K Number
K211649
Device Name
PCEA Syringe Set
Date Cleared
2021-11-23

(179 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.

Device Description

The proposed device is a patient controlled epidural analgesia (PCEA) syringe set. It is a single use disposable device intended for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device. It is a non-pyrogenic, sterile device that can be directly attached to a syringe. The PCEA syringe set consists of non-DEHP PVC tubing/bushing, female NRFit cap, female NRFit lock, anti-siphon valve, on-off clamp, male NRFit lock, and male NRFit cap. It is used to administer anesthetics and/or analgesics from a syringe to the patient's epidural space through an epidural access device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Baxter Healthcare Corporation PCEA Syringe Set (K211649), based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance CriteriaReported Device Performance
ISO 80369-6 Tests on Male NRFit (Neuraxial Fit) Lock ConnectorISO 80369-6: 2016, Clause 5 (as applicable), and Clauses 6.1 through 6.6All tests met the acceptance criteria.
ISO 80369-6 Tests on Female NRFit (Neuraxial Fit) Lock ConnectorISO 80369-6: 2016, Clause 5 (as applicable), and Clauses 6.1 through 6.6All tests met the acceptance criteria.
Tensile Strength TestISO 8536-9:2015, Clause 5.3All tests met the acceptance criteria.
Leak TestISO 8536-9:2015, Clause A.4All tests met the acceptance criteria.
Counter Flow TestISO 8536-12:2007+A1:2013, Clause A.4All tests met the acceptance criteria.
Blocking Performance TestISO 8536-12:2007+A1:2013, Clause A.6All tests met the acceptance criteria.
Opening Pressure TestISO 8536-12:2007+A1:2013, Clause A.7.1 and per Baxter Test MethodAll tests met the acceptance criteria.
Particulate Matter TestUSP Chapter <788>All tests met the acceptance criteria.
Clamp Activation Force TestPer Baxter Test MethodAll tests met the acceptance criteria.
Clamp Shut-Off TestISO 8536-14:2018, Clause A.1All tests met the acceptance criteria.
Non-DEHP Claim VerificationPer Baxter Test Method (as tested in K161808)All tests met the acceptance criteria.
BiocompatibilityISO-10993-1 categorization (for prolonged contact, external communicating device, tissue/bone/dentin) and FDA-2013-D-0350 Guidance. Specific assays: Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Intracutaneous (Irritation) Reactivity (ISO 10993-10), Acute Systemic Toxicity (ISO 10993-11), Material Mediated Pyrogen (ISO 10993-11), Genotoxicity (ISO 10993-3), Subacute Systemic Toxicity (ISO 10993-11).The device met the designated ISO 10993-1 categorization and is biocompatible and appropriate for the intended use.
SterilitySterility Assurance Level (SAL) of ≤ 10-6 using ANSI/AAMI/ISO 11137-2. Endotoxin limit of 2.15 EU/device per USP <161>.MSD (Minimum Sterilizing Dose) required to provide a 10-6 SAL was established and validated. Routine periodic pre-sterilization bioburden testing is performed. (Implied to have met criteria, as the conclusion states substantial equivalence).
Shelf LifeNot explicitly stated as a numerical acceptance criterion in the table, but "supports a shelf-life claim of 3 (three) years."Aging testing performed to support a 3-year shelf-life claim.

2. Sample Size Used for the Test Set and the Data Provenance:

The document describes "bench tests" and does not specify a separate "test set" in the context of clinical data, human readers, or image analysis. The tests appear to be conducted on the device itself. The provenance of the data is from Baxter Healthcare Corporation's internal testing. There is no information regarding country of origin or retrospective/prospective nature as this refers to device performance testing rather than clinical study data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

Not applicable. The "ground truth" for these tests relates to engineering and manufacturing standards (e.g., ISO standards, USP chapters) rather than human expert interpretation of medical images or conditions.

4. Adjudication Method for the Test Set:

Not applicable. The tests are based on objective measurements against established engineering and regulatory standards, not human opinion requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a medical administration set, not an AI-powered diagnostic or assistive tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" is established by:

  • International and national standards bodies (e.g., ISO 80369-6, ISO 8536-9, ISO 8536-12, ISO 8536-14, ISO 10993, ANSI/AAMI/ISO 11137-2)
  • Pharmacopeial standards (e.g., USP Chapter <788>, USP <161>)
  • Baxter Test Methods, which are company-specific internal procedures developed to ensure quality and compliance.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as no training set is used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 23, 2021

Baxter Healthcare Corporation Jeffrey Thompson Manager, Regulatory Affairs 25212 West Illinois Route 120 Round Lake, Illinois 60073

Re: K211649

Trade/Device Name: PCEA Syringe Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: PWH Dated: October 20, 2021 Received: October 25, 2021

Dear Jeffrey Thompson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211649

Device Name PCEA Syringe Set

Indications for Use (Describe)

For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is the main focus of the image and is positioned horizontally.

K211649

510(k) Summary

OWNER:

Baxter Healthcare Corporation 25212 W. IL Route 120 Round Lake, Illinois 60073

CONTACT PERSON:

Jeffrey E. Thompson Manager, Regulatory Affairs 25212 W. IL Route 120 Round Lake, Illinois 60073 Telephone: (224) 270 3806 Fax: (224) 270 4119

Date Preparation: November 23, 2021

SUBJECT DEVICE:

Trade/Device Name: PCEA Syringe Set Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5440 Common Name: Administrations Sets with Neuraxial Connectors Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: PWH

Code #Device Description
2P3334PCEA Syringe Set with Anti-Siphon Valve, 85" (216 cm),Vol 2.2 mLImage: [device]1: Female NRFit Cap2: Female NRFit Lock3: Anti-Siphon Valve4: On-Off Clamp5: Male NRFit Lock6: Male NRFit Cap7: Non-DEHP Tube Bushing8,9: Non-DEHP Tube

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is slightly italicized, giving the word a sense of movement. The word is the logo for the Baxter International Inc. company.

PREDICATE DEVICE:

DeviceCompanyPredicate 510(k)ClearanceDate
CADD Yellow Extension Setswith NRFit™ connectorSmiths Medical ASD, Inc.K162219July 20, 2017

Table 2: Predicate Device

REASON FOR SUBMISSION:

The basis for this premarket notification is the intent to market a patient controlled epidural analgesia (PCEA) syringe set. The proposed device in this submission is a single-use, disposable device, intended for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.

DESCRIPTION OF THE DEVICE:

The proposed device is a patient controlled epidural analgesia (PCEA) syringe set. It is a single use disposable device intended for the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device. It is a non-pyrogenic, sterile device that can be directly attached to a syringe.

The PCEA syringe set consists of non-DEHP PVC tubing/bushing, female NRFit cap, female NRFit lock, anti-siphon valve, on-off clamp, male NRFit lock, and male NRFit cap. It is used to administer anesthetics and/or analgesics from a syringe to the patient's epidural space through an epidural access device.

The anti-siphon valve component adds specific functionalities to facilitate the administration of anesthetics and/or analgesics to the patient's epidural access device, as described below:

  • Anti-siphon valve: reduces the risk of any inadvertent free flow of solution to the patient. The higher opening pressure does not allow administration of flow into the patient's epidural access device, unless the stipulated pressure is applied. Another safety feature is that the valve does not allow any backflow. This prevention of backflow is achieved through the valve's ability to withstand a counterflow pressure.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is horizontally oriented and takes up most of the frame.

INDICATIONS FOR USE:

For the administration of anesthetics and/or analgesics from a syringe into the patient's epidural space through an epidural access device.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:

The proposed device is substantially equivalent to the predicate device, previously cleared under 510(k) premarket notification K162219 on July 20, 2017. The intended use and function of the proposed device is equivalent to the predicate device.

Table 3 is a device comparison table outlining the differences between the predicate and proposed devices.

FeaturePredicateK162219Subject DeviceK211649Assessment of Differences
Indication for UseCADD YellowExtension Sets withNRFitTM connectorsare designed for useonly with CADDYellow MedicationCassette Reservoirswith NRFitTMconnectors for thedelivery of regionalanesthetics ornarcotics.For theadministration ofanesthetics and/oranalgesics from asyringe into thepatient's epiduralspace through anepidural accessdevice.Both the proposed andpredicate device are intendedto deliver similarmedications to the patientthrough NRFit connections.The specified fluidcontainers for the proposedand predicate devices aredifferent; however, theconnections to both fluidcontainers claim complianceto the same ISO 80369-6standard ("NRFit"). Designcontrol activities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed device doesnot raise different questionsof safety and effectiveness.
SterilitySterile; EthyleneOxideSterile; GammaIrradiationBoth the proposed andpredicate devices have beenvalidated to a sterilityassurance level (SAL) of ≤10-6 using ISO 11137 for theproposed device and ISO11135 for the predicatedevice. The sterilizationmodalities for the proposedand predicate devices are
FeaturePredicateK162219Subject DeviceK211649Assessment of Differences
different (gamma irradiationand ethylene oxide,respectively); however, thevalidated SAL requirementfor both devices aligns withthe same ANSI/AAMI ST67standard (Sections 4.2.1 and4.2.1.1), and thetechnological differencesbetween the sterilizationmodalities do not raisequestions of safety andeffectiveness.
Non-PyrogenicYesSameN/A
Single UseYesSameN/A
Length90"85" (216 cm)The proposed device lengthis nominally shorter than thepredicate device length.Design control activitieshave been conducted andhave confirmed that thedifferent technologicalcharacteristics of theproposed device does notraise different questions ofsafety and effectiveness.Lengths were calculatedusing nominal values fromcomponent and tubing partspecifications per Baxterprocedure.
Tubing Inner / OuterDiameterInformation notavailableInner Diameter - 1.1mmOuter Diameter - 2.5mmThe information is notavailable for the predicatedevice. Design controlactivities have beenconducted on the proposeddevice and have confirmedthat if there are differenttechnological characteristicsto the predicate device, theywould not raise differentquestions of safety andeffectiveness.Tensile strength and leakagewere evaluated per ISO8536-9.
FeaturePredicateK162219Subject DeviceK211649Assessment of Differences
Priming Volume2.3 mL2.2 mLThe proposed device primingvolume is nominally lessthan the predicate devicepriming volume. Designcontrol activities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed device doesnot raise different questionsof safety and effectiveness.Priming volume wasdetermined per Baxter testmethod.
Fluid Path Components/Materials
Anti-Siphon ValvePolycarbonate(Housing)Silicone(Membrane)SameN/A
TubingPolyvinyl ChlorideSameN/A
Tube BushingNot ApplicablePolyvinyl ChlorideThe predicate device doesnot have tube bushings.Design control activitieshave been conducted andhave confirmed that thedifferent technologicalcharacteristics of theproposed device does notraise different questions ofsafety and effectiveness.Biocompatibilityassessments were conductedper ISO-10993-1.
Female NRFit LockNot ApplicableCopolyesterThe predicate device doesnot have a female NRFit
FeaturePredicateK162219Subject DeviceK211649Assessment of Differences
lock. Design controlactivities have beenconducted and haveconfirmed that the differenttechnological characteristicsof the proposed device doesnot raise different questionsof safety and effectiveness.Biocompatibilityassessments were conductedper ISO-10993-1.Functional performance wasevaluated per ISO 80369-6.
Male NRFit LockAcrylonitrileButadiene StyreneCopolyesterThe predicate device doesnot have the same type ofmaterial formulation.Design control activitieshave been conducted andhave confirmed that thedifferent technological
Male NRFit CapLinear Low DensityPolyethyleneAcrylicThe predicate device doesnot have the same type ofmaterial formulation.Design control activitieshave been conducted andhave confirmed that thedifferent technologicalcharacteristics of theproposed device does notraise different questions ofsafety and effectiveness.Biocompatibilityassessments were conductedper ISO-10993-1.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slightly slanted to the right. The word is the logo for the Baxter International Inc. company.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right, giving it a dynamic appearance.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be italicized, with the letters slightly slanted to the right. The word is the logo for the Baxter International Inc. company.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Baxter" in a bold, italicized, sans-serif font. The color of the text is a dark blue. The word is the logo for the Baxter International Inc. company.

DISCUSSION OF NONCLINICAL TESTS:

Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet their acceptance criteria and support that the proposed device is appropriately designed for its intended use.

Performance Data:

The following bench tests (Table 4) were conducted to evaluate the functional performance of the proposed device:

TestAcceptance Criteria
ISO 80369-6 Tests on Male NRFit (Neuraxial Fit) LockConnectorISO 80369-6: 2016, Clause 5 (as applicable),ISO 80369-6:2016, Clause 6.1,ISO 80369-6:2016, Clause 6.2,ISO 80369-6:2016, Clause 6.3,ISO 80369-6:2016, Clause 6.4,ISO 80369-6:2016, Clause 6.5,ISO 80369-6:2016, Clause 6.6
ISO 80369-6 Tests on Female NRFit (Neuraxial Fit)Lock ConnectorISO 80369-6: 2016, Clause 5 (as applicable),ISO 80369-6:2016, Clause 6.1,ISO 80369-6:2016, Clause 6.2,ISO 80369-6:2016, Clause 6.3,ISO 80369-6:2016, Clause 6.4,ISO 80369-6:2016, Clause 6.5,ISO 80369-6:2016, Clause 6.6
Tensile Strength TestISO 8536-9:2015, Clause 5.3
Leak TestISO 8536-9:2015, Clause A.4
Counter Flow TestISO 8536-12:2007+A1:2013, Clause A.4
Blocking Performance TestISO 8536-12:2007+A1:2013, Clause A.6
Opening Pressure TestISO 8536-12:2007+A1:2013, Clause A.7.1and per Baxter Test Method
Particulate Matter TestUSP Chapter <788>
Clamp Activation Force TestPer Baxter Test Method
Clamp Shut-Off TestISO 8536-14:2018, Clause A.1
Non-DEHP Claim VerificationPer Baxter Test Method (as tested inK161808)

Table 4: Performance Data

All tests met the acceptance criteria.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and has a modern look.

Biocompatibility:

Biocompatibility assessments were conducted based on ISO-10993-1. Biological Evaluation of Medical Devices for prolonged contact duration, external communicating device, tissue/bone/dentin, and FDA-2013-D-0350 Guidance for Industry & FDA Staff, Use of ISO 10993-1, "Biological evaluation and medical devices - Part 1: Evaluation and testing within a risk management process, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing", as recommended in the Intravascular Administration Sets guidance, "Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)]". Biocompatibility assessments include the following assays:

  • Cytotoxicity ISO 10993-5
  • Sensitization ISO 10993-10
  • Intracutaneous (Irritation) Reactivity ISO 10993-10
  • Acute Systemic Toxicity ISO 10993-11
  • . Material Mediated Pyrogen ISO 10993-11
  • Genotoxicity ISO 10993-3
  • Subacute Systemic Toxicity ISO 10993-11

Based upon the results the device met the designated ISO 10993-1 categorization and is biocompatible and appropriate for the intended use.

Sterility:

The proposed device is sterilized with gamma radiation. The product is in the bioburden (sub) category "General Sets Labeled 'Sterile''. The Minimum Sterilizing Dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2, "Sterilization of health care products - Radiation-Part 2: Establishing the sterilization dose." The dose setting method used includes, but is not limited to, Method 1 or VDmax. Generally, the MSDs are between 14.2 - 25.0 kGy. The continued validity of the MSD for this (sub) category is confirmed via periodic dose audit studies. In addition, routine periodic pre-sterilization bioburden testing is performed for each (sub) category. The endotoxin limit is 2.15 EU/device per USP <161>.

Shelf Life:

Baxter has performed aging testing to support a shelf-life claim of 3 (three) years.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif and slightly italicized. The word is presented in a clean and professional manner.

CONCLUSION:

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate device that is legally marketed for the same intended use.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.