(125 days)
Not Found
No
The device is a standard examination glove and the summary contains no mention of AI or ML.
No
The device, "Comfortpro latex examination powder free gloves," is described as a disposable device intended for medical purposes worn on the hand to prevent contamination. Its function is protective, not therapeutic, as it does not treat or alleviate a disease or condition.
No
Explanation: The device is described as a glove intended to prevent contamination, and its performance studies focus on physical and biocompatibility properties, not on diagnosing medical conditions.
No
The device is a physical product (gloves) made of natural rubber latex, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the gloves are "worn on the examiner's hand to prevent contamination between patient and examiner." This describes a barrier device used for protection during medical examinations.
- Device Description: The description focuses on the physical properties and manufacturing standards of examination gloves made from natural rubber latex.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue, etc.) to provide information for diagnosis, monitoring, or screening. IVDs are typically reagents, instruments, or systems used for in vitro testing.
Therefore, the Comfortpro latex examination powder free gloves are classified as a medical device, specifically a barrier device, and not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
Comfortpro latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Product codes
LYY
Device Description
COMFORTPRO Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-19, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are Natural in color (No color added) and are powder free.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
hand
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Used by examiner for medical purpose.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench testing was conducted according to ASTM D3578-19, ASTM D5151-19, ASTM D6124-06 (Reapproved 2017), and ASTM D5712-15 (Reapproved 2020), covering dimensions (length, width, thickness), physical properties (tensile strength, ultimate elongation), presence of holes, residual powder, and extractable protein. All tested characteristics met the specified acceptance criteria.
Biocompatibility testing followed ISO 10993-10:2010(E) for irritation and skin sensitization, ISO10993-5:2009(E) for in vitro cytotoxicity, and ISO 10993-11:2017(E) for acute systemic toxicity. The results indicated that the device is not an irritant, not a sensitizer, and does not pose a systemic toxicity concern. While the initial cytotoxicity test showed cytotoxicity for 100% (neat) extract, a follow-up acute systemic toxicity test demonstrated no acute systemic toxicity concern.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 7, 2021
Thai Rubber Industry Company Limited % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114, Aurora, Illinois 60504
Re: K212438
Trade/Device Name: Comfortpro Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYY Dated: November 4, 2021 Received: November 4, 2021
Dear Manoj Zacharias:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
1
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K212438
Device Name Comfortpro
Indications for Use (Describe)
Comfortpro latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
A. APPLICANT INFORMATION
510(K) Owner's Name | THAI RUBBER INDUSTRY COMPANY LIMITED |
---|---|
Address | 738 MOO 5, MANAM KOO, |
PLUAKDAENG | |
RAYONG 21140 THAILAND | |
Phone | +66-81-6298773, +66-85-6000373 |
Fax | - |
chawalit@thairubberindustry.com | |
Contact Person | Mr. Chawalit Tiyadechachai |
Designation | Owner and Managing Director |
Contact Number | +66-81-6298773, +66-85-6000373 |
Contact Email | chawalit@thairubberindustry.com, |
Chawalit.tiya@gmail.com | |
Date Submitted | July 15, 2021 |
B. DEVICE IDENTIFICATION
Name of the device | COMFORTPRO |
---|---|
Product proprietary or trade name | COMFORTPRO |
Common or usualname | Latex examination powder free gloves |
Classification name | Latex Patient Examination Glove |
Device Classification | Class-1 |
Product Code | LYY |
Regulation Number | 21 CFR 880.6250 |
Review Panel | General Hospital |
C. PREDICATE DEVICE
Predicate Device | Hi-Care Thai Gloves Co. Ltd. |
---|---|
510(k) Number | K202377 |
Regulatory Class | Class 1 |
Product code | LYY |
D. DESCRIPTION OF THE DEVICE:
COMFORTPRO Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-19, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are Natural in color (No color added) and are powder free.
4
E. INDICATIONS FOR USE OF THE DEVICE:
COMFORTPRO latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | Comparison | ||
---|---|---|---|---|---|
PREDICATE | CURRENT | ||||
510(k) Number | -- | K202377 | K212438 | ||
Name of device | -- | Palm Care Latex | |||
Examination Powder | |||||
Free Gloves | ComfortPro latex Examination | ||||
Powder Free Gloves | -- | ||||
Dimensions- Length | ASTMD3578-19 | Length > 230 mm | Length > 230 mm | Similar | |
Size | Average | ||||
X-Small | 245 mm | ||||
Small | 238 mm | ||||
Medium | 233 mm | ||||
Large | 240 mm | ||||
Dimensions- Width | ASTMD3578-19 | Width Min 95+/- 10 | |||
mm (for medium | |||||
size) | Width Min 95+/-10 | ||||
mm (for medium size) | Similar | ||||
Size | Average | ||||
X-Small | 77 mm | ||||
Small | 81 mm | ||||
Medium | 93 mm | ||||
Large | 101 mm | ||||
Physical Properties- | |||||
Tensile Strength | ASTMD3578-19 | Before Ageing | |||
Tensile Strength |
18 Mpa After
Ageing Tensile
Strength
14 Mpa | Before Ageing
Tensile Strength > 18 Mpa | Similar | |
| | | | Size | Actual value | |
| | | | X-Small | 29.6 | |
| | | | Small | 25 | |
| | | | Medium | 26 | |
| | | | Large | 24.9 | |
| | | | After Ageing
Tensile Strength > 14 Mpa | | Similar |
| | | | Size | Actual
value | |
| | | | X-Small | 25.5 | |
| | | | Small | 24.1 | |
| | | | Medium | 22.2 | |
| | | | Large | 24.4 | |
5
CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | Comparison | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
PREDICATE | CURRENT | |||||||||||||||||||||||
510(k) Number | -- | K202377 | K212438 | |||||||||||||||||||||
Physical Properties- | ||||||||||||||||||||||||
Ultimate Elongation | ASTMD3578-19 | Before Ageing | ||||||||||||||||||||||
Ultimate | ||||||||||||||||||||||||
Elongation |
650%
After Ageing
Ultimate
Elongation
500% | Before Ageing
Ultimate Elongation > 650%
Size Actual value X-Small 810 Small 760 Medium 680 Large 780
After Ageing
Ultimate Elongation > 500% Size Actual value X-Small 820 Small 760 Medium 650 Large 760 | | | | | | | | | | | | | | | | | | | | | Similar |
| Thickness | ASTMD3578-19 | Palm > 0.08 mm
Finger > 0.08 mm | Palm > 0.08 mm
Finger > 0.08 mm Size Palm (Actual value) Finger (Actual value) X-Small 0.102 mm. 0.114 mm. Small 0.087 mm. 0.109 mm. Medium 0.10 mm. 0.122 mm. Large 0.108 mm. 0.121 mm. | | | | | | | | | | | | | | | | Similar | | | | | |
| Powder Free Residue | ASTMD3578-19 | $\leq$ 2 mg/glove | 0.68 mg/glove | Similar | | | | | | | | | | | | | | | | | | | | |
| Biocompatibility | Primary Skin
Irritation-ISO
10993-10:2010(E) | Under the
condition of
study, not an
irritant | Under the condition of study
not an irritant | Same | | | | | | | | | | | | | | | | | | | | |
| | Dermal
Sensitization-ISO
10993-10:2010(E) | Under the
conditions of the
study, not a
sensitizer | Under the conditions of the
study, not a sensitizer | Same | | | | | | | | | | | | | | | | | | | | |
| | In vitro
cytotoxicity
ISO10993-5
:2009(E) | Under the
conditions
of the study, non-
cytotoxic | Under the conditions of the
study, non-cytotoxic | Same | | | | | | | | | | | | | | | | | | | | |
| | Acute Systemic
Toxicity Test ISO
10993-11:2017(E) | Under the
conditions of study
the device extracts
do not pose a
systemic toxicity
concern | Under the conditions
of study the device extracts
do not pose a systemic
toxicity concern | same | | | | | | | | | | | | | | | | | | | | |
6
| CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON
ON |
|-----------------------|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| | | PREDICATE | CURRENT | |
| 510(k) Number | | K202377 | K212438 | |
| Water Tight (1000 ml) | ASTM D5151-19
AQL-2.5 | Passes | Passes | Similar |
| Indication for Use | | Latex Examination
Powder Free Gloves are
disposable devices intended
for medical purpose that are
worn on the examiner's
hand to prevent
contamination between
patient and examiner. | ComfortPro latex
examination powder free
gloves is disposable
device intended for
medical purpose that is
worn on the examiner's
hand to prevent
contamination
between patient and
examiner. | Same |
| Material | | Natural Latex | Natural Latex | Identical |
| Color | | Natural (No color is added) | Natural color (No color is
added) | Similar |
| Size | ASTMD3578-19 | X Small, Small, Medium,
Large | X Small, Small,
Medium, Large | Similar |
| Single Use | Medical Glove
Guidance
Manual-
Labeling | Single Use | Single Use | Same |
| Sterile/nonsterile | | Nonsterile | Nonsterile | Same |
| Powder/Powder
free | | Powder free | Powder free | Same |
| Label and
Labeling | FDA Label
requirements | Meets FDA's label and
labeling requirements | Meets FDA's
label and labeling
requirements | Same |
| Manufacturer(s) | | Hi-Care Thai Gloves Co.
Ltd. | THAI RUBBER
INDUSTRY COMPANY
LIMITED | --- |
Both devices meet the ASTM standard D3578.
7
G. COMPARISON BASED ON AN ASSESSMENT OF NON-CLINICAL PERFORMANCE DATA
BENCH TEST DATA
TEST METHOD | PURPOSE | ACCEPTANCE CRITERIA | RESULT | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
ASTM D3578-19 | ||||||||||||||||||
Standard | ||||||||||||||||||
Specification for | ||||||||||||||||||
Rubber Examination | ||||||||||||||||||
Gloves | To determine the | |||||||||||||||||
length of the gloves | Min 230 mm for all | |||||||||||||||||
sizes | X-Small : 245 mm | |||||||||||||||||
Small : 238 mm | ||||||||||||||||||
Medium : 233 mm | ||||||||||||||||||
Large : 240 mm | ||||||||||||||||||
ASTM D3578-19 | ||||||||||||||||||
Standard | ||||||||||||||||||
Specification for | ||||||||||||||||||
Rubber Examination | ||||||||||||||||||
Gloves | To determine the | |||||||||||||||||
width of the gloves | X-Small : 70+/-10 mm | |||||||||||||||||
Small : 80+/-10 mm | ||||||||||||||||||
Medium : 95+/-10 mm | ||||||||||||||||||
Large : 111+/-10 mm | X-Small : 77 mm | |||||||||||||||||
Small : 81 mm | ||||||||||||||||||
Medium : 93 mm | ||||||||||||||||||
Large : 101 mm | ||||||||||||||||||
ASTM D3578-19 | ||||||||||||||||||
Standard | ||||||||||||||||||
Specification for | ||||||||||||||||||
Rubber Examination | ||||||||||||||||||
Gloves | To determine the | |||||||||||||||||
thickness of the | ||||||||||||||||||
gloves | Palm | |||||||||||||||||
0.08 mm min for all | ||||||||||||||||||
sizes | ||||||||||||||||||
Finger | ||||||||||||||||||
0.08 mm min for all | ||||||||||||||||||
sizes | Size Palm Finger X-Small 0.102 mm 0.114 mm Small 0.087 mm 0.109 mm Medium 0.10 mm 0.122 mm Large 0.108 mm 0.121 mm | |||||||||||||||||
ASTM D3578-19 | ||||||||||||||||||
Standard | ||||||||||||||||||
Specification for | ||||||||||||||||||
Rubber Examination | ||||||||||||||||||
Gloves | To determine the | |||||||||||||||||
physical properties- | ||||||||||||||||||
Tensile strength | Before Ageing | |||||||||||||||||
Tensile Strength | ||||||||||||||||||
18MPa Min for all | ||||||||||||||||||
sizes | ||||||||||||||||||
After Ageing | ||||||||||||||||||
Tensile Strength | ||||||||||||||||||
14MPa Min for all | ||||||||||||||||||
sizes | Size Before | |||||||||||||||||
ageing After | ||||||||||||||||||
ageing X-Small 29.6 MPa 25.5 MPa Small 25 MPa 24.1 MPa Medium 26 MPa 22.2 MPa Large 24.9 MPa 24.4 MPa | ||||||||||||||||||
To determine the | ||||||||||||||||||
physical properties- | ||||||||||||||||||
Ultimate | ||||||||||||||||||
Elongation | Before Ageing | |||||||||||||||||
Ultimate Elongation | ||||||||||||||||||
650% Min for all sizes | ||||||||||||||||||
After Ageing | ||||||||||||||||||
Ultimate Elongation | ||||||||||||||||||
500% Min for all sizes | Size Before | |||||||||||||||||
ageing After | ||||||||||||||||||
ageing X-Small 810% 820% Small 760% 760% Medium 680% 650% Large 780% 760% |
8
| TEST METHOD | PURPOSE | ACCEPTANCE
CRITERIA | RESULT |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------|------------------------|
| ASTM D5151-19 Standard
Test Method for Detection
of Holes in Medical Gloves | To determine the holes
in the gloves | AQL 2.5 | Gloves Passes AQL 2.5 |
| ASTM D6124-06
(Reapproved 2017) Standard
Test Method for Residual
Powder on Medical Gloves | To determine the
residual powder in the
gloves | ≤2 mg/glove | Medium : 0.68 mg/glove |
| ASTM D5712-15
(Reapproved 2020) Standard
Test Method for Analysis of
Aqueous Extractable Protein
in Latex, Natural Rubber, and
Elastomeric Products Using
the Modified Lowry Method | To determine the
extractable protein in
the gloves. | 200 µg/ dm² Max | Medium : 187.1 µg/ dm² |
BIOCOMPATIBILITY DATA
| TEST METHOD | PURPOSE | ACCEPTANCE
CRITERIA | RESULT |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-10:2010(E)
Biological Evaluation Of
Medical Devices - Part 10,
Tests for Irritation and Skin
Sensitization.
Test done for irritation. | To evaluate the test
item, for skin irritation
test in New Zealand
White rabbits. | Under the condition
of study, not an
irritant | Under the condition of
study, not an irritant |
| ISO 10993 10:2010(E)
Biological Evaluation of
Medical Devices - Part 10,
Tests for Irritation and Skin
Sensitization.
Test done for skin sensitization | To evaluate the test
item, for the skin
sensitization in Guinea
pigs by maximization
test. | Under the conditions
of the study, not a
sensitizer | Under the conditions of
the study, not a
sensitizer |
| ISO10993-5:2009(E)
Biological Evaluation of
Medical Devices - Part 5,
Tests for In Vitro
Cytotoxicity. | To evaluate the test
item, for its ability to
induce cytotoxicity
using L-929 mouse
fibroblast cells by
Elution Method. | Under the conditions
of the study, non-
cytotoxic | Under the conditions of the
study cytotoxic for 100%
(neat) test item extract. As
a follow up, acute
systemic toxicity testing
was performed to
demonstrate the extract
did not present an acute
systemic toxicity concern. |
| ISO 10993-11:2017(E)
Biological Evaluation of
Medical Devices - Part 11,
Tests for Systemic Toxicity. | To evaluate the test
item, for acute systemic
toxicity in Swiss Albino
Mice. | Under the conditions
of study, the device
extracts do not pose a
systemic toxicity
concern | Under the conditions
of study the device
extracts do not pose a
systemic toxicity concern |
9
510(K) SUMMARY
AS REQUIRED BY: 21CFR§807.92(C)
The performance test data of the non-clinical tests meet following standards:
ASTM D3578-19 Standard Specification for Rubber Examination Gloves
ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medica1Gloves
ASTM D5712-15 (Reapproved 2020) Standard Test Method for Analysis of Aqueous Extractable Protein in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
ISO 10993-10:2010 (E)Biological Evaluation of Medical Devices - Part 10, Tests for Irritation and Skin Sensitization.
ISO 10993-5:2009(E) Biological Evaluation of Medical Devices - Part 5, Tests for In Vitro Cytotoxicity.
ISO 10993-11:2017 (E) Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity.
H. COMPARISON BASED ON ASSESSMENT OF CLINICAL PERFORMANCE DATA
Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
I. CONCLUSION
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission ComfortPro Latex Examination Powder Free Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicate device K202377.