K Number
K210445
Date Cleared
2021-06-29

(133 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nordiwell Medical Face Mask is intended to be worn to protect both patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate matters. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device(s) and provided non-sterile.

Device Description

The Nordiwell Medical Face Mask has a flat-folded design consisting of three layers that are comprised of an (polypropylene spunbond) inner and outer cover web and (polypropylene melt blown) filter web. The mask has earloops or tie strings to secure the mask over the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. This is a single use, disposable device, provided non-sterile.

AI/ML Overview

This document is a 510(k) Premarket Notification for a medical face mask. It does not describe an AI/ML powered medical device, but rather a physical product. Therefore, many of the requested criteria regarding AI/ML model evaluation (e.g., training set, ground truth acquisition, MRMC studies, standalone performance with human-in-the-loop, number of experts for ground truth) are not applicable.

However, I can extract the acceptance criteria and performance data for this physical medical device from the provided text.

Here's the information based on the provided document:

1. A table of acceptance criteria and the reported device performance

Test ItemAcceptance CriteriaReported Device PerformanceStudy Conclusion
Fluid Resistance Performance (ASTM F1862-17)≥ 29 of 32 pass at 120 mmHg31 of 32 pass at 120 mmHgPass
Particulate Filtration Efficiency (ASTM F2299-17)≥ 98%99.9%, 32 of 32 passPass
Bacterial Filtration Efficiency (ASTM F2101-19)≥ 98%99.9%, 32 of 32 passPass
Differential Pressure (Delta P) (EN 14683:2019, Annex C and ASTM F2100-19)< 6.0 mmH2O/cm²4.6 mmH2O/cm², 32 of 32 passPass
Flammability (16 CFR 1610)Class 1Class 1, 32 of 32 passPass
Biocompatibility (ISO 10993 series)Non-Cytotoxic, Non-Irritating, Non-SensitizingNon-Cytotoxic, Non-Irritating, Non-SensitizingPass

2. Sample size used for the test set and the data provenance

  • Sample Size for performance tests: For each performance test (Fluid Resistance, Particulate Filtration, Bacterial Filtration, Differential Pressure, Flammability), the sample size was 32 units. The document also states "Bench test was conducted on 3 nonconsecutive lots of Medical Face Mask". This implies 32 units per lot, or a total of 32 units drawn across 3 nonconsecutive lots. The wording "32 of 32 pass" for each specific test strongly suggests that the sample size for each test was 32.
  • Data Provenance: The tests were conducted internally by the manufacturer (Changzhou Combat Protective Equipment Co., Ltd.) in China, as indicated by the contact information and location. The studies are essentially prospective in nature, as they involve testing newly manufactured devices to verify performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is a physical product, not an AI/ML device that requires human expert annotation for ground truth. The "ground truth" here is the result of standardized physical and biological tests conducted against established ASTM and ISO standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. This is a physical product. Test results are objective measurements from laboratory equipment and procedures, not subjective human interpretations requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI/ML medical device. No human reader studies were conducted as the device is a medical face mask.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI/ML medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for this medical device's performance is based on established international and national standards (ASTM F1862-17, ASTM F2299-17, ASTM F2101-19, EN 14683:2019, 16 CFR 1610, ISO 10993 series). These standards define the test methodologies and acceptance criteria for medical face masks. The testing is performed in laboratory settings to objectively measure the device's physical and biological properties.

8. The sample size for the training set

  • Not Applicable. This is a physical product, not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

  • Not Applicable. This is a physical product, not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 29, 2021

Changzhou Combat Protective Equipment Co., Ltd. Xiaoqing Xue Quality Manager QingSiTang Village, Henglin Town, Wujin District Changzhou, Jiangsu 213101 China

Re: K210445

Trade/Device Name: Nordiwell Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 31, 2021 Received: March 31, 2021

Dear Xiaoqing Xue:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210445

Device Name Nordiwell Medical Face Mask

Indications for Use (Describe)

The Nordiwell Medical Face Mask is intended to be worn to protect both patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate matters. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device(s) and provided non-sterile.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K210445

1. Contact Information

1.1. Applicant

Applicant Name: Changzhou Combat Protective Equipment Co., Ltd. Address: Qingsitang, Henglin, Changzhou, Jiangsu, 213101 China Contact Person: Xiaoqing Xue Title: Quality Manager Telephone: +86-519-85190068; +86-15161196032 E-mail: xqxue2003@126.com

1.2. Consultant

Company: Sinow Medical AS Address: Vestre Fantoftåsen 44, 5072, Bergen, Norway Contact Person: Huifang Zhao Telephone: +86 13961151430 Email: zhao@bergemed.com Date of the 510(k) Summary Prepared: June 25, 2021

2. Device information

Trade Name: Nordiwell Medical Face Mask Common Name: Surgical Face Mask Regulatory information: Classification Name: Mask, Surgical Classification: II Product Code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: General Hospital

3. Legally Marketed Primary Predicate Device

Product name: 3MTM High Performance Surgical Mask 510(k) Number: K180874 Product Code: FXX Manufacture: 3M Health Care

{4}------------------------------------------------

4. Indication for use

The Nordiwell Medical Face Mask is intended to be worn to protect both patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate matters. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device(s) and provided non-sterile.

5. Device Description

The Nordiwell Medical Face Mask has a flat-folded design consisting of three layers that are comprised of an (polypropylene spunbond) inner and outer cover web and (polypropylene melt blown) filter web. The mask has earloops or tie strings to secure the mask over the users' mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. This is a single use, disposable device, provided non-sterile.

DeviceSubject DeviceK210445 Nordiwell MedicalFace MaskPredicate DeviceK180874 3M™ HighPerformance Surgical Mask1838RComparison
Intenduse/Indicationsfor UseThe Nordiwell Medical FaceMask is intended to be wornto protect both patient andhealthcare personnel fromtransfer of microorganisms,body fluids, and particulatematters. These face masks areintended for use in infectioncontrol practices to reducepotential exposure to bloodand body fluids. This is asingle use, disposabledevice(s) and provided non-sterile.3M™ High PerformanceSurgical Mask is intended to beworn to protect both patient andhealthcare personnel fromtransfer of microorganisms,body fluids, and particulatematerial. These face mask(s) areintended for use in infectioncontrol practices to reducepotential exposure to blood andbody fluids. This is a single use,disposable device(s) andprovided non-sterile.same

6. Technological Characteristic Comparison

{5}------------------------------------------------

Material
Outer CoverWebPolypropylene SpunbondPolypropylene Spunbond,(w Print) – Upper Top HalfPolypropylene Spunbond,White - Lower Bottom Halfsame
MiddleWeb/FilterPolypropylene Melt BlownPolypropylene Melt Blownsame
Inner CoverWebPolypropylene SpunbondPolypropylene Spunbondsame
Earloopsortie stringsSpandex and NylonPolypropylene Spunbond orPolyethylene TerephthalateDifferent¹
Nose PiecePolyethylene coated steelwirePolyethylene coated steelwiresame
Design Features
ColorsWhiteBlue - Outer Cover Web (Upper)White - Outer Cover Web (Lower)silimar
Mask Design
StyleFlat foldedDuckbillDifferent²
Single UseYesYessame
Sterility
SterileNon-SterileNon-Sterilesame
Dimension
Length160 mm ± 5 mm187 mm ± 4 mmDifferent²
Width106 mm ± 5mm107 mm ± 6 mmsimilar

{6}------------------------------------------------

Technological Characteristics:
Product Performance Specifications Per ASTM F2100 - Meets ASTM Level 2
FluidResistance31 of 32 pass at120 mmHgMeet ASTM F2100 level 2and ASTM F1862requirementsame
ParticulateFiltrationEfficiency99.9%,32 of 32 passMeet ASTM F2100 level 2and ASTM F2299requirementsame
BacterialFiltrationEfficiency99.9%,32 of 32 passMeet ASTM F2100 level 2and ASTM F2101requirementsame
Differentialpressure4.6 mmH20/cm²32 of 32 passMeet ASTM F2100 level 2and MIL-M36954Crequirementsame
FlammabilityClass 1,32 of 32 passMeet ASTM F2100 level 2and 16 CFR 1610 requirementsame
Shelf life2 yearsUnknown
Biocompatibility
ResultsBiocompatible,Non-Cytotoxic,Non-Irritating,Non-SensitizingBiocompatible,Non-Cytotoxic,Non- Irritating,Non-Sensitizingsame

Different': The patient contacting material for the proposed device is different from predicate device. However, biocompatibility test has been performed on the proposed device and the results does not show any adverse effect.

Different2:The design style and dimension for proposed device is different from predicate device, but the differences of design and size don't affect the indication for use.

7. Non-Clinical Test

Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

{7}------------------------------------------------

(1) Performance test:

Bench test was conducted on 3 nonconsecutive lots of Medical Face Mask for proposed device to determine if the subject device met the specifications in the standard. The bench tests include the following tests:

ItemProposed DeviceAcceptance CriteriaResult
Fluid Resistance PerformanceASTM F1862-1731 of 32 pass at120 mmHg≥29 of 32 pass at 120 mmHgPass
Particulate FiltrationEfficiency ASTM F2299-1799.9%,32 of 32 pass≥ 98%Pass
Bacterial Filtration EfficiencyASTM F2101-1999.9%,32 of 32 pass≥ 98%Pass
Differential Pressure(Delta P) EN 14683:2019,Annex C and ASTMF2100-19.4.6 mmH20/cm²32 of 32 pass<6.0 mmH20/cm²Pass
Flammability 16 CFR 1610Class 1,32 of 32 passClass 1Pass

(2) Biocompatibility evaluation and test

Biocompatibility evaluation conducted in accordance with FDA's 2020 guidance and

ISO10993-1:2018 supports that the subject devices are biocompatible.

The biocompatibility test includes the following tests:

In Vitro Cytotoxicity Test (ISO 10993-5: 2009)

Skin Irritation Test(ISO 10993-10:2010)

Skin Sensitization Test (ISO 10993-10:2010)

8. Clinical Test Conclusion

No clinical study is included in this submission.

9. Conclusion

The conclusion drawn from the non-clinical tests demonstrate that the subject device in 510(k) submission K210445 is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K180874.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.