K Number
K212515
Device Name
MOVIX DReamy
Manufacturer
Date Cleared
2021-11-03

(85 days)

Product Code
Regulation Number
892.1720
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
Device Description
These are modified versions of the predicate mobile digital diagnostic x-ray systems. They feature motorized movement and full battery operation. Various Canon digital X-ray panels are supplied with the system. The Mobile X-Ray Unit MOVIX DReamy has a Basic configuration or Advanced configuration. The Mobile X-Ray unit MOVIX DReamy is provided with touch screen to operate as a control console. The Digital Imaging System is composed by image receptors and application for image acquisition (control console & image processing controller). The Image acquisition software "CANON CXDI Control Software NE" runs on MOVIX DReamy and it is displayed on touch screen. The MOVIX DReamy Mobile X-Ray Unit is provided with separate battery packs for X-Ray generation and motorized movements of the unit can operate connected to mains or in stand-alone mode. New X-ray generator (model SHFM) is mounted on Battery Mobile X-Ray unit MOVIX DReamy. This X-Ray generator for MOVIX DReamy Mobile X-Ray Unit has a radiogenic unit mounted on head-assembly and comprising an X-ray tube with rotating anode and its circuit for high voltage. The electronic and associated software to control the X-ray generation are placed on mobile cart. The Battery Mobile X-Ray Unit MOVIX DReamy is provided with different output powers: 20 kW, 32 kW, 40 kW and 50 kW. There are available two X-ray tube inserts with rotating anode manufactured by CANON ELECTRON TUBES & DEVICES: XRR-3331 insert and E7886 insert. The Manual Beam Limiting Device is from Ralco, model: R108 F. External interface (controls) and covers are provided by Sedecal. There are two versions of collimator assembly, Basic and Advanced.
More Information

No
The summary describes a standard digital mobile X-ray system with image acquisition and processing software, but there is no mention of AI or ML capabilities, training data, or performance metrics typically associated with AI/ML algorithms.

No
The device is described as a "Digital Mobile Diagnostic X-Ray System" intended for "taking diagnostic exposures," not for treatment or therapy.

Yes

The "Intended Use / Indications for Use" section explicitly states that the system is "intended for use... for taking diagnostic exposures."

No

The device is a digital mobile diagnostic X-ray system that includes significant hardware components such as an X-ray generator, X-ray tube, collimator, digital X-ray panels, and a mobile cart with motorized movement and battery operation. While it includes software for image acquisition and control, it is not solely software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that the device is for taking diagnostic exposures of various body parts using X-rays. This is an in-vivo imaging procedure, not an in-vitro test.
  • Device Description: The description details an X-ray system, including an X-ray generator, tube, collimator, and digital imaging system for acquiring and processing X-ray images. This is consistent with an in-vivo imaging device.
  • Input Imaging Modality: The input modality is X-Ray, which is used for imaging inside the body (in-vivo).
  • Anatomical Site: The device is used to image various anatomical sites within the body.
  • No mention of in-vitro testing: There is no mention of analyzing samples of human origin (like blood, urine, tissue) outside the body, which is the defining characteristic of an IVD.

Therefore, this device is a diagnostic imaging system used for in-vivo procedures, not an in-vitro diagnostic device.

N/A

Intended Use / Indications for Use

This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Product codes

IZL, MQB

Device Description

These are modified versions of the predicate mobile digital diagnostic x-ray systems. They feature motorized movement and full battery operation. Various Canon digital X-ray panels are supplied with the system. The Mobile X-Ray Unit MOVIX DReamy has a Basic configuration or Advanced configuration. Advanced configurations include: Second screen on head assembly. Smooth movements of head assembly. Telescopic arm in four steps instead three steps for basic configuration.

The Mobile X-Ray unit MOVIX DReamy is provided with touch screen to operate as a control console. The Digital Imaging System is composed by image receptors and application for image acquisition (control console & image processing controller). The Image acquisition software "CANON CXDI Control Software NE" runs on MOVIX DReamy and it is displayed on touch screen. It is the user interface and compatible digital detectors are listed above. All have FDA Clearance. The Advanced configuration has a second Touch Screen Monitor located on head-assembly.

The MOVIX DReamy Mobile X-Ray Unit is provided with separate battery packs for X-Ray generation and motorized movements of the unit can operate connected to mains or in stand-alone mode, that is, operating without mains being present or unplugged from mains. The unit is connected to mains to charge the batteries; Rating: the input voltage range goes from 100 V~ to 240 V~, 1 phase, 1.1 kVA. New X-ray generator (model SHFM) is mounted on Battery Mobile X-Ray unit MOVIX DReamy This X-Ray generator for MOVIX DReamy Mobile X-Ray Unit has a radiogenic unit mounted on headassembly and comprising an X-ray tube with rotating anode and its circuit for high voltage. The electronic and associated software to control the X-ray generation are placed on mobile cart. The Battery Mobile X-Ray Unit MOVIX DReamy is provided with different output powers: 20 kW, 32 kW, 40 kW and 50 kW. There are available two X-ray tube inserts with rotating anode manufactured by CANON ELECTRON TUBES & DEVICES: XRR-3331 insert. E7886 insert.

The Manual Beam Limiting Device is from Ralco, model: R108 F. External interface (controls) and covers are provided by Sedecal. There are two versions of collimator assembly, Basic and Advanced.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-Ray

Anatomical Site

skull, spinal column, chest, abdomen, extremities, and other body parts

Indicated Patient Age Range

adult and pediatric subjects

Intended User / Care Setting

qualified/trained doctor or technician

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Summary of non-clinical testing: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. Firmware was validated according to the FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014. The digital panel software employed was already reviewed by FDA in K190368. Because the units are indicated for the pediatric population (as well as the adult population) we reviewed and took into account the FDA guidance Pediatric Information for X-ray Imaging Device Premarket Notifications.

The MOVIX DReamy Units have been tested to be in compliance with the following International Standards:
a) IEC 60601-1:2005+A1:2012 (Edition 3.1)
b) IEC 60601-1-2:2014 (Edition 4.0)
c) IEC 60601-1-3:2008+A1:2013 (Edition 2.1)
d) IEC 60601-2-54:2009+A1:2015 (Edition 1.1)
e) IEC 60601-2-28:2010 (Edition 2.0)
f) IEC 60601-1-6:2010 + A1:2013 (Edition 3.1)
g) IEC 62304:2006 + A1:2016 (Edition 1.1)
Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance.

Key Metrics

Not Found

Predicate Device(s)

K192011

Reference Device(s)

K131106, K133693, K170332, K171270, K190368

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.

November 3, 2021

Stephanix % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

Re: K212515

Trade/Device Name: MOVIX DReamy Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL, MQB Dated: August 5, 2021 Received: August 10, 2021

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

, for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K212515

Device Name MOVIX DReamy

Indications for Use (Describe)

This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary: 510(k) Number K212515

Image /page/3/Picture/1 description: The image shows the logo for Stephanix Medical Imaging Solutions. The logo features a blue square with a stylized "X" inside, followed by the word "STEPHANIX" in large, bold, blue letters. Below the company name, the words "MEDICAL IMAGING SOLUTIONS" are written in smaller, lighter blue letters.

Stephanix 10 Rue Jean Moulin La Ricamarie Auvergne-Rhone-Alpes, FR 42150 (Tel): +33 4 77 47 81 60 Registration Number: 3006972752 Date Prepared: October 22, 2021 Contact: Sandie Perret Quality Manager

    1. Identification of the Device: Trade/Device Name: MOVIX DReamy Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
    1. Equivalent legally marketed device: K192011 Trade/Device Name: PhoeniX Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
    1. Reference devices: We employ these cleared devices (K131106, K133693, K170332 and K171270 with software update cleared in K190368) without modification: Trade/Device Name: Canon Digital X-ray Panels CANON CXDI-401C Wireless CANON CXDI-701C Wireless CANON CXDI-801C Wireless CANON CXDI-710C Wireless CANON CXDI-810C Wireless CANON CXDI-410C Wireless Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
    1. Indications for Use: This Digital Mobile Diagnostic X-Ray System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts.

4

Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

    1. Description of the Device: These are modified versions of the predicate mobile digital diagnostic x-ray systems. They feature motorized movement and full battery operation. Various Canon digital X-ray panels are supplied with the system. (See list above.) The Mobile X-Ray Unit MOVIX DReamy has a Basic configuration or Advanced configuration. Advanced configurations include:
    • Second screen on head assembly.
    • Smooth movements of head assembly.
    • Telescopic arm in four steps instead three steps for basic configuration.

The Mobile X-Ray unit MOVIX DReamy is provided with touch screen to operate as a control console. The Digital Imaging System is composed by image receptors and application for image acquisition (control console & image processing controller). The Image acquisition software "CANON CXDI Control Software NE" runs on MOVIX DReamy and it is displayed on touch screen. It is the user interface and compatible digital detectors are listed above. All have FDA Clearance. The Advanced configuration has a second Touch Screen Monitor located on head-assembly.

The MOVIX DReamy Mobile X-Ray Unit is provided with separate battery packs for X-Ray generation and motorized movements of the unit can operate connected to mains or in stand-alone mode, that is, operating without mains being present or unplugged from mains. The unit is connected to mains to charge the batteries; Rating: the input voltage range goes from 100 V~ to 240 V~, 1 phase, 1.1 kVA. New X-ray generator (model SHFM) is mounted on Battery Mobile X-Ray unit MOVIX DReamy This X-Ray generator for MOVIX DReamy Mobile X-Ray Unit has a radiogenic unit mounted on headassembly and comprising an X-ray tube with rotating anode and its circuit for high voltage. The electronic and associated software to control the X-ray generation are placed on mobile cart. The Battery Mobile X-Ray Unit MOVIX DReamy is provided with different output powers: 20 kW, 32 kW, 40 kW and 50 kW. There are available two X-ray tube inserts with rotating anode manufactured by CANON ELECTRON TUBES & DEVICES:

  • XRR-3331 insert.
  • E7886 insert.

The Manual Beam Limiting Device is from Ralco, model: R108 F. External interface (controls) and covers are provided by Sedecal. There are two versions of collimator assembly, Basic and Advanced.

Image /page/4/Picture/10 description: The image is a comparison of two medical devices, labeled "Basic" and "Advanced". The "Basic" device has a white and gray body with a black control panel that has two knobs and a port. The "Advanced" device has a similar white and gray body, but it also has a display screen and additional components.

5

BasicAdvanced
Knobs to adjust the collimator blades.
Collimator light push button.
Rail system to install external additional
filtration.
Measuring tape to measure the SID.
Double laser (optional)
Motorized filter (optional)Knobs to adjust the collimator blades.
Collimator light push button.
Rail system to install external additional filtration.
Measuring tape to measure the SID.
Double laser (optional)
Motorized filter (optional)
Rear knobs to adjust the collimator blades (optional).
Additional options selectable with TFT:
Variable additional filtration.
Collimator blades aperture.

The Battery Mobile X-Ray Unit MOVIX DReamy is driven by holding the handlebar with the hands. It is provided to internal sensor (gauges) that control the direction and speed of each wheel depending the force applied by the operator.

The handlebar is capacitive handlebar and when operator put his handlebar the movements are enabled. The height of handlebar is adaptable to improve the user commodity.

Four buttons are provided on the hand grips placed on Head-Assembly. They control the motion of each driving wheel and allow fine positioning adjustment of the patient (movements with low speed because this control is not intended for long displacement).

The Battery Mobile X-Ray unit MOVIX DReamy supports:

  • Ethernet communication (with hospital and backup cable for digital detectors) ●
  • Wireless Wi-Fi communications (with digital detectors and remote hand switch). The wireless . communication is supported by Access Point and one single internal antenna.

Options:

Barcode scanner for register patient information. Dosimetry for Dose Area Product (DAP).

Tube Arm specifications:

-Counterbalanced tube arm.

  • Tube arm reach 1220 mm horizontally for both configurations (Advanced option with 4 steps and Basic option with 3 steps).

-Column rotation: ± 317 °.

  • Maximum SID to floor is 2020 mm for both configurations.

  • Minimum SID to floor is 530 mm for both configurations.

Mobile Unit:

  • Weight: the heaviest configuration 520 kg.
    -Size (L x W x H):

  • Mobile on parking position: 1220 x 540 x 1290 mm. o

  • Mobile fully extended (Telescopic column & Telescopic arm): 2290 x 1790 x 2230 mm. O

-Frontal anti-collision system.

  • Motorized unit up to 5,5 km/h.

  • Motor drive with two independent drive motors, one for each wheel (forward and reverse).

-Touch screen console:

  • o Main control console: 19 inch.
  • Additionally, second touch screen of 8.4 inch for Advanced configuration O

6

Substantial Equivalence Chart 6)

CharacteristicPhoenix K192011 (Sedecal)MOVIX DReamy
Indications
for Use:Intended for use by a qualified/trained
doctor or technician on both adult and
pediatric subjects for taking diagnostic
radiographic exposures of the skull, spinal
column, chest, abdomen, extremities, and
other body parts. Applications can be
performed with the patient sitting,
standing, or lying in the prone or supine
position. Not for mammographyThis Digital Mobile Diagnostic X-Ray System is
Intended for use by a qualified/trained doctor
or technician on both adult and pediatric
subjects for taking diagnostic radiographic
exposures of the skull, spinal column, chest,
abdomen, extremities, and other body parts.
Applications can be performed with the patient
sitting, standing, or lying in the prone or supine
position. Not for mammography. (SAME,
device description added)
ConfigurationMobile System with digital x-ray panel
and image acquisition computerSAME
X-ray
Generator(s)kW rating: 20 kW, 32 kW, 40 kW and 50 kW.
kV range: from 40 kV to 150 kV in 1 kV steps.
mA range: from 10 mA to 630 mA / 640 mA /
650 mA.SAME
CollimatorRalco R108FSAME
PhotosImage: Phoenix K192011Image: MOVIX DReamy
Digital X-ray
Panel
SuppliedCANON CXDI-401C Wireless
CANON CXDI-701C Wireless
CANON CXDI-801C Wireless
Plus:
CANON CXDI-710C Wireless
CANON CXDI-810C Wireless
CANON CXDI-410C WirelessSAME
CharacteristicPhoeniX K192011 (Sedecal)MOVIX DReamy
SoftwareCanon control software CXDI-NESAME
Panel
InterfaceEthernet or Wi-Fi wirelessSAME
Meets US
Performance
StandardYES 21 CFR 1020.30SAME
Power
SourceUniversal power supply, from 100 V~ to
240 V~. 1 phase, 1.1 kVASAME

7

    1. The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination.
    1. Summary of non-clinical testing: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. Firmware was validated according to the FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014. The digital panel software employed was already reviewed by FDA in K190368. Because the units are indicated for the pediatric population (as well as the adult population) we reviewed and took into account the FDA guidance Pediatric Information for X-ray Imaging Device Premarket Notifications.

The MOVIX DReamy Units have been tested to be in compliance with the following International Standards:

  • a) IEC 60601-1:2005+A1:2012 (Edition 3.1)
  • b) IEC 60601-1-2:2014 (Edition 4.0)
  • c) IEC 60601-1-3:2008+A1:2013 (Edition 2.1)
  • d) IEC 60601-2-54:2009+A1:2015 (Edition 1.1)
  • e) IEC 60601-2-28:2010 (Edition 2.0)
  • f) IEC 60601-1-6:2010 + A1:2013 (Edition 3.1)
  • g) IEC 62304:2006 + A1:2016 (Edition 1.1)
    1. Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance.
    1. Conclusion: After analyzing bench and clinical tests, it is the conclusion of STEPHANIX that the new MOVIX DReamy Digital Diagnostic Mobile X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.