K Number
K211794
Date Cleared
2021-09-28

(110 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. They are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The proposed device(s) are blue color, and flat pleated type mask, utilizing ear loops way for wearing, and they all have nose clips design for fitting the face mask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of Polypropylene non-woven fabric, and the middle layer is made of polypropylene melt-blown cloth. The model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of Elastic materials consisting of Nylon low elastic yarn and Spandex. The nose piece contained in the proposed device(s) is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Polypropylene and galvanized iron wire plastic materials. The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

AI/ML Overview

The document describes the non-clinical performance testing of a disposable surgical mask to demonstrate its substantial equivalence to a predicate device, rather than a study involving an AI-powered device or human readers. Therefore, much of the requested information regarding AI assistance, human readers, effect size, and ground truth establishment for AI models is not applicable to this document.

However, I can provide information based on the performance testing of the surgical mask.

1. A table of acceptance criteria and the reported device performance

ItemReported Device PerformanceAcceptance Criteria (Level 2)Result
Fluid Resistance Performance ASTM F186231 out of 32 Pass at 120 mmHg29 out of 32 Pass at 120mmHgPASS
Particulate Filtration Efficiency ASTM F229999.4%≥ 98%PASS
Bacterial Filtration Efficiency ASTM F210199.36%≥ 98%PASS
Differential Pressure (Delta P) EN 14683 Annex C4.43mmH2O/cm²< 6.0mmH2O/cm²PASS
Flammability 16 CFR 1610Class 1Class 1PASS
CytotoxicityNon-cytotoxicNon-CytotoxicPASS
IrritationNon-irritatingNon-IrritatingPASS
SensitizationNon-sensitizingNon-SensitizingPASS

2. Sample size used for the test set and the data provenance

  • Fluid Resistance Performance (ASTM F1862): 32 samples per lot, tested across three lots (though the table summarizes the results for the current device as "31 out of 32 Pass at 120 mmHg" without specifying if this is across multiple lots for the current device). For the predicate, it states "32 out of 32 pass at 120 mmHg".
  • Other performance tests (PFE, BFE, Differential Pressure, Flammability): The exact sample size for these specific tests is not explicitly stated as a number of individual masks, but rather a percentage or single value representing the performance of the device. These are typically performed on a representative sample of the manufactured product.
  • Biocompatibility (Cytotoxicity, Irritation, Sensitization): Not explicitly stated, but typically involves a defined number of samples according to ISO standards.
  • Data Provenance: The tests were conducted internally by SHANGHAI NO.9 KNITTING MILL CO., LTD., likely in a laboratory setting. The document does not specify the country of origin of the test samples beyond the manufacturing location (Shanghai, China). These are prospective tests on the manufactured product to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The "ground truth" here refers to established technical standards (like ASTM, ISO, EN, CFR) for material and product performance, not expert consensus on interpretations of data. The results are objective measurements against these predefined numerical and qualitative criteria.

4. Adjudication method for the test set

This question is not applicable. The test results are objective measurements (e.g., percentages, pressure readings, "pass/fail" based on established criteria) and do not involve human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not an AI device. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not an AI device. No standalone algorithm performance was assessed.

7. The type of ground truth used

The "ground truth" for the device's performance is based on established industry standards and regulatory requirements (e.g., ASTM F1862, ASTM F2299, ASTM F2101, EN 14683, 16 CFR 1610, ISO 10993). The acceptance criteria are derived directly from these technical specifications for medical face masks, specifically for Level 2 performance.

8. The sample size for the training set

This is not an AI device. There is no concept of a "training set" in this context.

9. How the ground truth for the training set was established

This is not an AI device. There is no concept of a "training set" or "ground truth establishment for a training set." The "ground truth" for the acceptance criteria for the surgical mask is established by widely recognized and published national and international standards for medical device performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

September 28, 2021

Shanghai No.9 Knitting Mill Co., Ltd. Ivy Wang Consultant Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K211794

Trade/Device Name: Disposable surgical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 30, 2021 Received: June 30, 2021

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211794

Device Name Disposable surgical mask

Indications for Use (Describe)

The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. They are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SHANGHAI NO.9 KNITTING MILL CO., LTD. NO.555 KANGWU ROAD, PUDONG NEW DISTRICT, SHANGHAI, China, 201315

510(K) Summary K211794

Date of Summary prepared: 2021-09-23

A. Applicant:

Name: SHANGHAI NO.9 KNITTING MILL CO., LTD. Address: NO.555 KANGWU ROAD, PUDONG NEW DISTRICT, SHANGHAI, China, 201315 Contact Person: WANG YIPEI Tel: 021-58128888, 86-13641936577 Email: jszxgongyi(@163.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com

B. Device:

Trade Name: Disposable surgical mask Common Name: Surgical face mask Model: Ear loops; Non-sterile

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

C. Predicate device:

K203426 Surgical Face Mask (Non-sterile) Manufactured by Nantong Taiweishi Medical Technology Co., Ltd.

D. Intended Use/Indications for Use:

The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. They are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

{4}------------------------------------------------

SHANGHAI NO.9 KNITTING MILL CO., LTD. NO.555 KANGWU ROAD, PUDONG NEW DISTRICT, SHANGHAI,China,201315

E. Device Description:

The proposed device(s) are blue color, and flat pleated type mask, utilizing ear loops way for wearing, and they all have nose clips design for fitting the face mask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of Polypropylene non-woven fabric, and the middle layer is made of polypropylene melt-blown cloth. The model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of Elastic materials consisting of Nylon low elastic yarn and Spandex. The nose piece contained in the proposed device(s) is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Polypropylene and galvanized iron wire plastic materials. The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

DeviceProposed DevicePredicate DeviceResult
ManufacturerSHANGHAI NO.9 KNITTINGMILL CO., LTD.Nantong Taiweishi MedicalTechnology Co., Ltd.-
510K numberK211794K203426-
NameDisposable surgical maskSurgical Face Mask (Non-sterile)Same
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21CFR878.4040)Same
Intendeduse/Indications forUseThe Disposable Surgical Mask isintended to be worn to protect boththe patient and healthcare personnelfrom transfer of microorganisms,body fluids and particulate material.They are intended for use in infectioncontrol practices to reduce thepotential exposure to blood and bodyfluids. This is a single use, disposabledevice(s), provided non-sterile.The Surgical Face Mask (non-sterile)intended to be worn to protect both thepatient and healthcare personnel fromtransfer of microorganisms, body fluidsand particulate material. These facemasks are intended for use in infectioncontrol practices to reduce the potentialexposure to blood and body fluids. Thisa single use, disposable device(s),provided non-sterile.Same
ModelEar loop, Flat pleates,3 layersEar loop, Flat pleated,3 layersSame
MaterialOuterlayerPolypropylene non-woven fabricSpun-bond polypropyleneSame
MiddlelayerPolypropylene melt-blown clothMelt blown polypropylene filterSame
InnerlayerPolypropylene non-woven fabricSpun-bond polypropyleneSame
NoseclipPolypropylene and Galvanized ironwirePolyethyleneDifferent
Ear loopsNylon low elastic yarn and spandexNylon and Spandex
ColorBlueBlueSame
Dimension175mm+/-8.75mm175mm+/-5%Similar
Dimension (Width)95mm+/-4.75mm95mm+/-5%Similar
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100 levelLevel 2Level 2Same
BiocompatibilityISO10993ISO10993Same
Cytotoxicity(ISO10993-5)Under the conditions of the study, thedevice is non-cytotoxic.Under the conditions of the study, thedevice is non-cytotoxic.Same
Irritation(ISO10993-10)Under the conditions of the study, thedevice is non-irritating.Under the conditions of the study, thedevice is non-irritating.Same
Sensitization(ISO10993-10)Under the conditions of the study, thedevice is non-sensitizingUnder the conditions of the study, thedevice is non-sensitizingSame
Fluid ResistancePerformanceASTM F186231 out of 32 pass at 120 mmHg in1st lot tested;32 out of 32 Pass at 120 mmHg in2nd lot tested:32 out of 32 Pass at 120 mmHg in3rd lot tested.32 out of 32 pass at 120 mmHgSimilar
ParticulateFiltration efficiencyASTM F229999.4%99.16%Similar
Bacterial FiltrationEfficiency ASTMF210199.36%99.74%Similar
DifferentialPressure (Delta P)EN 14683 Annex C4.43mmH2 O/cm²3.7mm H2O/cm²Similar
FlammabilityCFR 161016Class 1Class 1Same

F. Technological characteristics comparison between the subject and predicate devices Table 1 General Comparison

{5}------------------------------------------------

SHANGHAI NO.9 KNITTING MILL CO., LTD. NO.555 KANGWU ROAD, PUDONG NEW DISTRICT, SHANGHAI,China,201315

From the comparison we found the name of inner layer, out layer and Ear loops of the proposed device is different from the predicate device, but the material is the same. Only the material of the current nose clip is different from the predicate device.

The biocompatibility tests were conducted to both components to ensure their compliance to the ISO10993-5 and ISO10993-10. There is no new risk generated from the difference of the material.

For the Performance testing, the test results are not identical to each other, but they are similar. The predicate device and the proposed device meet the requirement of Level 2 medical mask according to the ASTM F 2100.

G. Summary of non-Clinical performance test

{6}------------------------------------------------

SHANGHAI NO.9 KNITTING MILL CO., LTD.

NO.555 KANGWU ROAD, PUDONG NEW DISTRICT, SHANGHAI,China,201315

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff:

Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

  • A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
  • A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
  • ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks

  • A ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity);
  • A EN 14683, Medical Face Masks-Requirements and Test Methods;
  • ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

  • ASTM F2299, stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;

  • A 16 CFR 1610, Standard for the Flammability of clothing textiles;
ItemProposed deviceAcceptance Criteria(level 2)Result
Fluid Resistance PerformanceASTM F186231 out of 32 Pass at 120mmHg29 out of 32 Pass at120mmHgPASS
Particulate FiltrationEfficiency ASTM F229999.4%$\u2265$ 98%PASS
Bacterial FiltrationEfficiency ASTM F210199.36%$\u2265$ 98%PASS
Differential Pressure (DeltaP) EN 14683 Annex C4.43mmH2 O/cm $\u00b2$< 6.0mmH2 O/cm $\u00b2$PASS
Flammability 16 CFR 1610Class 1Class 1PASS

Table 2 - Performance Testing

Table 3 Biocompatibility Comparison

ItemProposed deviceAcceptance CriteriaResult
CytotoxicityUnder the conditions of the study, the device is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder the conditions of the study, the device is non-irritating.Non-IrritatingPASS
SensitizationUnder the conditions of the study, the device is non-sensitizingNon-SensitizingPASS

H. Clinical Test Conclusion

No clinical study is included in this submission.

{7}------------------------------------------------

SHANGHAI NO.9 KNITTING MILL CO., LTD. NO.555 KANGWU ROAD, PUDONG NEW DISTRICT, SHANGHAI,China,201315

I. Conclusion

The nonclinical tests performed demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, K203426 Surgical Face Mask (Non-sterile) by Nantong Taiweishi Medical Technology Co., Ltd..

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.