K Number
K212180

Validate with FDA (Live)

Date Cleared
2022-02-11

(214 days)

Product Code
Regulation Number
884.5160
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wearable Breast Pump (Model S12) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

Device Description

The Wearable Breast Pump (Model S12), is an electrically powered wearable single breast pump consisting of the following key components: a flange, linker, silicone diaphragm, pump motor, USB cable, valve, milk collector, and bra adjustment buckle. It is designed to work in the user's bra and has a rechargeable battery so it can be used hands-free without any external power cords. The motor unit includes a press-button user interface, pump body, and LED display. Pumping can be performed on one breast (single pumping). The user interface allows the user to switch from stimulation to expression mode and control the vacuum levels within those modes. Both stimulation and expression mode consist of 9 vacuum levels. The S12 model is capable of providing vacuum levels from 40-105 mmHg with cycling rates from 75-109 cycles per minute in stimulation mode and vacuum levels from 40-245 mmHg with cycling rates from 28-85 cycles per minute in expression mode. The model S12 Wearable Breast Pump is charged with a 5 V DC adaptor and powered by an internal rechargeable lithium-ion polymer battery. The motor unit operates on embedded software updates by end-users are not supported. The subject device is for repeated use by a single user in a home environment. The device is provided not sterile.

The breast pump expresses by creating a seal around the nipple using the flange and applying and releasing suction to the nipple. The milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.

All other components (i.e., motor unit) of the subject device are not in contact with the breast.

The motor unit operates on a rechargeable battery and does not function when charging. The rechargeable battery can be charged from the external USB adapter if the motor unit is not in operation.

The subject device components are made of the following materials:

  • · Motor unit: acrylonitrile-butadiene-styrene (ABS) plastic
  • · Flange, tube, valve, diaphragm: Silicone
  • · Linker, milk collection container: Polypropylene

All milk contacting components are compliant with 21 CFR 174-179.

AI/ML Overview

The provided text describes the Wearable Breast Pump (Model S12) and its 510(k) submission to the FDA. However, it does not contain a study comparing the device's performance against specific acceptance criteria in a clinical setting with human subjects, nor does it detail a study that proves the device meets such criteria through statistical analysis or a clinical trial.

Instead, the document focuses on demonstrating substantial equivalence to a predicate device (Willow Wearable Breast Pump) through non-clinical performance testing and a comparison of technological characteristics.

Therefore, many of the requested items cannot be extracted from this document, as they pertain to clinical efficacy or diagnostic performance studies that are not present here.

Here's the information that can be extracted or inferred:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" for clinical performance. Instead, it details non-clinical performance tests conducted to ensure the device's basic functionality and safety. The performance reported is that the device "meets mode/cycle specifications" and functions as intended, rather than quantitative patient-centric outcomes.

Acceptance Criteria (Inferred from testing)Reported Device Performance (from "Summary of Non-Clinical Performance Testing")
BiocompatibilityNon-cytotoxic, non-irritating, and non-sensitizing for user-contacting materials (tested per ISO 10993-5:2009, ISO 10993-10:2010).
Electrical SafetyCompliant with AAMI ANSI ES60601-1:2005/(R)2012, A1:2012; IEC 62133-2:2017; and IEC 60601-1-11:2015.
Electromagnetic CompatibilityCompliant with IEC 60601-1-2:2014.
Software FunctionalityEvaluated as recommended in the 2005 FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (No specific performance metrics are given, only that it was evaluated per guidance).
Vacuum Level and Cycling Rate"Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications." (Specific mmHg and cycles/minute ranges are listed in the comparison table, and the testing confirmed these are met.) For Stimulation: -40 to -105 (±5) mmHg, 70 to 114 cycles/minute. For Expression: -40 to -245 (±5) mmHg, 23 to 90 cycles/minute.
Backflow Protection"Backflow protection testing was conducted to verify liquid does not backflow into the tubing." (Testing confirmed no backflow).
Use Life"Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life." (Confirmed device maintains specifications).
Battery Performance"Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life." (Confirmed battery functionality).
Battery Status Indicator"Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life." (Confirmed indicator functionality).

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. The document describes non-clinical engineering and safety tests (e.g., biocompatibility testing, electrical safety, vacuum verification), not studies involving human subjects or data sets in the context of clinical performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The testing described is primarily laboratory-based and engineering validation, not dependent on expert interpretation of medical data or images for ground truth establishment.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided, as it pertains to medical data interpretation, which is not part of this 510(k) submission's described testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study described in the document. The device is a breast pump, not an AI-assisted diagnostic or interpretive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance evaluation of the device's mechanical and electrical functions was done, as detailed in the "Summary of Non-Clinical Performance Testing". This includes vacuum level verification, backflow protection, use life, and battery performance testing. However, this is not "algorithm only" performance in the context of AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests:

  • Biocompatibility: Ground truth established against standardized laboratory testing protocols (e.g., ISO 10993).
  • Electrical Safety & EMC: Ground truth established against recognized international standards (e.g., IEC 60601 series).
  • Software: Ground truth established by compliance with FDA guidance.
  • Vacuum/Cycle, Backflow, Use Life, Battery: Ground truth established against engineering design specifications and internal requirements for functionality and durability.

There is no mention of expert consensus, pathology, or outcomes data, as these are typically used for clinical diagnostic or treatment efficacy studies, which are not detailed here.

8. The sample size for the training set

This information is not provided. The device does not appear to utilize machine learning algorithms that would require a "training set" in the traditional sense. Its software operates on embedded logic, not a learned model.

9. How the ground truth for the training set was established

This information is not applicable as there is no mention of a training set for machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

February 11, 2022

Shenzhen TPH Technology Co., Ltd. Peter Chen General Manager Bulan Road, Nanwan Community, Longgang District Shenzhen, Guangdong 518100 China

Re: K212180

Trade/Device Name: Wearable Breast Pump (Model S12) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: January 10, 2022 Received: January 14, 2022

Dear Peter Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212180

Device Name Wearable Breast Pump (Model S12)

Indications for Use (Describe)

The Wearable Breast Pump (Model S12) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K212180

1. Submitter Information

Applicant:Shenzhen TPH Technology Co., Ltd.
Address:5th Floor, Building No.29 East side,Lianchuang 2th technology park, BulanRoad, Nanwan Community, LonggangDistrict, Shenzhen, China

2. Correspondent Information

Contact:Peter Chen
General Manager
Phone:+ 86 (755) 827-03212
Email:peter@tph-tech.com

3. Date prepared: February 10, 2022

4. Device Information

Device Name:Wearable Breast Pump (Model S12)
Common Name:Powered Breast Pump
Regulation Number:21 CFR 884.5160
Regulation Name:Powered Breast Pump
Product Code:HGX (Pump, Breast, Powered)
Regulatory Class:Class II

5. Predicate Device Information

Device Name:Willow Wearable Breast Pump
510(k) Number:K191577
Manufacturer:Exploramed NC7, Inc.

The predicate device has not been subject to a design-related recall.

6. Device Description

The Wearable Breast Pump (Model S12), is an electrically powered wearable single breast pump consisting of the following key components: a flange, linker, silicone diaphragm, pump motor, USB cable, valve, milk collector, and bra adjustment buckle. It is designed to work in the user's bra and has a rechargeable battery so it can be used hands-free without any external power cords. The motor unit includes a press-button user interface, pump body, and LED display. Pumping can be performed on one breast (single pumping). The user interface allows the user to switch from stimulation to expression mode and control the vacuum levels within those modes. Both stimulation and expression mode consist of 9 vacuum levels. The S12 model is capable of providing vacuum levels from 40-105 mmHg with cycling rates from 75-109 cycles per minute in stimulation mode and vacuum levels from 40-245 mmHg with cycling rates from 28-85 cycles per minute in expression mode. The model S12 Wearable Breast Pump is charged with a 5 V DC adaptor and powered by an internal rechargeable lithium-ion polymer battery. The motor unit operates on embedded software updates by end-users are not supported. The subject device is for repeated use by a single user in a home environment. The device is provided not sterile.

The breast pump expresses by creating a seal around the nipple using the flange and applying and releasing suction to the nipple. The milk is collected in a milk collection container, which can be used for storage. To

{4}------------------------------------------------

prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.

All other components (i.e., motor unit) of the subject device are not in contact with the breast.

The motor unit operates on a rechargeable battery and does not function when charging. The rechargeable battery can be charged from the external USB adapter if the motor unit is not in operation.

The subject device components are made of the following materials:

  • · Motor unit: acrylonitrile-butadiene-styrene (ABS) plastic
  • · Flange, tube, valve, diaphragm: Silicone
  • · Linker, milk collection container: Polypropylene

All milk contacting components are compliant with 21 CFR 174-179.

7. Indications for Use

The Wearable Breast Pump (Model S12) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

Table 1: Comparator Table for Subject and Predicate Devices

Wearable Breast Pump,Model S12K212180Subject DeviceWillow WearableBreast Pump 2.0K191577Predicate DeviceComparison
Product NameWearable Breast PumpWillow Wearable Breast PumpN/A
Product CodeHGXHGXSame
Regulation Number21 CFR 884.516021 CFR 884.5160Same
Regulatory ClassClass IIClass IISame
Patient PopulationLactating WomenLactating WomenSame
Indications for UseThe Wearable Breast Pump(Model S12) is intended toexpress milk from lactatingwomen in order to collect milkfrom their breasts. The device isintended for a single user.The Willow Wearable BreastPump 2.0 is intended to expressmilk from lactating women inorder to collect milk from theirbreasts. The device is intendedfor a single user.Same
Pump OptionsSingleSingle or DoubleDifferent: The difference in pumpoptions do not raise differentquestions of safety andeffectiveness.
Cycling controlmechanismMicrocontrollerMicrocontrollerSame
Backflow ProtectionYesYesSame
Suction ModesStimulation Mode andExpression ModeStimulation Mode andExpression ModeSame
Suction levels97Different: The differences insuction levels for each mode do notraise different questions of safetyand effectiveness,
Adjustable suctionlevelsYesYesSame
Flange Size24 mm and 27 mm21 mm, 24 mm, and 27 mmDifferent: The difference in flangesizes do not raise differentquestions of safety andeffectiveness.
Vacuum range:Stimulation-40 to -105 (±5) mmHg-60 to -105 (±5) mmHgDifferent: The difference inspecification does not raisedifferent questions of safety andeffectiveness. Differences inmaximum suction pressure weresubstantiated by performancetesting below.
Vacuum range:Expression-40 to -245 (±5) mmHg-60 to -245 (±5) mmHgDifferent: The difference inspecification does not raisedifferent questions of safety andeffectiveness. Differences inmaximum suction pressure weresubstantiated by performancetesting below.
Cycle Speed:Stimulation70 to 114 cycles/minute90 cycles/minuteDifferent: The difference inspecification does not raisedifferent questions of safety andeffectiveness. Differences in cyclespeed were substantiated byperformance testing below.
Cycle Speed:Expression23 to 90 cycles/minute60 cycle/minuteDifferent: The difference inspecification does not raisedifferent questions of safety andeffectiveness. Differences in cyclespeed were substantiated byperformance testing below.
ControlsOn/Off button;Mode selectionIncrease/decrease vacuumbutton;On/Off button;Increase/decrease vacuumbutton:Similar: The subject and predicatehave similar controls. Differencesin controls do not raise differentquestions of safety andeffectiveness.
Power SupplyLi-Ion BatteryLi-Ion BatterySame
IndicatorsYes, LEDYesSimilar: The subject device has anLED display. Differences in visualindicators do not raise differentquestions of safety andeffectiveness.
MaterialsMilk Container: PolypropyleneFlange: SiliconePump Outer Housing:Acrylonitrile ButadieneStyrene (ABS) plasticMilk Container: Co-polymerpolypropylene and siliconeFlange: Polypropylene (gradechanged) and siliconePump Outer Housing:Polycarbonate andThermoplastic polyurethaneDifferent: The subject andpredicate devices are comprised ofdifferent materials. Differences incomposition do not raise differentquestions of safety andeffectiveness.

{5}------------------------------------------------

The indications for use of the subject and predicate device are identical.

{6}------------------------------------------------

The subject and predicate devices have similar technological features, including device design, user interface, overall vacuum pressure range, cycle speeds, materials, and power source. The different technological characteristics of the subject device do not raise different questions of safety and effectiveness.

9. Summary of Non-Clinical Performance Testing

Biocompatibility

Biocompatibility studies, including Skin Irritation Testing, and Skin Sensitization testing were performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993- 1:2009 as follows:

  • Cytotoxicity (ISO 10993-5:2009)
  • Skin Sensitization (ISO 10993-10:2010)
  • Skin Irritation (ISO 10993-10:2010) ●

The testing supports the biocompatibility of the device. The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing.

Electrical Safety

Testing was conducted in accordance with AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD), IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

Electromagnetic Compatibility

Testing was conducted in accordance with IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests

Software

Software was evaluated as recommended in the 2005 FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:

  • Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
  • Backflow protection testing was conducted to verify liquid does not backflow into the tubing. ●
  • . Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life.
  • Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life.
  • Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.

{7}------------------------------------------------

10. Conclusion

The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).