K Number
K212318
Device Name
IPL Hair Removal
Date Cleared
2022-01-14

(172 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IPL HAIR REMOVAL (Model: SG-8025) is an over the counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair.

Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6. 9. and 12 months after the completion of a treatment regimen.

Device Description

The IPL HAIR REMOVAL SG-8025 is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and it's IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device is for single-person use only.

AI/ML Overview

The provided document is an FDA 510(k) summary for an IPL (Intense Pulsed Light) Hair Removal device (Model SG-8025). This type of document is for regulatory clearance and focuses on demonstrating substantial equivalence to a predicate device, rather than proving a new device's efficacy through extensive clinical trials.

Therefore, the document explicitly states: "No clinical tests was performed on the subject device." This means there is no study described within this document that proves the device meets specific acceptance criteria based on its own clinical performance. Instead, the device's equivalence to a legally marketed predicate device (IPL Salon Hair Reduction System, K181568) is asserted through a comparison of technical and functional characteristics.

Because no clinical study was performed for the IPL HAIR REMOVAL SG-8025 to establish its own performance against acceptance criteria, much of the requested information cannot be provided from this document.

However, I can extract the comparative information and the non-clinical performance data provided to support the substantial equivalence claim.


1. A table of acceptance criteria and the reported device performance

Since no clinical study was performed for the subject device (IPL HAIR REMOVAL SG-8025) directly against acceptance criteria, this table will instead present the comparison of technological characteristics between the subject device and its predicate device, which is the basis of its FDA clearance for substantial equivalence. "Acceptance criteria" in this context refers to the FDA's determination that the differences between the subject device and the predicate device do not raise new questions of safety or effectiveness.

Comparison of Technological Characteristics: IPL HAIR REMOVAL SG-8025 vs. Predicate Device (K181568)

FeaturePredicate Device (K181568)Subject Device (IPL HAIR REMOVAL SG-8025)Comparison & Conclusion (from document)
Intended UseThe IPL Salon Hair Reduction System (Model: F60001) is an over the Counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regimen.The IPL Hair Removal (Model: SG-8025) is an over the Counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regimen.Identical. The document states: "The IPL Home Use Hair Removal Device has the same intended use... as the predicate devices."
Device TypeIntense Pulsed LightIntense Pulsed LightIdentical. The document states: "...mode of action and similar operational characteristics..." -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Delivery DeviceDirect illumination to tissueDirect illumination to tissueIdentical.
Power SourceSupplied by external adapterSupplied by external adapterIdentical.
Pulse ControlFinger switchFinger switchIdentical.
No. Output Intensity Levels5 level5 levelIdentical.
Wavelength range475-1200 nm530 nmDifference, but deemed acceptable: "Although the wavelength of subject device is a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-2-57 requirement. And the wavelength of subject device is in the range of the one of predicted device. So the differences of function specification will not raise any safety or effectiveness issue."
Spot sizeRegular window: 4.5cm², Facial adapter: 1.72cm²3 cm²Difference, but deemed acceptable: "There is minor difference in Spot size between the subject device and the predicate devices. And they all comply with IEC 60601-1, IEC60601-2-57 requirement. So the differences of Spot size will not raise any safety or effectiveness issue."
Pulse duration11-12 ms1msDifference, but deemed acceptable: "Although the Pulse Duration of subject device is less than the predicate device, the subject device comply with all electrical safety eye safety and EMC Standard, So the differences of pulse duration time will not raise any safety or effectiveness issue."
Max Energy densityUp to 5 J/cm²2.5 J/cm²Difference, but deemed acceptable: "Although the Max. Fluence of subject device is a little less than the the Predicate device, but they all comply with IEC 60601-1, IEC 60601-2-57 requirement. So the differences of function specification will not raise any safety or effectiveness issue."
Output energy (facial adapter)Level 1: 8.62J, Level 2: 9.45J, Level 3: 10.64J, Level 4: 11.48J, Level 5: 12.70JLevel 1: 7.5J, Level 2: 8.5J, Level 3: 9.5J, Level 4: 11J, Level 5: 12JDifference, but deemed acceptable. The document explicitly states the Max Energy Density difference is acceptable and the overall conclusion is that "Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness."
Weight650g220gDifference, but deemed acceptable: "Although it is a little different from the predicate device, it will not affect the main function and the intended use of the device."
Dimensions14369.543mm(HWD)2057656mm(HWD)Difference, but deemed acceptable: "Although it is a little different from the predicate device, it will not affect the main function and the intended use of the device." -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Software/Firmware/Microprocessor Control?YesYesIdentical.
60601 ComplianceYes (IEC 60601-1, IEC 60601-1-2, IEC60601-2-57)Yes (IEC 60601-1, IEC 60601-1-2, IEC60601-2-57)Identical compliance. The subject device also complies with IEC 60601-1-11 for home healthcare environment.
BiocompatibilityAll user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements.All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements.Identical compliance.
Electrical SafetyComply with IEC60601-1 and IEC60601-2-57Comply with IEC60601-1 and IEC60601-2-57Identical compliance, plus additional home healthcare standard.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size for test set: Not applicable. No clinical tests were performed on the subject device.
  • Data provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No clinical tests were performed to establish the device's efficacy. The substantial equivalence relies on technical comparison and compliance with recognized standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is an IPL hair removal device, not an AI diagnostic tool involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable for the subject device's efficacy. The "ground truth" for the clearance is the established safety and effectiveness profile of the predicate device, as well as compliance with various international standards for medical electrical equipment, photobiological safety, and biocompatibility.

8. The sample size for the training set

  • Not applicable. This is not an AI/machine learning device.

9. How the ground truth for the training set was established

  • Not applicable. This is not an AI/machine learning device.

Summary of Device Performance (Non-Clinical):

The document provides non-clinical performance data to support the safety aspects of the device, which contributes to the claim of substantial equivalence:

  • Biocompatibility Testing:
    • Conducted per ISO 10993-1.
    • Passed ISO 10993-5:2009/(R)2014 (Tests for In Vitro Cytotoxicity).
    • Passed ISO 10993-10:2010/(R)2014 (Tests for Irritation and Skin Sensitization).
  • Electrical Safety:
    • Passed IEC 60601-1-2 (Electromagnetic compatibility).
    • Passed IEC 60601-1 (General requirements for basic safety and essential performance).
    • Passed IEC 60601-1-11 (Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment).
    • Passed IEC 60601-2-57 (Particular requirements for non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use).
  • Eye Safety:
    • Compliance with IEC 62471 Photobiological safety of lamps and lamp systems.
    • Compliance with ISO 14971 Medical Devices Applications of Risk Management to Medical Devices.
  • Software Verification and Validation:
    • Software documentation consistent with a moderate level of concern was submitted.
    • System validation testing demonstrated all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

Conclusion from the document:

"IPL HAIR REMOVAL SG-8025 was found to be substantially equivalent to the predicate devices and shares the same or similar indications for use, design, operational and functional features as the predicate devices. Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 14, 2022

Dongguan Define Beauty Electronic Technology Co. Ltd % Helen Nan General Manager Cytech (Shenzhen) Enterprise Management Consulting Co.,Ltd. Room B204, Building 12 Hourui 2nd Industrial Zone, Bao'an District Shezhen, Guangdong 518128 China

Re: K212318

Trade/Device Name: IPL Hair Removal SG-8025 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: December 7, 2021 Received: December 14, 2021

Dear Helen Nan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212318

Device Name IPL HAIR REMOVAL SG-8025

Indications for Use (Describe)

The IPL HAIR REMOVAL (Model: SG-8025) is an over the counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair.

Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6. 9. and 12 months after the completion of a treatment regimen.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Research Use Only (21 CFR 820.30) Subject to Design Controls For In-Vitro Use Only (21 CFR 820.30) Subject to Design Controls
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a logo with the words "Define beauty" written in a cursive font. Above the English text, there are three Chinese characters in a stylized, calligraphic style. The overall design has a flowing, elegant feel, with the text and characters interconnected by curved lines.

Ongguan Define Beauty Electronic Technology Co. Ltd

510(k) Summary K212318 (As required by 21 CFR 807.92(a))

1.0 Submitter Information

Company:Dongguan Define Beauty Electronic TechnologyCo. Ltd.
AddressNo.10 Xiangya Road, Jichiling Village,Dalingshan Town, Dongguan, Guangdong,523812, CHINA
Phone:0086-0769-89367468
ContactShaowu Wang
Title:General Manager
• Date: January 13, 2022

2.0 Device Information

Trade/Device Name:IPL HAIR REMOVAL SG-8025
Model:SG-8025
Common Name:IPL Hair Removal Device
Device:Light Based Over-The-Counter Hair Removal
Definition:Over-the-counter device uses thermal energy tokill hair follicles for hair removal.
Review Panel:General & Plastic Surgery
Product Code:OHT
Submission Type:Traditional 510(k)
Regulation Number:CFR 878.4810
Device Class:Class II

3.0 Predicate Device Information

Trade/Device Name:IPL Salon Hair Reduction System
510k Number:K181568
Submitter:Medical Device Branch of Zhangzhou Easepal Industrial Co.,Ltd.

4.0 Device Description

The IPL HAIR REMOVAL SG-8025 is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and it's IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device is for single-person use only.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains a logo for a beauty company. The logo is pink and features the words "Define beauty" in a cursive font. Above the text is a Chinese character in a similar pink color. The logo is simple and elegant, and it conveys a sense of beauty and sophistication.

东莞市定义美电子科技有限公 Dongguan Define Beauty Electronic Technology Co. Ltd

5.0 Indications for Use

The IPL HAIR REMOVAL (Model: SG-8025) is an over the counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6. 9. and 12 months after the completion of a treatment regimen.

6.0 Comparison of Technological Characteristics with the Predicate Device
Table
Device FeaturePredicate DeviceSubject Device
Trade NameIPL Salon Hair Reduction SystemIPL HAIR REMOVALSG-8025
510k numberK181568K212318
Classification NameLight Based Over-The-Counter Hair Removal
Regulation NumberCFR 878.4810
Product codeOHT
UseOTC use
Indication for useThe IPL Salon Hair Reduction System (Model: F60001) is an over the Counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6. 9. and 12 months after the completion of a treatment regimen.The IPL Hair Removal (Model: SG-8025) is an over the Counter device intended for the removal of unwanted body and/or facial hair in adults. It is also intended for permanent reduction in unwanted hair. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-growing when measured at 6. 9. and 12 months after the completion of a treatment regimen.
Device TypeIntense Pulsed LightIntense Pulsed Light
Delivery deviceDirect illumination to tissueDirect illumination to tissue
Device Design
Source EnergySupplied by external adapterSupplied by external adapter
Wavelength range475-1200 nm530 nm
Spot sizeRegular window: 4.5cm²Facial adapter: 1.72cm²3 cm²
Pulse duration11-12 ms1ms
Pulse controlFinger switchFinger switch
Weight650g220g
Output IntensityLevel5 level5 level
Max EnergydensityUp to 5 J/cm²2.5 J/cm²
Output energywith facialadapterLevel 1: 8.62JLevel 2: 9.45JLevel 3: 10.64JLevel 4: 11.48JLevel 5: 12.70JLevel 1: 7.5JLevel 2: 8.5JLevel 3: 9.5JLevel 4: 11JLevel 5: 12J
Dimensions14369.543mm(HWD)2057656mm(HWD)
Software/Firmware/MicroprocessorControl?YesYes
Standards
60601 Compliancewith VoluntaryStandardsYesComply with IEC 60601-1and IEC 60601-1-2,IEC60601-2-57YesComply with IEC 60601-1and IEC 60601-1-2,IEC60601-2-57
BiocompatibilityAll user directly contactingmaterials are compliancewith ISO10993-5 andISO10993-10 requirements.All user directly contactingmaterials are compliancewith ISO10993-5 andISO10993-10 requirements.
Electrical SafetyComply with IEC60601-1and IEC60601-2-57Comply with IEC60601-1and IEC60601-2-57

Table 1 - Device Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo with the words "Define beauty" written in a cursive font. Above the English text, there are three Chinese characters in a stylized, calligraphic style. The logo is primarily pink or red, with the text and characters sharing the same color scheme. A decorative flourish underlines the text, adding to the overall aesthetic appeal of the logo.

市定义美甲子科技有限公

Dongguan Define Beauty Electronic Technology Co. Ltd

A brief summary of the similarities and differences between IPL Salon Hair Reduction System and IPL HAIR REMOVAL SG-8025 is included below:

Similarities:

The IPL Home Use Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the IPL HAIR REMOVAL may be found substantially equivalent to its predicate device.

Differences:

Basic Characteristics

"Weight", "Dimensions" is belong to basic characteristics. Although it is a little different from the predicate device, it will not affect the main function and the intended use of the device. They all also comply with IEC 60601-1 requirements. So the differences will not raise any safety or effectiveness issue.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains a logo with the words "Define beauty" written in a cursive font. Above the English text, there are three Chinese characters in a stylized, calligraphic style. The logo is primarily pink or red in color, with the Chinese characters appearing bolder and more prominent than the English text. The overall design has a feminine and elegant aesthetic.

ongguan Define Beauty Electronic Technology Co. Ltd

Max Energy Density

Although the Max. Fluence of subject device is a little less than the the Predicate device, but they all comply with IEC 60601-1, IEC 60601-2-57 requirement. So the differences of function specification will not raise any safety or effectiveness issue.

Pulse Duration

Although the Pulse Duration of subject device is less than the predicate device, the subject device comply with all electrical safety eye safety and EMC Standard, So the differences of pulse duration time will not raise any safety or effectiveness issue.

Wavelength

Although the wavelength of subject device is a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-2-57 requirement. And the wavelength of subject device is in the range of the one of predicted device. So the differences of function specification will not raise any safety or effectiveness issue.

Spot Size

There is minor difference in Spot size between the subject device and the predicate devices. And they all comply with IEC 60601-1, IEC60601-2-57 requirement. So the differences of Spot size will not raise any safety or effectiveness issue.

7.0 Non- Clinical Performance Data

The following performance data were provided in support of the substantial equivalence determination.

7.1 Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the IPL Home Use Hair Remover was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on June 16, 2016", as recognized by FDA. The battery of testing was performed to, and passed, including:

  • A ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices -Part 5: Tests for In Vitro Cytotoxicity
  • A ISO 10993-10:2010/(R)2014, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization

7.2 Electrical Safety

  • A Electrical safety and Eye safety testing was performed to, and passed, the following standards:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo with the words "Define beauty" written in a cursive font. Above the words, there are three Chinese characters in a stylized font. The logo is pink and has a decorative flourish above and below the text.

义美电子科技有限公

Dongguan Define Beauty Electronic Technology Co. Ltd

  • 60601-1-2 Medical electrical equipment -Part 1-2: General IEC requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility

  • IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance

  • A IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
  • A IEC 60601-2-57 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

7.3 Eve Safety

  • IEC 62471 Photobiological safety of lamps and lamp systems A
  • ISO 14971 Medical Devices Applications of Risk Management to A Medical Devices

7.4 Software Verification and Validation

  • Software documentation consistent with moderate level of concern was A submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
    Based on the above performance as documented in this application, IPL HAIR REMOVAL SG-8025 was found to have a safety and effectiveness profile that is similar to the predicate devices.

8.0 Clinical Performance Data

No clinical tests was performed on the subject device.

9.0 Conclusion:

IPL HAIR REMOVAL SG-8025 was found to be substantially equivalent to the predicate devices and shares the same or similar indications for use, design, operational and functional features as the predicate devices. Any difference in the technological characteristics does not raise any new issues or concerns of safety or effectiveness.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.