K Number
K211758
Date Cleared
2021-08-03

(57 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3.3 mg/ml)14.9
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)22.9
Vincristine Sulfate (1.0 mg/ml)>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 14.9 minutes. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa

Device Description

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

AI/ML Overview

The provided document describes the FDA 510(k) clearance for a "Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour)". This is not an AI/ML device but a physical medical device (examination gloves).

Therefore, the requested information (acceptance criteria and study details related to AI/ML device performance, ground truth, expert adjudication, MRMC studies, etc.) is not applicable to this document.

The document focuses on the glove's performance against chemotherapy drugs, as detailed in the "Indications for Use" section. It reports breakthrough times for various chemotherapy drugs based on ASTM D6978-05 (Reapproved 2019) standards.

Here's the relevant information that can be extracted from the provided text, modified to fit the closest possible interpretation of your request, even though it's not an AI/ML device:

1. Table of Acceptance Criteria and Reported Device Performance

Chemotherapy Drug and ConcentrationAcceptance Criteria (Minimum Breakthrough Detection Time)Reported Device Performance (Breakthrough Detection Time in Minutes)
Carmustine (3.3 mg/ml)(Implicitly, would ideally be high)14.9
Cisplatin (1.0 mg/ml)(Implicitly, would ideally be high)>240
Cyclophosphamide (20.0 mg/ml)(Implicitly, would ideally be high)>240
Dacarbazine (10.0 mg/ml)(Implicitly, would ideally be high)>240
Doxorubicin Hydrochloride (2.0 mg/ml)(Implicitly, would ideally be high)>240
Etoposide (20.0 mg/ml)(Implicitly, would ideally be high)>240
Fluorouracil (50.0 mg/ml)(Implicitly, would ideally be high)>240
Methotrexate (25.0 mg/ml)(Implicitly, would ideally be high)>240
Mitomycin C (0.5 mg/ml)(Implicitly, would ideally be high)>240
Paclitaxel (6.0 mg/ml)(Implicitly, would ideally be high)>240
Thiotepa (10.0 mg/ml)(Implicitly, would ideally be high)22.9
Vincristine Sulfate (1.0 mg/ml)(Implicitly, would ideally be high)>240

Note: The "acceptance criteria" here are implied by the standard ASTM D6978-05 for assessing permeation. The higher the breakthrough time, the better the performance. The document explicitly issues warnings for Carmustine and Thiotepa due to their low breakthrough times, suggesting these values did not meet an implicit desired criterion for safe use.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not explicitly stated in the document. The ASTM D6978-05 standard would define the sample size requirements for testing gloves against chemotherapy drugs.
  • Data Provenance: The document does not specify the country of origin of the data. The testing was conducted according to ASTM D6978-05 (Reapproved 2019), which is an American Society for Testing and Materials standard. The study is prospective in the sense that the gloves were specifically tested for this clearance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This is not an AI/ML device where expert consensus for ground truth is typically required. The "ground truth" here is the physical measurement of chemical permeation according to a standardized test method.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods are relevant for subjective interpretations, typically in AI/ML performance evaluation. The ASTM D6978-05 standard is a physical chemical test with objective measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/ML device; therefore, no MRMC study, human readers, or AI assistance is relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an algorithm or AI system. The testing describes the standalone performance of the physical glove.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this device's performance is established by objective chemical permeation measurements obtained through standardized laboratory testing as per ASTM D6978-05.

8. The sample size for the training set:

  • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. No training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2021

Hartalega Ngc Sdn. Bhd. Nurul Kong Deputy General Manager-Quality Assurance No.1 Persiaran Tanjung Sepang, Kawasan Perindustrian Tanjung 43900 Malaysia

Re: K211758

Trade/Device Name: Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZC, OPC Dated: June 3, 2021 Received: June 7, 2021

Dear Nurul Kong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211758

Device Name

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour)

Indications for Use (Describe)

Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
Carmustine (3.3 mg/ml)14.9
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin Hydrochloride (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)22.9
Vincristine Sulfate (1.0 mg/ml)>240

Please note that Carmustine and Thiotepa have extremely low permeation times of 14.9 minutes. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

× Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.