K Number
K212660
Manufacturer
Date Cleared
2021-09-22

(30 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Smylio Invisible Clear Aligners are indicated for the alignment of teeth during orthodontic treatment of malocclusion.

Device Description

The Smylio Invisible Clear Aligners are thermoformed plastic aligners designed to be worn in sequence to facilitate the movement to the final desired position. The sequential aligners introduce incremental movements that move teeth by way of gentle continuous force. The aligners are to be worn 20 to 22 hours a day and are to be removed for eating and for cleaning.

Smylio Invisible Clear Aligners are designed from digital scans of a patient's dentition submitted by a dental health professional (e.g. dentist or orthodontist). Using the scan, sequential dental models are designed and approved by the treating physician prior to manufacturing.

Once the treatment plan is reviewed and approved by a dental health professional, each 3D model from the treatment plan is manufactured. The aligner trays are then manufactured by thermoforming a dental thermoplastic sheet over each model. The aligner trays are then delivered to the patients' dental health professional then monitors their treatment of the first aligner to the delivery of the final aligner.

AI/ML Overview

The provided text is a 510(k) Premarket Notification summary for Smylio Invisible Clear Aligners. It focuses on demonstrating substantial equivalence to predicate devices based on indications for use, technological characteristics, and non-clinical performance testing. Crucially, the document explicitly states that "Clinical performance testing was not conducted for this 510(k) Notification." Therefore, there is no study described that proves the device meets acceptance criteria through clinical performance.

The "acceptance criteria" discussed in this document pertain to demonstrating substantial equivalence to existing legally marketed devices, rather than a direct evaluation of clinical efficacy or safety against performance metrics. The proof of meeting these "acceptance criteria" comes from the comparisons made in the tables and the statement that non-clinical testing was sufficient.

Here's an analysis based on the information provided, specifically addressing the questions you raised, while acknowledging the absence of a clinical study:

1. A table of acceptance criteria and the reported device performance:

Since no clinical performance study was conducted, there aren't "acceptance criteria" in the traditional sense of clinical endpoints (e.g., bone density improvement, disease detection accuracy). Instead, the "acceptance criteria" for this 510(k) submission are implied by the comparison to predicate devices, demonstrating equivalence in various specifications, materials, and non-clinical testing.

Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (Comparison to Predicates)
Regulation Number: Same regulation number as predicate (21 CFR 872.5470)Same (21 CFR 872.5470)
Classification Name: Same classification name as predicate (Orthodontic Plastic Bracket)Same (Orthodontic Plastic Bracket)
Product Code: Same product code as predicate (NXC)Same (NXC)
Classification: Same classification as predicate (Class II)Same (Class II)
OTC or Rx: Same prescription status as predicate (Rx)Same (Rx)
Material: Same as predicate (Thermoplastic polyurethane-polyester composite resin) or similar acceptable material.Same as primary predicate; material for reference predicate is Co-polyester or co-polymer, but stated "Same as predicate" due to the subject device matching the primary predicate.
Material Properties: Acceptable for use as an aligner.Acceptable materials properties established for use as an aligner (same as predicates).
Material Testing: Compliance with ASTM D790-10 for flexural properties.ASTM D790-10 testing conducted and found compliant (same as predicates).
Biocompatibility: Biocompatible according to ISO 10993-1.Biocompatible according to ISO 10993-1 (same as predicates). Additional cytotoxicity testing per ISO 10993-5:2009 performed on final product.
Sterility: Matching predicate status (Non-sterile).Non-sterile (same as predicates).
Device Description: Sequential thermoformed plastic aligners.Sequential thermoformed plastic aligners (same as predicates).
Patient Removable: Yes.Yes (same as predicates).
Indication for Use: Indicated for alignment of teeth during orthodontic treatment of malocclusion (same as predicate).Same as primary predicate. Differences from reference predicate ("alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion") do not raise substantial equivalence questions.
Intended Use: Orthodontic tooth movement.Same (same as predicates).
Mode of Action: Continuous gentle force applied to teeth.Same (same as predicates).
Durability Testing: Simulate real-world use to ensure suitability for prescribed period of use.Durability testing was conducted, and the aligner material and manufacturing process produced suitable aligners.
Manufacturing Validation: Robustness of the process from 3D printing through thermoforming.Internal manufacturing validation performed and robustness demonstrated.

2. Sample size used for the test set and the data provenance:

  • Test Set (Clinical): Not applicable. No clinical performance test set was used, as explicitly stated: "Clinical performance testing was not conducted for this 510(k) Notification."
  • Non-Clinical Testing: The document refers to "durability testing" and "material characterization" (ASTM D790-10, ISO 10993-1, ISO 10993-5) but does not provide specific sample sizes for these tests. Data provenance for non-clinical tests is internal to Smylio, Inc.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Since no clinical test set was used, there were no experts establishing ground truth for evaluating clinical performance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. No clinical test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a physical aligner, not an AI-assisted diagnostic or treatment planning tool for human readers. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a physical aligner. Performance is demonstrated through material properties and design, not an algorithm's standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For demonstrating substantial equivalence: The "ground truth" used are the established characteristics and performance metrics of the legally marketed predicate devices (Spark Clear Aligner System K203737 and Smylio Invisible Clear Aligners K173784).
  • For non-clinical testing: The "ground truth" is based on industry standards (ASTM D790-10, ISO 10993-1, ISO 10993-5) for material properties and biocompatibility, as well as internal validation procedures for manufacturing robustness.

8. The sample size for the training set:

  • Not applicable. This is a physical medical device, not a machine learning algorithm that requires a training set. The "design" is based on established orthodontic principles and the design of predicate devices. Patient-specific aligners are manufactured based on individual patient digital scans.

9. How the ground truth for the training set was established:

  • Not applicable. (See point 8).

In summary, the provided document describes a 510(k) submission where substantial equivalence for a physical medical device (clear aligners) is demonstrated through a comparison of design, materials, indications for use, and non-clinical testing to legally marketed predicate devices. No clinical performance study was conducted or deemed necessary for this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. The FDA logo is a recognizable symbol of the agency's role in protecting public health.

September 22, 2021

Smylio, Inc % Breanne Butler Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043

Re: K212660

Trade/Device Name: Smylio Invisible Clear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: August 19, 2021 Received: August 23, 2021

Dear Breanne Butler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212660

Device Name Smylio Invisible Clear Aligners

Indications for Use (Describe)

The Smylio Invisible Clear Aligners are indicated for the alignment of teeth during orthodontic treatment of malocclusion.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5. 510(k) Summary

510(k) SUMMARY

A summary of 510(k) safety and effectiveness information for this special 510(k) in accordance with the requirements of 21 CFR 807.92.

Submitter:Smylio, Inc.48890 Milmont Dr.Fremont, CA 94538
Company Contact Person:Phone:Email:Renjith S Menon, CEO(510) 868-2184rmenon@orthofx.com
Submission Correspondent:Address:Phone:Email:Breanne Butler, Regulatory Affairs Consultant1321 Upland Dr. Suite 6792 Houston, TX 7704860-810-5594bbutler@primepathmedtech.com
Date Prepared:August 2nd, 2021
Proprietary Name:Smylio Invisible Clear Aligners
Common Name:Orthodontic plastic bracket.
Product Code:NXC – Orthodontic plastic bracket.
Device Classification:Class II, 21 CFR 872.5470
Primary Predicate Device:Spark Clear Aligner System (K203737)
Reference Predicates:Invisalign System (K143630)Smylio Invisible Clear Aligners (K173784)

Device Description:

The Smylio Invisible Clear Aligners are thermoformed plastic aligners designed to be worn in sequence to facilitate the movement to the final desired position. The sequential aligners introduce incremental movements that move teeth by way of gentle continuous force. The aligners are to be worn 20 to 22 hours a day and are to be removed for eating and for cleaning.

Smylio Invisible Clear Aligners are designed from digital scans of a patient's dentition submitted by a dental health professional (e.g. dentist or orthodontist). Using the scan, sequential dental models are designed and approved by the treating physician prior to manufacturing.

Once the treatment plan is reviewed and approved by a dental health professional, each 3D model from the treatment plan is manufactured. The aligner trays are then manufactured by thermoforming a dental thermoplastic sheet over each model. The aligner trays are then delivered to the patients'

{4}------------------------------------------------

dental health professional then monitors their treatment of the first aligner to the delivery of the final aligner.

Indications for Use:

The Smylio Invisible Clear Aligners are indicated for the alignment of teeth during orthodontic treatment of malocclusion.

Comparison to Predicate Devices:

Smylio Invisible Clear Aligners are functionally equivalent to the following predicate device: Spark Clear Aligner System (K203737, cleared March 22, 2021) and to the previously cleared reference device, Smylio Invisible Clear Aligners (K173784). The following table demonstrates the functional specifications of Smylio Invisible Clear Aligners are substantially equivalent to the predicate devices.

Device Comparison Table
SpecificationSubject Device:Smylio Invisible ClearAlignersPredicate Device:Spark Clear AlignerSystem (K203737)Reference Device:Smylio Invisible ClearAligners (K173784)ComparisonResult
RegulationNumber21 CFR 872.547021 CFR 872.547021 CFR 872.5470Same
ClassificationNameOrthodontic PlasticBracketOrthodontic PlasticBracketOrthodontic PlasticBracketSame
Product CodeNXCNXCNXCSame
ClassificationClass IIClass IIClass IISame
OTC or RxRxRxRxSame
MaterialThermoplasticpolyurethane-polyester compositeresinThermoplasticpolyurethane-polyester compositeresinCo-polyester or co-polymerSame aspredicate
MaterialPropertiesAcceptable materialproperties establishedfor use as an aligner.Acceptable materialsproperties establishedfor use as an aligner.Acceptable materialsproperties establishedfor use as an aligner.Same
MaterialTestingASTM D790-10:Standard TestMethods for FlexuralProperties ofUnreinforced andReinforced Plastics andElectrical InsulatingMaterials.ASTM D790-10:Standard TestMethods for FlexuralProperties ofUnreinforced andReinforced Plastics andElectrical InsulatingMaterials.ASTM D790-10:Standard TestMethods for FlexuralProperties ofUnreinforced andReinforced Plastics andElectrical InsulatingMaterials.Same
BiocompatibleBiocompatibleaccording to ISO10993-1 (Biologicalevaluation of medicaldevices — Part 1:Evaluation and testingwithin a riskmanagement system).Biocompatibleaccording to ISO10993-1 (Biologicalevaluation of medicaldevices - Part 1:Evaluation and testingwithin a riskmanagement system).Biocompatibleaccording to ISO10993-1 (Biologicalevaluation of medicaldevices - Part 1:Evaluation and testingwithin a riskmanagement system).Same
SterileNon-sterileNon-sterileNon-sterileSame

{5}------------------------------------------------

SpecificationSubject Device:Smylio Invisible ClearAlignersPredicate Device:Spark Clear AlignerSystem (K203737)Reference Device:Smylio Invisible ClearAligners (K173784)ComparisonResult
DeviceDescriptionSequentialthermoformed plasticalignersSequentialthermoformed plasticalignersSequentialthermoformed plasticalignersSame
PatientRemovable?YesYesYesSame
Indication forUseThe Smylio InvisibleClear Aligners areindicated for thealignment of teethduring orthodontictreatment ofmalocclusion.The Spark™ ClearAligner System isindicated for thealignment of teethduring orthodontictreatment ofmalocclusion.The Smylio InvisibleClear Aligners isindicated for use in thealignment ofpermanent teeththrough orthodontictreatment ofmisalignmentand malocclusion.Same aspredicate.
Intended UseOrthodontic toothmovementOrthodontic toothmovementOrthodontic toothmovementSame
Mode ofActionContinuous gentleforce applied to teethfollowing theprescribed andapproved treatmentplan to achieveorthodontic movementContinuous gentleforce applied to teethfollowing theprescribed andapproved treatmentplan to achieveorthodontic movementContinuous gentleforce applied to teethfollowing theprescribed andapproved treatmentplan to achieveorthodontic movementSame

Comparison of Indications for Use to Predicate Devices and Previously Cleared Devices:

The indications for use of the Smylio Invisible Clear Aligners in this submission is the Spark Clear Aligner System (K203737) as they are both indicated for use in the alignment of teeth through orthodontic treatment of malocclusion. The differences between the indications for use of the subject device and the previously cleared device (K173784) do not raise questions of substantial equivalence. The aligners guide teeth to their final position by way of continuous gentle forces. Thus, the Smylio Invisible Clear Aligners can be considered substantially equivalent to its predicate device.

The difference between the Smylio Invisible Clear Aligners in this submission (Subject Device) and the Reference Device, Smylio Clear Aligners (K173784) is the material of the subject device is the same as the predicate device.

Comparison of Technological Characteristics to Predicate Devices:

Based on the above comparison, the design, construction, and performance characteristics of the Smylio Invisible Clear Aligners is the same as the predicate and similar to the reference device. Thus, the Smylio Invisible Clear Aligners can be considered substantially equivalent to its predicate device.

Non-clinical performance testing:

The use of thermoplastic materials for sequential aligners intended to treat malocclusions has been well documented in scientific literature regarding incremental tooth moving forces. However, durability

{6}------------------------------------------------

testing was conducted on the aligners. Real world use was simulated to ensure that the aligner material and manufacturing process produced aligners that were suitable for their prescribed period of use.

An internal manufacturing validation was performed to test the manufacturing process for Smylio Invisible Clear Aligners. The robustness of the process was demonstrated from 3D printing through thermoforming.

The thermoplastic material used for Smylio Invisible Clear Aligners has passed the required testing for material characterization. Material testing was conducted on the aligner material according to ASTM D790-10: Standard Test Methods for Flexural Properties of Unreinforced and Reinforced Plastics and Electrical Insulating Materials. Biocompatibility testing for the aligner material, the only patient contacting material, was conducted in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Additional cytotoxicity testing according to ISO 10993-5:2009 was performed on final manufactured Smylio Invisible Clear Aligners.

Clinical performance testing:

Clinical performance testing was not conducted for this 510(k) Notification. Clinical performance testing was not needed to establish substantial equivalence to the predicate device or to characterize performance. The above-mentioned non-clinical test data characterizes all performance aspects of the device based on well-established scientific and engineering principles. .

Conclusion:

Based on similarities in indications for use, technological characteristics, non-clinical performance testing, Smylio Invisible Clear Aligners are substantially equivalent to the Spark Aligner System and the previously cleared Smylio Invisible Clear Aligners (K173784).

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.