K Number
K210190
Date Cleared
2021-02-23

(29 days)

Product Code
Regulation Number
870.5150
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClotTriever Thrombectomy System is indicated for :

  • · The non-surgical removal of thrombi and emboli from blood vessels.
  • · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

Device Description

The ClotTriever Thrombectomy System is a single-use, sterile medical device designed for use in the peripheral vasculature. The ClotTriever Thrombectomy System consists of the ClotTriever Sheath and the ClotTriever Catheter. The ClotTriever Sheath comprises a reinforced polymeric sheath equipped with a self-expanding distal mesh funnel, a flush/aspiration port, and a proximal hemostatic valve. A dilator is provided to aid insertion. Other provided accessories include the funnel loading tool and a Large Bore 60 cc syringe. The ClotTriever Catheter comprises reinforced polymeric coaxial shafts terminating in an expandable coring element and thrombus collection bag. Two ports terminating in stopcocks are provided for de-airing the catheter shafts. To aid in fluoroscopic visualization, the Sheath dilator and ClotTriever Catheter distal tips are radiopaque.

AI/ML Overview

This document is a 510(k) summary for the Inari Medical ClotTriever Thrombectomy System. It outlines the FDA's determination of substantial equivalence for a modified version of the device.

Based on the provided text, the device in question is not an AI/ML powered device. It's a physical medical device designed for the non-surgical removal of thrombi and emboli. Therefore, the questions related to AI/ML device performance, such as acceptance criteria based on AI metrics, sample sizes for test/training sets for AI, expert adjudication, MRMC studies, and standalone algorithm performance, are not applicable to this document.

The document states:

  • "Non-Clinical Testing: Non-clinical testing was not required to support the change to the Indications for Use."
  • "Clinical Testing: The CLOUT registry clinical data supporting this labeling change was previously submitted in the ClotTriever Thrombectomy System 510(k) (K193462, cleared on September 9, 2020). No additional clinical data was deemed necessary to determine substantial equivalence of the subject and predicate devices."

This means that the current 510(k) submission (K210190) is for a labeling change (specifically, removing "chronic clot" from a contraindication) for an already cleared device, and no new performance testing (clinical or non-clinical) was conducted or required for this specific submission. The substantial equivalence determination is based on the previously submitted data for K193462.

Therefore, I cannot provide the requested information about acceptance criteria and study details for an AI/ML device from this document. The document describes a traditional medical device and a submission for a minor labeling modification.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 23, 2021

Inari Medical, Inc. Larry Boucher Regulatory Affairs Manager 9 Parker. Suite 100 Irvine, California 92618

Re: K210190

Trade/Device Name: ClotTriever Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: January 22, 2021 Received: January 25, 2021

Dear Larry Boucher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210190

Device Name ClotTriever Thrombectomy System

Indications for Use (Describe) The ClotTriever Thrombectomy System is indicated for :

  • · The non-surgical removal of thrombi and emboli from blood vessels.
  • · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210190

PAGE 1 OF 2

510(K) SUMMARY

Date preparedJanuary 22, 2021
NameInari Medical, Inc.9 Parker, Suite 100Irvine, CA 92618949.600.8433
Contact personLarry BoucherRegulatory Affairs Manager
Trade nameClotTriever Thrombectomy System
Common nameEmbolectomy catheter
Regulation nameEmbolectomy catheter
Classification number21 CFR 870.5150
Product codeQEW
Regulatory classII
Predicate deviceClotTriever Thrombectomy System (K193462)
DescriptionThe ClotTriever Thrombectomy System is a single-use, sterile medical devicedesigned for use in the peripheral vasculature. The ClotTriever ThrombectomySystem consists of the ClotTriever Sheath and the ClotTriever Catheter. TheClotTriever Sheath comprises a reinforced polymeric sheath equipped with a self-expanding distal mesh funnel, a flush/aspiration port, and a proximal hemostaticvalve. A dilator is provided to aid insertion. Other provided accessories includethe funnel loading tool and a Large Bore 60 cc syringe. The ClotTriever Cathetercomprises reinforced polymeric coaxial shafts terminating in an expandablecoring element and thrombus collection bag. Two ports terminating in stopcocksare provided for de-airing the catheter shafts. To aid in fluoroscopic visualization,the Sheath dilator and ClotTriever Catheter distal tips are radiopaque.
Indications for UseThe ClotTriever Thrombectomy System is indicated for:• The non-surgical removal of thrombi and emboli from blood vessels.• Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).
Device modificationsThe labeling has been changed by deleting “chronic clot” from the examples given “…for removal fibrous, adherent, or calcified material…” from the contraindications.
Summary of substantial equivalenceThe ClotTriever Thrombectomy System and predicate device has the same intended use: removal of thrombus and emboli from and infusion of fluids into the peripheral vasculature.Non-Clinical TestingNon-clinical testing was not required to support the change to the Indications for Use.Clinical TestingThe CLOUT registry clinical data supporting this labeling change was previously submitted in the ClotTriever Thrombectomy System 510(k) (K193462, cleared on September 9, 2020). No additional clinical data was deemed necessary to determine substantial equivalence of the subject and predicate devices.Removing “chronic clot” from the example “of fibrous, adherent, or calcified material” does not introduce new issues of safety or effectiveness.ConclusionRemoving “chronic clot” from the example “of fibrous, adherent, or calcified material” of the ClotTriever Thrombectomy System aligns with the currently cleared Indications for Use and does not raise new or different questions of safety or and effectiveness. It can be concluded that the ClotTriever System with the modified contraindication is substantially equivalent to the predicate device.

{4}------------------------------------------------

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).