer connector, handle, moldingmetal tube, Inner sheathconnection limit tube, Fixingsleeve, Outer sheath, Innersheath, connection tube, metalcap, needleSimilar
PhotographsSclerotherapy Needle:Image: [Sclerotherapy Needle A Standard type]1. Luer connector 2. Handle 3. Drive Pipe 4. SheathImage: [Sclerotherapy Needle B simple type]1. Luer connector 2. Handle 3. Drive Pipe 4. SheathDisposable Sclerotherapy Needle:Image: [Disposable Sclerotherapy Needle]5311 Series ( with tip forming ) :Similar
Zhejiang Chuangxiang Medical Technology Co., LTD.Page: 3 / 5
Image: Diagram of tip structure with labels "Infusion tube", "connection tube Needle", and "Fixed button".The tip structureWith fixed button:Image: Syringe with fixed buttonImage: Syringe with fixed button in packagingImage: Syringe with metal cap5311 Series (with metal cap):Image: Syringe with metal cap
Without fixed button:Image: Syringe without fixed buttonImage: Syringe without fixed buttonImage: Green syringe with metal cap5313 Series (with metal cap):Image: Green syringe with metal cap
Sheath diameter1.8mm,2.4mm2.4mmSimilar
Needle size:19G,21G, 23G,25G21G, 22G, 23G, 24G, 25GSimilar
Outer tubingPTFEFor 5311 series: PP;For 5313 series: SUS 304 covered by PE.Similar: We conducted a Biocompatibility
Page: 4 / 5
evaluation of thedevice. The resultsshow the device issafe in the aspect ofbiocompatibilityevaluation
WorkingLength1200mm,1800mm, 2000mm,2300mm1600mm, 2000mm, 2300mmSimilar
PackagingSingle-use EO sterilizedpaper-plastic pouch with onedevice per pouch.Single-use EO sterilized Tyvekpouch with one device per pouch.Similar
MaterialsLuer connector: ABSHandle: ABSInfusion tube: PPSheath: PTFEFixed button: Y12Cr18Ni9Cu3Connection tube:SUS304(06Cr19Ni10 )Needle:X5CrNi18-9Luer connector: PC;Handle: PC;Molding metal tube: SUS304;Inner sheath connection limittube: PTFE;Fixing sleeve: ABS;Outer sheath: PP( 5311series);SUS 304 and PE (5313 series);Inner sheath: PP;Connection tube: SUS 304;Metal cap: SUS303;Needle: SUS304.The material expected to comeinto contact with the patient isSUS304, SUS303 and PP, orSUS304、SUS303 and PE.Similar
Principle ofoperationThe catheter sheath of theproduct is inserted into theendoscope channel. Whenthe front part of the cathetersheath is placed on the lesionsite, push the Luer connectorfor injection, the needle isexposed to the catheter sheath,and the needle is inserted intothe lesion site, then druginjection.The outer sheath of the product isinserted into the endoscopeclamp. When the front part of theouter sheath tube is placed on thelesion site, push the Luerconnector, the needle tube isexposed to the outer sheath, andthe needle is inserted into thelesion site, then normal salineinjection.Similar

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains a logo with a geometric shape on the left, followed by text. The geometric shape is a hexagon with a gradient fill, transitioning from green at the bottom to blue at the top. To the right of the shape, there are Chinese characters in blue, followed by the word "MEDNOVA" in blue and green. The word "MED" is in blue, while "NOVA" is in green.

Zhejiang Chuangxiang Medical Technology Co., LTD.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains a logo with a geometric shape on the left and text on the right. The geometric shape is a hexagon with a smaller hexagon inside, and it is colored with a gradient from green to blue. To the right of the shape, there are Chinese characters in blue, followed by the word "MEDNOVA" in blue and green. The logo appears to represent a company or organization, possibly in the medical or technology field.

Zhejiang Chuangxiang Medical Technology Co., LTD.

{7}------------------------------------------------

K212668 Page 5 of 5

Image /page/7/Picture/1 description: The image contains a logo for Zhejiang Chuangxiang Medical Technology Co., LTD. The logo features a geometric shape in shades of green and blue, along with the text "MEDNOVA" in green. The company name is written below the logo in black text.

5 / 5 Page: Page

8. Applicable Guidance Document

NA

9. Performance Data

The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing", ISO 11135:2014 "Sterilization of Health Care products Ethylene Oxide - Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals"

The following bench tests were performed on Sclerotherapy Needle: Appearance, Physical properties. The results of all testing were passing.

10. Clinical Test Conclusion

No clinical study is included in this submission.

11. Conclusions

Chuangxiang medical has demonstrated that the proposed device Sclerotherapy Needle is substantially equivalent to Hangzhou AGS MedTech CO., Ltd. currently marketed Disposable Sclerotherapy Needle (K190032).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.