(86 days)
Not Found
Yes
The device description explicitly states that the software algorithm is based on machine learning.
No
The device is a computer-aided triage and notification software that flags and communicates suspected findings; it does not directly treat or diagnose.
No
The text explicitly states: "The StrokeSENS LVO device... is not intended to be used as a diagnostic device." It clarifies that notifications are "for informational purposes only and are not intended for diagnostic use beyond notification." Instead, it is intended for "workflow triage by flagging and communication of suspected positive findings."
Yes
The device is described as a software system consisting of a processing engine and a compatible radiological software platform, both of which are software components. It analyzes image data and sends notifications, without mentioning any dedicated hardware components included with the device.
Based on the provided text, this device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs analyze biological samples: In Vitro Diagnostics are designed to examine specimens taken from the human body, such as blood, urine, tissue, etc., to provide information about a person's health.
- This device analyzes medical images: The StrokeSENS LVO device analyzes CTA head images, which are medical images, not biological samples.
- The intended use is image analysis for triage and notification: The stated intended use is to assist in workflow triage by flagging and communicating suspected findings in images. It explicitly states it is not intended for diagnostic use beyond notification and does not alter the original medical image and is not intended to be used as a diagnostic device.
Therefore, based on the definition and intended use described, StrokeSENS LVO falls under the category of medical imaging software, specifically a computer-aided triage and notification (CADt) device, rather than an In Vitro Diagnostic.
No
The letter does not explicitly state that the FDA has reviewed and approved or cleared a PCCP for this specific device.
Intended Use / Indications for Use
StrokeSENS LVO is a radiological computer-aided triage and notification (CADt) software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVO) in head CTA images.
StrokeSENS LVO uses a software algorithm to identify suspected LVO findings. In the case of a suspected LVO, the system will send a notification to a pre-configured destination(s), notifying the clinicians of the existence of a suspected LVO that requires review. The notification system is intended to be used in parallel to the standard of care workflow to notify clinicians of the existence of a potential LVO earlier than being notified as part of the standard of care workflow.
Notifications may include a compressed preview of images. Notifications are meant for informational purposes only and are not intended for diagnostic use beyond notification. The StrokeSENS LVO device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of StrokeSENS LVO are intended to be used in conjunction with other patient information and based on professional judgement, to assist with triage / prioritization of medical images. Notified clinicians are responsible for viewing full images per standard of care.
Product codes
QAS
Device Description
StrokeSENS LVO is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVO) in head CTA images. StrokeSENS LVO uses a software algorithm based on machine learning to identify suspected LVO findings. In the case of a suspected LVO, the system will send a notification to a preconfigured destination(s), notifying the clinicians of a suspected LVO that requires review.
After a CT Angiography (CTA) head scan is performed, the images are automatically routed to StrokeSENS LVO where they are processed and analyzed for characteristics suggestive of LVO. The notification system is comprised of an on-screen notice of a suspected LVO to the user and an outbound notification to a neurovascular specialist. The notification states that a suspected LVO is present, and that the user should review the images in a diagnostic radiological viewer. In the case of a negative finding, no notification is sent, and the standard of care workflow remains in place.
StrokeSENS LVO DICOM-compliant software system consists of two main components: 1) the StrokeSENS LVO Processing Engine and 2) a compatible Radiological Software Platform:
- The StrokeSENS LVO Processing Engine is responsible for receiving, processing, and analyzing image data and communicating results. Primarily, the Processing Engine consists of a software algorithm (a sequence of instructions/operations) that is responsible for analyzing contrast-enhanced CT (CTA) image data of the head to identify characteristics that are consistent with LVO. The software algorithm is a binary classifier, providing a binary output of either positive or negative for suspected LVO, based on a pre-defined threshold. The output is returned to the Radiological Software Platform for the purposes of triage and notification. The Processing Engine is integrated into, or installed adjacent to, a compatible Radiological Software Platform.
- The compatible Radiological Software Platform is configured to retrieve/receive contrast-enhanced head CT (CTA) images from the CT scanner or PACS, and automatically transmit, or make available, a copy of the image data for processing and analysis by the LVO Processing Engine. After successful processing of a case via the StrokeSENS LVO Processing Engine, the results are returned to the Radiological Software Platform for the intended purpose of triage and notification.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Yes
Input Imaging Modality
CTA data in DICOM format (vendor independent)
Anatomical Site
Head
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Hospital Networks and trained clinicians
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Retrospective case study on a heterogenous dataset of 400 independent studies (217 LVO cases and 183 non-LVO cases). The dataset had sufficient representation of relevant parameters such as age, sex, scanner vendor, and slice thickness. The dataset had a reasonably balanced prevalence of LVO (54.3%; 217 positive cases & 183 negative cases) and includes challenging cases (small/other occlusions and intracranial hemorrhage) that are representative of the clinical population undergoing baseline imaging for suspected acute ischemic stroke.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance validation testing and software verification activities were conducted to comply with specified design requirements in accordance with applicable consensus standards and to satisfy the special controls of the device classification.
Performance testing was conducted to verify compliance with specified design requirements in accordance with ISO 13485:2016, IEC 62304:2015, IEC 62366:2015 and ISO 14971:2019. DICOM conformance testing was performed to verify compliance with NEMA 3.1-3.20 (2011) standards. Verification and validation testing were conducted to ensure specifications and performance of the device and were performed per the FDA Guidance documents "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices". Software performance, verification and validation testing demonstrated that the StrokeSENS system met all design requirements and specifications.
To demonstrate the standalone performance of StrokeSENS LVO in accordance with the 892.2080 special controls, a retrospective case study was conducted to assess the sensitivity and specificity of StrokeSENS LVO for detecting Large Vessel Occlusion (LVO). Performance was reported on a heterogenous dataset of 400 independent studies (217 LVO cases and 183 non-LVO cases).
The StrokeSENS LVO module met the performance targets intended for the clinical performance assessment, as described in table 10.1-3. The device achieved a mean sensitivity of 89.4% CI = [85.3%, 93.5%], and mean specificity of 87.4% CI = [82.6%, 92.2%] for the binary LVO detection task on the test set (N=400, LVO=217. Non-LVO=183). In addition, an analysis of time to notify of suspicious cases was conducted by evaluating the average time for the StrokeSENS LVO device to process the CTA image and generate a notification (for LVO positive cases). The device achieved a mean value of 0.75 minutes (S.D: +/-0.17 mins, Min: 0.46 mins), which exceeded the goal of
§ 892.2080 Radiological computer aided triage and notification software.
(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Circle Neurovascular Imaging, Inc % Kyle Mayr Official Correspondent Circle Cardiovascular Imaging, Inc. 800 5th Ave SW Suite 1100 Calgary, Alberta T2P 3T6 Canada
October 21, 2021
Re: K212261
Trade/Device Name: StrokeSENS LVO Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS
Dear Kyle Mayr:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 14, 2021. Specifically, FDA is updating this SE Letter to indicate the correct official correspondent contact name and address as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jessica Lamb, OHT7: Office of In Vitro Diagnostics and Radiological Health, 301-796-6167, iessica.lamb(@fda.hhs.gov.
Sincerely.
Jessica Lamb
For
Jessica Lamb Assistant Director Mammography Ultrasound and Imaging Software Branch Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
1
Image /page/1/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Circle Neurovascular Imaging, Inc. % John Smith Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004
October 14, 2021
Re: K212261
Trade/Device Name: StrokeSENS LVO Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: OAS Dated: July 16, 2021 Received: July 20, 2021
Dear John Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
2
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
510(k) Number (if known) K212261
Device Name StrokeSENS LVO
Indications for Use (Describe)
StrokeSENS LVO is a radiological computer-aided triage and notification (CADt) software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained clinicians in workflow trage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVO) in head CTA images.
StrokeSENS LVO uses a software algorithm to identify suspected LVO findings. In the case of a suspected LVO, the system will send a notification to a pre-configured destination(s), notifying the clinicians of the existence of a suspected L VO that requires review. The notification system is intended to the standard of care workflow to notify clinicians of the existence of the case earlier that they may have been notified as part of the standard of care workflow.
Notifications may include a compressed preview of images. Notifications are meant for informational purposes only and are not intended for diagnostic use beyond notification. The StrokeSENS LVO device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of StrokeSENS LVO are intended to be used in conjunction with other patient information and based on professional judgement, to assist with triage / prioritization of medical images. Notified clinicians are responsible for viewing full images per standard of care.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED. |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Image /page/4/Picture/0 description: The image shows the logo for Circle Neurovascular Imaging. The logo features a brain-shaped design made of green circles on the left side. To the right of the brain design, the word "CIRCLE" is written in large, bold, black letters. Below "CIRCLE", the words "NEUROVASCULAR IMAGING" are written in smaller, black letters.
5.1 510k Summary - StrokeSENS LVO
I. SUBMITTER
Submitter's Name: | |
---|---|
Address: | |
Date Prepared: | |
Telephone Number: | |
Fax Number: | |
Contact Person: | |
Email: |
Circle Neurovascular Imaging, Inc. Suite 1100 - 800 5th Ave SW, Calgary, AB, Canada, T2P 3T6 July 16, 2021 +1 403 338 1870 +1 403 338 1895 Kyle Mayr kyle.mayr@circlecvi.com
II. DEVICE
Name of the Device: Short Brand Name: Common or Usual Name: Classification Name: Proposed Classification:
StrokeSENS LVO StrokeSENS LVO Medical Image Processing Software Radiological Computer-Assisted Triage And Notification Software Device Class II Special Controls Product Code: QAS Regulation Number: 21 CFR 892.2080
III. PREDICATE DEVICE
ContaCT, manufactured by Viz.Al (DEN170073)
IV. DEVICE DESCRIPTION
StrokeSENS LVO is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVQ) in head CTA images. StrokeSENS LVO uses a software alqorithm based on machine learning to identify suspected LVO findings. In the case of a suspected LVO, the system will send a notification to a preconfigured destination(s), notifying the clinicians of a suspected LVQ that requires review.
Clinical Characteristics and Procedures of Use
5
Image /page/5/Picture/0 description: The image is a logo for Circle Neurovascular Imaging. The logo features a cluster of green circles on the left side, resembling a brain. To the right of the circles, the word "CIRCLE" is written in large, bold, black letters. Below "CIRCLE", the words "NEUROVASCULAR IMAGING" are written in smaller, black letters.
After a CT Angiography (CTA) head scan is performed, the images are automatically routed to StrokeSENS LVO where they are processed and analyzed for characteristics suggestive of LVO. The notification system is comprised of an on-screen notice of a suspected LVQ to the user and an outbound notification to a neurovascular specialist. The notification states that a suspected LVO is present, and that the user should review the images in a diagnostic radiological viewer. In the case of a neqative finding, no notification is sent, and the standard of care workflow remains in place.
Technological Characteristics
StrokeSENS LVO DICOM-compliant software system consists of two main components: 1) the StrokeSENS LVO Processing Engine and 2) a compatible Radiological Software Platform:
-
- The StrokeSENS LVO Processing Engine is responsible for receiving, processing, and analyzing image data and communicating results. Primarily, the Processing Engine consists of a software algorithm (a sequence of instructions/operations) that is responsible for analyzing contrast-enhanced CT (CTA) image data of the head to identify characteristics that are consistent with LVO. The software algorithm is a binary classifier, providing a binary output of either positive or negative for suspected LVO, based on a pre-defined threshold. The output is returned to the Radiological Software Platform for the purposes of triage and notification. The Processing Engine is integrated into, or installed adjacent to, a compatible Radiological Software Platform.
-
- The compatible Radiological Software Platform is configured to retrieve/receive contrast-enhanced head CT (CTA) images from the CT scanner or PACS, and automatically transmit, or make available, a copy of the image data for processing and analysis by the LVO Processing Engine. After successful processing of a case via the StrokeSENS LVO Processing Engine, the results are returned to the Radiological Software Platform for the intended purpose of triage and notification.
V. INDICATIONS FOR USE/ INTENDED USE
1. Indications for use
StrokeSENS LVO is a radiological computer-aided triage and notification (CADt) software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communication of suspected positive findings of Large Vessel Occlusion (LVO) in head CTA images.
StrokeSENS LVO uses a software algorithm to identify suspected LVO findings. In the case of a suspected LVQ, the system will send a notification to a pre-configured destination(s), notifying the clinicians of the existence of a suspected LVO that requires review. The notification system is
6
Image /page/6/Picture/0 description: The image is a logo for Circle Neurovascular Imaging. The logo features a cluster of green circles on the left side, resembling a brain. To the right of the circles, the word "CIRCLE" is written in large, bold, black letters. Below "CIRCLE", the words "NEUROVASCULAR IMAGING" are written in smaller, black letters.
intended to be used in parallel to the standard of care workflow to notify clinicians of the existence of a potential LVO earlier than being notified as part of the standard of care workflow.
Notifications may include a compressed preview of images. Notifications are meant for informational purposes only and are not intended for diagnostic use beyond notification. The StrokeSENS LVO device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of StrokeSENS LVO are intended to be used in conjunction with other patient information and based on professional judgement, to assist with triage / prioritization of medical images. Notified clinicians are responsible for viewing full images per standard of care.
2. Special Conditions for Use Statement(s):
For prescription use only. US Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.
VI. COMPARISON WITH PREDICATE DEVICE
StrokeSENS LVO is considered to be substantially equivalent to Viz.Al: ContaCT (hereafter "ContaCT"), a commercially available device manufactured by Viz.Al, Inc. The subject and predicate devices are both radiological computer-assisted triage and notification software programs. Both devices use machine learning software implementations to analyze CTA images to aid in the prioritization, triage, and notification of suspected large vessel occlusion (LVQ) cases. Both software devices notify a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis performed by the device's algorithm, in parallel to the standard of care workflow. The subject and predicate devices are similar in that neither device alters the original image database or marks up / alters the CTA input images. The subject and predicate device differ in that the subject device generates an email notification that links to the StrokeSENS platform radiological viewing software or other compatible radiological viewing software whereas the predicate generates a notification that is sent to a mobile application.
StrokeSENS LVO has the same general intended use and similar indications, technological characteristics, and principles of operation as the previously cleared predicate device, with minor differences in the notification mechanism. The minor differences do not raise new questions on the safety and effectiveness of the device; therefore the subject device is as safe and effective as the predicate device. A summary substantial equivalence chart comparing the similarities and differences between the StrokeSENS LVO and its predicate device is provided below, as table 5.1-1.
| Feature /
Characteristic | Subject Device | Predicate Device | |
---|---|---|---|
General information | |||
Device name | StrokeSENS LVO | ContaCT | |
Manufacturer | Circle Neurovascular Imaging | Viz.Al | |
510(k) number | K212261 | DEN170073 | |
Device Class | II | II | |
Device classification | QAS | QAS | |
Regulation Name | Radiological Computer-Assisted | ||
Triage And Notification Software | Radiological Computer-Assisted Triage | ||
And Notification Software | |||
Regulation number | 21 CFR 892.2080 | 21 CFR 892.2080 | |
Indications for Use / | |||
Intended Use | StrokeSENS LVO is a | ||
radiological computer-aided | |||
triage and notification (CADt) | |||
software indicated for use in the | |||
analysis of CTA head images. | |||
The device is intended to assist | |||
hospital networks and trained | |||
clinicians in workflow triage by | |||
flagging and communication of | |||
suspected positive findings of | |||
Large Vessel Occlusion (LVO) in | |||
head CTA images. |
StrokeSENS LVO uses a
software algorithm to identify
suspected LVO findings. In the
case of a suspected LVO, the
system will send a notification to
a pre-configured destination(s),
notifying the clinicians of the
existence of a suspected LVO
that requires review. The
notification system is intended to
be used in parallel to the standard
of care workflow to notify
clinicians of the existence of the
case earlier that they may have
been notified as part of the
standard of care workflow.
Notifications may include a
compressed preview of images.
Notifications are meant for
informational purposes only and
are not intended for diagnostic
use beyond notification. The
StrokeSENS LVO device does
not alter the original medical | ContaCT is a notification-only, parallel
workflow tool for use by hospital networks
and trained clinicians to identify and
communicate images of specific patients
to a specialist, independent of standard of
care workflow.
ContaCT uses an artificial intelligence
algorithm to analyze images for findings
suggestive of a pre-specified clinical
condition and to notify an appropriate
medical specialist of these findings in
parallel to standard of care image
interpretation. Identification of suspected
findings is not for diagnostic use beyond
notification. Specifically, the device
analyzes CT angiogram images of the
brain acquired in the acute setting, and
sends notifications to a neurovascular
specialist that a suspected large vessel
occlusion has been identified and
recommends review of those images.
Images can be previewed through a
mobile application. Images that are
previewed through the mobile application
are compressed and are for informational
purposes only and not intended for
diagnostic use beyond notification.
Notified clinicians are responsible for
viewing non-compressed images on a
diagnostic viewer and engaging in
appropriate patient evaluation and
relevant discussion with a treating
physician before making care-related
decisions or requests. ContaCT is limited
to analysis of imaging data and should not
be used in-lieu of full patient evaluation or
relied upon to make or confirm diagnosis. | |
| | image and is not intended to be
used as a diagnostic device.
The results of StrokeSENS LVO
are intended to be used in
conjunction with other patient
information and based on
professional judgement, to assist
with triage / prioritization of
medical images. Notified
clinicians are responsible for
viewing full images per standard
of care. | | |
| Clinical Characteristics | | | |
| User Population | Hospital Networks and trained
clinicians | Same | |
| Clinical
application/Anatomical
Region | Acute Stroke / Head | Same | |
| Relationship to
standard of care
workflow | In Parallel / Concurrently | Same | |
| Technological Characteristics | | | |
| Input data type | CTA data in DICOM format
(vendor independent) | CTA data in DICOM format (vendor
independent) | |
| Algorithm
Implementation | Artificial Intelligence / Machine
Learning
Algorithms are static and locked.
Algorithms are not dynamic or
learning while in the market. | Same | |
| Alteration of original
image database | No | Same | |
| Notification / Workflow | Email, Workstation, Mobile
Notification message of
Suspected LVO | Mobile
Notification message of Suspected LVO | |
| DICOM compliant | Yes | Same | |
Table 5.1-1. Feature comparison table of StrokeSENS LVO with the predicate device, ContaCT.
7
Image /page/7/Picture/0 description: The image is a logo for Circle Neurovascular Imaging. The logo features a cluster of green circles on the left side, resembling a brain. To the right of the circles, the word "CIRCLE" is written in large, bold, black letters. Below "CIRCLE", the words "NEUROVASCULAR IMAGING" are written in smaller, black letters.
8
Image /page/8/Picture/0 description: The image shows the logo for Circle Neurovascular Imaging. The logo features a green brain-shaped design made of circles on the left side. To the right of the brain design is the word "CIRCLE" in large, bold, black letters, and below that is the phrase "NEUROVASCULAR IMAGING" in smaller, black letters.
9
Image /page/9/Picture/0 description: The image is a logo for Circle Neurovascular Imaging. The logo features a cluster of green circles on the left side, resembling a brain. To the right of the circles is the word "CIRCLE" in large, bold, black letters. Below the word "CIRCLE" is the phrase "NEUROVASCULAR IMAGING" in smaller, black letters.
VII. PERFORMANCE DATA
Performance validation testing and software verification activities were conducted to comply with specified design requirements in accordance with applicable consensus standards and to satisfy the special controls of the device classification.
Performance testing was conducted to verify compliance with specified design requirements in accordance with ISO 13485:2016, IEC 62304:2015, IEC 62366:2015 and ISO 14971:2019. DICOM conformance testing was performed to verify compliance with NEMA 3.1-3.20 (2011) standards. Verification and validation testing were conducted to ensure specifications and performance of the device and were performed per the FDA Guidance documents "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices". Software performance, verification and validation testing demonstrated that the StrokeSENS system met all design requirements and specifications.
To demonstrate the standalone performance of StrokeSENS LVO in accordance with the 892.2080 special controls, a retrospective case study was conducted to assess the sensitivity and specificity of StrokeSENS LVO for detecting Large Vessel Occlusion (LVO). Performance was reported on a heterogenous dataset of 400 independent studies (217 LVO cases and 183 non-LVO cases). The dataset had sufficient representation of relevant parameters such as age, sex, scanner vendor, and slice thickness. The dataset had a reasonably balanced prevalence of LVO (54.3%; 217 positive cases & 183 negative cases) and includes challenging cases (small/other occlusions and intracranial hemorrhage) that are representative of the clinical population undergoing baseline imaging for suspected acute ischemic stroke.
The StrokeSENS LVO module met the performance targets intended for the clinical performance assessment, as described in table 10.1-3. The device achieved a mean sensitivity of 89.4% Cl = [85.3%, 93.5%], and mean specificity of 87.4% C1 = [82.6%, 92.2%] for the binary LVO detection task on the test set (N=400, LVO=217. Non-LVO=183). In addition, an analysis of time to notify of suspicious cases was conducted by evaluating the average time for the StrokeSENS LVO device to process the CTA image and generate a notification (for LVO positive cases). The device achieved a mean value of 0.75 minutes (S.D: ±0.17 mins, Min: 0.46 mins), which exceeded the qoal of 80 % | 89.4%, (95% CI = 85.3, 93.5) |
10
Image /page/10/Picture/0 description: The image is a logo for Circle Neurovascular Imaging. The logo features a cluster of green circles on the left, resembling a brain. To the right of the circles is the word "CIRCLE" in large, bold, black letters, and below that is the phrase "NEUROVASCULAR IMAGING" in smaller, black letters.
Specificity | Specificity > 80 % | 87.4%, (95% CI = 82.6, 92.2) |
---|---|---|
Processing Time |