K Number
K210150
Date Cleared
2021-06-02

(132 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Disposable Medical Masks are single use, three-layer, flat –folded masks with ear loops and nose clip. The Disposable Medical Masks are manufactured with three layers, the inner and outer layers are made of non-woven Spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two strings welded to the facemask. The ear loops are made of 17cm spandex elastic strings(performance of the spandex elastic: Tensile strength greater than or equal to 100N; Elasticity greater than or equal to 2.8times; Breaking tension greater than 100N). The nose clip in the layers of facemask is to allow the user to fit the facemask around their nose, which is not touch with the users' skin directly. The Disposable Medical Masks will be provided in blue. The Disposable Medical Masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

This 510(k) premarket notification describes the non-clinical performance evaluation of Hubei Wanli Protective Products Co., Ltd.'s "Disposable Medical Mask" (K210150) to demonstrate its substantial equivalence to the predicate device, the Kimberly-Clark KC100 Mask (K110455).

  1. Acceptance Criteria and Reported Device Performance:

    ItemAcceptance Criteria (Level 2)Reported Device Performance (Lot2020120501, Lot2020120101, Lot2020120801)
    Synthetic Blood Performance (ASTM F1862)29 out of 32 pass at 120 mmHg32 out of 32 pass at 120 mmHg
    Particulate Filtration Efficiency (ASTM F2299)≥ 98%≥ 98%
    Bacterial Filtration Efficiency (ASTM F2101-19, EN 14683:2019 Annex B)≥ 98%≥ 98%
    Differential Pressure (Delta P) (ASTM F2100-19, EN 14683:2019 Annex C)< 6.0 mmH₂O/cm²< 6.0 mmH₂O/cm²
    Flammability (16 CFR 1610)Class 1 (Burn time ≥ 3.5 seconds)Class I
    Cytotoxicity (ISO 10993-5)Non-CytotoxicNo Cytotoxic
    Irritation (ISO 10993-10)Non-IrritatingNo irritating
    Sensitization (ISO 10993-10)Non-SensitizingNo Sensitizing
  2. Sample size used for the test set and the data provenance:

    • Synthetic Blood Performance (ASTM F1862): 32 samples were tested for each of the three lots (Lot2020120501, Lot2020120101, Lot2020120801).
    • Other performance tests: The exact sample size for other performance tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability) is not explicitly stated as a number of discreet items, but results are provided for three distinct lots of the device.
    • Biocompatibility tests: Not explicitly stated how many samples were used for Cytotoxicity, Irritation, and Sensitization.
    • Data Provenance: The document does not specify the country of origin of the data or whether the tests were retrospective or prospective, but it's reasonable to infer they were specifically conducted for this submission (prospective) and likely conducted by or for the manufacturer in China, given the manufacturer's location.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this involves laboratory performance testing against established standards, not interpretation by human experts.
  4. Adjudication method for the test set:

    • Not applicable as this involves objective laboratory measurements against set standards, not subjective expert judgment.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No. This type of study is not applicable to a medical mask, which is evaluated based on physical performance characteristics rather than human interpretation of medical images or data.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is not an AI/algorithm-based diagnostic or assistive system. Its performance is intrinsic to its physical properties.
  7. The type of ground truth used:

    • Objective performance standards: The "ground truth" for the performance tests consists of established, internationally recognized standards and test methods (e.g., ASTM F2100, ASTM F1862, ASTM F2101, ASTM F2299, EN 14683, 16 CFR 1610, ISO 10993). The device's performance is measured against the specified acceptance criteria defined within these standards.
  8. The sample size for the training set:

    • Not applicable. This refers to training data for machine learning models, which is irrelevant for the evaluation of a medical mask.
  9. How the ground truth for the training set was established:

    • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 2, 2021

Hubei Wanli Protective Products Co., Ltd. % Eva Li Consultant Shanghai Sungo Management Consulting Company Limited Room 1309, Dongfang Building, 1500#Century Ave Shanghai, Shanghai 200122 China

Re: K210150

Trade/Device Name: Disposable Medical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 21, 2021 Received: January 21, 2021

Dear Eva Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K210150

Device Name Disposable Medical Mask

Indications for Use (Describe)

The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Image /page/2/Picture/19 description: The image shows the text "PSC Publishing Services (301) 443-6740". The text appears to be a company name and phone number. The area code is 301 and the phone number is 443-6740.

EF

{3}------------------------------------------------

Hubei Wanli Protective Products Co., Itd. Yuanshi, Ganhe, Xiantao, Hubei, China

K210150

510(K) Summary

A. Applicant

Hubei Wanli Protective Products Co., Itd. Address: Yuanshi, Ganhe, Xiantao, Hubei, China Contact Person: Andy Wen Tel: 0086-728-3227299 Email: sale01@hbwanli.com Date Prepared: April 27, 2021

Submission Correspondent Primary contact: Ms. Eva Li Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: eatereva@hotmail.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device

Trade Name: Disposable Medical Mask

ModelDescription
WLM2002Ear Loops, Flat Pleated, 3 layers, 17.5cm*9.5cm

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

C. Predicate device:

K110455 Kimberly-Clark KC100 Mask Kimberly-Clark

D. Intended use of the device:

The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce

{4}------------------------------------------------

Hubei Wanli Protective Products Co., Itd.

Yuanshi, Ganhe, Xiantao, Hubei, China

the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

E. Device Description:

The Disposable Medical Masks are single use, three-layer, flat –folded masks with ear loops and nose clip.

The Disposable Medical Masks are manufactured with three layers, the inner and outer layers are made of non-woven Spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.

The ear loops are held in place over the users' mouth and nose by two strings welded to the facemask. The ear loops are made of 17cm spandex elastic strings(performance of the spandex elastic: Tensile strength≥100N; Elasticity≥2.8times; Breaking tension> 100N)

The nose clip in the layers of facemask is to allow the user to fit the facemask around their nose, which is not touch with the users' skin directly.

The Disposable Medical Masks will be provided in blue. The Disposable Medical Masks are sold non-sterile and are intended to be single use, disposable devices.

F. Comparison with predicate device

Table 1 General Comparison

DeviceProposed DevicePredicate DeviceComparison
ManufacturerHubei Wanli Protective ProductsCo., ltd.Kimberly-Clark
510(K) numberK210150K110455
Model NameDisposable Medical MaskKimberly-Clark KC100 Mask
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21CFR878.4040)same
Intend useThe Disposable Medical Masks areintended to be worn to protect boththe patient and healthcarepersonnel from transfer ofmicroorganisms, body fluids andparticulate material. These facemasks are intended for use ininfection control practices to reducethe potential exposure to blood andbody fluids. This is a single use,disposable device(s), provided non-sterile.The Kimberly-Clark KC100Procedure Mask(s) is intended tobe worn to protect both thepatient and healthcare personnelfrom transfer of microorganisms,body fluids and particulatematerial. These face masks areintended for use in infectioncontrol practices to reduce thepotential exposure to blood andbody fluids. The Kimberly-ClarkKC100 Procedure Mask(s) is asingle use, disposable devices,provided non-sterile.same
ModelEar Loops, Flat Pleated, 3 layersEar Loops, Tie-On, Flat Pleated, 3layerssame

{5}------------------------------------------------

Hubei Wanli Protective Products Co., Itd. Yuanshi, Ganhe, Xiantao, Hubei, China

Outerfacinglayer
Materialnon-woven Spun-bondpolypropylenenon-woven Spun-bondpolypropyleneSpun-bond polypropylenesame
MiddlelayerMelt blown polypropylene filterMelt blown polypropylene filtersame
Innerfacinglayernon-woven Spun-bondpolypropyleneSpun-bond polypropylenesame
Earloops/tiesSpandex elasticPolyester/lycra knittedsimilar
ColorBlueVariety (include blue)same
Dimension(length)$175mm \pm 5mm$$165 \pm 19mm$Similar
Dimension(width)$95 mm \pm 5mm$$102 \pm 19mm$Similar
OTC useYesYessame
SterilityNon-SterileNon-Sterilesame
UseSingle use, DisposableSingle use, Disposablesame
ASTM F2100LevelLevel 2Level 1Similar*
BiocompatibilityISO10993ISO10993same

*Similar Discussion :

The proposed device conducted the test and the pass the Level 3 Acceptance Criteria, the predicate device pass the level 1 Acceptance Criteria. The test and the acceptance is following:

G. Summary of Non-Clinical Performance Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks

ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity);

EN 14683, Medical Face Masks-Requirements and Test Methods; A

  • A ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency

{6}------------------------------------------------

Hubei Wanli Protective Products Co., ltd.

Yuanshi, Ganhe, Xiantao, Hubei, China

(BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

ASTM F2299, Standard test method for determining the initial efficiency of

materials used in medical face masks to penetration by particulates using latex spheres;

A 16 CFR 1610, Standard for the Flammability of clothing textiles;

Table 2-performance Testing

ItemAcceptanceCriteria (level2)Result ofLot2020120501Result ofLot2020120101Result ofLot2020120801
Synthetic BloodPerformanceASTM F186229 out of 32pass at 120mmHg32 out of 32pass at 120mmHg32 out of 32pass at 120mmHg32 out of 32pass at 120mmHg
Particulate FiltrationEfficiencyASTM F2299≥ 98%≥ 98%≥ 98%≥ 98%
Bacterial FiltrationEfficiencyASTM F2101-19EN 14683:2019 Annex B≥ 98%≥ 98%≥ 98%≥ 98%
DifferentialPressure(Delta P)ASTM F2100-19EN 14683:2019 Annex C( mmH₂O/cm²)< 6.0< 6.0< 6.0< 6.0
Flammability16 CFR 1610(IBE=Test Article ignited,but extinguished)Class 1 (Burntime ≥3.5seconds)Class IClass IClass I

Table3 Biocompatibility Comparison

ItemProposed deviceAcceptanceCriteriaResult
CytotoxicityUnder the conditions ofthe study, the device isnon-cytotoxic.Non-CytotoxicNo Cytotoxic
IrritationUnder the conditions ofthe study, the device isnon-irritating.Non-IrritatingNo irritating
SensitizationUnder the conditions ofthe study, the device isnon-sensitizingNon-SensitizingNo Sensitizing

{7}------------------------------------------------

Hubei Wanli Protective Products Co., ltd. Yuanshi, Ganhe, Xiantao, Hubei, China

H. Clinical Test Conclusion

No clinical study is included in this submission.

I. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Kimberly-Clark KC100 Mask cleared under K110455.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.