K Number
K210767
Date Cleared
2021-06-10

(87 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Surgical Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Medical Procedure Masks are manufactured with three layers. The outer layer is made of double spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) non-woven fabric. The inner layer which contacts with face is made of double spun-bonded polypropylene (PP) non-woven fabric. The Surgical Masks are single use, disposable device, provided non-sterile.

AI/ML Overview

This document is a 510(k) Summary for a Surgical Mask, not an AI/ML medical device. Therefore, a significant portion of the requested information regarding AI/ML device acceptance criteria and study design (e.g., sample size for test/training sets, expert ground truth, MRMC studies, standalone performance, etc.) is not applicable to this submission.

The acceptance criteria and performance data provided in this document pertain to the physical and functional characteristics of the surgical mask, as tested against ASTM F2100-19 standards for medical face masks.

Here's a breakdown of the relevant information from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Methodology (Test Standard)PurposeAcceptance CriteriaResults
ASTM F1862M-17Fluid Resistance Performance29 out of 32 pass at 160mmHgPass at 160mmHg
ASTM F2299Particulate Filtration Efficiency$\geq$ 98%Pass
ASTM F2101-19Bacterial Filtration Efficiency$\geq$ 98%Pass
EN 14683:2019 Annex CDifferential Pressure< 6.0mmH2O/cm²Pass
16 CFR 1610FlammabilityClass I non-flammablePass

2. Sample size used for the test set and the data provenance:

  • Test set sample size: Not explicitly stated for each test, but for Fluid Resistance (ASTM F1862M-17), the acceptance criterion mentions "29 out of 32 pass," indicating a test of at least 32 samples.
  • Data Provenance: The tests were non-clinical performance tests conducted by the manufacturer, Hunan Triplex Precision Medical Devices Co., Ltd., based in Changsha, Hunan, China. The data would be considered prospective for the purpose of demonstrating compliance for this specific device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This is a physical product, not an AI/ML device relying on human expert annotations for ground truth. Testing involves laboratory measurements against established standards.

4. Adjudication method for the test set:

  • None applicable. This is non-clinical laboratory testing, not a human reader study requiring adjudication. The results are quantitative measurements against predefined thresholds.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI/ML algorithm.

7. The type of ground truth used:

  • For the non-clinical tests, the "ground truth" is defined by the established performance standards (e.g., ASTM F2100-19, EN 14683:2019, 16 CFR 1610) and the specific methodologies outlined within those standards for measuring characteristics like fluid resistance, filtration efficiency, differential pressure, and flammability.

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device, so there is no "training set."

9. How the ground truth for the training set was established:

  • Not applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 10, 2021

Hunan Triplex Precision Medical Devices Co., Ltd % Amber Pang Director Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 703, 705, Building 1, West Guangzhong Road 555, Shanghai, 200072 China

Re: K210767

Trade/Device Name: Surgical Masks, Model:FE-1 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 9, 2021 Received: March 15, 2021

Dear Amber Pang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210767

Device Name Surgical Masks

Indications for Use (Describe)

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 3-510(k) Summary

l. Submitter

Hunan Triplex Precision Medical Devices Co., Ltd. Room 612 & 613, Building No.5~8, SUNVO Health & Medicine Industrial Park, No.1048 Zhongqing Road, Kaifu District, Changsha, Hunan, China

Establishment Registration Number: 3017420299

Contact person: Huangkai Position: Manager Tel.: +86 0731-86780069 Fax: +86-0731-86780069 E-mail: Hwangkay@tom.com

Preparation date: Jun. 01, 2021

II. Proposed Device

Trade Name of Device:Surgical Masks
Common name:Surgical Mask
Regulation Number:21 CFR 878.4040
Regulatory Class:Class II
Product code:FXX
Review PanelGeneral Hospital

III. Predicate Devices

510(k) Number:K160269
Trade name:Surgical Face masks (Ear loops and Tie-on)
Common name:Surgical Mask
Classification:Class II
Product Code:FXX
ManufacturerSan-M Package Co., Ltd.

IV. Device Description

The Surgical Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The

{4}------------------------------------------------

Medical Procedure Masks are manufactured with three layers. The outer layer is made of double spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) non-woven fabric. The inner layer which contacts with face is made of double spun-bonded polypropylene (PP) non-woven fabric.

The Surgical Masks are single use, disposable device, provided non-sterile.

V. Indication for Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

VI. Comparison of Technological Characteristics with the Predicate Devices

ItemProposed DevicePredicate Device(K160269)
Trade NameSurgical MasksSurgical Face masks (Earloops and Tie-on)
Product CodeFXXFXX
Regulation No.21 CFR 878.404021 CFR 878.4040
ClassClass IIClass II
Mask StyleFlat-pleated, ear loops,3 layersFlat-pleated, ear loops ortie-on, 4 layers
Indication forUseThe Surgical Masks areintended to be worn to protectboth the patient and healthcarepersonnel from transfer ofmicroorganisms, body fluidsand particulate material. Theseface masks are intended foruse in infection controlpractices to reduce thepotential exposure to bloodand body fluids. This is a singleuse, disposable device(s),The Surgical facemasks areintended to be worn to protectthe patient and healthcarepersonnel from transfer ofmicroorganisms, blood fluid,and particulate material. Theseface masks are intended foruse in infection controlpractices to reduce thepotential exposure to bloodand body fluid. This is asingle-use, disposable devices

Table 10-1 General Comparison

{5}------------------------------------------------

provided non-sterile.provided non-sterile.
MaterialInnerlayerWhite double spun-bondpolypropylenePolypropylene
MiddlelayerMelt blown polypropylene filter1. Polypropylene spun-bond2. Polypropylene melt blown
OuterlayerBlue double spun-bondpolypropylenePolypropylene
Ear loops85% Polyester +15% Spandex3.0 Latex – free white elasticbandEar loops: Polyester, polyurethane;Side tapes: Polyester spun-bond (ear loops mask only)Ties tapes: Polypropylene spun-bond or polyester spun-bond
NosepieceIron wire covered bypolypropylenePolyethylene coatedsteel wire
ColorBlueBlue ,white
Dimension17.5cmx9.5cm17.5cmx9.0cm18.0cmx9.0cm
OTC UseYesYes
SterileNon-sterileNon-sterile
For single UseYesYes
ASTM F2100LevelLevel 3Level 1, Level 2, Level 3
BiocompatibilityConfirm to the requirements ofISO 10993 series standardsConfirm to the requirements ofISO 10993 series standards

VII. Non-Clinical Testing

Non clinical performance tests were conducted to verify that the proposed device met all design specifications. The below table shows the test results of test article, which demonstrated that the proposed device complies with the standards of ASTM F2100-19:

MethodologyPurposeAcceptance CriteriaResults
ASTM F1862M-17Fluid ResistancePerformance29 out of 32 passat 160mmHgPass at 160mmHg
ASTM F2299Particulate Filtration$\geq$ 98%Pass

{6}------------------------------------------------

Efficiency
ASTM F2101-19Bacterial FiltrationEfficiency≥98%Pass
ENAnnex C14683:2019DifferentialPressure< 6.0mmH2O/cm²Pass
16 CFR 1610FlammabilityClass I nonflammablePass

VIII. Clinical Testing

No clinical study is included in this submission.

IX. Conclusion

The conclusion drawn from the non-clinical performance testing data demonstrates that the subject device is as safe, as effective, and performs as well as or better than the predicate device, Surgical Face masks (K160269).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.