(62 days)
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- The extensive treatment of EPA potable water consists of:
-
- Pre-filtration through two pre-filters:
- · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
- · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
-
- UV Irradiation:
- · During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > or equal to 6-log reduction of MS2 virus is delivered to the water.
- 3.0.1 micron filtration:
- · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria. fungi and protozoa > 0.1 micron.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution(> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable sensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The current submission is provided to describe modifications for:
- . Firmware update
- . Obsolescence and replacement of varistor
The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system. utilizing peracetic acid to process totally immersible semi-critical medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe modifications for:
- Firmware update ●
- Obsolescence and replacement of varistor ●
The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible semi-critical heat-sensitive medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe the modification for obsolescence and replacement of varistor (the firmware update is not required on this model).
The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. After LCS processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels.
The provided documentation outlines acceptance criteria and the results of a study for several STERIS Liquid Chemical Sterilant Processing Systems: SYSTEM 1E, SYSTEM 1 endo Model P6800, and SYSTEM 1 endo Model P6900. These systems utilize the S40 Sterilant Concentrate. The submission is a 510(k) premarket notification, indicating a claim of substantial equivalence to a predicate device (K192929), largely due to software/firmware updates and component obsolescence/replacement.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
For all three models (SYSTEM 1E, SYSTEM 1 endo Model P6800, and SYSTEM 1 endo Model P6900), the testing described is limited to evaluating the impact of minor modifications (firmware update and/or varistor replacement). The core performance criteria and germicidal efficacy are based on the predicate device's established performance.
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Software validation per IEC 62304 | The firmware was validated and determined to operate effectively and as designed. | Pass |
| Functional test with replacement varistor | The modification does not affect the operation of the device (or ensures operation). | Pass |
Note: The tables in the document (Table 1, Table 2 in each section concerning the systems and S40 Sterilant Concentrate) extensively compare the proposed device features and performance to the predicate device, indicating "Identical" across various parameters such as germicidal claim, exposure time, use temperature, reuse, active ingredient, mode of action, rinses, microbial efficacy (sporicidal, fungicidal, bactericidal, viricidal, tuberculocidal, simulated-use, clinical in-use), biocompatibility, residue reduction, and device material compatibility. This implies that the acceptance criteria for these aspects are the same as those met by the predicate device, and the proposed devices are considered to perform equivalently.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Functional & Software Testing: The document does not specify the exact sample size (e.g., number of devices or test cycles) used for the software validation or functional tests with the replacement varistor. It simply states the tests were performed and passed.
- Data Provenance: The new tests (software validation and functional tests) would likely be internal studies conducted by STERIS Corporation. The data on microbial efficacy and material compatibility for the predicate device (which are claimed as identical for the proposed devices) stem from various in vitro tests and clinical in-use studies, as indicated by the listed citations (Block, Clapp et al., Maillard et al., McDonnell et al.). The country of origin for these foundational efficacy tests is not specified, but they refer to internationally recognized methods (e.g., AOAC Official Methods, EPA Viricidal Testing). These previous studies would be considered retrospective for the current submission, as they establish the performance of the predicate. The current evaluation of modifications would be prospective relative to these specific changes.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- The document primarily describes engineering verification and validation activities (software validation, functional testing). These types of tests do not typically involve human experts for establishing "ground truth" in the same way clinical or diagnostic studies do. Instead, the "truth" is established by adherence to design specifications, regulatory standards, and verified operational parameters.
- For the microbial efficacy (sporicidal, fungicidal, bactericidal, viricidal, tuberculocidal, simulated-use, clinical in-use), the "ground truth" implicitly relies on established microbiological methods and their respective acceptance criteria. The experts involved in the original studies for the predicate device would have been microbiologists and other scientific professionals, as indicated by the referenced scientific literature. The number and specific qualifications are not detailed in this submission.
4. Adjudication Method for the Test Set
- Given the nature of the tests mentioned (software validation, functional tests), an adjudication method (like 2+1, 3+1 often used in medical image analysis) is not applicable. These are objective engineering and performance verification tests with clear pass/fail criteria.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The device is a sterilizer, not a diagnostic imaging device typically associated with human reader performance. The improvements described are related to maintenance and software updates, not human interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- The device itself is a standalone system for liquid chemical sterilization; it does not involve an "algorithm" in the sense of AI performing a diagnostic task. Its performance is evaluated intrinsically through its ability to achieve sterilization (microbial kill) and maintain operational parameters. The studies verifying its efficacy (e.g., sporicidal activity, simulated-use tests) are standalone evaluations of the system's performance.
7. Type of Ground Truth Used
- For Software and Functional Testing: The ground truth is defined by the device's design specifications and regulatory requirements (e.g., correct operation of firmware, proper function after component replacement).
- For Microbial Efficacy (Predicate Device): The ground truth was established through rigorous microbiological testing using standardized methods (e.g., AOAC Official Methods, EPA Viricidal Testing). This includes:
- Culture-based assays: Demonstrating log reduction of various microorganisms (Bacillus subtilis, Clostridium sporogenes, Trichophyton mentagrophytes, Salmonella choleraesuis, Staphylococcus aureus, Pseudomonas aeruginosa, Herpes simplex Type 1, Adenovirus Type 5, Poliovirus Type 1, Mycobacterium terrae).
- Simulated-use tests: Demonstrating log reduction of Geobacillus stearothermophilus spores under manual application conditions.
- Clinical In-Use: Absence of surviving microorganisms on representative medical devices.
8. Sample Size for the Training Set
- This information is not applicable. The described devices are physical sterilizers, not AI/machine learning models that require a "training set" in the conventional sense. The "firmware update" is likely a software change to address bugs or improve existing functionality, not a re-training of a learning algorithm.
9. How the Ground Truth for the Training Set Was Established
- This information is not applicable, as there is no "training set" for an AI algorithm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 12, 2021
STERIS Corporation Jennifer Nalepka Lead Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K210737
Trade/Device Name: SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid Chemical Sterilants/High Level Disinfectants Regulatory Class: Class II Product Code: MED Dated: April 9, 2021 Received: April 12, 2021
Dear Jennifer Nalepka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K210737
Device Name
SYSTEM 1E Liquid Chemical Sterilant Processing System
Indications for Use (Describe)
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- The extensive treatment of EPA potable water consists of:
-
- Pre-filtration through two pre-filters:
- · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
- · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
-
- UV Irradiation:
- · During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > or equal to 6-log reduction of MS2 virus is delivered to the water.
- 3.0.1 micron filtration:
- · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria. fungi and protozoa > 0.1 micron.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800
Indications for Use (Describe)
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution(> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900
Indications for Use (Describe)
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable sensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is in a bold, serif font. Below the word is a graphic of several horizontal, wavy lines in blue, resembling water.
510(k) Summary For SYSTEM 1E Liquid Chemical Sterilant Processing System
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact: Jennifer Nalepka Lead Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198
Summary Date: April 9, 2021
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{6}------------------------------------------------
1. Device Name
| Trade Name: | SYSTEM 1E Liquid Chemical Sterilant ProcessingSystem |
|---|---|
| Device Classification: | Class II |
| Common/usual Name: | Liquid Chemical Sterilizer |
| Classification Name: | Sterilant, Medical devices, Liquid ChemicalSterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED |
2. Predicate Device
SYSTEM 1E Liquid Chemical Sterilant Processing System, K192929
3. Description of Device
The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The current submission is provided to describe modifications for:
- . Firmware update
- . Obsolescence and replacement of varistor
The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed
{7}------------------------------------------------
to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.
4. Indications for Use
The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semicritical heat-sensitive medical devices in healthcare facilities.
The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid). liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.
The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
- 米 The extensive treatment of EPA potable water consists of:
-
- Pre-filtration through two pre-filters:
- Pre-filter A is a gross depth filter that removes approximately 2.5 ● micron or larger particles/contaminants.
- . Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
-
- UV Irradiation:
- During transit through the UV water treatment chamber, a UV dose . sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.
-
- 0.1 micron filtration:
- The water prepared by pre-filtration and UV irradiation is filtered . through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.
-
ട. Technological Characteristic Comparison Table
The SYSTEM 1E Liquid Chemical Sterilant Processing System is the same as the predicate device; the specific modifications described in this submission are for a firmware update and obsolescence to components.
{8}------------------------------------------------
| Feature | ProposedSYSTEM 1E Processor | Predicate (K192929)SYSTEM 1E Processor | Comparison |
|---|---|---|---|
| Indications forUse | The SYSTEM 1E LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization ofcleaned, immersible, andreusable critical and semi-critical heat-sensitive medicaldevices in healthcare facilities.The SYSTEM 1E Processordilutes the S40 Sterilant | The SYSTEM 1E LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization of cleaned,immersible, and reusable criticaland semi-critical heat-sensitivemedical devices in healthcarefacilities.The SYSTEM 1E Processordilutes the S40 Sterilant | |
| Concentrate to its use dilution(>1820 mg/L peracetic acid),liquid chemically sterilizes theload during a controlled 6-minute exposure at 45.5 to60°C, and rinses the load withextensively treated* potablewater. After completion of acycle, critical devices should beused immediately; semi-criticaldevices should be usedimmediately or may be handledand stored in a manner similarto that of high level disinfectedendoscopes. Critical devices notused immediately should beprocessed again before use. | Concentrate to its use dilution(>1820 mg/L peracetic acid),liquid chemically sterilizes theload during a controlled 6-minute exposure at 45.5 to 60°C,and rinses the load withextensively treated* potablewater. After completion of acycle, critical devices should beused immediately; semi-criticaldevices should be usedimmediately or may be handledand stored in a manner similar tothat of high level disinfectedendoscopes. Critical devices notused immediately should beprocessed again before use. | Identical | |
| The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquidchemically sterilize medicaldevices.* The extensive treatment ofEPA potable water consists of:1. Pre-filtration through twopre-filters:• Pre-filter A is a grossdepth filter that removesapproximately 2.5micron or largerparticles/contaminants.• Pre-filter B is a surfacefilter that removesparticles/contaminants >0.1 micron. | The SYSTEM 1E Processor usesonly S40 Sterilant Concentrate toliquid chemically sterilizemedical devices.* The extensive treatment ofEPA potable water consists of:1. Pre-filtration through two pre-filters:• Pre-filter A is a gross depthfilter that removesapproximately 2.5 micronor largerparticles/contaminants.• Pre-filter B is a surfacefilter that removesparticles/contaminants >0.1 micron. | ||
| Feature | ProposedSYSTEM 1E Processor | Predicate (K192929)SYSTEM 1E Processor | Comparison |
| 2. UV Irradiation:• During transit throughthe UV water treatmentchamber, a UV dosesufficient to achieve a ≥6-log reduction of MS2virus is delivered to thewater.3. 0.1 micron filtration:• The water prepared bypre-filtration and UVirradiation is filteredthrough redundant, 0.1-micron (absolute rated)membranes to removebacteria, fungi andprotozoa > 0.1 micron. | 2. UV Irradiation:• During transit through theUV water treatmentchamber, a UV dosesufficient to achieve a ≥ 6-log reduction of MS2 virusis delivered to the water.3. 0.1 micron filtration:• The water prepared by pre-filtration and UV irradiationis filtered throughredundant, 0.1-micron(absolute rated) membranesto remove bacteria, fungiand protozoa > 0.1 micron. | ||
| OperatingPrinciples/Technology | A microprocessor controlledunit with interchangeableprocessing trays/containers.The processor lid opens toreveal the processing chamberin which the load is placed.Devices with internal lumensare interfaced with theprocessor using connectors.Sterilant Concentrate is placedin a specialized compartmentand when the processor fillswith water, it creates thesterilant use dilution from thesingle use sterilant cup. Theprocessor monitors and controlsthe use dilution temperature andcontact time. The processorautomatically rinses the loadwith extensively treated waterto remove sterilant residuals. | A microprocessor controlled unitwith interchangeable processingtrays/containers. The processorlid opens to reveal theprocessing chamber in which theload is placed. Devices withinternal lumens are interfacedwith the processor usingconnectors. SterilantConcentrate is placed in aspecialized compartment andwhen the processor fills withwater, it creates the sterilant usedilution from the single usesterilant cup. The processormonitors and controls the usedilution temperature and contacttime. The processorautomatically rinses the loadwith extensively treated water toremove sterilant residuals. | Identical |
| ProcessParameters | Standardized cycle parameterscannot be altered by operator.The critical process parametersare:• Contact Time• Use Dilution Temperature• Peracetic acidconcentration• Bacterial retentive waterfilter integrity | Standardized cycle parameterscannot be altered by operator.The critical process parametersare:• Contact Time• Use Dilution Temperature• Peracetic acid concentration• Bacterial retentive water• filter integrity• UV irradiation | Identical |
| Feature | ProposedSYSTEM 1E Processor | Predicate (K192929)SYSTEM 1E Processor | Comparison |
| UV irradiation | |||
| ProcessMonitors | Cycle Printout documents successful cycle completion or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification | Cycle Printout documents successful cycle completion or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification | Identical |
| Feature | ProposedSYSTEM 1E Processor | Predicate (K192929)SYSTEM 1E Processor | Comparison |
| DesignFeatures | Microprocessor controlledunalterable andstandardized liquidchemical sterilantprocessing and Diagnosticcycles Intended for use with S40Sterilant Concentrate Processor provides dual0.1 micron filtered, UVtreated water for liquidchemical sterilantprocessing and rinsing Automated dilution anddelivery of sterilant Make up air for processorduring drain sequences isfiltered through a 0.2-micron membrane airfilter. | Microprocessor controlledunalterable andstandardized liquidchemical sterilantprocessing and Diagnosticcycles Intended for use with S40Sterilant Concentrate Processor provides dual 0.1micron filtered, UV treatedwater for liquid chemicalsterilant processing andrinsing Automated dilution anddelivery of sterilant Make up air for processorduring drain sequences isfiltered through a 0.2-micron membrane air filter. | Identical |
| Processing Cycle | |||
| Incoming watertemperature | $\geq$ 43°C | $\geq$ 43°C | Identical |
| Temperature tostart exposurephase | $\geq$ 46°C | $\geq$ 46°C | Identical |
| Temperaturealarm pointduring theexposure phase | <45.5 or >60°C | <45.5 or >60°C | Identical |
| Temperaturerange during atypical LiquidChemicalSterilantProcessingCycle | 46 - 55°C | 46 - 55°C | Identical |
| ExposureTime | 6 minutes | 6 minutes | Identical |
| Rinse waterpreparation | Hot potable tap water is:pre-filtered flowed through a UV Lighttreatment chamber toachieve $\geq$ a 6-log reductionof virus Filtered through redundant0.1-micron filtermembranes | Hot potable tap water is:pre-filtered flowed through a UV Lighttreatment chamber toachieve $\geq$ a 6-log reductionof virus Filtered through redundant0.1-micron filtermembranes | Identical |
| Number of | 2 | 2 | Identical |
| Feature | ProposedSYSTEM 1E Processor | Predicate (K192929)SYSTEM 1E Processor | Comparison |
| rinses | |||
| Air Purge | Aids in removing excess waterfrom instrument lumens afterrinsing | Aids in removing excess waterfrom instrument lumens afterrinsing | Identical |
| Water FilterIntegrity Test | Conducted at the end of everyliquid chemical sterilantprocessing cycle and during theDiagnostic cycle | Conducted at the end of everyliquid chemical sterilantprocessing cycle and during theDiagnostic cycle | Identical |
| ApproximateCycle time | 25 minutes | 25 minutes | Identical |
| DiagnosticCycle | Performs 15 tests onprocessor's systems confirmingproper function (same tests aspredicate device except for anadded UV monitor test).Recommended to performevery 24 hours. After a failedDiagnostic cycle, a liquidchemical sterilant processingcycle cannot be performed untilthe problem is rectified and asuccessful Diagnostic cycle hasbeen completed. | Performs 15 tests on processor'ssystems confirming properfunction (same tests as predicatedevice except for an added UVmonitor test). Recommended toperform every 24 hours. After afailed Diagnostic cycle, a liquidchemical sterilant processingcycle cannot be performed untilthe problem is rectified and asuccessful Diagnostic cycle hasbeen completed. | Identical |
| Interchangeable Components | |||
| ProcessingTray /Containers | Uses interchangeableprocessing trays/containers• Universal Flexible ProcessingTray (C1160E)• General ProcessingContainer/Tray (C1200)• Directed Flow ProcessingContainer/Tray (C1220)• Flexible EndoscopeProcessing Container / Tray(C1140)• Ultrasound Processing Tray(C3000XL) | Uses interchangeable processingtrays/containers• Universal Flexible ProcessingTray (C1160E)• General ProcessingContainer/Tray (C1200)• Directed Flow ProcessingContainer/Tray (C1220)• Flexible EndoscopeProcessing Container / Tray(C1140)• Ultrasound Processing Tray(C3000XL) | Identical |
| Accessories | |||
| SterilantConcentrate | Uses S40 Sterilant Concentrate | Uses S40 Sterilant Concentrate | Identical |
| QuickConnects | Uses Quick Connects to adaptinstrument lumens to theTray/Container ports | Uses Quick Connects to adaptinstrument lumens to theTray/Container ports | Identical |
| ChemicalIndicator | VERIFY Chemical Indicatorfor the S40 SterilantConcentrate | VERIFY Chemical Indicator forthe S40 Sterilant Concentrate | Identical |
| Spore TestStrip | VERIFY Spore Test Strip forS40 Sterilant Concentrate | VERIFY Spore Test Strip forS40 Sterilant Concentrate | Identical |
| Feature | ProposedSYSTEM 1E Processor | Predicate (K192929)SYSTEM 1E Processor | Comparison |
| OperatorMaintenanceRequirements | Periodic replacement of printertape, water filters and air filter | Periodic replacement of printertape, water filters and air filter | Identical |
Table 1. Processor Device Comparison Table
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
Table 2. S40 Sterilant Concentrate Device Comparison Table
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
|---|---|---|---|
| Indications forUse | The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquid chemicallysterilize medical devices. | The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquid chemicallysterilize medical devices. | Identical |
| Germicidalclaim | Liquid Chemical Sterilant | Liquid Chemical Sterilant | Identical |
| GermicideExposureTime (min) forintended use | 6 | 6 | Identical |
| UseTemperature | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at <43°C | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at <43°C | Identical |
| Reuse | Single use | Single use | Identical |
| HumanFactors | Dispensed ready to use.Container is opened and dilutedby the processor, thus limitinguser exposure to the activeingredient | Dispensed ready to useContainer is opened and dilutedby the processor, thus limitinguser exposure to the activeingredient | Identical |
| ActiveIngredient | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the SYSTEM 1EProcessor. | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the SYSTEM 1EProcessor. | Identical |
| Mode ofAction | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydryl and sulfurbonds in proteins and enzymes,particularly in the cell walls 1-hydroxyl radicals producedfrom PAA are bactericidal 2-PAA damages the viral capsidand viral nucleic acid3,4. | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydryl and sulfurbonds in proteins and enzymes,particularly in the cell walls1-hydroxyl radicals produced fromPAA are bactericidal2-PAA damages the viral capsidand viral nucleic acid3,4 | Identical |
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
| Rinses | Automatic, UV-irradiated, dual0.1 micron filtered, potable hotwater. | Automatic, UV-irradiated, dual0.1 micron filtered, potable hotwater. | Identical |
| Microbial Efficacy | |||
| SporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacy requirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacy requirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| FungicidalActivity ofDisinfectantsAOACOfficialMethod 955.17 | Solution is fungicidal.Trichophyton mentagrophytesTesting conducted in vitro | Solution is fungicidal.Trichophyton mentagrophytesTesting conducted in vitro | Identical |
| Use DilutionMethodAOAC,OfficialMethods955.14, 955.15,964.02 | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Identical |
| EPA ViricidalTesting(DIS/TSS-7,Nov. 1981) | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Identical |
| TuberculocidalActivityAscenziQuantitativeSuspensionTest | Solution is tuberculocidalMycobacterium terraeTesting conducted in vitro | Solution is tuberculocidalMycobacterium terraeTesting conducted in vitro | Identical |
| Simulated-UseTest | Meets efficacy requirement.> 6 log reduction Geobacillusstearothermophilus spores in amanual application | Meets efficacy requirement.> 6 log reduction Geobacillusstearothermophilus spores in amanual application | Identical |
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
| Clinical In-Use | No surviving microorganismson representative medicaldevices tested | No surviving microorganisms onrepresentative medical devicestested | Identical |
| Biocompatibility | |||
| CytotoxicityDevice Extracts | Two rinses with UV treated,dual 0.1-micron membranefiltered water effectively reducesterilant residues to safe levels. | Two rinses with UV treated, dual0.1-micron membrane filteredwater effectively reduce sterilantresidues to safe levels. | Identical |
| ResidueReduction | Automatic within the SYSTEM1E Processor:Two rinses with UV treated,dual 0.1-micron membranefiltered water effectively reducesterilant residues to safe levels. | Automatic within the SYSTEM1E Processor:Two rinses with UV treated, dual0.1-micron membrane filteredwater effectively reduce sterilantresidues to safe levels. | Identical |
| DeviceMaterialCompatibility | Compatible with medicaldevices as established bytesting finished flexibleendoscopes through 300 cyclesand rigid devices through 150cycles.No functional changes haveoccurred to flexible devices.Some materials show cosmeticchanges such as fading of blackanodized aluminum withoutharm to the base material. | Compatible with medicaldevices as established by testingfinished flexible endoscopesthrough 300 cycles and rigiddevices through 150 cycles.No functional changes haveoccurred to flexible devices.Some materials show cosmeticchanges such as fading of blackanodized aluminum withoutharm to the base material. | Identical |
1 Block, S. ed., Disinfection, Sterilization, and Preservation. 5th edition, 2001.
2 Clapp et al., Free Rad. Res., (1994) 21:147-167.
3 Maillard et. al., J. Med. Microbiol (1995) 42:415-420.
4 Maillard et. al., J. Appl Bacteriol (1996) 80:540-554.
{14}------------------------------------------------
5 McDonnell et al., J. AOAC International (2000) 83:269-276.
{15}------------------------------------------------
The proposed device and its predicate have identical intended use and technological characteristics. New testing was performed to evaluate the modified device and the results are summarized in Table 3.
Summary of Non-Clinical Testing 6.
Shown in Table 3 is the new testing that was performed to evaluate the modified device.
| Test | Acceptance Criteria | Result |
|---|---|---|
| Software validation per IEC 62304 | The firmware was validated and determined to operate effectively and as designed | Pass |
| Functional test with replacement varistor | The modification does not affect the operation of the device. | Pass |
Table 3. Summary of verification activities.
{16}------------------------------------------------
7. Summary of Clinical test
Clinical test is not applicable.
8. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K192929), Class II (21 CFR 880.6885), product code MED.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the STERIS logo. The word STERIS is in all caps and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines in blue.
510(k) Summary For SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact:
Jennifer Nalepka Lead Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198
Summary Date: April 9, 2021
{18}------------------------------------------------
Device Name 1.
| Trade Name: | SYSTEM 1 endo Liquid Chemical SterilantProcessing System, Model P6800 |
|---|---|
| Device Class: | Class 2 |
| Common/usual Name: | Liquid Chemical Sterilizer |
| Classification Name: | Sterilant, Medical devices, Liquid ChemicalSterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED |
2. Predicate Device
SYSTEM 1endo Liquid Chemical Sterilant Processing System, Model P6800, K192929
3. Description of Device
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system. utilizing peracetic acid to process totally immersible semi-critical medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe modifications for:
- Firmware update ●
- Obsolescence and replacement of varistor ●
The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels. Table 1 compares the proposed and predicate devices.
{19}------------------------------------------------
4. Intended Use
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 55°C, and rinses the load with 0.2 micron filtered potable water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
ട്. Description of Technological Similarities and Differences
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is identical to the predicate device. A comparison between the proposed and predicate devices is included in Table 1 and Table 2. Since there are no technological differences between the proposed and predicate devices, there are no new concerns of safety and effectiveness.
| Feature | Proposed DeviceSYSTEM 1 endo Processor,Model P6800 | Predicate Device (K192929)SYSTEM 1 endo Processor,Model P6800 | Comparison |
|---|---|---|---|
| The SYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization ofcleaned, immersible, andreusable semi-critical medicaldevices and their accessoriesin healthcare facilities. | The SYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization of cleaned,immersible, and reusable semi-critical medical devices and theiraccessories in healthcarefacilities. | ||
| Intended UseIndications forUse | The SYSTEM 1 endoProcessor automaticallydilutes the S40 SterilantConcentrate to its use dilution(>1820 mg/L peracetic acid),liquid chemically sterilizes theload during a controlled 6-minute exposure at 45.5 to60°C, and rinses the load with0.2 micron filtered water. | The SYSTEM 1 endo Processorautomatically dilutes the S40Sterilant Concentrate to its usedilution (>1820 mg/L peraceticacid), liquid chemically sterilizesthe load during a controlled 6-minute exposure at 45.5 to 60°C,and rinses the load with 0.2micron filtered water.The SYSTEM 1 endo Processoruses only S40 Sterilant | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor,Model P6800 | Predicate Device (K192929)SYSTEM 1 endo Processor,Model P6800 | Comparison |
| The SYSTEM 1 endoProcessor uses only S40Sterilant Concentrate to liquidchemically sterilize medicaldevices. | Concentrate to liquid chemicallysterilize medical devices. | ||
| OperatingPrinciples /Technology | A microprocessor controlledunit with interchangeableprocessing trays/containers.The processor lid opens toreveal the processing chamberin which the load is placed.Devices with internal lumensare interfaced with theprocessor using connectors,i.e. Quick Connects. S40Sterilant is placed in aspecialized compartment andwhen the processor fills withwater, it creates the sterilantuse dilution from the singleuse sterilant cup. Theprocessor monitors andcontrols the use dilutiontemperature and contact time.The processor automaticallyrinses the load with 0.2 micronfiltered water to removesterilant residuals. | A microprocessor controlled unitwith interchangeable processingtrays/containers. The processorlid opens to reveal the processingchamber in which the load isplaced. Devices with internallumens are interfaced with theprocessor using connectors, i.e.Quick Connects. S40 Sterilant isplaced in a specializedcompartment and when theprocessor fills with water, itcreates the sterilant use dilutionfrom the single use sterilant cup.The processor monitors andcontrols the use dilutiontemperature and contact time.The processor automaticallyrinses the load with 0.2 micronfiltered water to remove sterilantresiduals. | Identical |
| ProcessParameters | Standardized cycle parameterscannot be altered by theoperator. The critical processparameters are:• Use dilution contact time• Use dilution temperature• Peracetic acidconcentration• Bacterial retentive waterfilter integrity | Standardized cycle parameterscannot be altered by the operator.The critical process parametersare:• Use dilution contact time• Use dilution temperature• Peracetic acid concentration• Bacterial retentive waterfilter integrity | Identical |
| ProcessMonitors | • Cycle Printout documentssuccessful cyclecompletion or identifiesfault if cycle aborts• Alarms if thermocouplesindicate temperature outof specification | • Cycle Printout documentssuccessful cycle completionor identifies fault if cycleaborts• Alarms if thermocouplesindicate temperature out ofspecification | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor,Model P6800 | Predicate Device (K192929)SYSTEM 1 endo Processor,Model P6800 | Comparison |
| Alarms if pressure switch indicates that highpressure pump is not operating Alarms if conductivity probe indicatedconductivity specification not met Alarms if pressure transducer indicatescirculation pressure is out of specification inDiagnostic cycle Alarms if pressure transducer indicatesinternal water filter failed integrity test | Alarms if pressure switch indicates that high pressurepump is not operating Alarms if conductivity probe indicated conductivityspecification not met Alarms if pressure transducer indicatescirculation pressure is out of specification in Diagnosticcycle Alarms if pressure transducer indicates internalwater filter failed integrity test | ||
| DesignFeatures | Microprocessor controlled unalterable andstandardized liquid chemical sterilization andDiagnostic cycles Intended for use only with S40 SterilantConcentrate Automated dilution and delivery of S40 Sterilant Processor provides 0.2 micron filtered water forliquid chemical sterilant process and rinsing Make-up air for processor during drain sequences isfiltered through a 0.2 micron membrane filter | Microprocessor controlled unalterable and standardizedliquid chemical sterilization and Diagnostic cycles Intended for use only with S40 Sterilant Concentrate Automated dilution and delivery of S40 Sterilant Processor provides 0.2 micron filtered water forliquid chemical sterilant process and rinsing Make-up air for processor during drain sequences isfiltered through a 0.2 micron membrane filter | Identical |
| Cycle Parameters | Comparison | ||
| Incoming watertemp. | ≥ 43°C | ≥ 43°C | Identical |
| Temperature tostart sterilantexposure | ≥ 46°C | ≥ 46°C | Identical |
| Temperaturealarm pointduring LCSexposure | < 45.5°C or > 60°C | < 45.5°C or > 60°C | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor,Model P6800 | Predicate Device (K192929)SYSTEM 1 endo Processor,Model P6800 | Comparison |
| Temperaturerange of typicalLCS cycle | 46-55°C | 46-55°C | Identical |
| Rinse waterpreparation | Hot potable water• is pre-filtered• is filtered through 0.2 micron bacterial retentivefilter | Hot potable water• is pre-filtered• is filtered through 0.2 micron bacterial retentivefilter | Identical |
| Number ofrinses | 2 | 2 | Identical |
| Air Purge | Aids in removing excess waterfrom instrument lumens afterrinsing | Aids in removing excess waterfrom instrument lumens afterrinsing | Identical |
| Internal waterfilter integritytest | Conducted during theDiagnostic cycle | Conducted during the Diagnosticcycle | Identical |
| Approximatecycle time | 18 - 20 minutes | 18 - 20 minutes | Identical |
| DiagnosticCycle | Performs 14 tests onprocessor's systemsconfirming proper function.Recommended to performeach day of use. After a failedDiagnostic cycle a liquidchemical sterilization cyclecannot be performed until theproblem is rectified and asuccessful Diagnostic cyclehas been completed. | Performs 14 tests on processor'ssystems confirming properfunction.Recommended to performeach day of use. After a failedDiagnostic cycle a liquidchemical sterilization cyclecannot be performed until theproblem is rectified and asuccessful Diagnostic cyclehas been completed. | Identical |
| Accessories | |||
| Sterilant | Uses S40 Sterilant Concentrate- See Table 2 | Uses S40 Sterilant Concentrate -See Table 2 | Identical |
| ProcessingTrays andContainers | Uses interchangeableprocessing trays/containers• Universal FlexibleProcessing Tray• General ProcessingContainer and Tray• Directed Flow ProcessingContainer and Tray | Uses interchangeable processingtrays/containers• Universal FlexibleProcessing Tray• General ProcessingContainer and Tray• Directed Flow ProcessingContainer and Tray | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo Processor,Model P6800 | Predicate Device (K192929)SYSTEM 1 endo Processor,Model P6800 | Comparison |
| Flexible Endoscope Processing Container and Tray Ultrasound Processing Tray | Flexible Endoscope Processing Container and Tray Ultrasound Processing Tray | ||
| Quick Connects | Uses Quick Connects to attach instrument lumens to the Tray/Container ports | Uses Quick Connects to attach instrument lumens to the Tray/Container ports | Identical |
| ChemicalIndicator | VERIFY Chemical Indicator for S40 Sterilant is available for use in SYSTEM 1 endo LCSPS | VERIFY Chemical Indicator for S40 Sterilant is available for use in SYSTEM 1 endo LCSPS | Identical |
| Spore Test Strip | VERIFY Spore Test Strip for S40 Sterilant for use in SYSTEM 1 endo LCSPS | VERIFY Spore Test Strip for S40 Sterilant for use in SYSTEM 1 endo LCSPS | Identical |
| OperatorMaintenance | Periodic replacement of printer tape, water filters and air filter | Periodic replacement of printer tape, water filters and air filter | Identical |
Table 1. Processor Comparison Table
{20}------------------------------------------------
{21}------------------------------------------------
{22}------------------------------------------------
{23}------------------------------------------------
Table 2. S40 Sterilant Concentrate Comparison Table
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
|---|---|---|---|
| Indications forUse | The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquid chemicallysterilize medical devices. | The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquid chemicallysterilize medical devices. | Identical |
| Germicidalclaim | Liquid Chemical Sterilant | Liquid Chemical Sterilant | Identical |
| GermicideExposureTime (min) forintended use | 6 | 6 | Identical |
| UseTemperature | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at <43°C | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at <43°C | Identical |
| Reuse | Single use | Single use | Identical |
| HumanFactors | Dispensed ready to use.Container is opened and dilutedby the processor, thus limitinguser exposure to the activeingredient | Dispensed ready to useContainer is opened and dilutedby the processor, thus limitinguser exposure to the activeingredient | Identical |
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
| ActiveIngredient | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the SYSTEM 1EProcessor. | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the SYSTEM 1EProcessor. | Identical |
| Mode ofAction | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydryl and sulfurbonds in proteins and enzymes,particularly in the cell walls1-hydroxyl radicals producedfrom PAA are bactericidal2-PAA damages the viral capsidand viral nucleic acid3,4. | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydryl and sulfurbonds in proteins and enzymes,particularly in the cell walls1-hydroxyl radicals produced fromPAA are bactericidal2-PAA damages the viral capsidand viral nucleic acid3,4 | Identical |
| Rinses | Automatic, UV-irradiated, dual0.1 micron filtered, potable hotwater. | Automatic, UV-irradiated, dual0.1 micron filtered, potable hotwater. | Identical |
| Microbial Efficacy | |||
| SporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacy requirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacy requirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| FungicidalActivity ofDisinfectantsAOACOfficialMethod 955.17 | Solution is fungicidal.Trichophyton mentagrophytesTesting conducted in vitro | Solution is fungicidal.Trichophyton mentagrophytesTesting conducted in vitro | Identical |
| Use DilutionMethodAOAC,Official | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosa | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosa | Identical |
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
| Methods955.14, 955.15,964.02 | Testing conducted in vitro | Testing conducted in vitro | |
| EPA ViricidalTesting(DIS/TSS-7,Nov. 1981) | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Identical |
| TuberculocidalActivityAscenziQuantitativeSuspensionTest | Solution is tuberculocidalMycobacterium terraeTesting conducted in vitro | Solution is tuberculocidalMycobacterium terraeTesting conducted in vitro | Identical |
| Simulated-UseTest | Meets efficacy requirement.>= 6 log reduction Geobacillusstearothermophilus spores in amanual application | Meets efficacy requirement.>= 6 log reduction Geobacillusstearothermophilus spores in amanual application | Identical |
| Clinical In-Use | No surviving microorganismson representative medicaldevices tested | No surviving microorganisms onrepresentative medical devicestested | Identical |
| Biocompatibility | |||
| CytotoxicityDevice Extracts | Two rinses with UV treated,dual 0.1-micron membranefiltered water effectively reducesterilant residues to safe levels. | Two rinses with UV treated, dual0.1-micron membrane filteredwater effectively reduce sterilantresidues to safe levels. | Identical |
| ResidueReduction | Automatic within the SYSTEM1E Processor:Two rinses with UV treated,dual 0.1-micron membranefiltered water effectively reducesterilant residues to safe levels. | Automatic within the SYSTEM1E Processor:Two rinses with UV treated, dual0.1-micron membrane filteredwater effectively reduce sterilantresidues to safe levels. | Identical |
| DeviceMaterialCompatibility | Compatible with medicaldevices as established bytesting finished flexibleendoscopes through 300 cyclesand rigid devices through 150cycles.No functional changes haveoccurred to flexible devices.Some materials show cosmeticchanges such as fading of blackanodized aluminum withoutharm to the base material. | Compatible with medicaldevices as established by testingfinished flexible endoscopesthrough 300 cycles and rigiddevices through 150 cycles.No functional changes haveoccurred to flexible devices.Some materials show cosmeticchanges such as fading of blackanodized aluminum withoutharm to the base material. | Identical |
{24}------------------------------------------------
1 Block, S. ed., Disinfection, Sterilization, and Preservation. 5th edition, 2001.
2 Clapp et al., Free Rad. Res., (1994) 21:147-167.
3 Maillard et. al., J. Med. Microbiol (1995) 42:415-420.
4 Maillard et. al., J. Appl Bacteriol (1996) 80:540-554.
5 McDonnell et al., J. AOAC International (2000) 83:269-276.
{25}------------------------------------------------
{26}------------------------------------------------
6. Description of Safety and Substantial Equivalence
The SYSTEM endo Liquid Chemical Sterilant Processing System is the same as the predicate device as identified in Tables 1 and 2.
The proposed device and its predicate have identical intended use and technological characteristics. Testing was performed to evaluate the modifications and demonstrate substantial equivalence to the predicate as summarized in Table 3.
Table 3. Performance Testing
| Test | Acceptance Criteria | Result |
|---|---|---|
| Software validation perIEC 62304 | The firmware was validated anddetermined to operate effectively and asdesigned | Pass |
| Functional Tests withreplacement varistor | The modifications do not affect theoperation of the device. | Pass |
7. Summary of Clinical test
Clinical test is not applicable.
8. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective, and performs at least as well or better than the legally marketed predicate device (K192929), Class II (21 CFR 880.6885), product code MED.
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image shows the logo for STERIS. The word "STERIS" is written in a serif font in all capital letters. Below the word is a graphic of several horizontal wavy lines in blue. The lines are stacked on top of each other and appear to represent water.
510(k) Summary For SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Jennifer Nalepka Contact: Lead Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198
Summary Date: April 9, 2021
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{28}------------------------------------------------
1. Device Name
| Trade Name: | SYSTEM 1 endo Liquid Chemical SterilanProcessing System, Model P6900 |
|---|---|
| Device Class: | Class 2 |
| Common/usual Name: | Liquid Chemical Sterilizer |
| Classification Name: | Sterilant, Medical devices, Liquid ChemicalSterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885 |
| Product Code: | MED |
2. Predicate Device
SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, K192929.
3. Description of Device
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible semi-critical heat-sensitive medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe the modification for obsolescence and replacement of varistor (the firmware update is not required on this model).
The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. After LCS processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.
The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels. Tables 1 and 2 compare the proposed and predicate devices.
{29}------------------------------------------------
Indications for Use 4.
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical heatsensitive medical devices and their accessories in healthcare facilities.
The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.
The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.
5. Technological Characteristic Comparison Table
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System (LCSPS) is the same as the predicate device; the submission is for modifications to the SYSTEM 1 endo LCSPS software. A comparison between the proposed and predicate devices can be found in Table 1 and Table 2 below.
| Feature | Proposed DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900 | Predicate DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900(K192929) | Comparison |
|---|---|---|---|
| IntendedUseIndicationsfor Use | The SYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization of cleaned,immersible, and reusable semi-critical heat-sensitive medicaldevices and their accessories inhealthcare facilities.The SYSTEM 1 endo Processorautomatically dilutes the S40Sterilant Concentrate to its usedilution (> 1820 mg/L peraceticacid), liquid chemically sterilizesthe load during a controlled 6-minute exposure at 45.5 to 60°C,and rinses the load with 0.2micron filtered water. | The SYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem is intended for liquidchemical sterilization of cleaned,immersible, and reusable semi-critical heat-sensitive medicaldevices and their accessories inhealthcare facilities.The SYSTEM 1 endo Processorautomatically dilutes the S40Sterilant Concentrate to its usedilution (> 1820 mg/L peraceticacid), liquid chemically sterilizesthe load during a controlled 6-minute exposure at 45.5 to 60°C,and rinses the load with 0.2micron filtered water. | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900 | Predicate DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900(K192929) | Comparison |
| Concentrate to liquid chemicallysterilize medical devices. | Concentrate to liquid chemicallysterilize medical devices. | ||
| OperatingPrinciples/Technology | A microprocessor controlled unit with interchangeable processing trays/containers. The processor lid opens to reveal the processing chamber in which the load is placed. Devices with internal lumens are interfaced with the processor using connectors, i.e. Quick Connects. S40 Sterilant is placed in a specialized compartment and when the processor fills with water, it creates the sterilant use dilution from the single use sterilant cup. The processor monitors and controls the use dilution temperature and contact time. The processor automatically rinses the load with 0.2 micron filtered water to remove sterilant residuals. | A microprocessor controlled unit with interchangeable processing trays/containers. The processor lid opens to reveal the processing chamber in which the load is placed. Devices with internal lumens are interfaced with the processor using connectors, i.e. Quick Connects. S40 Sterilant is placed in a specialized compartment and when the processor fills with water, it creates the sterilant use dilution from the single use sterilant cup. The processor monitors and controls the use dilution temperature and contact time. The processor automatically rinses the load with 0.2 micron filtered water to remove sterilant residuals. | Identical |
| ProcessParameters | Standardized cycle parameters cannot be altered by operator. The critical process parameters are:Use dilution contact time Use dilution temperature Peracetic acid concentration Integrity of the internal water filter (tested by the system) | Standardized cycle parameters cannot be altered by operator. The critical process parameters are:Use dilution contact time Use dilution temperature Peracetic acid concentration Integrity of the internal water filter (tested by the system) | Identical |
| ProcessMonitors: | Cycle Printout documents successful cycle completion or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating | Cycle Printout documents successful cycle completion or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900 | Predicate DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900(K192929) | Comparison |
| DesignFeatures | • Alarms if conductivity probeindicated conductivityspecification not met• Alarms if pressure transducerindicates circulation pressureis out of specification inDiagnostic cycle• Alarms if pressure transducerindicates internal water filterfailed integrity test• Microprocessor controlledunalterable and standardizedliquid chemical sterilizationand Diagnostic cycles• Intended for use only withS40 Sterilant Concentrate• Automated dilution anddelivery of S40 Sterilant• Processor provides 0.2 micronfiltered water for liquidchemical sterilization andrinsing• Make-up air for processorduring drain sequences isfiltered through a 0.2 micronmembrane air filter• Includes a barcode scanner;employs touchscreen displayinterface; has USB drive forelectronic cycle download;facilitates use of a web-baseddata management system.• Separate, optional printer | • Alarms if conductivity probeindicated conductivityspecification not met• Alarms if pressure transducerindicates circulation pressureis out of specification inDiagnostic cycle• Alarms if pressure transducerindicates internal water filterfailed integrity test• Microprocessor controlledunalterable and standardizedliquid chemical sterilizationand Diagnostic cycles• Intended for use only withS40 Sterilant Concentrate• Automated dilution anddelivery of S40 Sterilant• Processor provides 0.2 micronfiltered water for liquidchemical sterilization andrinsing• Make-up air for processorduring drain sequences isfiltered through a 0.2 micronmembrane air filter• Includes a barcode scanner;employs touchscreen displayinterface; has USB drive forelectronic cycle download;facilitates use of a web-baseddata management system.• Separate, optional printer | Identical |
| Incomingwater temp. | $≥ 43°C$ | $≥ 43°C$ | Identical |
| Temperatureto startsterilantexposure | $≥ 46°C$ | $≥ 46°C$ | Identical |
| Temperaturealarm pointduring LCSexposure | $< 45.5 or > 60°C$ | $< 45.5 or > 60°C$ | Identical |
| Temperature | 46 - 55°C | 46 - 55°C | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900 | Predicate DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900(K192929) | Comparison |
| range oftypical LCScycle | |||
| ExposureTime - S40use dilution | 6 minutes | 6 minutes | Identical |
| Rinse waterpreparation | Hot potable tap water• is pre-filtered• is filtered through 0.2 micronbacterial retentive membranefilter | Hot potable tap water• is pre-filtered• is filtered through 0.2 micronbacterial retentive membranefilter | Identical |
| Number ofrinses | 2 | 2 | Identical |
| Air Purge | Aids in removing excess waterfrom instrument lumens afterrinsing | Aids in removing excess waterfrom instrument lumens afterrinsing | Identical |
| InternalWater FilterIntegrity Test | Conducted during the Diagnosticcycle | Conducted during the Diagnosticcycle | Identical |
| ApproximateCycle Time | 18 - 20 minutes | 18 - 20 minutes | Identical |
| DiagnosticCycle | Performs 14 tests on processor'ssystems confirming properfunction. Recommended toperform each day of use. After afailed Diagnostic cycle, a liquidchemical sterilization cyclecannot be performed until theproblem is rectified and asuccessful Diagnostic cycle hasbeen completed. | Performs 14 tests on processor'ssystems confirming properfunction. Recommended toperform each day of use. After afailed Diagnostic cycle, a liquidchemical sterilization cyclecannot be performed until theproblem is rectified and asuccessful Diagnostic cycle hasbeen completed. | Identical |
| Accessories | |||
| Sterilant | Uses S40 Sterilant Concentrate | Uses S40 Sterilant Concentrate | Identical |
| ProcessingTrays andContainers | Uses interchangeable processingtrays/containers• Universal Flex Processing Tray• General Processing Container& Tray• Directed Flow ProcessingContainer & Tray• Flexible Endoscope ProcessingContainer & Tray• Ultrasound Processing Tray | Uses interchangeable processingtrays/containers• Universal Flex Processing Tray• General Processing Container& Tray• Directed Flow ProcessingContainer & Tray• Flexible Endoscope ProcessingContainer & Tray• Ultrasound Processing Tray | Identical |
| QuickConnects | Uses Quick Connects to attachinstrument lumens to the | Uses Quick Connects to attachinstrument lumens to the | Identical |
| Feature | Proposed DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900 | Predicate DeviceSYSTEM 1 endo LiquidChemical Sterilant ProcessingSystem, Model P6900(K192929) | Comparison |
| Tray/Container ports | Tray/Container ports | ||
| ChemicalIndicator | VERIFY Chemical Indicator forS40 Sterilant is available for usein SYSTEM 1 endo LCSPS | VERIFY Chemical Indicator forS40 Sterilant is available for usein SYSTEM 1 endo LCSPS | Identical |
| Spore TestStrip | VERIFY Spore Test Strip for S40Sterilant for use in SYSTEM 1endo LCSPS | VERIFY Spore Test Strip for S40Sterilant for use in SYSTEM 1endo LCSPS | Identical |
| OperatorMaintenance | Periodic replacement of waterfilters and air filter.Periodic replacement of printertape, if using the external printeroption. | Periodic replacement of waterfilters and air filter.Periodic replacement of printertape, if using the external printeroption. | Identical |
Table 1. Processor Comparison Table
{30}------------------------------------------------
{31}------------------------------------------------
{32}------------------------------------------------
{33}------------------------------------------------
Table 2. S40 Sterilant Concentrate Comparison Table
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
|---|---|---|---|
| Indications forUse | The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquid chemicallysterilize medical devices. | The SYSTEM 1E Processoruses only S40 SterilantConcentrate to liquid chemicallysterilize medical devices. | Identical |
| Germicidalclaim | Liquid Chemical Sterilant | Liquid Chemical Sterilant | Identical |
| GermicideExposureTime (min) forintended use | 6 | 6 | Identical |
| UseTemperature | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at <43°C | 45.5-60°C – allowable46-55°C - typicalPotency and simulated useevaluations conducted at <43°C | Identical |
| Reuse | Single use | Single use | Identical |
| HumanFactors | Dispensed ready to use.Container is opened and dilutedby the processor, thus limitinguser exposure to the activeingredient | Dispensed ready to useContainer is opened and dilutedby the processor, thus limitinguser exposure to the activeingredient | Identical |
| ActiveIngredient | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the SYSTEM 1EProcessor. | 35% peroxyacetic (peracetic)acid automatically diluted foruse in the SYSTEM 1EProcessor. | Identical |
| Mode ofAction | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydryl and sulfur | It is believed that peracetic acidexerts its germicidal effect byseveral mechanisms:-oxidizing sulfhydryl and sulfur | Identical |
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
| bonds in proteins and enzymes,particularly in the cell walls 1-hydroxyl radicals producedfrom PAA are bactericidal 2-PAA damages the viral capsidand viral nucleic acid3,4. | bonds in proteins and enzymes,particularly in the cell walls1-hydroxyl radicals produced fromPAA are bactericidal2-PAA damages the viral capsidand viral nucleic acid3,4 | ||
| Rinses | Automatic, UV-irradiated, dual0.1 micron filtered, potable hotwater. | Automatic, UV-irradiated, dual0.1 micron filtered, potable hotwater. | Identical |
| Microbial Efficacy | |||
| SporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacy requirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements5.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| ConfirmatorySporicidalActivity ofDisinfectantsAOACOfficialMethod 966.04 | Meets efficacy requirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Meets efficacy requirements6.Bacillus subtilisClostridium sporogenesTesting conducted in vitro | Identical |
| FungicidalActivity ofDisinfectantsAOACOfficialMethod 955.17 | Solution is fungicidal.Trichophyton mentagrophytesTesting conducted in vitro | Solution is fungicidal.Trichophyton mentagrophytesTesting conducted in vitro | Identical |
| Use DilutionMethodAOAC,OfficialMethods955.14, 955.15,964.02 | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Solution is bactericidal.Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaTesting conducted in vitro | Identical |
| EPA ViricidalTesting(DIS/TSS-7,Nov. 1981) | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Solution is viricidal.Herpes simplex Type 1Adenovirus Type 5Poliovirus Type 1Testing conducted in vitro | Identical |
| Feature | Proposed DeviceS40 Sterilant Concentrate | Predicate DeviceS40 Sterilant Concentrate(K192929) | Comparison |
| ActivityAscenziQuantitativeSuspensionTest | Mycobacterium terraeTesting conducted in vitro | Mycobacterium terraeTesting conducted in vitro | |
| Simulated-UseTest | Meets efficacy requirement.> 6 log reduction Geobacillusstearothermophilus spores in amanual application | Meets efficacy requirement.> 6 log reduction Geobacillusstearothermophilus spores in amanual application | Identical |
| Clinical In-Use | No surviving microorganismson representative medicaldevices tested | No surviving microorganisms onrepresentative medical devicestested | Identical |
| Biocompatibility | |||
| CytotoxicityDevice Extracts | Two rinses with UV treated,dual 0.1-micron membranefiltered water effectively reducesterilant residues to safe levels. | Two rinses with UV treated, dual0.1-micron membrane filteredwater effectively reduce sterilantresidues to safe levels. | Identical |
| ResidueReduction | Automatic within the SYSTEM1E Processor:Two rinses with UV treated,dual 0.1-micron membranefiltered water effectively reducesterilant residues to safe levels. | Automatic within the SYSTEM1E Processor:Two rinses with UV treated, dual0.1-micron membrane filteredwater effectively reduce sterilantresidues to safe levels. | Identical |
| DeviceMaterialCompatibility | Compatible with medicaldevices as established bytesting finished flexibleendoscopes through 300 cyclesand rigid devices through 150cycles.No functional changes haveoccurred to flexible devices.Some materials show cosmeticchanges such as fading of blackanodized aluminum withoutharm to the base material. | Compatible with medicaldevices as established by testingfinished flexible endoscopesthrough 300 cycles and rigiddevices through 150 cycles.No functional changes haveoccurred to flexible devices.Some materials show cosmeticchanges such as fading of blackanodized aluminum withoutharm to the base material. | Identical |
{34}------------------------------------------------
1 Block, S. ed., Disinfection, Sterilization, and Preservation. 5th edition, 2001.
2 Clapp et al., Free Rad. Res., (1994) 21:147-167.
3 Maillard et. al., J. Med. Microbiol (1995) 42:415-420.
4 Maillard et. al., J. Appl Bacteriol (1996) 80:540-554.
- Mainard on an, of Fippi Baoterior (1990) 0018 10 08 10 08 11
5 McDonnell et al., J. AOAC International (2000) 83:269-276.
{35}------------------------------------------------
6. Description of Safety and Substantial Equivalence
The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 is the same as the predicate device described in this submission and identified in Tables 1 and 2.
{36}------------------------------------------------
The proposed device and its predicate have identical intended use and technological characteristics. New testing was performed to evaluate the modified device and the results are summarized in Table 3.
| Test | Acceptance Criteria | Result |
|---|---|---|
| Functional tests withreplacement varistor | The modification does not affect theoperation of the device. | Pass |
Table 3. Summarv of verification activities
7. Summary of Clinical test
Clinical test is not applicable.
8. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective, and performs at least as well or better than the legally marketed predicate device (K192929), Class II (21 CFR 880.6885), product code MED.
§ 880.6885 Liquid chemical sterilants/high level disinfectants.
(a)
Identification. A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.(b)
Classification. Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.