K Number
K211380
Device Name
Elexir
Manufacturer
Date Cleared
2021-07-30

(87 days)

Product Code
Regulation Number
882.5891
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

THE ELEXIR is indicated for the acute treatment of migraine (program 1) and the prophylactic treatment of episodic migraine (Program 2) in patients 18 years of age or older

Device Description

The ELEXIR device is a transcutaneous electrical nerve stimulator (TENS) that is applied to the forehead using a self-adhesive electrode positioned over the upper branches of the trigeminal verve bilaterally. It is intended to stimulate the upper branches of the trigeminal verve in order to reduce the frequency of migraine attack.

AI/ML Overview

The provided text does not contain specific acceptance criteria or performance data for the ELEXIR device itself. Instead, it states that the device's safety and effectiveness are supported by non-clinical (biocompatibility, electrical safety, performance, usability, software) tests and a clinical evaluation report based on data from the predicate device, CEFALY® Dual. The document claims that the ELEXIR device is "not applied with clause b.6) of 21CFR882.5981," which likely refers to a requirement for clinical performance data for devices to treat headache. However, it then states that a clinical evaluation report was prepared using collected clinical data of the predicate device, concluding that the subject device is safe and effective.

Therefore, I cannot provide a table of acceptance criteria and reported device performance from the provided text, nor can I answer questions 2, 3, 4, 5, 6, 7, 8, and 9 for the subject device (ELEXIR). The document explicitly mentions that clinical performance data for the subject device are not available and that its safety and effectiveness are inferred from the predicate device's data.

Here's a summary of what can be extracted:

  • Acceptance Criteria for Non-Clinical Tests: The acceptance criteria for the non-clinical tests are implicitly the successful fulfillment of the listed international and FDA guidance standards.
  • Study Proving Device Meets Acceptance Criteria (Non-Clinical): Various non-clinical tests were performed.

1. A table of acceptance criteria and the reported device performance

The document does not provide specific numerical acceptance criteria or performance results for the ELEXIR device in a table format. It states that non-clinical tests were performed in accordance with various standards, implying that meeting these standards constitutes the acceptance criteria.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not provided for ELEXIR. For the clinical evaluation, data was collected from the predicate device. No details on its sample size or provenance are given for the predicate device's clinical data, other than it was "collected clinical data."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no direct clinical study data for the subject device (ELEXIR) is presented. The clinical evaluation relies on data from the predicate device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no direct clinical study data for the subject device (ELEXIR) is presented.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a transcutaneous electrical nerve stimulator, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a transcutaneous electrical nerve stimulator.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For the clinical evaluation, the safety and effectiveness for headache treatment were established using "collected clinical data" of the predicate device. The specific type of "ground truth" (e.g., patient-reported outcomes, physician diagnosis, etc.) for the predicate device's data is not detailed in this document.

8. The sample size for the training set

Not applicable, as this is not an AI/ML device that requires a training set. If referring to the predicate device's clinical data used for evaluation, no sample size is provided.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/ML device. If referring to the predicate device's clinical data, the method for establishing ground truth is not specified.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 30, 2021

Nu Eyne Co., Ltd. Dong Lee Manager #608, 28, Digital-ro 30-gil, Guro-gu Seoul, 08389 Korea. South

Re: K211380

Trade/Device Name: Elexir Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous electrical nerve stimulator to treat headache Regulatory Class: Class II Product Code: PCC Dated: April 28, 2021 Received: May 4, 2021

Dear Dong Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Patrick Antkowiak Assistant Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected network of blue shapes resembling molecules or connections, with two dark blue circles within the network. Below the network graphic, the company name "Nu Eyne" is written in a modern font, with "Nu" in a darker blue and "Eyne" in a lighter blue. Underneath the company name, the text "A MEDICAL R&D COMPANY" is written in smaller, all-caps letters.

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: 06/30/2023See PRA Statement below.
Indications for Use

510 K211380

Device Name ELEXIR (MODEL: ALLIVE2)

Indications for Use (Describe)

THE ELEXIR is indicated for the acute treatment of migraine (program 1) and the prophylactic treatment of episodic migraine (Program 2) in patients 18 years of age or older

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6744

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NuEyene, a medical R&D company. The logo features a stylized, interconnected network of blue shapes resembling molecules or connections. Above the company name are two dark blue circles. The company name, "Nu Eyne," is written in a modern font, with "Nu" in a darker shade of blue and "Eyne" in a lighter shade. Below the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.

510(k) SUMMARY

This summary of 510(k) -safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: APRIL, 28, 2021

1. INFORMATION

1.1 Submitter Information

  • Submitter Name: Nu Eyne Co., Ltd.
  • . Address
    • : #608, 28, Digital-ro 30-gil, Guro-gu, Seoul, 08389, Republic of Korea
  • Telephone Number: +82-2-6953-8120 . Fax: +82-303-3447-0017
  • 트 Email: dongdeong.lee@nueyne.com

1.2 Contact Person

  • . Name: Dong Seong Lee (Manager / Nu Eyne Co.,Ltd.)
  • Address: #608, 28, Digital-ro 30-gil, Guro-gu, Seoul, 08389, Republic of Korea

Fax: +82-303-3447-0017

2. DEVICE INFORMATION

  • 2.1 Trade Name / Proprietary Name: ELEXIR (Model: ALLIVE2)
  • 2.2 Common Name: Transcutaneous electrical nerve stimulator to treat headache
  • 2.3 Classification Name: Stimulator, Nerve, Electrical, Transcutaneous, For Migraine
  • 2.4 Product Code: PCC
  • 2.5 Classification Regulation: 21CFR 882.5891
  • 2.6 Device Class: Class II (Special Control)
  • 2.7 Classification Panel: Neurology

3. PREDICATE DEVICE

Predicate Device
ManufacturerCEFALY Technology
Device Name (Trade Name)CEFALY® Dual
510(k) NumberK173006

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected network of blue shapes above the company name. The name "Nu" is in a darker blue, while "Eyne" is in a lighter blue, creating a visual distinction. Below the name, the words "A MEDICAL R&D COMPANY" are written in smaller, gray font.

4. SUBJECT DEVICE DESCRIPTION

The ELEXIR device is a transcutaneous electrical nerve stimulator (TENS) that is applied to the forehead using a self-adhesive electrode positioned over the upper branches of the trigeminal verve bilaterally. It is intended to stimulate the upper branches of the trigeminal verve in order to reduce the frequency of migraine attack.

5. INTENDED USE

The ELEXIR is indicated for the acute treatment of migraine (program 1) and the prophylactic treatment of episodic migraine (program 2) in patients 18 years of age or older.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for NuEyne, a medical R&D company. The logo features a stylized, interconnected blue design resembling a network or neural connections. Below the logo, the company name "NuEyne" is written in a dark blue, sans-serif font, with the "E" stylized to resemble an eye. Underneath the company name, the text "A MEDICAL R&D COMPANY" is written in a smaller, sans-serif font.

6. SUBSTANTIAL EQUIVALENCE

ItemsSubject DevicePredicate DeviceComparison Result
ManufacturerNu Eyne Co., Ltd.CEFALY TechnologyDifferent
DeviceStimulator, Nerve, Electrical,Transcutaneous, For MigraineStimulator, Nerve, Electrical,Transcutaneous, For MigraineSame
Trade/Device NameELEXIR/ALLIVE2CEFALY® DualDifferent
K NumberNoneK173006Different
Regulation Number21CFR 882.589121CFR 882.5891Same
Regulation DescriptionTranscutaneous electrical nervestimulator to treat headacheTranscutaneous electrical nervestimulator to treat headacheSame
Regulatory ClassClass IIClass IISame
Product CodePCCPCCSame
DefinitionUsed to apply an electrical currentto a patient's cranium throughelectrodes placed on the skin.Used to apply an electrical currentto a patient's cranium throughelectrodes placed on the skin.Same
Review PanelNeurologyNeurologySame
Physical StateElectrical stimulation unit withleads and cutaneous electrodes.Electrical stimulation unit withleads and cutaneous electrodes.Same
Technical MethodApplies an electrical currentthrough electrodes on patient'sskin.Applies an electrical currentthrough electrodes on patient's skin.Same
Target AreaAfferent cranial nerves.Afferent cranial nerves.Same
Intended useThe indications for use of theELEXIR are:- The acute treatment of migrainewith or without aura in 18 years ofage or older;- The prophylactic treatment ofepisodic migraine in patients 18years of age or older.The indications for use of theCefaly® Dual are:- The acute treatment of migrainewith or without aura in 18 years ofage or older;- The prophylactic treatment ofepisodic migraine in patients 18years of age or older.Same
PictureImage: Drawing of a person with a device on their foreheadImage: Photo of a person with a device on their foreheadDifferent
Power Source1rechargeable Lipo 3.7 V battery1rechargeable Lipo 3.7 V batterySame
Channels11Same
ComputerizedYesYesSame
S/W provided2 fixed program:- 1 fixed program for the acutetreatment of migraine attacks(Program 1)2 fixed program:- 1 fixed program for the acutetreatment of migraine attacks(Program 1)Same
- 1 fixed program for prophylactic- 1 fixed program for prophylactictreatment of migraine (Program 2)treatment of migraine (Program 2)
Constant currentYesYesSame
Constant voltageNoNoSame
Max output current16mA16mASame
Max output voltage (2kOhm)32 Volts32 VoltsSame
Patient Override Control MethodOn/Off buttonOn/Off buttonSame
Max Leakage CurrentNone (battery operated)None (battery operated)Same
ElectrodeSelf-adhesiveSelf-adhesiveSame
Indicator display: Unit functioningYesYesSame
Low battery indicatorYesYesSame
StandardsIEC 60601-1IEC 60601-1-2IEC 60601-1-6IEC 60601-1-11IEC 60601-2-10IEC 62366-1IEC 62304ISO 10993-1ISO 10993-5ISO 10993-10IEC 60601-1IEC 60601-1-2IEC 60601-1-6IEC 60601-1-11IEC 60601-2-10IEC 62366-1IEC 62304ISO 10993-1ISO 10993-5ISO 10993-10Same
Timer SettingYesYesSame
Weight20.71g12gDifferent
DeviceDimensions60.00mm x 44.00mm x 17.60mm55mm x 35mm x 15mmDifferent
Expected Service Life2 years7 yearsDifferent

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Nu Eyne, a medical R&D company. The logo features a stylized blue symbol resembling interconnected nodes or molecules, with two dark blue circles within the design. Below the symbol, the company name "Nu Eyne" is written in a modern, sans-serif font, with the tagline "A MEDICAL R&D COMPANY" appearing in smaller letters underneath.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected network of blue shapes resembling molecules or connections. Below the graphic is the company name, "Nu Eyne," in a modern, sans-serif font, with the first letter in a slightly darker shade of blue. Underneath the company name, in smaller letters, is the tagline "A MEDICAL R&D COMPANY."

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains a logo for a medical research and development company. The logo features a stylized, interconnected blue design resembling a network or neural connections above the company name. The company name, "Nu Eyne," is written in a modern, sans-serif font, with the first letter in uppercase and the rest in lowercase. Below the company name, the text "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.

IP ClassificationIP22IP22Same
Electrical ProtectionType BFType BFSame
Battery TypeLithium ion BatteryLithium ion BatterySame
BatteryExpected Service Life2 years (75Cycles of completedischarge)7 years (300Cylces of completedischarge)Different
Maximum input voltage(USB connector)5.25 Vdc5.25 VdcSame
ElectrodeDimensions90mm (W) x 31.9mm (H)94mm (W) x 30mm (H)Different
Expected Service Life20 times20 timesSame
Special Requirement b.3) in accordance with 21CFR 882.5891
WaveformBiphasicBiphasicSame
Maximum output voltage(V)500 ohms88Same
2,000 ohms3232Same
10,000 ohms6560Same
Maximum output current(mA)500 ohms1616Same
2,000 ohms1616Same
10,000 ohms6.56Same
Pulse duration (µS)505505Same
FrequencyProgram 1100Hz fixedProgram 1100Hz fixedSame
Program 260Hz fixedProgram 260Hz fixed
Net charge (µC) per pulse00Same
Duration of the primary (depolarizing) phase250250Same

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for NuEyene, a medical R&D company. The logo features a stylized blue symbol resembling interconnected nodes, possibly representing neural connections or networks. Below the symbol, the company name "NuEyene" is written in a modern font, with the "N" in a darker blue and the rest of the name in a lighter shade. Underneath the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.

(µS)
Standby duration between the two phase (µS)55Same
Maximum current density(mA/cm2, r.m.s.)500 ohms2.372.37Same
Maximum average power density(W/cm2)500 ohms0.0000170.000017Same
Maximum average current(Average absolute value, mA)500 ohms0.480.48Same
Other Technical Item
Wave ShapeRectangularFull compensatedSymmetricalRectangularFull compensatedSymmetricalSame
MethodSame charge quantity on positiveand negative impulseSame charge quantity on positiveand negative impulseSame
Max phase amplitude16mA with a load of a 4.7 µFcapacitor parallel with 2.2K ohmsresistance16mA with a load of a 4.7 µFcapacitor parallel with 2.2K ohmsresistanceSame
Phase rise time2 µS2 µSSame
Phase decay time2 µS2 µSSame
Interphase intervalnilnilSame
Burst modeNoNoSame
Maximum Phase Charge (µS) at 500 ohms44Same
Amplitude0 to 16 mA FixedSame
Modulation optionFrequency
Program 1 100 Hz fixed Program 2 60Hz fixedProgram 1 100 Hz fixed Program 2 60Hz fixed

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for NuEyene, a medical R&D company. The logo features a stylized, interconnected blue design above the company name. Below the name, the text "A MEDICAL R&D COMPANY" is written in smaller, capitalized letters.

Duration
250 µS250 µSSame
Ramp ModulationRamp up14 min14 min.Same
Ramp down1 min1 minSame
Steady timeProgram 145 minProgram 145 minSame
Program 25 minProgram 25 min
Session DurationProgram 160 minProgram 160 minSame
Program 220 minProgram 220 min.
Amplitude modulationAmplitude is adjusted by the userAmplitude is adjusted by the userSame
Material
Device housing materialsPlastic ABSPlastic ABSSame
Electrode top layerPolyethylene terephthalatePolyurethaneDifferent
Electrode intermediate layerConductive silver carbon inkCarbon with silver coatingDifferent
Electrode bottom layerxxSame
Central pinN/AN/ASame
Metallic parts for magnetic attractionTinplateTinplateSame

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized graphic of interconnected circles in a light blue color, resembling a network or molecular structure. Below the graphic, the company name "Nu Eyne" is written in a modern font, with "Nu" in a darker blue and "Eyne" in a lighter blue. Underneath the name, the text "A MEDICAL R&D COMPANY" is written in smaller, sans-serif font.

7. NON-CLINICAL DATA

7.1 Safety Test

  1. Biocompatibility

The biocompatibility tests were performed to protect patients from undue risks arise from biological hazards associated with materials of manufacture and final device. The tests were performed in accordance with the following standards and FDA Guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

No.Test ItemsStandards
1CytotoxicityISO 10993-5:2009
2SensitizationISO 10993-10:2010
3Intracutaneous Reactivity TestISO 10993-10:2010

2) Electrical Safety and EMC

The electrical safety tests were performed to protect patients from undue risks arise from any

hazards associated with final device. The tests were performed in accordance with the following standards.

No.Test ItemsStandards
1General requirement for basic safety andessential performance- IEC 60601-1:2005+A1: 2012(AAMI/ANSI ES 60601-1: 2005+A1: 2012)
2General requirement for safety -Electromagnetic disturbances- IEC 60601-1-2:2014
3General requirement for safety - Medicalelectrical equipment used in the homehealthcare environment- IEC 60601-1-11:2015 and- FDA Guidance ("Design Considerations forDevices Intended for Home Use")
4Particular requirement for safety – Nerveand muscle stimulators- IEC 60601-2-10:2012/Amd1:2016

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected network of blue shapes resembling molecules or connections, positioned above the company name. The text "Nu" is in a dark blue, blocky font, while "Eyne" is in a lighter blue, more stylized font. Below the company name, in smaller font, is the text "A MEDICAL R&D COMPANY".

7.2 Performance Test

The following tests were performed to assess effectiveness of performance of the device. The tests were performed in accordance with following standards.

No.Test ItemsStandards
1Particular requirement for safety - Nerveand muscle stimulators- IEC 60601-2-10:2012/Amd1:2016
2Technical Test- IEC 60601-2-10:2012/Amd1:2016

7.3 Usability V&V

The following tests were performed to assess effectiveness of usability of the device. The test was performed in accordance with following standards.

No.Test ItemsStandards
1General requirement for safety – Usability- IEC 60601-1-6:2013- IEC 62366-1:2015 and- FDA Guidance (“Applying Human Factors andUsability Engineering to Medical Devices”)

7.4 Software

The following tests were performed to assess effectiveness of software of the device. The test

was performed in accordance with following standards.

No.Test ItemsStandards
1General requirement for safety -Programmable electrical medical systems(PEMS)- IEC 62304:2006/Amd1:2015- FDA Guidance ("Guidance for the Content ofPremarket Submissions for Software Contained inMedical Devices")

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features the company name in two lines, with "Nu" in a darker blue and "Eyne" in a lighter blue. Above the name is a stylized graphic of interconnected circles, also in the lighter blue color. Below the name, in smaller font, is the text "A MEDICAL R&D COMPANY".

8. CLINICAL DATA

Although clinical performance data are required to demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population on the special control according to clause b.6) of 21CFR882.5981, we consider that the subject device (ELEXIR) is not applied with clause b.6) of 21CFR882.5981.

Although there are not the clinical performance data of the subject device, we prepare the clinical evaluation report by using the collected clinical data of the predicate device.

In this evaluation report, the subject device is safe and effective as a treatment for headache in the indicated patient population.

The clinical evaluation was performed in accordance with following standards.

No.Test ItemsStandards
1Clinical Evaluation- MEDDEV 2.7.1. rev.4

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains the logo for "Nu Eyne, A Medical R&D Company". The logo features a stylized blue graphic above the company name. The graphic appears to be a network of connected nodes, possibly representing a biological or technological network. The text "Nu Eyne" is in a modern sans-serif font, with "Nu" in a darker shade of blue and "Eyne" in a lighter shade. Below the company name, the text "A MEDICAL R&D COMPANY" is printed in a smaller, simpler font.

9. CONCLUSION

Under the comparing substantial equivalence between the subject device and the predicate device, there are the same points such as general information, some technical and material information. Although there are some differences, the safety and performance test reports are supported to the safety and effectiveness of the subject device.

In this regard, we conclude that the subject device is substantially equivalent to the predicate device.

§ 882.5891 Transcutaneous electrical nerve stimulator to treat headache.

(a)
Identification. A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.