(25 days)
The Freudenberg Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.
The Freudenberg Surgical Mask is composed of three-layers and is flat-pleated. The Freudenberg Surgical Mask materials consist of an outer cover web (polypropylene spunbond, white), insertion layer (polypropylene, melt-blown, white), and inner cover web (polypropylene spunbond, white). Each mask contains polyester spandex blend ear loops to secure the mask over the user's mouth and nose and includes a polyethylene coated wire nosepiece to provide a firm fit over the nose. This face mask is a single use, disposable device, provided non-sterile. This device is not made with natural rubber latex.
This document is a 510(k) Premarket Notification for a Freudenberg Surgical Mask, not an AI/ML medical device. Therefore, the information requested about acceptance criteria and study data for an AI/ML device is not present in this document. This document pertains to the physical and performance characteristics of a surgical mask, compared to a predicate surgical mask.
However, I can extract the acceptance criteria and performance data for the surgical mask as presented in the document.
Here's the information about the surgical mask:
1. A table of acceptance criteria and the reported device performance
The document compares the "Proposed Device (K210063) Freudenberg Surgical Mask" to a "Predicate Device (K182514) SURGICAL FACE MASK" based on several performance metrics. The acceptance criteria for the proposed device are implicitly defined by meeting or exceeding the performance of the predicate device and conforming to relevant ASTM/EN standards, particularly for Level 3 categorization.
| Performance Metric | Acceptance Criteria (ASTM/EN Standard or Predicate Performance) | Reported Device Performance (Freudenberg Surgical Mask - K210063) |
|---|---|---|
| Fluid Resistance | Predicate: 32/32 passed at 120 mmHg - ASTM F1862. Goal: Exceed predicate/meet Level 3 (160 mmHg, AQL 4.0). | Three non-sequential lots of 32 (total of 96, AQL 4.0) passed at 160mmHg - ASTM F1862.Lot 698002: 31/32 passedLot 697715: 32/32 passedLot 697944: 31/32 passed(Comment: "Similar, subject device exceeded fluid resistance of predicate device") |
| Bacterial Filtration Efficiency (BFE) | Predicate: pass at 99.6% - ASTM F2101. Goal: ≥98% (for Level 3 per ASTM F2100-19) | Three non-sequential lots of 32 (total of 96, AQL 4.0) passed at ≥98% - ASTM F2101.Lot 698002: 32/32 passed at 99.9%Lot 697715: 32/32 passed at 99.9%Lot 697944: 32/32 passed at 99.9%(Comment: "Similar, both devices meet ASTM F2101") |
| Particulate Filtration Efficiency (PFE) | Predicate: pass at 99.88% - ASTM F2299. Goal: ≥98% (for Level 3 per ASTM F2100-19) | Three non-sequential lots of 32 (total of 96, AQL 4.0) passed at ≥98% - ASTM F2299.Lot 698002: 32/32 passed at 99.6±0.1%Lot 697715: 32/32 passed at 99.5±0.1%Lot 697944: 32/32 passed at 99.7±0.1%(Comment: "Similar, both devices meet ASTM F2299") |
| Differential Pressure (Delta P) | Predicate: pass at 3.0 mmH₂O/cm² - MIL-M036954C. Goal: <6.0 H₂O/cm² (for Level 3 per ASTM F2100-19 / EN 14683:2019, Annex C) | Three non-sequential lots of 32 (total of 96, AQL 4.0) passed at <6.0 H₂O/cm² - ASTM F2100 / EN 14683:2019, Annex C.Lot 698002: 29/32 passed at 5.3mm H₂O / cm²Lot 697715: 30/32 passed at 5.3mm H₂O / cm²Lot 697944: 31/32 passed at 5.7mm H₂O / cm²(Comment: "Similar, subject device utilized ASTM F2100-19 test methods") |
| Flammability | Predicate: Class 1 16 CFR 1610. Goal: Class 1 16 CFR 1610. | Three non-sequential lots of 32 (total of 96, AQL 4.0) passed Class 1 16 CFR 1610.Lot 698002: 32/32 passedLot 697715: 32/32 passedLot 697944: 32/32 passed(Comment: "Same") |
| Biocompatibility - Cytotoxicity | Predicate: non-cytotoxic. Goal: non-cytotoxic (ISO 10993-5). | Under the conditions of the study, the subject device extract was determined to be non-cytotoxic.(Comment: "Same") |
| Biocompatibility - Irritation | Predicate: non-irritating. Goal: non-irritating (ISO 10993-10). | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.(Comment: "Same") |
| Biocompatibility - Sensitization | Predicate: non-sensitizing. Goal: non-sensitizing (ISO 10993-10). | Under the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing.(Comment: "Same") |
2. Sample size used for the test set and the data provenance
- Sample Size for performance testing: For fluid resistance, BFE, PFE, differential pressure, and flammability, 96 samples were tested (3 non-sequential lots of 32 samples each, with an Acceptable Quality Level (AQL) of 4.0).
- Sample Size for biocompatibility: The document does not specify the exact number of samples for biocompatibility testing, but it refers to "the study" for each of cytotoxicity, irritation, and sensitization, indicating that tests were conducted. Biocompatibility tests usually involve a small number of samples but are well-controlled in vitro or in vivo studies.
- Data Provenance: Not explicitly stated, but based on the nature of medical device testing for FDA clearance, it would be prospective testing conducted in controlled laboratory environments. The country of origin of the data is not specified, but the applicant's address is in Durham, NC, USA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable as the device is a physical surgical mask, not an AI/ML medical device requiring expert interpretation for ground truth establishment. The performance is measured against established physical and biological standards (e.g., fluid penetration, filtration efficiency, flammability, biocompatibility).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable for a physical device. Performance is determined by objective laboratory measurements and adherence to specified test methods (e.g., ASTM F1862 for fluid resistance, ASTM F2101 for BFE).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable as the device is a physical surgical mask and does not involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is not applicable as the device is a physical surgical mask and does not involve algorithms.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device is based on objective, internationally recognized laboratory test standards and methods for material performance, filtration, and biocompatibility, such as:
- ASTM F2100 (Standard Specification for Performance of Materials Used in Medical Face Masks) - defines performance levels.
- ASTM F1862 (Resistance to Penetration by Synthetic Blood)
- ASTM F2299 (Efficiency to Penetration by Particulates Using Latex Spheres)
- ASTM F2101 (Bacterial Filtration Efficiency)
- EN 14683 (Differential Pressure)
- 16 CFR Part 1610 (Flammability)
- ISO 10993-1, ISO 10993-5, ISO 10993-10 (Biocompatibility: Cytotoxicity, Irritation, Sensitization)
8. The sample size for the training set
This section is not applicable as the device is a physical surgical mask and does not involve AI/ML models requiring training data.
9. How the ground truth for the training set was established
This section is not applicable as the device is a physical surgical mask and does not involve AI/ML models requiring training data.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
February 5, 2021
Freudenberg Performance Materials LP % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K210063
Trade/Device Name: Freudenberg Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: January 8, 2021 Received: January 11, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Elizabeth F. Claverie-Williams. MS. CAPT, USPHS-CC Assistant Director DHT4B: Division of Infection Controland Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K210063
Device Name Freudenberg Surgical Mask
Indications for Use (Describe)
The Freudenberg Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.
Type of Use (Select one or both, as applicable)
| X | X | X | |
| X | X | ||
| X |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
FREUDENBERG PERFORMANCE MATERIALS
Image /page/3/Picture/1 description: The image shows the Freudenberg company logo. The logo consists of the company name "FREUDENBERG" in blue, with a stylized wave graphic to the left. Below the company name is the tagline "innovating together" also in blue.
510(k) Summary
| 510(k) Number: | K210063 |
|---|---|
| Sponsor Information: | Freudenberg Performance Materials LP3500 Industrial DriveDurham, NC 27704United States |
| Sponsor Contact:Eberhard LinkTechnical Market Manager - Medical919-349-8389eberhard.link@freudenberg-pm.com | |
| Contact Person: | Eberhard LinkTechnical Market Manager - Medical919-349-8389eberhard.link@freudenberg-pm.com |
| Date of SummaryPreparation: | February 2, 2021 |
| Trade Name: | Freudenberg |
| Common Name: | Surgical Mask |
| Classification Name : | Surgical Apparel |
| Proprietary Name: | Freudenberg |
| Regulation MedicalSpecialty: | General and Plastic Surgery |
| Review Panel: | General Hospital |
| Product Code: | FXX |
| Device Classification: | Class II per 21 CFR §878.4040 |
| Predicate Device: | SURGICAL FACE MASK (K182514) - Model Ear Loop |
| Intended Use: | The Freudenberg Surgical Mask is intended to be worn toprotect both the patient and healthcare personnel fromtransfer of microorganisms, body fluids and particulatematerial. This face mask is intended for use in infectioncontrol practices to reduce the potential exposure to bloodand body fluids. This is a single use, disposable deviceprovided non-sterile. |
| Device Description: | The Freudenberg Surgical Mask is composed of three-layersand is flat-pleated. The Freudenberg Surgical Mask materialsconsist of an outer cover web (polypropylene spunbond,white), insertion layer (polypropylene, melt-blown, white), andinner cover web (polypropylene spunbond, white). Each mask |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Freudenberg. The logo consists of a blue graphic on the left and the word "FREUDENBERG" in blue on the right. Below the word "FREUDENBERG" is the phrase "INNOVATING TOGETHER" in a smaller font.
contains polyester spandex blend ear loops to secure the mask over the user's mouth and nose and includes a polyethylene coated wire nosepiece to provide a firm fit over the nose. This face mask is a single use, disposable device, provided non-sterile. This device is not made with natural rubber latex.
Available Model Number:
FPMD2020
Surgical Mask
Comparison of Proposed and Predicate Devices:
| ltem(s) | Proposed Device(K210063)Freudenberg SurgicalMask | Predicate Device(K182514)SURGICAL FACE MASK | Comparison |
|---|---|---|---|
| Intended Use /Indications for Use | The Freudenberg SurgicalMask is intended to beworn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. This face mask isintended for use ininfection control practicesto reduce the potentialexposure to blood andbody fluids. This is asingle use, disposabledevice provided non-sterile. | The Disposable SurgicalFace Masks are intendedto be worn to protect boththe patient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. These facemasks are intended foruse in infection controlpractices to reduce thepotential exposure toblood and body fluids. Thisis a single use, disposabledevice provided non-sterile. | Same, minordifferences tocorrect grammar |
| Type of Use | Over-The-Counter Use (21CFR 801 Subpart C) | Over-The-Counter Use (21CFR 801 Subpart C) | Same |
| Outer facing layer | Spun-bond polypropylene | Spun-bond polypropylene | Same |
| Middle layer | Melt blown polypropylenefilter | Melt blown polypropylenefilter | Same |
| Inner facing layer | Spun-bond polypropylene | Spun-bond polypropylene | Same |
| Nose piece | Malleable polyethylene-coated wire | Malleable aluminum wire | Similar, bothnose pieces arecomposed ofmalle able metal. |
| Nose piece length | 10.0cm ±1cm | Not provided | - |
| Item(s) | Proposed Device(K210063)Freudenberg SurgicalMask | Predicate Device(K182514)SURGICAL FACE MASK | Comparison |
| Ear loops | Polyester spandex blend | Polyester | Similar, bothmask ear loopsuse polyester |
| Ear loop length | 14.5cm -1/+2.5cm | Not provided | - |
| Color | White | White | Same |
| Mask Style | Flat Pleated | Flat Pleated | Same |
| Length | 17.5cm±1cm | 17.5cm±1cm | Same |
| Width | 9.5cm ±1cm | 9.5cm ±1cm | Same |
| Sterile | Non-Sterile | Non-Sterile | Same |
| Use | Single Use | Single Use | Same |
| PerformanceTesting | Level 3 - ASTMF2100-19 | Level 2 - ASTMF2100-11 | Different,subject devicesexceededpredicate devicein oneperformancetest. ASTMF2100-19 usesdifferent testmethod formeasuringDelta-P(H2O/cm²) |
| Fluid ResistancePerformance | Three non-sequential lotsof 32 (total of 96, AQL4.0) passed at 160mmHg- ASTM F1862Lot 698002: 31/32 passedLot 697715: 32/32 passedLot 697944: 31/32 passed | 32/32 passed at 120mmHg - ASTM F1862 | Similar,subject deviceexceeded fluidresistance ofpredicate device |
| BacterialFiltrationEfficiency | Three non-sequential lotsof 32 (total of 96, AQL4.0) passed at ≥98% -ASTM F2101Lot 698002: 32/32 passedat 99.9%Lot 697715: 32/32 passedat 99.9% Lot 697944:32/32 passed at 99.9% | pass at 99.6%- ASTMF2101 | Similar, bothdevices meetASTMF2101 |
| Item(s) | Proposed Device(K210063)Freudenberg SurgicalMask | Predicate Device(K182514)SURGICAL FACE MASK | Comparison |
| ParticulateFiltrationEfficiency | Three non-sequential lotsof 32 (total of 96, AQL4.0) passed at ≥98% -ASTM F2299Lot 698002: 32/32 passedat 99.6±0.1%Lot 697715: 32/32 passedat 99.5±0.1%Lot 697944: 32/32 passedat 99.7±0.1% | pass at 99.88% - ASTMF2299 | Similar, bothdevices meetASTMF2299 |
| DifferentialPressure (DeltaP) | Three non-sequential lotsof 32 (total of 96, AQL4.0) passed at <6.0H₂O/cm² - ASTM F2100 /EN 14683:2019, Annex CLot 698002: 29/32 passedat 5.3mm H₂O / cm²Lot 697715: 30/32 passedat 5.3mm H₂O / cm²Lot 697944: 31/32 passedat 5.7mm H₂O / cm² | pass at 3.0mmH₂O/cm² -MIL-M036954C | Similar, subjectdevice utilizedASTMF2100-19 testmethods (FRRecognitionNumber 6-425) |
| Flammability | Three non-sequential lotsof 32 (total of 96, AQL4.0) passed Class 1 16CFR 1610Lot 698002: 32/32 passedLot 697715: 32/32 passedLot 697944: 32/32 passed | Class 1 16 CFR 1610 | Same |
| Biocompatibility | |||
| Cytotoxicity | Under the conditions ofthe study, the subjectdevice extract wasdetermined to be non-cytotoxic. | Under the conditions ofthe study, the subjectdevice extract wasdetermined to be non-cytotoxic. | Same |
| Irritation | Under the conditions ofthe study, the subjectdevice non-polar and polarextracts were determinedto be non-irritating. | Under the conditions ofthe study, the subjectdevice non-polar and polarextracts were determinedto be non-irritating. | Same |
| Sensitization | Under the conditions ofthe study, the subjectdevice non-polar and polar | Under the conditions ofthe study, the subjectdevice non-polar and polar | Same |
| Item(s) | Proposed Device(K210063)Freudenberg SurgicalMask | Predicate Device(K182514)SURGICAL FACE MASK | Comparison |
| extracts were determinedto be non-sensitizing. | extracts were determinedto be non-sensitizing. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "FREUDENBERG" in a bold, sans-serif font. To the left of the word is a blue graphic consisting of several curved lines. The text is in a dark blue color.
Innovating Together
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image features the Freudenberg logo. The logo consists of a blue graphic element on the left, resembling stylized waves or flowing lines. To the right of the graphic is the word "FREUDENBERG" in a bold, sans-serif, blue font.
Innovating Together
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Freudenberg. The logo consists of a blue graphic on the left and the word "FREUDENBERG" in blue on the right. Below the word "FREUDENBERG" is the phrase "INNOVATING TOGETHER" in a smaller font.
NNOVATING TOGETHE
Summary of Non-Clinical Performance Testing:
The following standards have been used to evaluate the Freudenberg Surgical Mask:
| ASTM F2100 | Standard Specification for Performance of Materials Used in Medical Face Masks |
|---|---|
| ASTM F1862 | Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity) |
| ASTM F2299 | Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres |
| ASTM F2101 | Standard Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus |
| EN 14683 | Standard Test Method for Differential Pressure |
| 16 CFR Part 1610 | Standard for the Flammability of Clothing |
| ISO 10993-1 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process |
| ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity of medical devices |
| ISO 10993-10 | Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization |
Non-clinical tests were conducted on three non-consecutive lots to verify that the proposed device met all design specifications as the predicate device, or better. The test results demonstrate that the proposed device conforms to the recognized standards ASTM F2100, ASTM F1862, ASTM F2299, ASTM F2101, EN 14683, 16 CFR Part 1610, and ISO 10993 in addition to the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004.
Summary of Clinical Performance Test:
No clinical study is included in this submission.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Freudenberg. The logo consists of a blue wave-like graphic on the left, followed by the company name "FREUDENBERG" in bold, blue letters. Below the company name, the tagline "INNOVATING TOGETHER" is written in smaller, blue letters.
Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device K210063 is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K182514.
§ 878.4040 Surgical apparel.
(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.