(81 days)
Not Found
No
The document describes a sterilization tray for surgical instruments and does not mention any AI or ML capabilities.
No.
The device is a sterilization tray intended to hold and protect surgical instruments during storage and sterilization. It does not directly provide therapy to a patient.
No
This device is a sterilization tray used to enclose and protect surgical instruments for storage and sterilization; it does not perform any diagnostic function.
No
The device description clearly states it is a "rigid containment device consisting of a base with lid," indicating it is a physical hardware product, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to enclose and protect surgical instruments for storage and sterilization. This is a function related to the reprocessing of medical devices, not the diagnosis of disease or other conditions.
- Device Description: The device is described as a rigid containment device for reprocessing surgical instruments.
- Lack of Diagnostic Function: There is no mention of the device being used to examine specimens from the human body to provide information for diagnosis, monitoring, or screening.
IVD devices are specifically designed to perform tests on biological samples (like blood, urine, tissue) to provide diagnostic information. This device's function is entirely focused on the sterilization and storage of surgical tools.
N/A
Intended Use / Indications for Use
The UroLift System Rigid Retrieval Kit Sterilization Tray is intended to enclose and protect surgical instruments optionally used for the UroLift System procedure for storage and sterilization in a prevacuum steam sterilizer. The following instruments are intended to be loaded into the UroLift System Rigid Retrieval Kit Sterilization Tray: · UL-SCOPE4-0 / UL-SCOPE4-0-FE* - Cystoscope, 4.0 mm diameter x 302 mm length, 0° angle of view, wide angle - · UL-GRASP-R Optical Graspers, double action jaws - · UL-SCI-R Optical Scissors, double action jaws * Factory Exchange: Devices with a part number ending in "FE" have been repaired to meet product specifications. The tray by itself is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, 510(k) cleared sterilization wrap to maintain sterility of the enclosed instruments when sterilized using the following prevacuum steam sterilization cycles: Prevacuum Steam Temperature: 132°C (270°F) Sterilization time: 4 minutes Minimum dry time: 30 minutes Maximum weight: 5 lbs (2.7 kg)
Product codes (comma separated list FDA assigned to the subject device)
KCT
Device Description
The UroLift System Rigid Retrieval Kit Sterilization Tray is a rigid containment device consisting of a base with lid which enables reprocessing of surgical instruments that are optionally used in the UroLift System Procedure.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Usability Testing: The user is able to load and unload the tray, perform the reprocessing workflow (Clean the tray, inspect the tray, clean and inspect the instruments, and load the tray and sterilize), and determine the end of the serviceable life of the tray. Results: Pass.
Cytotoxicity Testing: The test sample meets the requirements of the test if the biological response is less than or equal to grade 2 (mild). Results: Pass.
Sensitization Testing: Grades of 1 or greater observed in the test group generally indicate sensitization, provided that grades of less than 1 are observed on the control animals. If grades of 1 or greater are noted on control animals, then the reactions of test animals that exceeded the most severe control reaction will be considered to be due to sensitization. Results: Pass.
Intracutaneous Reactivity Testing: The difference of overall mean of the test group to the control group on erythema and edema score must be less than 1. Results: Pass.
Useful Life Testing: The tray must pass the inspection criteria on the IFU after 100 reprocessing cycles. Results: Pass.
Cleaning Validation: Per the protocol, there were three acceptance criteria; protein residual analysis, hemoglobin residual analysis, and visual inspection. The protein residual benchmark level was 6.4µg/cm² and the hemoglobin benchmark level was 2.2µg/cm². The tray must pass the visual inspection in the IFU. Results: Pass.
Sterilization Validation: The minimum sterility assurance level (SAL) of 10-6 can be achieved if the sterilization instructions in the IFU were followed. Results: Pass.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 22, 2021
NeoTract, Inc. Brian Gall Senior Manager of Regulatory Affairs 4155 Hopvard Road Pleasanton, California 94588
Re: K212396
Trade/Device Name: UroLift System Rigid Retrieval Kit Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: September 22, 2021 Received: September 23, 2021
Dear Brian Gall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K212396
Device Name
UroLift® System Rigid Retrieval Kit Sterilization Tray
Indications for Use (Describe)
The UroLift System Rigid Retrieval Kit Sterilization Tray is intended to enclose and protect surgical instruments optionally used for the UroLift System procedure for storage and sterilization in a prevacuum steam sterilizer. The following instruments are intended to be loaded into the UroLift System Rigid Retrieval Kit Sterilization Tray: · UL-SCOPE4-0 / UL-SCOPE4-0-FE* - Cystoscope, 4.0 mm diameter x 302 mm length, 0° angle of view, wide angle
- · UL-GRASP-R Optical Graspers, double action jaws
- · UL-SCI-R Optical Scissors, double action jaws
- Factory Exchange: Devices with a part number ending in "FE" have been repaired to meet product specifications.
The tray by itself is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, 510(k) cleared sterilization wrap to maintain sterility of the enclosed instruments when sterilized using the following prevacuum steam sterilization cycles:
Prevacuum Steam Temperature: 132°C (270°F) Sterilization time: 4 minutes Minimum dry time: 30 minutes Maximum weight: 5 lbs (2.7 kg)
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) SUMMARY K212396
COMPANY INFORMATION
NeoTract, Inc. 4155 Hopyard Road Pleasanton, CA 94588 Registration Number: 3015181082
SUBMISSION CORRESPONDENT
Brian Gall Senior Regulatory Affairs Manager, Interventional Urology NeoTract, Inc. 4155 Hopyard Road Pleasanton, CA 94588
Telephone - 925.329.6547 E-mail – brian.gall@teleflex.com
DATE PREPARED
20 October 2021
DEVICE INFORMATION
Trade Name: | UroLift® System Rigid Retrieval Kit Sterilization Tray (UL-RRKTRAY) |
---|---|
Common Name: | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories |
Classification Name: | Sterilization Wrap |
Product Code: | KCT |
Regulation Number: | 880.6850 |
Classification: | II |
Classification Panel: | Surgical and Infection Control Devices (OHT4) Infection Control and Plastic and Reconstructive Surgery (DHT4B) |
DEVICE DESCRIPTION
The UroLift System Rigid Retrieval Kit Sterilization Tray is a rigid containment device consisting of a base with lid which enables reprocessing of surgical instruments that are optionally used in the UroLift System Procedure.
INDICATIONS FOR USE
The UroLift System Rigid Retrieval Kit Sterilization Tray is intended to be used to enclose and protect surgical instruments optionally used for the UroLift System procedure for storage and sterilization in a pre-vacuum steam sterilizer. The following instruments are intended to be loaded into the UroLift System Rigid Retrieval Kit Sterilization Tray:
- UL-SCOPE4-0 / UL-SCOPE4-0-FE Cystoscope, 4.0 mm diameter x 302 mm . length, 0° angle of view, wide angle
- UL-GRASP-R Optical Graspers, double action jaws ●
Image /page/3/Picture/16 description: The image shows the logos for "NeoTract" and "Teleflex" side by side, separated by a vertical line. Below the logos, the text "INTERVENTIONAL UROLOGY" is displayed in a smaller font size. The logos and text are likely associated with a medical company or division specializing in interventional urology.
4
- UL-SCI-R Optical Scissors, double action jaws .
- Factory Exchange: Devices with a part number ending in "FE" have been repaired to meet product specifications.
The tray by itself is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, 510(k) cleared sterilization wrap to maintain sterility of the enclosed instruments when sterilized using the following prevacuum steam sterilization cycles:
Pre-Vacuum Steam | |
---|---|
Temperature | 132°C (270°F) |
Sterilization Time | 4 minutes |
Minimum Dry Time | 30 minutes |
Maximum Weight | 2.27 kg / 5.0 lbs |
CONTRAINDICATIONS
There are no known contraindications.
PREDICATE DEVICE
The predicate device is the UroLift System Procedure Kit Sterilization Tray, NeoTract / Teleflex (K192781).
Trade Name: | UroLift System Procedure Kit Sterilization Tray (UL-PKTRAY) |
---|---|
Common Name: | Sterilization Wrap Containers, Trays, Cassettes & Other |
Accessories | |
Product Code: | KCT |
Regulation Number: | 880.6850 |
Classification: | II |
Classification Panel: | Surgical and Infection Control Devices (OHT4) |
Infection Control and Plastic and Reconstructive Surgery (DHT4B) |
COMPARISON WITH THE PREDICATE DEVICE
The legally-marketed UroLift System Procedure Kit Sterilization Tray cleared via K192781 (14 April 2020) is the predicate for the UroLift System Rigid Retrieval Kit Sterilization Tray since both are metal sterilization trays with similar indications for use. Also, the predicate device and the proposed sterilization tray can undergo pre-vacuum steam sterilization with similar cycles. The primary differences are the size of the tray and the validated dry time. The differences do not affect the safety and effectiveness of the device as demonstrated through the cleaning and sterilization validation.
Technological Comparison Table:
| Characteristic | Subject Device
UroLift® System Rigid Retrieval Kit
Sterilization Tray [K212396] | Predicate Device
UroLift System Procedure Kit Sterilization
Tray [K192781] | Comparison |
|------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------|
| Device Name | UroLift® System Rigid Retrieval Kit
Sterilization Tray | UroLift System Procedure Kit Sterilization
Tray | N/A |
| 510(k)
Number | K212396 | K192781 | N/A |
Image /page/4/Picture/13 description: The image features the logos of NeoTract and Teleflex, two companies in the interventional urology field. The NeoTract logo is on the left, and the Teleflex logo is on the right. Below the logos, the words "INTERVENTIONAL UROLOGY" are written in smaller letters.
5
Product Code | KCT | KCT | Same | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Product Class | Class II | Class II | Same | |||||||||||
Regulation | ||||||||||||||
Number | 21 CFR 880.6850 | 21 CFR 880.6850 | Same | |||||||||||
Indications for | ||||||||||||||
Use | The UroLift System Rigid Retrieval Kit | |||||||||||||
Sterilization Tray is intended to be used to | ||||||||||||||
enclose and protect surgical instruments | ||||||||||||||
optionally used for the UroLift System | ||||||||||||||
procedure for storage and sterilization in a | ||||||||||||||
prevacuum steam sterilizer. The following | ||||||||||||||
instruments are intended to be loaded into | ||||||||||||||
the UroLift System Rigid Retrieval Kit | ||||||||||||||
Sterilization Tray: |
- UL-SCOPE4-0 / UL-SCOPE4-0-FE* -
Cystoscope, 4.0 mm diameter x 302 mm
length, 0° angle of view, wide angle - UL-GRASP-R - Optical Graspers, double
action jaws - UL-SCI-R - Optical Scissors, double
action jaws
- Factory Exchange: Devices with a part
number ending in "FE” have been repaired
to meet product specifications. | The UroLift System Procedure Kit
Sterilization Tray is intended to be used to
enclose and protect surgical instruments
used for the UroLift System Procedure for
storage and sterilization in a pre-vacuum
steam sterilizer.
The following instruments are intended to
be loaded into the UroLift System
Procedure Kit Sterilization Tray:
UL-SCOPE / UL-SCOPE-FE - Cystoscope,
2.9 mm diameter x 365 mm length, 0° angle
of view, 85° field of view
UL-SHEATH / UL-SHEATH-FE – Sheath,
20 Fr., with 2 tube connectors (Luer lock
and Luer lock with stopcock)
UL-VO / UL-VO-FE – Visual obturator,
20 Fr.
The tray by itself is not intended to maintain
sterility; it is intended to be used in
conjunction with a legally marketed,
validated, 510(k) cleared sterilization wrap
to maintain sterility of the enclosed
instruments when sterilized using the
following pre-vacuum steam sterilization
cycles: | Different | | | | | | | | | | | |
| | The tray by itself is not intended to maintain
sterility; it is intended to be used in
conjunction with a legally marketed,
validated, 510(k) cleared sterilization wrap
to maintain sterility of the enclosed
instruments when sterilized using the
following pre-vacuum steam sterilization
cycles:
Pre-vacuum Steam
Temperature: 132°C (270°F)
Sterilization time: 4 minutes
Minimum dry time: 30 minutes
Maximum weight: 5 lbs (2.7 kg) | Pre-Vacuum Steam Temperature 132°C
(270°F) 134°C
(273°F) Sterilization Time 4 minutes 3 minutes Minimum Dry Time 30 minutes 20 minutes Maximum Weight 1.8 kg / 4.0 lbs | | | | | | | | | | | | |
| Reusability | Reusable | Reusable | Same | | | | | | | | | | | |
| Patient
Contact | No direct patient contact | No direct patient contact | Same | | | | | | | | | | | |
| Design | Rigid containment device consisting of a
base with lid which can be fastened by a
latching mechanism. The device is
perforated in order to enable reprocessing
of enclosed medical devices held in place
by silicone retainers. | Rigid containment device consisting of a
base with lid which can be fastened by a
latching mechanism. The device is
perforated in order to enable
reprocessing of enclosed medical
devices held in place by silicone
retainers. | Same | | | | | | | | | | | |
| Device Image | Image: Urolift Sterilization Tray | Image: Urolift Sterilization Tray | Similar | | | | | | | | | | | |
Image /page/5/Picture/1 description: The image features the logos of "neotract" and "Teleflex" side by side, separated by a vertical line. Below the logos, the words "INTERVENTIONAL UROLOGY" are written in smaller, capitalized letters. The logos and text are presented in a clean and professional manner.
6
| Materials of
Construction | Stainless steel, silicone, Radel | Stainless steel, silicone, Radel | Same |
---|---|---|---|
Dimensions | 5.3 (W) x 20.7 (L) x 1.75 (H) inches | 4.5 (W) x 20 (L) x 2.125 (H) inches | Similar |
Air | |||
Permeance | Yes | Yes | Same |
Material | |||
Biocompatibility | Materials are biocompatible. Tested per | ||
ISO 10993-1 and Use of International | |||
Standard ISO 10993-1, "Biological | |||
evaluation of medical devices - Part 1: | |||
Evaluation and testing within a risk | |||
management process" - Guidance for | |||
Industry and Food and Drug Administration | |||
Staff issued 16 June 2016 | Materials are biocompatible. Tested per | ||
ISO 10993-1 and Use of International | |||
Standard ISO 10993-1, "Biological | |||
evaluation of medical devices - Part 1: | |||
Evaluation and testing within a risk | |||
management process" – Guidance for | |||
Industry and Food and Drug Administration | |||
Staff issued 16 June 2016 | Same | ||
Sterilization | |||
Method | Pre-vacuum Steam Sterilization | Pre-vacuum Steam Sterilization | Same |
Sterilization | |||
Parameters | 132 °C Pre-vacuum Steam Cycle | ||
Exposure temperature: 270 °F (132 °C) | |||
Exposure time: 4 minutes | |||
Vacuum dry time: 30 minutes | 132 °C Pre-vacuum Steam Cycle | ||
Exposure temperature: 270 °F (132 °C) | |||
Exposure time: 4 minutes | |||
Vacuum dry time: 30 minutes | |||
134 °C Pre-vacuum Steam Cycle | |||
Exposure temperature: 273 °F (134 °C) | |||
Exposure time: 3 minutes Vacuum dry time: | |||
20 minutes | Similar |
SUMMARY OF NON-CLINICAL TESTING
Title of test | Purpose of test | Acceptance Criteria / Source of references | Results |
---|---|---|---|
Usability Testing | Usability testing assured that the user could perform the steps of the Instructions for Use to meet the intended use of the sterilization tray and assured that the sterilization tray performed as intended to securely store and sterilize the instruments | The user is able to load and unload the tray, perform the reprocessing workflow (Clean the tray, inspect the tray, clean and inspect the instruments, and load the tray and sterilize), and determine the end of the serviceable life of the tray. | |
ANSI/AAMI ST77:2013 – Containment devices for reusable medical device sterilization | Pass | ||
Cytotoxicity Testing | The purpose of this study is to evaluate the cytotoxicity of a test article extract using an in vitro mammalian cell culture test. | The test sample meets the requirements of the test if the biological response is less than or equal to grade 2 (mild). | |
Cytotoxicity testing per ISO 10993-5:2009 – Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity | Pass | ||
Sensitization Testing | The purpose of this study is to evaluate the potential of the test | Grades of 1 or greater observed in the test group generally indicate | Pass |
article to cause delayed dermal | |||
contact sensitization in the guinea | |||
pig maximization test. | sensitization, provided that grades | ||
of less than 1 are observed on the | |||
control animals. If grades of 1 or | |||
greater are noted on control | |||
animals, then the reactions of test | |||
animals that exceeded the most | |||
severe control reaction will be | |||
considered to be due to | |||
sensitization. |
Sensitization testing per ISO
10993-10:2010 – Biological
evaluation of medical devices Part
10: Tests for irritation and skin
sensitization | |
| Intracutaneous
Reactivity
Testing | The purpose of this study is to
evaluate the local dermal irritation
of a test article extract following
intracutaneous injection in rabbits. | The difference of overall mean of
the test group to the control group
on erythema and edema score
must be less than 1.
Intracutaneous Reactivity testing
per ISO 10993-10:2010 – Biological
evaluation of medical devices Part
10: Tests for irritation and skin
sensitization | Pass |
| Useful Life
Testing | The tray was subject to 100
cleaning and sterilization cycles per
the Instructions for Use
reprocessing directions. The
purpose of the test was to
determine the serviceable life span
of the tray. | The tray must pass the inspection
criteria on the IFU after 100
reprocessing cycles.
ANSI/AAMI ST77:2013 –
Containment devices for reusable
medical device sterilization | Pass |
| Cleaning
Validation | The purpose of this test is to
validate that the cleaning
instructions listed in the Instructions
for Use appropriately clean the tray
to ensure the sterilization cycle will
be effective. | Per the protocol, there were three
acceptance criteria; protein residual
analysis, hemoglobin residual
analysis, and visual inspection. The
protein residual benchmark level
was 6.4µg/cm² and the hemoglobin
benchmark level was 2.2µg/cm².
The tray must pass the visual
inspection in the IFU.
AAMI TIR30: 2011/(R)2016, A
compendium of processes,
materials, test methods, and
acceptance criteria for cleaning
reusable medical devices revised
15 December 2016 and
Reprocessing Medical Devices in
Health Care Settings: Validation
Methods and Labeling Guidance for
Industry and Food and Drug | Pass |
| | | Administration Staff issued 17
March 2015 | |
| Sterilization
Validation | The purpose of this test is to
validate that the sterilization
instructions listed in the Instructions
for use appropriately sterilize the
tray and contents. | The minimum sterility assurance
level (SAL) of 10-6 can be achieved if
the sterilization instructions in the
IFU were followed.
ANSI/AAMI ST77:2013,
Containment devices for reusable
medical device sterilization | Pass |
Image /page/6/Picture/3 description: The image features the logos of "neotract" and "Teleflex" side by side, separated by a vertical line. The "neotract" logo is in a sans-serif font with the "neo" part in black and the "tract" part in blue, underlined by a black line. The "Teleflex" logo is in a bold, sans-serif font, also in blue, with a registered trademark symbol. Below both logos, in smaller, all-caps letters, is the text "INTERVENTIONAL UROLOGY".
7
nectract | ገfeleflex® INTERVENTIONAL UROLOGY
8
PERFORMANCE TESTING
The design requirements for the UroLift System Rigid Retrieval Kit Sterilization Tray were reviewed and non-clinical design verification testing was required to assure that the device met the intended use. Non-clinical testing included usability testing, and cleaning / sterilization testing. The usability testing assured that the user could perform the steps of the Instructions for Use to meet the intended use of the sterilization tray and assured that the sterilization tray performed as intended to securely store and sterilize the instruments.
BIOCOMPATIBILITY TESTING
The biocompatibility of the UroLift System Rigid Retrieval Kit Sterilization tray is equivalent to the predicate. Both trays are manufactured using identical materials and equivalent processing methods. The type and duration of contact of the proposed and predicate tray are identical, and the instruments contained in both trays are identical duration and contact type. As such, the biocompatibility of the predicate is considered representative of the proposed tray and no additional biocompatibility testing was required.
The predicate device (UroLift System Procedure Kit Sterilization Tray) has been tested for biocompatibility and passed the relevant tests according to ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" – Guidance for Industry and Food and Drug Administration Staff issued September 2020. Biocompatibility testing was performed on worst case sterilized devices and included:
- Cytotoxicity testing per ISO 10993- 5:2009 - Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- Sensitization testing per ISO 10993- 10:2010 Biological evaluation of medical ● devices Part 10: Tests for irritation and skin sensitization
- Intracutaneous Reactivity testing per ISO 10993- 10:2010 - Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
CLEANING / STERILIZATION TESTING
The UroLift System Rigid Retrieval Kit Sterilization Tray is intended to be sold non-sterile and used during the reprocessing of the surgical instruments that are optionally used in the UroLift System Procedure. As such, cleaning and sterilizations were performed to ensure the tray could withstand multiple reprocessing cycles without adverse reaction or
Image /page/8/Picture/11 description: The image features the logos of "neotract" and "Teleflex" side by side, separated by a vertical line. The "neotract" logo is in black and blue, with a horizontal line underneath. The "Teleflex" logo is in blue. Below the logos, the words "INTERVENTIONAL UROLOGY" are written in a smaller, gray font.
9
degradation.
The cleaning validation was designed to simulate the worst case scenario conforming to several industry standards includinq AAMI TIR30: 2011/(R)2016, A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices revised 15 December 2016 and Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff issued 17 March 2015. The cleanliness of the tray was assessed by protein and hemoglobin residuals, and all residuals met predefined acceptance criteria demonstrating the adequate cleanliness of the tray using the cleaning instructions.
The sterilization validation, designed based on the FDA-recognized standard ANSI/AAMI ST77:2013, Containment devices for reusable medical device sterilization, demonstrated the minimum sterility assurance level (SAL) of 10-6 can be achieved if the sterilization instructions in the instructions for use (IFU) were followed.
The reuse life of the UroLift System Rigid Retrieval Kit Sterilization tray is equivalent to the predicate. Both trays are manufactured using identical materials and equivalent processing methods, and subject to equivalent cleaning and sterilization cycles. As such, the reuse life of the predicate (100 cycles) is considered adequate for the proposed tray and no additional reuse life testing was required. Inspection criteria are the same as the predicate and described in the instructions for use (IFU).
CONCLUSION
The conclusions drawn from the non-clinical tests demonstrate the UroLift System Rigid Retrieval Kit Sterilization Tray is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Image /page/9/Picture/6 description: The image features the logos of two companies, "NeoTract" and "Teleflex", side by side. Below the logos, the words "Interventional Urology" are written in smaller font. The logos and text are all in a professional, corporate style.