(428 days)
The disposable surgical face mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.
The proposed device, Disposable Surgical Face Mask is a three-layer, flat-folded mask. The mask body is made of 25g/m2 polypropylene (PP) non-woven cloth. The mask contains tie strings or ear loop to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. Ear loops are made of Polyurethane and spandex, and tie strings are made of polypropylene nonwoven cloth. The nose clip which is made of Iron-Zinc strip covered by polyethylene terephthalate (PET) covering. The disposable surgical face mask is available in two different specifications: 17.5×9.5cm and 16.5×8.5cm. The device is single use and provided sterile.
The document describes the testing and acceptance criteria for a "Disposable Surgical Face Mask" (K212649). This is a medical device, not an AI/ML device, therefore, many of the requested categories such as "Number of experts used to establish the ground truth," "Adjudication method," "MRMC comparative effectiveness study," and "sample size for the training set" are not applicable.
Here's the information extracted and organized as requested, with "N/A" for sections not relevant to a non-AI medical device submission:
1. Table of Acceptance Criteria and Reported Device Performance
| No | Item | Test Method/Standard | Type | Acceptance Criteria | Reported Performance for Ear Loop Type | Reported Performance for Tie-on Type |
|---|---|---|---|---|---|---|
| 1 | Particulates Filtration Efficiency | ASTM F2299/F2299M-03(2017) | Ear loop | ≥98% | Lot 1: 98.81%Lot 2: 98.78%Lot 3: 98.8% | N/A (listed separately below) |
| Tie-on | ≥98% | N/A (listed above) | Lot 1: 98.82%Lot 2: 98.78%Lot 3: 98.79% | |||
| 2 | Bacterial Filtration Efficiency | ASTM F2101-19 | Ear loop | ≥98% | Lot 1: 99.9%Lot 2: 99.91%Lot 3: 99.91% | N/A (listed separately below) |
| Tie-on | ≥98% | N/A (listed above) | Lot 1: 99.91%Lot 2: 99.9%Lot 3: 99.91% | |||
| 3 | Fluid Resistance | ASTM F1862/F1826M-17 | Ear loop | 120 mmHg | Lot 1: No penetration at 120 mmHgLot 2: No penetration at 120 mmHgLot 3: No penetration at 120 mmHg | N/A (listed separately below) |
| Tie-on | 120 mmHg | N/A (listed above) | Lot 1: No penetration at 120 mmHgLot 2: No penetration at 120 mmHgLot 3: No penetration at 120 mmHg | |||
| 4 | Differential Pressure | EN 14683:2019 | Ear loop | △P<6.0 mm H2O/cm2 | Lot 1: 3.59Lot 2: 3.59Lot 3: 3.59 | N/A (listed separately below) |
| Tie-on | △P<6.0 mm H2O/cm2 | N/A (listed above) | Lot 1: 3.61Lot 2: 3.59Lot 3: 3.61 | |||
| 5 | Flammability | ASTM F2100-19 | Ear loop | Class 1 | Lot 1: Class 1Lot 2: Class 1Lot 3: Class 1 | N/A (listed separately below) |
| Tie-on | Class 1 | N/A (listed above) | Lot 1: Class 1Lot 2: Class 1Lot 3: Class 1 |
Biocompatibility Tests:
| No | Item | Test Method/Standard | Acceptance Criteria | Results |
|---|---|---|---|---|
| 1 | Cytotoxicity | ISO 10993-5 | Non-cytotoxic | Pass (No potential toxicity) |
| 2 | Irritation | ISO 10993-10 | Non-irritating | Pass (No Irritation) |
| 3 | Skin Sensitization | ISO 10993-10 | Non-sensitizing | Pass (No Sensitization) |
2. Sample Size Used for the Test Set and the Data Provenance
For all performance tests (Particulates Filtration Efficiency, Bacterial Filtration Efficiency, Fluid Resistance, Differential Pressure, Flammability):
- Sample Size: 3 non-consecutive lots were tested for each type (Ear loop and Tie-on). For each lot, a sample size of 125 was used.
- Data Provenance: Not explicitly stated as country of origin, but the submission is from Lyncmed Medical Technology (Beijing) Co., Ltd. in China, implying testing was likely conducted in or for the company's region. The tests are described as "Non-Clinical Test Conclusion," meaning they were laboratory tests, not clinical studies involving human patients. The report does not specify if the provenance is retrospective or prospective for the samples used for testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- N/A. This is a physical medical device, and the acceptance criteria are based on standardized performance tests (e.g., ASTM, EN, ISO standards), not on human expert interpretation of data or images.
4. Adjudication Method for the Test Set
- N/A. As above, the acceptance criteria are based on objective physical and material performance standards.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- N/A. This is a physical medical device, not an AI/ML device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- N/A. This is a physical medical device. No algorithm is involved.
7. The Type of Ground Truth Used
- The "ground truth" for the device's performance is defined by the objective measurement criteria and threshold values specified in various international and national standards (e.g., ASTM F2299, ASTM F2101, ASTM F1862, EN 14683, ASTM F2100 for performance and ISO 10993 for biocompatibility). These standards set the quantitative benchmarks that the device must meet.
8. The Sample Size for the Training Set
- N/A. This is a physical medical device, not an AI/ML device. There is no "training set."
9. How the Ground Truth for the Training Set Was Established
- N/A. This is a physical medical device. There is no "training set" and therefore no ground truth establishment for it.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 25, 2022
Lyncmed Medical Technology (Beijing) Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai. 200120 China
Re: K212649
Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 26, 2022 Received: September 30, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K212649
Device Name Disposable Surgical Face Mask
Indications for Use (Describe)
The disposable surgical face mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.
Type of Use (Select one or both, as applicable)
_ | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K212649
-
- Date of Preparation: 10/24/2022
-
- Sponsor Identification
Lyncmed Medical Technology (Beijing) Co., Ltd.
Room 1601, Building No.2, Zhubang 2000 Business Building, Balizhuang Xili 99, Chaoyang District, 100022 Beijing, PEOPLE'S REPUBLIC OF CHINA
Establishment Registration Number: 3013678400
Contact Person: Joyce Jiang Position: RA Manager Tel: +86-10-8646 8058 Email: joycejiang@lyncmed.com
-
- Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
- Designated Submission Correspondent
Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
4. Identification of Proposed Device
Trade Name: Disposable Surgical Face Mask Common Name: Surgical mask
Regulatory Information Classification Name: Mask, Surgical Classification: II Product Code: FXX
Regulation Number: 21CFR 878.4040 Review Panel: General Hospital
Indication for use:
The disposable surgical face mask is intended for single use by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and particulate materials.
Device Description:
The proposed device, Disposable Surgical Face Mask is a three-layer, flat-folded mask. The mask body is made of 25g/m2 polypropylene (PP) non-woven cloth. The mask contains tie strings or ear loop to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. Ear loops are made of Polyurethane and spandex, and tie strings are made of polypropylene nonwoven cloth. The nose clip which is made of Iron-Zinc strip covered by polyethylene terephthalate (PET) covering. The disposable surgical face mask is available in two different specifications: 17.5×9.5cm and 16.5×8.5cm. The device is single use and provided sterile.
-
న్. Identification of Predicate Device
510(k) Number: K202029 Product Name: Medical Surgical Mask -
- Comparison of Technology with Predicate Device
| ITEM | Proposed Device K212649 | Predicate Device K202029 | Remark |
|---|---|---|---|
| Product Code | FXX | FXX | Same |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Indication for Use | The disposable surgical face mask is intended for single use by operating room personnel and | The medical surgical mask is intended for single use by operating room personnel and | Same |
| other general healthcare workersto protect both patients andhealthcare workers against transferof microorganisms, blood andbody fluids, and particulatematerials. | other general healthcare workersto protect both patients andhealthcare workers againsttransfer of microorganisms,blood and body fluids, andparticulate materials. | ||
| Mask style | Flat pleated | Flat pleated | Same |
| Mask color | Blue | Blue | Same |
| Design feature | Ear loop or tie-on | Ear loop or tie-on | Same |
| Dimension | 175mm×95 mm165mm×85mm | 175mm×95mm145mm×90mm120mm×70mm | Difference |
| Level | Level II | Level II | Same |
| Fluid Resistance | Pass at 120mmHg | Pass at 120mmHg | Same |
| Particulate efficiency level | Average 99.9% | Average 98.87% | Difference |
| Bacterial filtration level | Average 99.9% | Average 99.46% | Difference |
| Differential pressure | Average 3.63mmH2O/cm2 | Average 3.72mmH2O/cm2 | Difference |
| Flammability | Class 1 | Class 1 | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| Patient Contacting Material | |||
| Ear loops | Polyurethane and spandex | Nylon and polyurethane (PU) | Difference |
| Tie strings | PP non-woven cloth | Nylon | |
| Nose clip | Iron, zinc and polyethylene (PE)crosslinking material | Iron and polyethylene | |
| Mask body | 25g/m2 Polypropylene (PP) non-woven cloth | 25g/m2 Polypropylene (PP) non-woven cloth | |
| Biocompatibility | |||
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Sensitization | No Sensitization | No Sensitization | Same |
| Irritation | No Irritation | No Irritation | Same |
| Sterilization | |||
| Method | EO sterilized | EO sterilized | Same |
| SAL | 10-6 | 10-6 | |
| Ethylene oxide residuals | EO<4mg/device, ECH<9mg/device | EO<4mg/device, ECH<9mg/device |
Table 1 Comparison of Surgical Mask
{5}------------------------------------------------
Difference- Dimension
The dimension for the proposed device is different from predicate device. This difference does not affect intended use and will not raise any safety issues. In addition, the mask dimension of the proposed device is within the size range of the predicate device.
{6}------------------------------------------------
Difference -Particulate efficiency level
The test result for particulate efficiency for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask.
Difference-Bacterial filtration level
The test result for bacteria efficiency for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask.
Difference-Differential pressure
The test result for different pressure for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask.
Difference-Patient Contacting Material
The patient contact material for the proposed device is different from predicate device. However, biocompatibility test has been conducted on the proposed device and the test result does not show any adverse effect.
-
- Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- Non-Clinical Test Conclusion
-
ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
-
ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization
-
ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)
-
A ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
-
ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
-
ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks
-
ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
-
ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres
-
ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
-
EN 14683:2019 Medical face masks-Requirements and test methods
-
CPSC CS-191-53 Flammability Test Method (16 CFR 1610)
-
ISO 10993-7:2008 Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals
{7}------------------------------------------------
| No | Item | Test Method/Standard | Type | AcceptanceCriteria | Results |
|---|---|---|---|---|---|
| 1 | ParticulatesFiltrationEfficiency | ASTM F2299/F2299M-03(2017) | Ear loop | ≥98% | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: 98.81%Lot 2: 98.78%Lot 3: 98.8% |
| Tie-on | ≥98% | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: 98.82%Lot 2: 98.78%Lot 3: 98.79% | |||
| 2 | Bacterialfiltrationefficiency | ASTM F2101-19 | Ear loop | ≥98% | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: 99.9%Lot 2: 99.91%Lot 3: 99.91% |
| Tie-on | ≥98% | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: 99.91%Lot 2: 99.9%Lot 3: 99.91% | |||
| 3 | FluidResistance | ASTM F1862/F1826M-17 | Ear loop | 120 mmHg | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: No penetration at 120 mmHgLot 2: No penetration at 120 mmHgLot 3: No penetration at 120 mmHg |
| Tie-on | 120 mmHg | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: No penetration at 120 mmHgLot 2: No penetration at 120 mmHgLot 3: No penetration at 120 mmHg | |||
| 4 | DifferentialPressure | EN 14683:2019 | Ear loop | △P<6.0 mmH2O/cm2 | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: 3.59Lot 2: 3.59Lot 3: 3.59 |
| Tie-on | △P<6.0 mmH2O/cm2 | Passed3 non-consecutive lots tested, using asample size of 125/lot |
| Table 2 Summary of the Non-clinical Testing | |
|---|---|
| --------------------------------------------- | -- |
{8}------------------------------------------------
| sample size of 125/lotLot 1: 3.61Lot 2: 3.59Lot 3: 3.61 | |||||
|---|---|---|---|---|---|
| 5 | Flammability | ASTM F2100-19 | Ear loop | Class 1 | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: Class 1Lot 2: Class 1Lot 3: Class 1 |
| Tie-on | Class 1 | Passed3 non-consecutive lots tested, using asample size of 125/lotLot 1: Class 1Lot 2: Class 1Lot 3: Class 1 |
Biocompatibility
In accordance with ISO10993-1 the syringe and needle are classified as: contacting intact skin, Limited Contact (< 24hrs). The following testing was conducted:
- Cytotoxicity (ISO 10993-5)
- Irritation (ISO 10993-10)
- Skin Sensitization (ISO 10993-10)
| No | Item | Test Method/Standard | AcceptanceCriteria | Results |
|---|---|---|---|---|
| 1 | Cytotoxicity | ISO 10993-5 | Non-cytotoxic | PassUnder the condition of this study,the device has no potential toxicity. |
| 2 | Irritation | ISO 10993-10 | Non-irritating | PassUnder the condition of this study,the device has no Irritation. |
| 3 | Skin Sensitization | ISO 10993-10 | Non-sensitizing | PassUnder the condition of this study,the device has no Sensitization. |
Table 3 Summary of the Biocompatibility testing
The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10 · EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 2 year shelf-life.
{9}------------------------------------------------
8. Clinical Test Conclusion
No clinical study is included in this submission.
- Substantially Equivalent (SE) Conclusion 9.
The conclusion drawn from the nonclinical tests demonstrates that the subject device in the 510(k) submission, the Disposable Surgical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K202029.
§ 878.4040 Surgical apparel.
(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.