K Number
K212312

Validate with FDA (Live)

Date Cleared
2022-01-27

(188 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a digital monitor for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 inches to 7 1/4 inches (12.8 cm to 18.5 cm).

Device Description

RA-T59 Wrist blood pressure monitor is designed to measure the blood pressure and pulse rate of adults within the range of the designated wrist cuff (12.8 to 18.5 cm). RA-T59 Wrist blood pressure monitor is based on an oscillometric method to measure blood pressure and pulse rate from wrist. The device uses automatic inflation mode. It starts to inflate from 0 mmHg and automatically stops the motor after 25 seconds and quickly deflates, with a steady inflation speed detecting the measurement during inflation. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The signal from the electronic pressure sensor is conditioned with a circuit before data conversion by an analog-to-digital converter (ADC). The systolic pressure, diastolic pressure, and pulse rate are then calculated in the digital domain. The resulting systolic, diastolic, and pulse-rate measurements are displayed on a LCD and stored in the device's memory.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the RA-T59 Wrist blood pressure monitor, based on the provided document:

Acceptance Criteria and Reported Device Performance

The device's performance for blood pressure measurement is evaluated against the requirements of ISO 81060-2:2018. The document does not explicitly present a table of "acceptance criteria" and "reported device performance" in a quantitative manner as one might find in a detailed study report. However, it states that "The results of this clinical investigation show that the proposed device fulfills the requirement of ISO 81060-2:2018."

The key performance criterion for blood pressure monitors, based on the Accuracy or pressure indicator and Accuracy Pulse Rate sections in the comparison table (page 4), is:

Acceptance CriteriaReported Device Performance
Accuracy (Pressure): ±3 mmHg or 2% of readingMeets: The device was validated to ISO 81060-2:2018, which includes these accuracy requirements.
Accuracy (Pulse Rate): ±5%Meets: The device was validated to ISO 81060-2:2018, which includes these accuracy requirements.

Note: While the document confirms compliance with ISO 81060-2:2018, specific numerical results (e.g., mean difference and standard deviation between device and reference measurements) that demonstrate this compliance are not provided within this summary for either systolic or diastolic blood pressure, or pulse rate.


Study Information

2. Sample size used for the test set and the data provenance:

  • Sample Size: 85 qualified healthy adult subjects.
  • Data Provenance: Not explicitly stated regarding country of origin. The study is described as a "clinical investigation," which generally implies a prospective study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. For blood pressure clinical validation studies like ISO 81060-2, ground truth is typically established by trained and certified observers using a validated reference sphygmomanometer, not usually by "experts" in the context of interpretation (like radiologists). The standard specifies requirements for the observers.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not provided in the document. For blood pressure measurements in clinical validation, ground truth is typically established by two (or more) independent observers. The standard ISO 81060-2:2018 outlines the methodology for taking concurrent auscultatory measurements by these observers.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, an MRMC comparative effectiveness study was not done. This device is an automated blood pressure monitor and does not involve human readers interpreting results or AI assistance in that context. The study performed was a clinical validation of the device's measurement accuracy.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance study was done. The clinical investigation directly assessed the RA-T59 Wrist blood pressure monitor's (algorithm and hardware) ability to accurately measure blood pressure and pulse rate against a reference standard in a clinical setting, without human interpretation of the device's output.

7. The type of ground truth used:

The type of ground truth used is implied by the adherence to ISO 81060-2:2018. This standard typically establishes ground truth through concurrent auscultatory measurements performed by trained observers using a reference sphygmomanometer. This is a form of expert-derived, objective measurement used as the "truth" against which the automated device's readings are compared.

8. The sample size for the training set:

The document does not mention a training set sample size. This is a clinical validation document for a medical device that measures physiological parameters, not a machine learning or AI algorithm in the typical sense that would have an explicit "training set" for model development. The "clinical investigation" mentioned is for validation, not training.

9. How the ground truth for the training set was established:

As no training set is mentioned in the context of this device, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

January 27, 2022

LotusNine Medical Limited % Yuhua Chen Manager PuHsu Consulting Ltd. 7F., No.272, Jiankang Rd., Zhonghe Dist. New Taipei City, 235042 Taiwan

Re: K212312/S001

Trade/Device Name: RA-T59 Wrist blood pressure monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: December 17, 2021 Received: December 29, 2021

Dear Yuhua Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K212312

Device Name

RA-T59 Wrist blood pressure monitor

Indications for Use (Describe)

The device is a digital monitor for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 inches to 7 1/4 inches (12.8 cm to 18.5 cm).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

Date prepared: July 15, 2021

1. Submitter's Information

Submission Submitter

Company NameLotusNine Medical Limited
AddressNo. 2, Ln. 223, Sec. 1, Taiping Rd., TsaotunTownship, Nantou County 542007, Taiwan
Contact PersonYating Chang
Phone886-966512879
Emailytc@lotus9.xyz

Submission Correspondent

Company NamePuHsu Consulting Ltd.
Address7F., No.272, Jiankang Rd., Zhonghe Dist., NewTaipei County 235042, Taiwan
Contact PersonYuhua Chen
Phone886-965650265
Emailyuhua@puhsuconsult.com

2. Subject Device Information

Proprietary/Trade NameRA-T59 Wrist blood pressure monitor
Regulation NameNoninvasive blood pressure measurement system
Regulation Number870.1130
Product CodeDXN
Device ClassificationII
PanelCardiovascular

3. Predicate Device Information

510(k)NumberTrade NameManufacturerType
K131742Model HEM-6131Omron Healthcare, Inc.PredicateDevice

{4}------------------------------------------------

4. Device Description

RA-T59 Wrist blood pressure monitor is designed to measure the blood pressure and pulse rate of adults within the range of the designated wrist cuff (12.8 to 18.5 cm). RA-T59 Wrist blood pressure monitor is based on an oscillometric method to measure blood pressure and pulse rate from wrist. The device uses automatic inflation mode. It starts to inflate from 0 mmHg and automatically stops the motor after 25 seconds and quickly deflates, with a steady inflation speed detecting the measurement during inflation. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The signal from the electronic pressure sensor is conditioned with a circuit before data conversion by an analog-to-digital converter (ADC). The systolic pressure, diastolic pressure, and pulse rate are then calculated in the digital domain. The resulting systolic, diastolic, and pulse-rate measurements are displayed on a LCD and stored in the device's memory.

5. Indications for Use

The device is a digital monitor for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 inches to 7 1/4 inches (12.8 cm to 18.5 cm).

Proposed DevicePredicate DeviceRemark
DeviceRA-T59 Wrist blood pressure monitorModel HEM-6131
ManufacturerLotusNine Medical LimitedOmron Healthcare, Inc.--
510(k) NumberN/AK131742--
Product CodeDXNDXNSame
RegulationNumber870.1130870.1130Same
Indications forUseThe device is a digital monitorfor use in measuring bloodpressure and pulse rate inadult patient population withwrist circumference rangingfrom 5 inches to 7 1/4 inches(12.8 cm to 18.5 cm).The device is a digital monitorfor use in measuring bloodpressure and pulse rate inadult patient population withwrist circumference rangingfrom 5 1/4 inches to 8 1/2inches (13.5 cm to 21.5 cm).Different(Note 1)
Proposed DevicePredicate DeviceRemark
DeviceRA-T59 Wrist blood pressuremonitorModel HEM-6131
PatientPopulationAdultAdultSame
PrescriptiveOTCOTCSame
Anatomical SiteWristWristSame
Environment ofUseHomeHomeSame
MeasuringMethodOscillometricOscillometricSame
MeasurementRangeSystolic blood pressure: 60 -230 mmHgDiastolic blood pressure: 40 -130 mmHgRated range of CUFFpressure: 300 mmHgPulse: 40 to 180 beatsPressure: 0-299 mmHgPulse rate: 40 to 80 bpmDifferent(Note 2)
Accuracy orpressureindicator±3 mmHg or 2% of reading±3 mmHg or 2% of readingSame
Accuracy PulseRate±5%±5%Same
Inflation andDeflationAutomaticAutomaticSame
Display TypeLCDLCDSame
Power SourceUSB type C chargingAAA batteriesDifferent(Note 3)
OperationConditionsTemperature: 10°C to 40°CHumidity: 15 to 85% RHTemperature: 10°C to 40°CHumidity: 15 to 85% RHSame
StorageConditionsTemperature: -20°C to 70°CHumidity: 10 to 90% RHTemperature: -20°C to 60°CHumidity: 10 to 95% RHDifferent(Note 4)
Memory25 Measurements60 MeasurementsDifferent(Note 5)
DeviceProposed DevicePredicate DeviceRemark
RA-T59 Wrist blood pressuremonitorModel HEM-6131
Electrical safetyIEC 60601-1-2IEC60601-1IEC 60601-1-2IEC60601-1Same
BiocompatibilityISO 10993 seriesISO 10993 seriesSame

6. Comparison to the Predicate Device

{5}------------------------------------------------

{6}------------------------------------------------

LotusNine Medical Limited 510(k) Notification

K212312

Comparison

Note 1:

The substantial difference of the intended use is the wrist circumference ranging. The proposed device is narrower than the predicate device. The proposed device was validated according to ISO 80601-2-30 and ISO 81060-2. The performance data can demonstrate this difference does not raise different questions of safety and effectiveness.

Note 2:

The proposed device's systolic and diastolic blood pressure measurement range is restricted from 40 to 230 mmHg. The measuring rang of the proposed device was validated according to ISO 80601-2-30. The different does not raise different questions of safety and effectiveness.

Note 3:

Although the power source is different, IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2, and IEC 62133-2 can demonstrate that the proposed device can maintain the safety and performance when the proposed device is charged. Thus, this difference does not raise different questions of safety and effectiveness.

Note 4:

Compared with the predicate device, the proposed device requires different storage environment. IEC 60601-1-11 test report can demonstrate that the proposed device can maintain the safety and performance within the storage environment. Thus, this difference does not raise different questions of safety and effectiveness.

Note 5:

The memory size of the proposed device is less than the predicate device. The difference is very slight and it will not affect the main function and the intended use of the proposed

{7}------------------------------------------------

device. Therefore, this difference will not raise any safety or effectiveness issue.

7. Performance Testing

Performance testing has been carried out to demonstrate that the proposed device meets the performance specifications for its intend use.

The proposed device was tested to ISO 81060-2: 2018 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The study population consisted of 85 qualified healthy adult subjects. The results of this clinical investigation show that the proposed device fulfills the requirement of ISO 81060-2:2018.

8. Non-Clinical Testing

The subject device has been tested according to the following standards:

  • A. Safety Test
    • . IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
    • . IEC 60601-1-11, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
    • IEC 62133-2, Secondary cells and batteries containing alkaline or other nonacid electrolytes - Safety requirements for portable sealed secondary lithium cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
  • B. EMC Test
    • . IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • C. Biocompatibility testing
    • ISO- 10993, Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing
    • . ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
    • 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

{8}------------------------------------------------

  • . ISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
  • Performance Test D.
    • IEC 80601-2-30, Medical electrical equipment Particular requirements for the . basic safety and essential performance of automated non-invasive sphygmomanometers.
  • E. Software Verification and Validation
    • . FDA Guidance for the Content of Pre-Market Submission for Software Contained in Medical Devices
    • . IEC 62304, Medical device software - Software life cycle processes
  • F. Usability
    • IEC 60601-1-6, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
    • IEC 62366-1, Medical devices - Application of usability engineering to medical devices.

9. Conclusion

In conclusion, it shows that the proposed device and the predicate device has the same intended use, and the difference in technological features of the proposed devices and the predicate devices do not raise different questions of safety and effectiveness. The result of all testing conducted was found acceptable to support the claim of substantial equivalence.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).