K Number
K211045
Date Cleared
2021-06-23

(76 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the FDA for "Synthetic Vinyl Examination Gloves." This type of document is for medical devices that are deemed substantially equivalent to a predicate device, and thus, typically does not involve the rigorous clinical studies that would require extensive details on acceptance criteria and study methodologies as one might find for a novel device or software.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (K153028) by comparing performance against recognized ASTM and ISO standards for examination gloves. Therefore, the questions related to AI performance, reader studies, ground truth establishment, and training sets are not applicable to this device.

Here's an analysis of the available information regarding acceptance criteria and testing, tailored to the context of a medical glove:


Acceptance Criteria and Reported Device Performance

The acceptance criteria for the "Synthetic Vinyl Examination Gloves" are based on recognized industry standards for medical gloves, primarily ASTM D5250, ASTM D5151, and ASTM D6124, along with biocompatibility standards ISO 10993-5 and ISO 10993-10. The device's reported performance demonstrates compliance with these standards.

Acceptance Criteria (Standard & Requirement)Reported Device Performance (Subject Device K211045)
Physical Dimensions (ASTM D5250)
Length, mm: min 230230 (all sizes XS-XL)
Width, mm: $\pm$5 from specified target78-115 ($\pm$5) (meets specification for each size)
Thickness, mm (Finger): min 0.080.08
Thickness, mm (Palm): min 0.080.08
Physical Properties (ASTM D5250)
Tensile Strength (Before Aging): 11 MPa, minComply with ASTM D5250 (Average 16.9MPa for predicate, subject device also complies)
Ultimate Elongation (Before Aging): 300% minComply with ASTM D5250 (Average 550% for predicate, subject device also complies)
Tensile Strength (After Aging): 11 MPa, minComply with ASTM D5250 (Average 14.4MPa for predicate, subject device also complies)
Ultimate Elongation (After Aging): 300% minComply with ASTM D5250 (Average 550% for predicate, subject device also complies)
Freedom from Holes (ASTM D5151)
AQL=2.5 when tested in accordance with ASTM D5151Be free from holes when tested in accordance with ASTM D5151 AQL=2.5
Powder Content (ASTM D6124)
Meet the requirements of ASTM D6124Average 0.02 mg per glove
Biocompatibility (ISO 10993-10)
Not an irritantUnder the conditions of the study, not an irritant
Not a sensitizerUnder conditions of the study, not a sensitizer
Biocompatibility (ISO 10993-5)
Did not show potential toxicity to L-929 cellsUnder conditions of the study, did not show potential toxicity to L-929 cells

Study Details:

  1. Sample size used for the test set and the data provenance:

    • The document does not specify the exact sample sizes used for each of the non-clinical tests (e.g., number of gloves tested for tensile strength, holes, etc.). The tests refer to compliance with ASTM and ISO standards, which typically define their own sampling plans.
    • Data Provenance: The tests were conducted by the manufacturer, Jiangsu Huayuan Medical Technology Co., Ltd. The document does not explicitly state the country of origin of the data beyond the manufacturer's location in China. The testing appears to be prospective for the subject device to demonstrate compliance with the stated standards.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. For non-clinical tests of physical properties and biocompatibility of medical gloves, "ground truth" is established by laboratory measurement against predefined objective criteria in the standards, not by expert consensus or subjective assessment.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjective diagnostic interpretations, especially in image analysis where human readers evaluate cases. For objective laboratory testing, results are determined by instrumentation and adherence to standard protocols.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This device is a medical glove, not an AI software or a device that involves human reader interpretation in a diagnostic context.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This device is a physical medical glove and does not involve any algorithms or AI for standalone performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the physical and performance tests, the "ground truth" is defined by the objective, quantitative specifications outlined in the referenced national and international standards (ASTM D5250, ASTM D5151, ASTM D6124).
    • For biocompatibility tests, the "ground truth" is established by the biological response observed in standardized in vitro (cytotoxicity) and in vivo (irritation, sensitization) tests as per ISO 10993-5 and ISO 10993-10, evaluated against predefined criteria for an acceptable biological reaction.
  7. The sample size for the training set:

    • Not Applicable. This device is a medical glove. It does not involve a "training set" in the context of machine learning or AI.
  8. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 23, 2021

Jiangsu Huayuan Medical Technology Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K211045

Trade/Device Name: Synthetic Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: March 26, 2021 Received: April 8, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Ryan Ortega, Ph. D Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211045

Device Name Synthetic Vinyl Examination Gloves

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K211045)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Jiangsu Huayuan Medical Technology Co.,Ltd. Address: No.36 Nanzhuang Road, High-Tech Industrial Development Zone of Dongtai City,Jiangsu Province,China Phone Number: +86-18252909158 Contact: Yun Guo Date of Preparation: 06/23/2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Synthetic Vinyl Examination Gloves Common name: Synthetic Vinyl Examination Gloves Classification name: Non-powdered Patient Examination Glove Model(s): XS,S, M, L, XL

3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Zibo Huiying Medical Products, Co. Ltd. Device: Synmax Synthetic Patient Examination Vinyl Gloves,Powder Free,Blue 510(k) number: K153028

{4}------------------------------------------------

5.0 Indication for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free vinyl synthetic patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject devicePredicate deviceComparison
510(k) numberK211045K153028/
Product CodeLYZLYZSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examinationglove is a disposabledevice intended formedical purposes that isworn upon theexaminer's hands toprevent contaminationbetween patient andexaminer.A patient examinationglove is a disposabledevice intended formedical purposes thatis worn upon theexaminer's hands orfingers to preventcontamination betweenpatient and examiner.Same
Powered or Powered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
Labeling InformationSingle use, powder free,device color, devicename, glove size andquantity, product name,Non-SterileSingle use, powderfree, device color,device name, glovesize and quantity,product name, Non-SterileSimilar

Table1-General Comparison

Table2 Device Dimensions Comparison

PredicateDevice(K153028)DesignationSizeTolerance
Length, mmAverage over 234 on M size-
Width, mmAverage over 96 on M size-
Thickness, mm:

{5}------------------------------------------------

FingerAverage 0.096-
PalmAverage 0.098-
Subject Device(K211045)DesignationSizeTolerance
XSSMLXL
Length, mm230230230230230min
Width, mm788595105115$\pm 5$
Thickness, mm:
Finger0.08min
Palm0.08min
RemarkSimilar

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250,so the differences do not raise any new safety or performance questions.

ItemSubject device(K211045)Predicate device(K153028)Comparison
ColorantBlueBlueSame
PhysicalPropertiesBeforeAgingTensileStrength11MPa, minAverage 16.9MPaAnalysis
UltimateElongation300%minAverage 550%Analysis
AfterAgingTensileStrength11MPa, minAverage 14.4MPa, minAnalysis
UltimateElongation300%minAverage 550%Analysis
Comply with ASTM D5250Comply with ASTM D5250Same
Freedom from HolesBe free from holes when tested in accordance with ASTM D5151 AQL=2.5Be free from holes when tested in accordance with ASTM D5151 AQL=2.5Same
Powder Contentaverage 0.02 mg per gloveMeet the requirements of ASTM D6124Similar

Table3 Performance Comparison

Analysis: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250,so the differences do not raise any new safety or performance questions.

Table4 Safety Comparison

ItemSubject device(K211045)Predicate device(K153028)Comparison
MaterialPoly Vinyl ChloridePolyurethanePoly Vinyl ChloridePolyurethaneSimilar

{6}------------------------------------------------

NitrileDi-(2-ethylhexyl)Terephthalate(DOTP)Diisononyl Phthalate(DINP)
BiocompatibilityIrritationUnder the conditions of thestudy, not an irritantComply withISO10993-10SAME
SensitizationUnder conditions of thestudy, not a sensitizer.
CytotoxicityUnder conditions of thestudy, did not show potentialtoxicity to L-929 cells./SIMILAR
Label and LabelingMeet FDA's RequirementMeetRequirementSAME

Analysis: The materials of the subject device are little different with that of the predicate, but they all meet the performance requirements of ASTM D5250,also biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards. Therefore, the differences will not raise any safety and effectiveness issues on performance and biocompatibility.

8.0 Discussion of Non-clinical and Clinical Test Performed

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTMD5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

9.0 Clinical Test Conclusion

Clinical testing is not needed for this device.

10.0Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.