K Number
K210168
Date Cleared
2021-05-10

(108 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diode Laser Therapy Systems are intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

The proposed device, Diode Laser Therapy System, is a surgical device. It utilizes a semiconductor diode as a laser source (808nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a footswitch.

The treatment can be applied on different Fitzpatrick skin type, including I (White), II (White with pigment), III (Yellow), IV (Yellow with pigment), V (Brown) and VI (Black); in addition, the treatment can also be applied to different parts of the body, such as arm, chest, leg, underarm etc.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called "Diode Laser Therapy Systems," which is intended for hair reduction. It details the device, its indications for use, and a comparison to predicate and reference devices.

However, the document does not contain information about acceptance criteria, the study design, ground truth establishment, or performance metrics typically associated with proving a device meets acceptance criteria through clinical or standalone studies for an AI/algorithm-based medical device.

Specifically, the document states: "No clinical study is included in this submission." This means there is no clinical performance data for this specific device in this submission to draw from to describe acceptance criteria or study results as requested in the prompt. The "study" mentioned in the prompt refers to a performance evaluation (often a clinical trial or a robust standalone study for AI devices).

The comparison table (Table 1: Comparison of Technology Characteristics) on page 6 primarily focuses on technical specifications (e.g., laser wavelength, spot size, fluence, electrical safety) and regulatory compliance with standards like ISO and IEC, rather than clinical performance metrics or AI algorithm performance.

Therefore, I cannot fulfill most of the request based on the provided text, as the information regarding clinical performance studies, acceptance criteria for such studies, and all related details (sample size, experts, ground truth, MRMC, standalone performance, training set details) is not present.

The document concludes that the device is "Substantially Equivalent (SE)" based on non-clinical tests and technological comparisons to predicate devices, not on the basis of a clinical performance study with defined acceptance criteria and tested results.

To answer your request based on the provided text, I must state that the information is absent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

May 10, 2021

Beijing Kes Biology Technology CO., LTD. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai, 200120 China

Re: K210168

Trade/Device Name: Diode Laser Therapy Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 30, 2021 Received: April 2, 2021

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K210168

Device Name Diode Laser Therapy Systems

Indications for Use (Describe)

The Diode Laser Therapy Systems are intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K210168

    1. Date of Preparation: 3/30/2021
    1. Sponsor Identification

BEIJING KES BIOLOGY TECHNOLOGY CO., LTD.

No.15, Zone 2, Mafang Industrial Park, PingguyuanZhongguancun Science And Technology Park Pinggu District, Beijing CHINA 101204

Establishment Registration Number: 3009444646

Contact Person: Wei Meng Meng Position: QS Specialist Tel: +86-10-60998698 Fax: +86-10-61227575 Email: weimengmeng@keslaser.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Diode Laser Therapy Systems Common Name: Powered Laser Surgical Instrument

Regulatory Information

Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810 Review Panel: General & Plastic Surgery;

Indication for use:

The Diode Laser Therapy Systems are intended for hair removal permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description:

The proposed device, Diode Laser Therapy System, is a surgical device. It utilizes a semiconductor diode as a laser source (808nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a footswitch.

The treatment can be applied on different Fitzpatrick skin type, including I (White), II (White with pigment), III (Yellow), IV (Yellow with pigment), V (Brown) and VI (Black); in addition, the treatment can also be applied to different parts of the body, such as arm, chest, leg, underarm etc.

  • રું ર Identification of Predicate Device
    510(k) Number: K180353 Product Name: Diode laser hair removal device

  • б. Identification of Reference Device
    Reference Device 1 510K Number: K182924 Trade Name: Diode Laser Treatment System

{5}------------------------------------------------

Manufacturer: Weifang KM Electronics Co., Ltd.

Reference Device 2 510K Number: K193426 Trade Name: Elite iQ™ Manufacturer: Cynosure, LLC

Note: The Elite iQTM including two models, 755nm and 1064nm. The 755nm of Elite iQM was selected as reference device.

    1. Non-Clinical Test Conclusion
      Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
  • ISO 10993-5:2009 Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity.

  • ISO 10993-10: 2010 Biological Evaluation of Medical Device, Part 10-Test for irritation and skin sensitization.

  • ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and A2:2010/ (R) 2012 Medical electrical equipment - Part1: General requirements for basic safety and essential performance

  • IEC 60601-2-22:2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.

  • A IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements.

  • IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

  • Clinical Test Conclusion 8.

No clinical study is included in this submission.

{6}------------------------------------------------

9. Substantially Equivalent (SE) Comparison

ITEMProposed DevicePredicate DeviceReference Device 1Reference Device 2
Product CodeGEXK180353K182924K193426
RegulationNo.GEXGEXGEXGEX
21 CFR 878.481021 CFR 878.481021 CFR 878.481021 CFR 878.4810
ClassIIIIIIII
Indication forUseThe Diode Laser TherapySystems are intended for hairremoval permanent hairreduction on all skin types(Fitzpatrick skin type IVI),including tanned skin.Permanent hair reduction isdefined as the long-term,stable reduction in the numberof hairs regrowing whenmeasured at 6, 9, and 12months after the completion ofa treatment regime.The Diode Laser HairRemoval device is intendedfor hair removal, permanenthair reduction on all skin types(Fitzpatrick skin type I-VI),including tanned skin.Permanent hair reduction isdefined as the long-term,stable reduction in the numberof hairs regrowing whenmeasured at 6, 9, and 12months after the completion ofa treatment regime.The Diode Laser TreatmentSystem is intended for hairremoval, permanent hairreduction on all skin types(Fitzpatrick skin type I-VI),including tanned skin.Permanent hair reduction isdefined as the long-term,stable reduction in the numberof hairs regrowing whenmeasured at 6, 9, and 12months after the completion ofa treatment regimeThe Elite iQ Laser System is indicatedfor stable long-term or permanent hairreduction. Permanent hair reduction isdefined as long-term stable reduction inthe number of hairs regrowing whenmeasured at 6, 9 or 12 months after thecompletion of a treatment regime. It isused for skin types (Fitzpatrick I-VI)including tanned skin. It is alsoindicated for the treatment of vascularlesions, benign pigmented lesions, andwrinkles.
ConfigurationMain UnitMain UnitMain UnitMain Unit
HandpieceHandpieceHandpieceHandpiece
Foot ControlFoot Control/Foot Control and Finger Control
Laser TypeDiode LaserDiode LaserDiode LaserAlexandrite
LaserClass IVClass IVClass IVUnknown
Classification
LaserWavelength808 nm808 nm808nm755nm
Spot Size12 mm×12 mm=1.44 cm²1.44 cm²1.44 cm²2.5mm, 5mm, 7mm, 10mm, 12mm,15mm, 18mm, 20mm, 22mm, & 24mm
Fluence10-125 J/cm²1-100 J/ cm²2-120 J/ cm²Up to 600J/ cm²
Frequency1-10 Hz1-20 Hz1-10Hz0.5~10Hz
PulseDuration10-400 ms10-400 ms10-300ms0.5~300ms
Power Supply99V-121V, 50/60Hz1400VAAC 110V-230V/50-60Hz2000VA100-240V AC, 50/60Hz208V240V/5060Hz5500VA
Dimension450×430×1000mm560×380×1180mm600×420×380mm600×420×350mm600×420×400mm550×420×300mmUnknown
Weight52kg60 Kg35KgUnknown
PatientContactingMaterialAluminum alloy, ABSSapphire in handpieceSapphire in handpiece andhandpiece tip (stainless steel)316 Stainless Steel
Biocompatibility
CytotoxicityNo CytotoxicityNo CytotoxicityNo CytotoxicityComply with ISO 10993-1
SensitizationNo evidence of SensitizationNo evidence of SensitizationComply with ISO 10993-1Comply with ISO 10993-1
IrritationNo evidence of IrritationNo evidence of IrritationComply with ISO 10993-1Comply with ISO 10993-1
Electrical Safety and EMC
ElectricalSafetyComply withIEC 60601-1IEC 60601-2-22Comply withIEC 60601-1IEC 60601-2-22Comply withIEC 60601-1IEC 60601-2-22Comply withIEC 60601-1IEC 60601-2-22
IEC 60825-1IEC 60825-1IEC 60825-1IEC 60825-1
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Comply with IEC 60601-1-2Comply with IEC 60601-1-2

Table 1 Comparison of Technology Characteristics

{7}------------------------------------------------

510(k) Summary

{8}------------------------------------------------

510(k) Summary

  1. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.