(60 days)
No
The document mentions image processing and data analysis but does not explicitly mention AI, ML, or related terms like deep learning or neural networks. The description of the "Smart Positioning" workflow does not provide enough detail to confirm the use of AI/ML.
No.
The device is described as a medical tool intended to aid in detecting, localizing, and diagnosing cardiac diseases, not for providing therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states that the systems are a "medical tool intended for use... to aid in detecting, localizing and diagnosing of cardiac diseases and heart function of diseases, abnormalities, and disorders."
No
The device description explicitly states that the systems include a nuclear medicine imaging system using a CZT-based multi-detector array, patient and equipment supports, components and accessories, indicating the presence of significant hardware components beyond just software.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The GE MyoSPECT and MyoSPECT ES systems perform Nuclear Medicine (NM) imaging procedures. This involves detecting radioisotope tracer uptake within the patient's body, not analyzing samples taken from the body.
- Intended Use: The intended use is to aid in detecting, localizing, and diagnosing cardiac diseases and heart function by imaging the distribution of a tracer within the patient. This is an in-vivo (within the living organism) diagnostic method, not in-vitro (in glass or outside the living organism).
Therefore, based on the provided information, the GE MyoSPECT and MyoSPECT ES systems are not IVDs. They are medical imaging devices used for in-vivo diagnosis.
N/A
Intended Use / Indications for Use
The GE MyoSPECT and MyoSPECT ES systems are a medical tool intended for use by appropriately trained healtheare professionals to aid in detecting, localizing and diagnosing of cardiac diseases and heart function of diseases, abnormalities, and disorders. The systems output can also be used by the physician for planning, guiding, and monitoring.
MyoSPECT and MyoSPECT ES perform Nuclear Medicine (NM) imaging procedures for the detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types.
The GE MyoSPECT and MyoSPECT ES systems include signal analysis and display equipment, patient and equipment supports, components and accessories. The systems may include data and image processing to produce reconstructed trans-axial images. The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation.
The systems may be used for patients of all ages.
Product codes (comma separated list FDA assigned to the subject device)
KPS
Device Description
The GE MyoSPECT and MyoSPECT ES are Single Photon Emission Computed Tomography (SPECT) systems intended for nuclear cardiology imaging. MyoSPECT and MyoSPECT ES include a nuclear medicine imaging system using a CZT-based multi-detector array. MyoSPECT ES are identical systems that are differentiated by the number of detectors within the multi-detector array. MyoSPECT is offered with 19 detectors as opposed to 9 detectors for MyoSPECT ES is upgradable to MyoSPECT. The MyoSPECT and MyoSPECT ES systems do not introduce any new Intended Use.
The multi-detector array on MyoSPECT ES has a multi-pinhole collimator providing each detector with a pinhole collimator that is focused on a volume in space, all together forming the Quality Field of View (Q.FOV) in which the patient heart is positioned for a cardiac scan.
MyoSPECT and MyoSPECT ES add an option for Extended Field of View (EFOV) processing that enables more flexible positioning. The predicate device patient positioning workflow is also enhanced to provide users with visual means of the Q.FOV and EFOV boundaries, and recommendations on the use of Q.FOV or EFOV. The enhanced workflow is named "Smart Positioning" workflow. MyoSPECT's and MyoSPECT ES's patient table is the same as the one used on the NM 830 reference device with slight adaptations.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Nuclear Medicine (NM) imaging (Single Photon Emission Computed Tomography (SPECT))
Anatomical Site
Cardiac / Heart
Indicated Patient Age Range
All ages
Intended User / Care Setting
appropriately trained healthcare professionals
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
A clinical reader study using 28 studies for MyoSPECT and 32 studies for MyoSPECT ES was performed by three experience NM physicians. The exams constituted a clinically representative sample for evaluation of MyoSPECT's and MyoSPECT ES's performance. The three experienced NM physicians scored the images using 5-point Likert scales for overall image quality. All the physicians attested that their assessments demonstrated acceptable diagnostic results.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Clinical Reader Study:
- Sample size: 28 studies for MyoSPECT and 32 studies for MyoSPECT ES.
- Key results: The three experienced NM physicians scored the images using 5-point Likert scales for overall image quality. All the physicians attested that their assessments demonstrated acceptable diagnostic results. The results of the study support the determination of substantial equivalence.
Non-Clinical, Design Control Testing:
- Successfully completed design control testing per quality system.
- Conforms with IEC 60601-1 Ed. 3.1 and its applicable Collateral and Particular Standards.
- Quality assurance measures included: Requirement Definition, Risk Analysis, Technical Design Reviews, Formal Design Reviews, Software Development Lifecycle, Testing on unit level (Module verification), Integration testing (System verification), System Testing (Safety Testing, System and Image Performance Testing, Simulating Use Testing).
- Additional engineering bench testing: SPECT Resolution, Image Uniformity and Contrast, Energy resolution and simultaneous multi-isotope acquisition, Extended Field of View, Smart Positioning Q.FOV / EFOV Workflow.
- No additional hazards identified, no unexpected test results, no new or different questions of safety and effectiveness observed.
- Substantial equivalence concluded based on these tests and software documentation for "Moderate" level of concern.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1200 Emission computed tomography system.
(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.
0
August 27, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare % Mr. George Mashour Regulatory Affairs Manager 4 Hayozma Street Tirat Hacarmel, 30200 ISRAEL
Re: K212004
Trade/Device Name: MyoSPECT, MyoSPECT ES Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS Dated: June 24, 2021 Received: June 28, 2021
Dear Mr. Mashour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
1
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K212004
Device Name MyoSPECT MyoSPECT ES
Indications for Use (Describe)
The GE MyoSPECT and MyoSPECT ES systems are a medical tool intended for use by appropriately trained healtheare professionals to aid in detecting, localizing and diagnosing of cardiac diseases and heart function of diseases, abnormalities, and disorders. The systems output can also be used by the physician for planning, guiding, and monitoring.
MyoSPECT and MyoSPECT ES perform Nuclear Medicine (NM) imaging procedures for the detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types.
The GE MyoSPECT and MyoSPECT ES systems include signal analysis and display equipment, patient and equipment supports, components and accessories. The systems may include data and image processing to produce reconstructed trans-axial images. The images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation.
The systems may be used for patients of all ages.
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D) |
---|
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script in the center. There are decorative swirls around the letters. The logo is simple and recognizable.
510(k) SUMMARY
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
Date: | June 24th, 2021 |
---|---|
Submitter: | GE Medical Systems Israel, Functional Imaging (GE Healthcare) |
4 Havozma Street | |
Tirat Hacarmel, 30200, Israel | |
Primary Contact: | George Mashour |
Regulatory Affairs Manager | |
GE Medical Systems Israel, Functional Imaging | |
Tel: +972-4-8563666 ext. 639 | |
Email: george.mashour@ge.com | |
Secondary Contact: | John Jaeckle |
Chief Regulatory Affairs Engineer and Strategist | |
GE Healthcare | |
Tel: 262-424-9547 | |
email: john.jaeckle@ge.com | |
Device Trade Name: | MyoSPECT |
MyoSPECT ES | |
Device Classification: | Class II |
Regulation Number: | 21CFR 892.1200 |
Product Codes: | 90 KPS |
4
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. There are decorative flourishes resembling water droplets or stylized leaves surrounding the circle. The logo is simple and recognizable, representing the brand identity of General Electric.
Predicate Device Information | |
---|---|
Device Name: | Ventri 1.1 |
Manufacturer: | GE Medical Systems Israel, Functional Imaging |
510(k) Number: | K080124 |
Regulation Number / | |
Product Code: | 21CFR 892.1200 |
90 KPS |
Note: The current production of Ventri 1.1 is commercially named Discovery NM 530c. Through this summary, the commercial name Discovery NM 530c is used to refer to the predicate device.
Reference Device Information | |
---|---|
Device Name: | Discovery NM 630 |
Manufacturer: | GE Medical Systems Israel, Functional Imaging |
510(k) Number: | K111445 |
Regulation Number/Product Code: | 21CFR 892.1200 |
90 KPS |
Note: The current production of Discovery NM 630 is commercially named NM 830. Through this summary, the commercial name NM 830 is used to refer to the reference device.
Marketed Devices
MyoSPECT and MyoSPECT ES are a modification to the predicate Discovery NM 530c. The primary changes are the introduction of nine (9) detectors configuration on MyoSPECT ES, an option for Extended Field of View (EFOV) processing, enhanced patient positioning workflow, and replacing the patient table with the patient table from the reference NM 830 device. A nineteen (19) detector configuration remains on MyoSPECT, as the current production version of the predicate device.
Device Description
The GE MyoSPECT and MyoSPECT ES are Single Photon Emission Computed Tomography (SPECT) systems intended for nuclear cardiology imaging. MyoSPECT and MyoSPECT ES include a nuclear medicine imaging system using a CZT-based multi-detector array. MyoSPECT ES are identical systems that are differentiated by the number of detectors within the multi-detector array. MyoSPECT is offered with 19 detectors as opposed to 9 detectors for MyoSPECT ES is upgradable to MyoSPECT. The MyoSPECT and MyoSPECT ES systems do not introduce any new Intended Use.
The multi-detector array on MyoSPECT ES has a multi-pinhole collimator providing each detector with a pinhole collimator that is focused on a volume in space, all together forming the Quality Field of View (Q.FOV) in which the patient heart is positioned for a cardiac scan.
MyoSPECT and MyoSPECT ES add an option for Extended Field of View (EFOV) processing that enables more flexible positioning. The predicate device patient positioning workflow is also enhanced to provide users with visual means of the Q.FOV and EFOV boundaries, and recommendations on the use of Q.FOV
5
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, decorative elements, giving the logo a classic and recognizable appearance.
or EFOV. The enhanced workflow is named "Smart Positioning" workflow. MyoSPECT's and MyoSPECT ES's patient table is the same as the one used on the NM 830 reference device with slight adaptations.
Intended Use
The GE MyoSPECT and MyoSPECT ES systems are intended for Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. MyoSPECT and MyoSPECT ES include a Nuclear Medicine system using tomographic scanning mode supported by various acquisition types.
Indications for Use
The GE MyoSPECT and MyoSPECT ES systems are a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing and diagnosing of cardiac diseases and heart function for the evaluation of diseases, abnormalities, and disorders. The systems output can also be used by the physician for planning, guiding, and monitoring.
MyoSPECT and MyoSPECT ES perform Nuclear Medicine (NM) imaging procedures for the detection of radioisotope tracer uptake in the patient body, using tomographic scanning of single or multi-isotopes with either single or multi energy peaks. The tomographic scanning is supported by various acquisition types.
The GE MyoSPECT and MyoSPECT ES systems include signal analysis and display equipment, patient and equipment supports, components and accessories. The systems may include data and image processing to produce reconstructed trans-axial images can also be post processed to obtain additional images, imaging planes, analysis results and uptake quantitation.
The systems may be used for patients of all ages.
6
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are decorative swirls around the circle, giving it a classic and recognizable appearance.
Technological Characteristics
MyoSPECT and MyoSPECT ES employ the same fundamental scientific technology as the predicate and reference devices. The three systems share the same CZT multi-detector array, except for difference in the number of detectors within the array. The detectors acquire projections simultaneously, without system motion. The acquired projections are processed into trans-axial images using the same methods for the three systems. The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
| Subsystem /
Specification | Predicate Device
Discovery NM 530c (K080124) | Proposed Device
MyoSPECT, MyoSPECT ES |
|---------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Gantry | Multi-detector array gantry for
SPECT imaging.
70 cm bore. | Multi-detector array gantry for
SPECT imaging.
70 cm bore. |
| Detection | Multi-detector array
19 CZT Detectors | Multi-detector array
19 CZT Detectors (MyoSPECT)
9 CZT Detectors (MyoSPECT ES) |
| Collimator | Multi-Pinhole
Collimator | Multi-Pinhole
Collimator |
| Field of View | Q.FOV | Q.FOV
Extended FOV |
| Energy Range | 40 -200 keV | 60 -200 keV |
| Energy Resolution | Tc99m FWHM ≤ 7.0% | Tc99m FWHM ≤ 6.2% |
| Patient Positioning
Workflow | Yes | Yes |
| Standards Conformance | IEC 60601-1 and applicable
Collateral and Particular
Standards. | IEC 60601-1 and applicable
Collateral and Particular
Standards. |
| Image Processing | Xeleris Workstation | Xeleris Workstation and Smart
Console |
| Software Level of Risk | Moderate | Moderate |
MyoSPECT's and MyoSPECT ES's technological characteristics do not create new questions of safety or effectiveness, and did not introduce any new risks/hazards, warnings, or limitations.
7
Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative white swirls around the perimeter, giving it a dynamic and recognizable appearance.
Determination of Substantial Equivalence
Summary of Non-Clinical, Design Control Testing
MyoSPECT and MyoSPECT ES have successfully completed the design control testing per our quality system. No additional hazards were identified, and no unexpected test results were observed. MyoSPECT and MyoSPECT ES were designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GE believes that the extensive bench testing and the physician evaluations performed are sufficient for FDA's substantial equivalence determination.
MyoSPECT and MyoSPECT ES have has been independently tested and conforms with IEC 60601-1 Ed. 3.1 and its applicable Collateral and Particular Standards.
The following quality assurance measures have been applied to the development of the system:
- Requirement Definition
- Risk Analysis
- Technical Design Reviews
- Formal Design Reviews
- · Software Development Lifecycle
- Testing on unit level (Module verification)
- · Integration testing (System verification)
- · System Testing:
- o Safety Testing (Verification)
- System and Image Performance Testing (Verification)
- o Simulating Use Testing (Validation)
The testing and results did not raise new or different questions of safety and effectiveness than associated with predicate device. We consider the proposed device is substantially equivalent to the predicate device.
The substantial equivalence is also based on the software documentation for a "Moderate" level of concern. GE believes that MyoSPECT and MyoSPECT ES is of comparable type and substantially equivalent to the predicate device.
Additional Non-Clinical Testing
Engineering bench testing was performed to support substantial equivalence, demonstrate performance, and substantiate the product claims. This included testing for:
- I SPECT Resolution
- Image Uniformity and Contrast
- Energy resolution and simultaneous multi-isotope acquisition
- Extended Field of View
- Smart Positioning Q.FOV / EFOV Workflow
8
Image /page/8/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The letters are white, and the circle has a thin white border, giving it a clean and recognizable appearance.
Clinical Testing
A clinical reader study using 28 studies for MyoSPECT and 32 studies for MyoSPECT ES was performed by three experience NM physicians. The exams constituted a clinically representative sample for evaluation of MyoSPECT's and MyoSPECT ES's performance. The results of the study support the determination of substantial equivalence. The three experienced NM physicians scored the images using 5-point Likert scales for overall image quality. All the physicians attested that their assessments demonstrated acceptable diagnostic results.
Substantial Equivalence Conclusion
The changes associated with MyoSPECT and MyoSPECT ES do not create a new Intended Use and represent equivalent technological characteristics, with no impact on the control mechanisms, operating principle, and energy type. GE's quality system's design verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
Based on development under GE Healthcare's quality system, the successful system and software verification and validation testing, conformance to standards, the additional engineering bench testing, and the clinical reader study demonstrates that MyoSPECT and MyoSPECT ES are substantially equivalent to, and hence as safe and as effective for its Intended Use, as the legally marketed predicate device.