K Number
K211098
Device Name
Breast Volume Navigator (BVN) Model G-2000
Manufacturer
Date Cleared
2021-05-26

(43 days)

Product Code
Regulation Number
892.1560
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Breast Volume Navigator (BVN) is an add-on accessory for existing ultrasound imaging systems and is intended to control position and movement of ultrasound transducers for the systematic acquisition of 2-dimensional image slices throughout a volume of interest. The BVN is intended to acquire, analyze, store and retrieve digital ultrases for computerized 3-dimensional image processing. The BVN is intended to assist health care providers trained in breast ultrasound with tools for electromagnetic tracking of instruments with respect of breast ultrasound images generated from FDA cleared hand-held ultrasound devices. The BVN is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast when used with a linear array transducer. The BVN is indicated for use as an adjunct to hand-held breast ultrasound to assist the health care providers trained in breast ultrasound in their scanning process. The scanning paths are displayed on a route map and provide quality control of scanning to provide an overall observation of scanning process. The BVN Software Application is intended as a standalone software device installed on a windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. The BVN Software Application provides the capability to visualize ultrasound images along with the scanning paths and position information of the probe that is stored in the DICOM file in advance. The BVN will allow exporting to any third-party application that has the appropriate level of DICOM compliance. The BVN is intended as a general-purpose digital 3D breast ultrasound image processing tool for radiology and surgery. The device is not intended to be used in the environment of strong magnetic fields, such as in Magnetic Resonance Imaging (MRI) room. The device is not intended to be used as a replacement for screening mammography.
Device Description
The Breast Volume Navigator (BVN™) Model G-2000, hereinafter maybe referred to as the BVN™ G-2000 System or the BVN™ System, comprises hardware components and a software element, including the following components: a magnetic position tracking device, sensor attaching pieces used to attach magnetic sensors to the skin and ultrasound probe, a central control unit computer, and software for controlling the system, collecting, and processing ultrasound images and positional data, and performing automated annotations. The BVN™ System has a touch-screen user interface and push-button for power on the system. The Use Interface (touchscreen display and central control unit computer) are placed on a medical grade cart that can be positioned next to a standard adjustable examination table for ease of use and ergonomic adaptation. The BVN™ System has a USB port available for transferring files via USB Memory Stick. The BVN™ System has an Ethernet port for connection to a PACS system, using DICOM. The BVN™ System has a VGA/DVI Input Ports for capturing images from an ultrasound imaging scanner. The BVN™ System receives ultrasound DICOM images from the US machine via the network connection and telemetry data from a position tracking system. The BVN™ automatically detects when the image is being frozen on the US machine and takes a snapshot of the telemetry data at that time. Later, when the BVN™ receives the DICOM image, it associates the telemetry data to the image from the time when the image was frozen on the US machine. A Completeness-of-Scan Assessment ensures that all breast tissue has been sufficiently covered, the BVN™ System will achieve this by tracking the real-time position of the probe relative to the breast and chest wall to ensure the entire breast volume has been covered with sufficient detail to allow the detection of sub centimeter lesions. As the probe is swept over the breast region, the technician will be provided a completeness map highlighting the portions of the breast that have not been sufficiently scanned. The BVN™ System will provide the operator a display of regions with insufficient coverage, prompting the operator to rescan these regions. The BVN™ System can retrieve previous examinations DICOM images allowing the user to enter positional data from another/previous ultrasound examination for follow-up purposes. Any anomalies discovered during the follow-up process can be evaluated using the localization and measurement tools included in the software. The customer's existing ultrasound probe securely attaches to the BVN™ System probe sensor. During a scan, the operator applies constant pressure to the transducer against the patient's breast tissue and can rotate the transducer (pitch and roll) to accommodate for the physical characteristics of the breast. Exam data is subsequently reviewed on standard radiological viewing stations.
More Information

No
The summary describes image processing, tracking, and navigation functionalities, but does not mention AI or ML.

No
The device is described as an accessory for imaging systems, intended to control transducer movement for systematic acquisition of 2D image slices, analyze, store, and retrieve digital ultrasounds for 3D image processing. Its purpose is to assist healthcare providers with tools for tracking instruments with respect to breast ultrasound images and to aid in scanning processes and image review. The device's primary function is for imaging and image processing, not for treating any condition or disease.

Yes

The "Intended Use / Indications for Use" section explicitly states that the BVN is intended to "assist health care providers trained in breast ultrasound with tools for electromagnetic tracking of instruments with respect of breast ultrasound images" and "is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast". This indicates its purpose is to aid in the assessment and imaging of breast tissue for diagnostic purposes, rather than treatment. Additionally, it mentions assisting in the "detection of sub centimeter lesions" and allowing for the evaluation of "anomalies discovered during the follow-up process," which are diagnostic functions.

No

The device description explicitly states that the system "comprises hardware components and a software element," listing specific hardware components like a magnetic position tracking device, sensors, and a central control unit computer.

Based on the provided information, the Breast Volume Navigator (BVN) is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information about a person's health. This testing is performed outside of the living body (in vitro).
  • BVN's Function: The BVN is an accessory for ultrasound imaging systems. It assists in the acquisition, processing, and visualization of ultrasound images of the breast. It does not analyze biological specimens taken from the patient.
  • Intended Use: The intended use clearly states that the BVN is for controlling transducer position, acquiring and processing ultrasound images, and assisting healthcare providers with scanning and reviewing breast ultrasound images. This is all related to in vivo imaging.
  • Device Description: The device description details hardware and software components for tracking transducer position, collecting image and positional data, and processing this information for visualization and analysis. Again, this is focused on imaging, not laboratory testing of specimens.

Therefore, the BVN falls under the category of a medical imaging device accessory, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Breast Volume Navigator (BVN) is an add-on accessory for existing ultrasound imaging systems and is intended to control position and movement of ultrasound transducers for the systematic acquisition of 2-dimensional image slices throughout a volume of interest. The BVN is intended to acquire, analyze, store and retrieve digital ultrases for computerized 3-dimensional image processing.

The BVN is intended to assist health care providers trained in breast ultrasound with tools for electromagnetic tracking of instruments with respect of breast ultrasound images generated from FDA cleared hand-held ultrasound devices. The BVN is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast when used with a linear array transducer.

The BVN is indicated for use as an adjunct to hand-held breast ultrasound to assist the health care providers trained in breast ultrasound in their scanning process. The scanning paths are displayed on a route map and provide quality control of scanning to provide an overall observation of scanning process.

The BVN Software Application is intended as a standalone software device installed on a windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. The BVN Software Application provides the capability to visualize ultrasound images along with the scanning paths and position information of the probe that is stored in the DICOM file in advance.

The BVN will allow exporting to any third-party application that has the appropriate level of DICOM compliance. The BVN is intended as a general-purpose digital 3D breast ultrasound image processing tool for radiology and surgery.

The device is not intended to be used in the environment of strong magnetic fields, such as in Magnetic Resonance Imaging (MRI) room.

The device is not intended to be used as a replacement for screening mammography.

Product codes (comma separated list FDA assigned to the subject device)

IYO, ITX, LLZ

Device Description

The Breast Volume Navigator (BVN™) Model G-2000, hereinafter maybe referred to as the BVN™ G-2000 System or the BVN™ System, comprises hardware components and a software element, including the following components: a magnetic position tracking device, sensor attaching pieces used to attach magnetic sensors to the skin and ultrasound probe, a central control unit computer, and software for controlling the system, collecting, and processing ultrasound images and positional data, and performing automated annotations.

The BVN™ System has a touch-screen user interface and push-button for power on the system.

The Use Interface (touchscreen display and central control unit computer) are placed on a medical grade cart that can be positioned next to a standard adjustable examination table for ease of use and ergonomic adaptation.

The BVN™ System has a USB port available for transferring files via USB Memory Stick.

The BVN™ System has an Ethernet port for connection to a PACS system, using DICOM.

The BVN™ System has a VGA/DVI Input Ports for capturing images from an ultrasound imaging scanner.

The BVN™ System receives ultrasound DICOM images from the US machine via the network connection and telemetry data from a position tracking system. The BVN™ automatically detects when the image is being frozen on the US machine and takes a snapshot of the telemetry data at that time. Later, when the BVN™ receives the DICOM image, it associates the telemetry data to the image from the time when the image was frozen on the US machine.

A Completeness-of-Scan Assessment ensures that all breast tissue has been sufficiently covered, the BVN™ System will achieve this by tracking the real-time position of the probe relative to the breast and chest wall to ensure the entire breast volume has been covered with sufficient detail to allow the detection of sub centimeter lesions. As the probe is swept over the breast region, the technician will be provided a completeness map highlighting the portions of the breast that have not been sufficiently scanned. The BVN™ System will provide the operator a display of regions with insufficient coverage, prompting the operator to rescan these regions. The BVN™ System can retrieve previous examinations DICOM images allowing the user to enter positional data from another/previous ultrasound examination for follow-up purposes. Any anomalies discovered during the follow-up process can be evaluated using the localization and measurement tools included in the software.

The customer's existing ultrasound probe securely attaches to the BVN™ System probe sensor. During a scan, the operator applies constant pressure to the transducer against the patient's breast tissue and can rotate the transducer (pitch and roll) to accommodate for the physical characteristics of the breast.

Exam data is subsequently reviewed on standard radiological viewing stations.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Ultrasound

Anatomical Site

breast

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Intended User: health care providers trained in breast ultrasound, qualified medical imaging personnel that are familiar with traditional breast ultrasound procedures and techniques, physicians
Care Setting: ultrasound examinations and procedures

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Study Type: Design Validation Tests - Non-Clinical Testing
Sample Size: Not specified (testing using phantoms)
Key Results:

  1. Mapping the accuracy and precision of the BVN™ System position tracking capability within its sensing volume using a Volume Mapping Performance Test Phantom with 9 test locations and 3 layers above the test phantom:
    • Linear Accuracy and Linear Precision of Distance to Nipple: accuracy and precision of ≤ 5 mm for measuring the distance of a target in a breast to nipple of the same breast.
    • Accuracy and Precision of Clock Face Angle: accuracy and precision of ≤ 5 degrees for measuring the Clock Face Angle of a target in a breast.
    • Accuracy and Precision of Patient Body Angles: accuracy and precision of ≤ 5 degrees for measuring the angles of the patient body relative to the examination bed in all three anatomical planes, i.e., Coronal, Transverse, and Sagittal planes.
    • Accuracy and Precision of Linear Accuracy: accuracy and precision of ≤ 1 mm for measuring the linear accuracy between two points in the ultrasound image at one standard deviation.
  2. The completeness of scan accuracy has been tested using a Rigid Test Phantom:
    • Completeness of Scanning Breast Contour Accuracy: BVN™ accurately captured and represented the hand drawn breast contour with a maximum positional error of +/- 5 mm in the X-Y plane.
    • BVN™ correctly calculated the overlap between adjacent scanning bands and identified the occurrence of when this overlap is less than the set threshold: BVN™ was able to discern when overlaps of adjacent bands are less than a set threshold within a range of 5 to 30 mm ± 2 mm of overlap.
    • 3D Point Coordinates of Image Corners Velocity: BVN™ detected 3D Point coordinates of image corners velocity above and below a set threshold within a range of 15 to 50 mm/s +/- 10% at 1 standard deviation as to ensure breast tissue is not missed between images.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Linear Accuracy and Linear Precision of Distance to Nipple: accuracy and precision of ≤ 5 mm
Accuracy and Precision of Clock Face Angle: accuracy and precision of ≤ 5 degrees
Accuracy and Precision of Patient Body Angles: accuracy and precision of ≤ 5 degrees
Accuracy and Precision of Linear Accuracy: accuracy and precision of ≤ 1 mm at one standard deviation
Completeness of Scanning Breast Contour Accuracy: maximum positional error of +/- 5 mm in the X-Y plane.
Overlap detection: BVN™ was able to discern when overlaps of adjacent bands are less than a set threshold within a range of 5 to 30 mm ± 2 mm of overlap.
3D Point Coordinates of Image Corners Velocity: detected above and below a set threshold within a range of 15 to 50 mm/s +/- 10% at 1 standard deviation.

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K141870

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K171309, K171709

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

0

May 26, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

MetriTrack, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114

Re: K211098

Trade/Device Name: Breast Volume Navigator (BVN™) Model G-2000 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX, LLZ Dated: May 11, 2021 Received: May 12, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

1

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K211098

Device Name Breast Volume Navigator (BVN™) Model G-2000

Indications for Use (Describe)

The Breast Volume Navigator (BVN) is an add-on accessory for existing ultrasound imaging systems and is intended to control position and movement of ultrasound transducers for the systematic acquisition of 2-dimensional image slices throughout a volume of interest. The BVN is intended to acquire, analyze, store and retrieve digital ultrases for computerized 3-dimensional image processing.

The BVN is intended to assist health care providers trained in breast ultrasound with tools for electromagnetic tracking of instruments with respect of breast ultrasound images generated from FDA cleared hand-held ultrasound devices. The BVN is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast when used with a linear array transducer.

The BVN is indicated for use as an adjunct to hand-held breast ultrasound to assist the health care providers trained in breast ultrasound in their scanning process. The scanning paths are displayed on a route map and provide quality control of scanning to provide an overall observation of scanning process.

The BVN Software Application is intended as a standalone software device installed on a windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. The BVN Software Application provides the capability to visualize ultrasound images along with the scanning paths and position information of the probe that is stored in the DICOM file in advance.

The BVN will allow exporting to any third-party application that has the appropriate level of DICOM compliance. The BVN is intended as a general-purpose digital 3D breast ultrasound image processing tool for radiology and surgery.

The device is not intended to be used in the environment of strong magnetic fields, such as in Magnetic Resonance Imaging (MRI) room.

The device is not intended to be used as a replacement for screening mammography.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) SummaryBreast Volume Navigator (BVN™) Model G-2000
510(k) Number:K211098
Date Prepared:February 5th, 2021
Manufacturer:MetriTrack, Inc.
Address:4415 W. Harrison Street, Suite 243
Hillside, IL 60162
Primary Contact Person:Mirela Wohlford
Director of Operations and Compliance
MetriTrack, Inc.
Tel. (708) 498-3578
Email: mirela@metritrack.com
Secondary Contact Person:Calin Caluser
CEO
MetriTrack, Inc.
Tel. (708) 498-3578
Email: calin@metritrack.com
Name of Device:Breast Volume Navigator (BVN™) Model G-2000
Classification Name:Ultrasonic pulsed echo imaging system
Classification Regulation:21 CFR 892.1560 Ultrasonic pulsed echo imaging system
21 CFR 892.1570 Diagnostic ultrasonic transducer
21 CFR 892.2050 Medical image management and
processing system
Classification Panel:Radiology
Device Class:II
Primary Product CodeIYO, ITX, LLZ

Predicate Devices:

Primary predicate: MetriTrack, LLC Breast Volume Navigator (BVN™) (K141870) Reference device: TaiHao Medical Inc. BR-FHUS Navigation 1.0 (K171309) Reference device: TaiHao Medical Inc. BR-FHUS Viewer (K171709)

Indications for Use / Intended Use: The Breast Volume Navigator (BVN™) is an add-on accessory for existing ultrasound imaging systems and is intended to control position and movement of ultrasound transducers for the systematic acquisition of 2-dimensional image slices throughout a volume of interest. The BVN™ is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing.

4

The BVN™ is intended to assist health care providers trained in breast ultrasound with tools for electromagnetic tracking of instruments with respect of breast ultrasound images generated from FDA cleared hand-held ultrasound devices.

The BVN™ is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast when used with a linear array transducer.

The BVN™ is indicated for use as an adjunct to hand-held breast ultrasound to assist the health care providers trained in breast ultrasound in their scanning process. The scanning paths are displayed on a route map and provide quality control of scanning to provide an overall observation of scanning process.

The BVN™ Software Application is intended as a standalone software device installed on a windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. The BVN™ Software Application provides the capability to visualize ultrasound images along with the scanning paths and position information of the probe that is stored in the DICOM file in advance.

The BVN™ will allow exporting to any third-party application that has the appropriate level of DICOM compliance.

The BVN™ is intended as a general-purpose digital 3D breast ultrasound image processing tool for radiology and surgery.

The device is not intended to be used in the environment of strong magnetic or electromagnetic fields, such as in Magnetic Resonance Imaging (MRI) room.

The device is not intended to be used as a replacement for screening mammography.

Technological Characteristics

The Breast Volume Navigator (BVN™) Model G-2000, hereinafter maybe referred to as the BVN™ G-2000 System or the BVN™ System, comprises hardware components and a software element, including the following components: a magnetic position tracking device, sensor attaching pieces used to attach magnetic sensors to the skin and ultrasound probe, a central control unit computer, and software for controlling the system, collecting, and processing ultrasound images and positional data, and performing automated annotations.

The BVN™ System has a touch-screen user interface and push-button for power on the system.

The Use Interface (touchscreen display and central control unit computer) are placed on a medical grade cart that can be positioned next to a standard adjustable examination table for ease of use and ergonomic adaptation.

5

The BVN™ System has a USB port available for transferring files via USB Memory Stick.

The BVN™ System has an Ethernet port for connection to a PACS system, using DICOM.

The BVN™ System has a VGA/DVI Input Ports for capturing images from an ultrasound imaging scanner.

The BVN™ System receives ultrasound DICOM images from the US machine via the network connection and telemetry data from a position tracking system. The BVN™ automatically detects when the image is being frozen on the US machine and takes a snapshot of the telemetry data at that time. Later, when the BVN™ receives the DICOM image, it associates the telemetry data to the image from the time when the image was frozen on the US machine.

A Completeness-of-Scan Assessment ensures that all breast tissue has been sufficiently covered, the BVN™ System will achieve this by tracking the real-time position of the probe relative to the breast and chest wall to ensure the entire breast volume has been covered with sufficient detail to allow the detection of sub centimeter lesions. As the probe is swept over the breast region, the technician will be provided a completeness map highlighting the portions of the breast that have not been sufficiently scanned. The BVN™ System will provide the operator a display of regions with insufficient coverage, prompting the operator to rescan these regions. The BVN™ System can retrieve previous examinations DICOM images allowing the user to enter positional data from another/previous ultrasound examination for follow-up purposes. Any anomalies discovered during the follow-up process can be evaluated using the localization and measurement tools included in the software.

The customer's existing ultrasound probe securely attaches to the BVN™ System probe sensor. During a scan, the operator applies constant pressure to the transducer against the patient's breast tissue and can rotate the transducer (pitch and roll) to accommodate for the physical characteristics of the breast.

Exam data is subsequently reviewed on standard radiological viewing stations.

Pursuant to 809.92(a)(6), MetriTrack, Inc. claims that the Breast Volume Navigator (BVNTM) Model G-2000 system is substantially equivalent to the devices previously cleared by FDA in K141871, K171309, and K171709 MetriTrack, Inc. claims this equivalence because the proposed devices have an equivalent intended use, manufacturing materials, operating principles, and physical and operational specifications compared to the predicate devices:

  • lt has the same intended use as the predicate devices (ultrasound . image capture).
  • lt has the same technological characteristics as the predicate . devices (use of magnetic sensors to track spatial relationships).
  • The BVN™ System and the predicated devices listed above are ● accessories to an Ultrasonic Pulsed Echo Imaging System that have a Moderate Level of Concern.

6

The MetriTrack BVN™ G-2000 Software Application is Safety Class B according to ANSI/AAMI/IEC 62304 Ed. 1.1:2015 Medical device software - Software life cycle processes. Determination of the LOC and Safety Class is the result of risk assessment activities per ISO 14971. The Primary risks of the BVN™ G-2000 Software Application are related to the consequences of clinical decisions based on false positive and false negative results for a patient due to inaccurate result. For a false positive result, the risks could include unnecessary testing or inappropriate treatment related to an inaccurate result. For a false negative result, the risk could include a missed or delayed diagnosis, the following measures are intended to mitigate the risks and shall be communicated to the users:

    1. The BVN™ G-2000 System is not intended to be used as the sole means for patient diagnosis or patient management decisions. The results should only be used in conjunction with the patient's clinical history.
    1. All software-aided results must be reviewed and confirmed by a qualified physician. It is the responsibility of a qualified physician to employ appropriate procedures and safeguards to assure the validity of the interpretation of results obtained. No patient report can be created if the qualified physician does not confirm the result.

Further, MetriTrack, Inc. has determined that its device advances the field without deviating from the scope and spirit of the Act by enabling the user to visualize the ultrasound probe position and orientation over the patient's breast diagram in real time and perform automatic annotation in a stand-alone device that works with any standard ultrasound imaging system which have been previously cleared by FDA under the 510(k) process.

7

| Substantial Equivalence Parameters | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
2000)
Subject
Device | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
1000) –
(K141870)
Predicate
Device | TaiHao
Medical Inc.
BR-FHUS
Navigation
1.0
(K171309)
Reference
Device | BR-FHUS
Viewer 1.0
(K171709)
Reference
Device | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Regulation
Number -
Name | 21 CFR 892.1560 - Ultrasonic
Pulsed Echo Imaging System | Yes | Yes | Yes | Yes |
| | 21 CFR 892.1570 Diagnostic
ultrasonic transducer | Yes | Yes | Yes | No |
| | 21 CFR 892.2050 Medical
image management and
processing system | Yes | No | No | Yes |
| Product
Code | IYO | Yes | Yes | Yes | No |
| ITX | Yes | Yes | Yes | No | |
| LLZ | Yes | No | No | Yes | |
| Intended Use | The Breast Volume Navigator
(BVN™) is an add-on accessory
for existing ultrasound imaging
systems and is intended to
control position and movement
of ultrasound transducers for the
systematic acquisition of 2-
dimensional image slices
throughout a volume of interest.
The BVN™ is intended to
acquire, analyze, store and
retrieve digital ultrasound
images for computerized 3-
dimensional image processing. | Yes | Yes | No | No |
| Substantial Equivalence Parameters | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
2000)
Subject
Device | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
1000) -
(K141870)
Predicate
Device | TaiHao
Medical Inc.
BR-FHUS
Navigation
1.0
(K171309)
Reference
Device | BR-FHUS
Viewer 1.0
(K171709)
Reference
Device | |
| The BVN™ is intended to assist
health care providers trained in
breast ultrasound with tools for
electromagnetic tracking of
instruments with respect of
breast ultrasound images
generated from FDA cleared
handheld ultrasound devices. | Yes | Yes | Yes | No | |
| The BVN™ is indicated for use
as an adjunct to mammography
for B-mode ultrasonic imaging of
a patient's breast when used
with a linear array transducer. | Yes | Yes | No | No | |
| The BVN™ is indicated for use
as an adjunct to handheld breast
ultrasound to assist the health
care providers trained in breast
ultrasound in their scanning
process. The scanning paths are
displayed on a route map and
provide quality control of
scanning to provide an overall
observation of scanning
process. | Yes | No | Yes | No | |
| The BVN™ Software
Application is intended as a
standalone software device | Yes | No | No | Yes | |
| Substantial Equivalence Parameters | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
2000)
Subject
Device | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
1000) –
(K141870)
Predicate
Device | TaiHao
Medical Inc.
BR-FHUS
Navigation
1.0
(K171309)
Reference
Device | BR-FHUS
Viewer 1.0
(K171709)
Reference
Device | |
| installed on a windows-based computer to assist physicians with manipulation and analysis tools in reviewing breast ultrasound images. Images and data are previously recorded from various imaging systems and other sources such as calibrated spatial positioning devices. The BVNTMTM Software Application provides the capability to visualize ultrasound images along with the scanning paths and position information of the probe that is stored in the DICOM file in advance. | | | | | |
| The BVNTMTM will allow exporting to any third-party application that has the appropriate level of DICOM compliance. | Yes | Yes | No | No | |
| The BVNTMTM is intended as a general-purpose digital 3D breast ultrasound image processing tool for radiology and surgery. | Yes | Yes | No | No | |
| The device is not intended to be used in the environment of | Yes | Yes | Yes | No | |
| | Substantial Equivalence Parameters | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
2000)
Subject
Device | MetriTrack -
Breast
Volume
Navigator
(BVNTMG-
1000) -
(K141870)
Predicate
Device | TaiHao
Medical Inc.
BR-FHUS
Navigation
1.0
(K171309)
Reference
Device | BR-FHUS
Viewer 1.0
(K171709)
Reference
|
| | strong magnetic or
electromagnetic fields, such as
in Magnetic Resonance Imaging
(MRI) room. | | | | |
| | The device is not intended to be
used as a replacement for
screening mammography. | Yes | Yes | No | No |
| Patient
Population | This Device is intended to be
used primarily upon female
patients.
This device is intended to be
used on breast tissue. | Yes | Yes | No (Breast) | No (Breast) |
| Design | Accessory to System, Imaging,
Pulsed Echo, Ultrasonic | Yes | Yes | Yes | No |
| | Ultrasonic Pulsed Echo Imaging
System. | Yes | Yes | Yes | No |
| | Off-the-Shelf PC Computer | Yes | Yes | Yes | Yes |
| | Operating System: Windows | Yes | Yes | Yes | Yes |
| | Off-the-shelf Image Capture
Device: USB frame grabber
device connected to computer
by cable | Yes | Yes | Yes | No |
| | Substantial Equivalence Parameters | MetriTrack -
Breast
Volume
Navigator
(BVN-G-
2000)
Subject
Device | MetriTrack -
Breast
Volume
Navigator
(BVN-G-
1000) —
(K141870)
Predicate
Device | TaiHao
Medical Inc.
BR-FHUS
Navigation
1.0
(K171309)
Reference
Device | BR-FHUS
Viewer 1.0
(K171709)
Reference
Device |
| | Using a magnetic position
sensing unit to collect 3D spatial
position information. | Yes | Yes | Yes | No |
| | Software: Position Sensor
Monitoring, Image presentation
and recording, user interface. | Yes | Yes | Yes | Yes |
| Where Used | Ultrasound examinations and
procedures | Yes | Yes | Yes | Yes |
| End Users | Qualified medical imaging
personnel that are familiar with
traditional breast ultrasound
procedures and techniques. | Yes | Yes | Skilled
medical
professionals | Skilled
medical
professionals |
| Performance
Testing | Results from System
verification and validation
testing performed per internal
procedures. | Yes | Yes | Yes | Yes |

SUBSTANTIAL EQUIVALENCE CHART

8

9

10

Breast Volume Navigator (BVN™) Model G-2000

11

Breast Volume Navigator (BVN™) Model G-2000

12

Electrical Testing

EMC testing per IEC 60601-1-2:2014 Class A, Group 2 for Emissions, Immunity for Professional Healthcare Facility Environment FCC Part 18.305:2020 and Electrical Safety Testing per NSI/AAMI ES60601-1 Edition 3.1 (revision 2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012). Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, technically the same as IEC 60601-1 with US National deviations) were performed for the BVN™ G-2000 System. All test results were acceptable for the BVN™G-2000 System.

Performance Data

Extensive Performance, Verification and Validation testing for the BVN™ G-2000 System was performed per internal procedures, which are compliant with 21 CFR Part 820.30, to ensure that all functional requirements have been met, and that core functions execute as expected.

Testing was conducted in-house by trained personnel in a simulated work-environment using phantoms to obtain the functional, accuracy and precision test results.

DESIGN VERIFICATION TESTS: The design specifications of the BVN™ G-2000 System were tested and verified to confirm that the product performance fulfilled those specification requirements.

DESIGN VALIDATION TESTS - NON-CLINICAL TESTING: System Validation testing was performed as follow:

    1. Mapping the accuracy and precision of the BVN™ System position tracking capability within its sensing volume using a Volume Mapping Performance Test Phantom with 9 test locations and 3 layers above the test phantom, there layers were intended to cover the height of the operative volume which is 35 cm. In this environment the BVN™ System has achieved the following:
    • Linear Accuracy and Linear Precision of Distance to Nipple: accuracy . and precision of ≤ 5 mm for measuring the distance of a tarqet in a breast to nipple of the same breast.
    • Accuracy and Precision of Clock Face Angle: accuracy and precision of ≤ 5 degrees for measuring the Clock Face Angle of a target in a breast.
    • . Accuracy and Precision of Patient Body Angles: accuracy and precision of ≤ 5 degrees for measuring the angles of the patient body relative to the examination bed in all three anatomical planes, i.e., Coronal, Transverse, and Sagittal planes.
    • Accuracy and Precision of Linear Accuracy: accuracy and precision of ≤ . 1 mm for measuring the linear accuracy between two points in the ultrasound image at one standard deviation.
    1. The completeness of scan accuracy has been tested using a Rigid Test Phantom to track the real-time position of the ultrasound probe relative to the region scanned to ensure that the entire scanning volume has been sufficiently scanned. In this environment the BVN™ System has achieved the following:

13

  • Completeness of Scanning Breast Contour Accuracy: BVN™ accurately ● captured and represented the hand drawn breast contour with a maximum positional error of +/- 5 mm in the X-Y plane.
  • . BVN™ correctly calculated the overlap between adjacent scanning bands and identified the occurrence of when this overlap is less than the set threshold as to ensure breast tissue is not missed between images: BVN™ was able to discern when overlaps of adjacent bands are less than a set threshold within a range of 5 to 30 mm ± 2 mm of overlap.
  • 3D Point Coordinates of Image Corners Velocity: BVN™ detected 3D . Point coordinates of image corners velocity above and below a set threshold within a range of 15 to 50 mm/s +/- 10% at 1 standard deviation as to ensure breast tissue is not missed between images.

Substantial Equivalence

The Breast Volume Navigator (BVN™) Model G-2000 is as safe and effective as the predicate devices: MetriTrack, LLC Breast Volume Navigator (BVNT™) Model G-1000 (K141870), TaiHao Medical Inc. BR-FHUS Navigation 1.0 (K171309), and TaiHao Medical Inc. BR-FHUS Viewer (K171709). The minor technological differences between the Breast Volume Navigator and its predicate devices raise no new issues of safety and effectiveness.

Performance data demonstrate that the Breast Volume Navigator (BVN™) Model G-2000 is as safe and effective as its predicate devices.