K Number
K211417
Manufacturer
Date Cleared
2022-02-04

(273 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The F3D-C2 Cervical Stand-Alone System is a Stand-Alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one or two contiguous disc levels depending on the assembly. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The F3D-C2 Cervical Stand-Alone System is used to facilitate intervertebral body fusion in the cervical spine and is placed via an anterior approach at one- or two-disc levels (C2-T1) depending on the assembly. The interior of the spacers can be packed with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft and/or demineralized allograft bone marrow aspirate as an adjunct to fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment.

The F3D-C2 Cervical Stand-Alone System is an interbody fusion device intended to be used with two titanium alloy screws and/or anchors which accompany the implants. When used with screws, the F3D-C2 Cervical Stand-Alone System is intended for use at one or two levels of the cervical spine (C2-T1) and requires no additional fixation. When used with one or more anchors, the F3D-C2 Cervical Stand-Alone System is intended for use at one level of the cervical spine (C2-T1) and requires additional supplemental fixation such as posterior cervical screw fixation.

Device Description

The CoreLink F3D Cervical Stand-Alone Interbody Fusion System is a collection of additively manufactured and machined implants and associated instruments for surgical site preparation and implantation to provide mechanical support to the cervical spine while arthrodesis occurs. The subject cages are additively manufactured from Ti-6Al-4V per ASTM F3001. Integration consists of additive Ti-6Al-4V (ASTM F3001) anchors (also referred to as nails) or machined Ti-6A-4V (ASTM F136 and ISO 5832-3) screws. The spacer and screw components are available in an assortment of dimensional combinations to accommodate the individual anatomic and clinical circumstances of each patient. The basic shape of the spacer is a trapezoidal column to provide surgical stabilization of the spine. The inferior/superior aspects of the spacer incorporate a vertical cavity which can be packed with bone graft.

AI/ML Overview

The provided text describes a medical device, the F3D-C2 Cervical Stand-Alone System, and its FDA 510(k) clearance. However, it does not include information about acceptance criteria, device performance in terms of metrics like sensitivity, specificity, or accuracy, or any study involving human readers or AI.

Therefore, I cannot provide the requested information about acceptance criteria and a study proving the device meets them because such details are not present in the provided document.

The document discusses the regulatory clearance process for a spinal implant and focuses on:

  • Device Description: What the F3D-C2 Cervical Stand-Alone System is made of and its components.
  • Indications for Use: The medical conditions and patient population for which the device is intended.
  • Technological Characteristics: How it compares to predicate devices.
  • Non-Clinical Performance Data: A list of mechanical tests performed to demonstrate its safety and effectiveness compared to predicate devices. These tests are physical and mechanical, not related to AI or diagnostic accuracy.

The performance data mentioned are for mechanical tests such as static axial compression, dynamic axial compression, subsidence, and pullout, all referencing ASTM standards. These are physical properties of the implant, not performance metrics for a diagnostic or AI-driven medical device.

To directly answer your numbered points based only on the provided text, I would have to state that the information is not present for most of them:

  1. A table of acceptance criteria and the reported device performance: Not provided. The text lists types of mechanical tests (e.g., Static axial compression per ASTM F2077) but does not provide specific acceptance criteria values or reported performance values for these tests.
  2. Sample sized used for the test set and the data provenance: Not provided. The mechanical tests typically involve testing multiple units of the device but the sample size is not stated. Data provenance refers to the origin of clinical data, which is not applicable here as these are non-clinical mechanical tests.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These are mechanical tests, not clinical studies involving expert interpretation.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is about a physical medical implant, not an AI or diagnostic device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context you're asking. For mechanical tests, the "ground truth" would be the engineering specifications and established test methods, but not expert consensus or pathology in a clinical sense.
  8. The sample size for the training set: Not applicable. This device is not an AI algorithm requiring a training set.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 4, 2022

CoreLink, LLC % Nathan Wright, MS Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K211417

Trade/Device Name: F3D-C2 Cervical Stand-Alone System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: December 30, 2021 Received: January 5, 2022

Dear Mr. Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K211417

Device Name

F3D-C2 Cervical Stand-Alone System

Indications for Use (Describe)

The F3D-C2 Cervical Stand-Alone System is a Stand-Alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one or two contiguous disc levels depending on the assembly. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The F3D-C2 Cervical Stand-Alone System is used to facilitate intervertebral body fusion in the cervical spine and is placed via an anterior approach at one- or two-disc levels (C2-T1) depending on the assembly. The interior of the spacers can be packed with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft and/or demineralized allograft bone marrow aspirate as an adjunct to fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment.

The F3D-C2 Cervical Stand-Alone System is an interbody fusion device intended to be used with two titanium alloy screws and/or anchors which accompany the implants. When used with screws, the F3D-C2 Cervical Stand-Alone System is intended for use at one or two levels of the cervical spine (C2-T1) and requires no additional fixation. When used with one or more anchors, the F3D-C2 Cervical Stand-Alone System is intended for use at one level of the cervical spine (C2-T1) and requires additional supplemental fixation such as posterior cervical screw fixation.

Type of Use (Select one or both, as applicable) 図 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submitter's Name:CoreLink, LLC
Submitter's Address:2072 Fenton Logistics ParkSt. Louis, MO 63026
Submitter's Telephone:888-349-7808
Contact Person:Nathan Wright MSEmpirical Testing Corp.719-351-0248nwright@empiricaltech.com
Image: EMPIRICAL TESTING CORP logo
Date Summary was Prepared:May 6, 2021
Trade or Proprietary Name:F3D-C2 Cervical Stand-Alone System
Common or Usual Name:Intervertebral Fusion Device with Integrated Fixation, Cervical
Classification:Class II per 21 CFR §888.3080
Product Code:OVE
Classification Panel:Orthopedic

K211417 510(K) SUMMARY

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The CoreLink F3D Cervical Stand-Alone Interbody Fusion System is a collection of additively manufactured and machined implants and associated instruments for surgical site preparation and implantation to provide mechanical support to the cervical spine while arthrodesis occurs. The subject cages are additively manufactured from Ti-6Al-4V per ASTM F3001. Integration consists of additive Ti-6Al-4V (ASTM F3001) anchors (also referred to as nails) or machined Ti-6A-4V (ASTM F136 and ISO 5832-3) screws. The spacer and screw components are available in an assortment of dimensional combinations to accommodate the individual anatomic and clinical circumstances of each patient. The basic shape of the spacer is a trapezoidal column to provide surgical stabilization of the spine. The inferior/superior aspects of the spacer incorporate a vertical cavity which can be packed with bone graft.

INDICATIONS FOR USE

The F3D-C2 Cervical Stand-Alone System is a Stand-Alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one or two contiguous disc levels depending on the assembly. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The F3D-C2 Cervical Stand-Alone System is used to facilitate intervertebral body fusion in the cervical spine and is placed via an anterior approach at one- or two-disc levels (C2-T1) depending on the assembly. The interior of the spacers can be packed with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft and/or demineralized allograft bone with bone marrow aspirate as an adjunct to fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment.

{4}------------------------------------------------

The F3D-C2 Cervical Stand-Alone System is an interbody fusion device intended to be used with two titanium alloy screws and/or anchors which accompany the implants. When used with screws, the F3D-C2 Cervical Stand-Alone System is intended for use at one or two levels of the cervical spine (C2-T1) and requires no additional fixation. When used with one or more anchors, the F3D-C2 Cervical Stand-Alone System is intended for use at one level of the cervical spine (C2-T1) and requires additional supplemental fixation such as posterior cervical screw fixation.

TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:

  • Indications for Use ●
  • Materials of manufacture ●
  • Sizes ●
  • Biocompatibility ●
  • . Mechanical Performance
510k NumberTrade or Proprietary orModel NameManufacturerPredicate Type
K200087F3D Cervical Stand-AloneInterbody Fixation SystemCoreLink, LLCPrimary
K173115MIS COALITION® SpacersGlobus MedicalAdditional
K152793Unison-C Anterior CervicalFixation SystemRTI Surgical, Inc.Additional
K191489Genesys Spine 3DP CervicalInterbody SystemGenesys SpineAdditional

Table 5-1 Predicate Devices

PERFORMANCE DATA

The F3D-C2 Cervical Stand-Alone System has been tested in the following test modes:

  • . Static axial compression per ASTM F2077
  • Static compression shear per ASTM F2077 ●
  • . Static torsion per ASTM F2077
  • . Dynamic axial compression per ASTM F2077
  • . Dynamic compression shear per ASTM F2077
  • . Dynamic torsion per ASTM F2077
  • Subsidence per ASTM F2267 ●
  • Expulsion ●
  • Pullout per ASTM F543 ●

{5}------------------------------------------------

  • Static cantilever bending per ASTM F2193 .
  • Dynamic cantilever bending per ASTM F2193 ●
  • . Anchor backout

The results of this non-clinical testing show that the performance of the F3D-C2 Cervical Stand-Alone System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the F3D-C2 Cervical Stand-Alone System is substantially equivalent to the predicate device.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.