K Number
K210055

Validate with FDA (Live)

Device Name
Cios Alpha
Date Cleared
2021-02-05

(28 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastrointestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

Device Description

The Cios Alpha (VA30) mobile fluoroscopic C-arm X-ray System is designed for the surgical environment. The Cios Alpha provides comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:
a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The c-arm can be angulated in both planes and be lifted vertically, shifted to the side and move forward/backward by an operator.
b) The second unit is the image display station with a moveable trolley for the image processing and storage system, image display and documentation. Both units are connected to each other with a cable.
The main unit is connected to the main power outlet and the trolley is connected to a data network.

AI/ML Overview

The Siemens Cios Alpha (VA30) is a mobile X-ray system. This submission describes modifications to the predicate device software.

Here's a breakdown of the acceptance criteria and study information:

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of quantitative acceptance criteria for each software modification with corresponding device performance metrics. Instead, it states that "The Cios Alpha software (VA30) was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process." and "The testing results support that all the software specifications have met the acceptance criteria."

The modifications are primarily software enhancements aimed at improving user interaction and safety features. The "reported device performance" is qualitative, asserting that the modified features function as intended and are comparable to or improved from the predicate devices.

Feature / ModificationAcceptance Criteria (Implied)Reported Device Performance (Implied)
Target PointerFunctionality equivalent to or improved from the predicate device (Cios Flow K203504).Same: Target Pointer has the same functionality as cleared in the Secondary Predicate Device Cios Flow K202504. No technological differences.
Interactive User Touch ControlFunctionality equivalent to or improved from the predicate device (Cios Flow K203504) for Collimation, Brightness/Contrast, Rotate/Flip, Zoom/Pan, and Spot Adapt.Same: Functionality is the same as cleared in the Secondary Predicate device (Cios Flow K203504). Functionality of these features has not changed.
Dose Regulation IndicatorDose regulation equal to the predicate device (Cios Flow K203504).Same: The Dose regulation is equal to the Secondary Predicate Cios Flow K203504.
New Sound Radiation DelaySound during Radiation Delay equivalent to the predicate device (Cios Flow K203504).Same: The Sound during Radiation Delay is equal to the Secondary Predicate Cios Flow K203504.
New Product Software SecurityUpdated Product Software Security functionality equivalent to the predicate device (Cios Flow K181560).Same: The updated Product Software Security functionality is the same as cleared in the Secondary Predicate device Cios Flow K181560. (Note: K181560 is the primary predicate, not Cios Flow here)

2. Sample size used for the test set and the data provenance (country of origin of the data, retrospective or prospective)

The document does not specify the sample size for any test set or the data provenance (country, retrospective/prospective). It mentions "verification and validation testing" and "clinical use tests with customer report and feedback form" but provides no details on these.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number or qualifications of experts used to establish ground truth for any test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not mention any specific adjudication method for the test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study is mentioned in the provided text, nor is any AI component or human reader improvement with AI assistance discussed. The device is an X-ray system with software modifications, not an AI-powered diagnostic tool for interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is an X-ray system, not a standalone algorithm. The "software modifications" refer to direct user interface and system control features, not an independent algorithm for diagnostic interpretation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The document does not explicitly state the type of ground truth used. However, given the nature of the software modifications (Target Pointer, Interactive User Touch Control, Dose Regulation Indicator, Sound Radiation Delay, Product Software Security), the ground truth for their effectiveness would likely be based on functional testing, compliance with specifications, and user feedback/system behavior validation, rather than medical ground truth like pathology or outcomes data. The Human Factor Usability Validation suggests user feedback played a role.

8. The sample size for the training set

This is not applicable since the modifications are for a medical device's operating software, not for an AI/machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. The full name of the agency is "U.S. Food & Drug Administration".

February 5, 2021

Siemens Medical Solutions USA, Inc. % Mr. Cordell Fields, Esq. Technical Specialists, Regulatory Submissions 40 Liberty Blvd., 65-1A MALVERN PA 19355

Re: K210055

Trade/Device Name: Cios Alpha Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: January 7, 2021 Received: January 8, 2021

Dear Mr. Fields:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210055

Device Name Cios Alpha

Indications for Use (Describe)

The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastrointestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

Type of Use (Select one or both, as applicable):

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logos for Siemens and Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.

510(k) Summary: Cios Alpha

K210055

Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355

Date Prepared: January 07, 2021

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

    1. General Information: lmporter / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
      Manufacturing Site: Siemens Healthcare GmbH Roentgenstrasse 19 - 21 95478 Kemnath, Germany Establishment Registration Number: 3002466018

2. Contact Person:

Mr. Cordell Fields, Esq. Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 306-3167 Email: cordell.fields@siemens-Healthineers.com

3. Device Name and Classification:

Trade Name: Classification Name: Classification Panel: Requlation Number: Device Class: Product Codes: Subsequent Product Code:

Cios Alpha Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II OWB, OXO JAA

  • Legally Marketed Predicate Devices 4.
    Legally Marketed Primary Predicate Device Trade Name: Cios Alpha 510(k) Clearance K181560

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

Special 510(k) Submission: Cios Alpha (VA30)

Special 510(k) Submission: Cios Alpha (VA30)
Clearance DateOctober 24, 2018
Classification Name:Image-intensified fluoroscopic x-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1650
Device Class:Class II
Product Codes:OWB, OXO
Subsequent Product Codes:JAA
Total Product Life Cycle:All product Recall incidents are considered during theDesign Input phase of development to ensure the latestmodels will not be affected by any of the applicableissues.
Legally Marketed Secondary Predicate Device
Trade Name:Cios Flow
510(k) ClearanceK203504
Clearance DateDecember 22, 2020
Classification Name:Image-intensified fluoroscopic x-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1650
Device Class:Class II
Product Codes:OWB, OXO
Subsequent Product Codes:JAA
Total Product Life Cycle:All product Recall incidents are considered during theDesign Input phase of development to ensure the latestmodels will not be affected by any of the applicableissues.

5. Device Description:

The Cios Alpha (VA30) mobile fluoroscopic C-arm X-ray System is designed for the surgical environment. The Cios Alpha provides comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:

  • a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The c-arm can be angulated in both planes and be lifted vertically, shifted to the side and move forward/backward by an operator.
  • b) The second unit is the image display station with a moveable trolley for the image processing and storage system, image display and documentation. Both units are connected to each other with a cable.

The main unit is connected to the main power outlet and the trolley is connected to a data network.

The following modifications were made to the predicate device Cios Alpha. Siemens Medical Solutions USA, Inc. submits this Special 510(k) to request clearance for the Subject Device Cios Alpha (VA30) for the following device modifications made to the Predicate Device Cios Alpha (VA30).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.

Special 510(k) Submission: Cios Alpha (VA30)

This 510(k) submission, Subject Device "Cios Alpha" with software version VA30 the following categories of modifications made to the Subject Device in comparison to Predicate Devices:

Modified Software:

  • Table 1: Overview of Software Modifications supported by Software Version VA30 .

Table 1: Software Modifications for Cios Alpha

Subject Device: Cios Alpha (VA30) Modifications/Features
Software changes specific to New System Software VA30
1.Target Pointer
2.Interactive User Touch Control (Software component of VA30)
A. Collimation Controls
B. Brightness Contrast Controls
C. Rotate and Flip Controls
D. Zoom and Pan Controls
E. Spot Adapt
3.Dose Regulation Indicator
4.New Sound Radiation Delay
5.New Product Software Security
Other Device Modifications
6.Update 510(k) Information

6. Indications for Use:

The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

7. Substantial Equivalence:

The Cios Alpha (VA30) system is substantial equivalent to the legally marketed predicates listed in Table 2 below:

Predicate Device Name andManufacturer510(k)NumberClearanceDateComparable Properties
Primary PredicateK18156010/24/2018Indications for use
Siemens' Cios AlphaSystem for Image Acquisition●
Post-processing Software
Examination Settings●
Secondary PredicateK20350412/22/2020Target Pointer●
Siemens' Cios FlowInteractive User Touch Control●
Dose Regulation Indicator
Sound Radiation Delay
Product Software Security●

Table 2: Predicate Device Comparable Properties for Subject Device Modifications:

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

Summary of Technological Characteristics of the Subject Device as Cor 8. the Predicate Device:

The Cios Alpha (VA30) is substantially equivalent to the commercially available Siemens Cios Alpha (VA30), cleared with K181560.

Indication for use remains unchanged and technology and design of the Cios Alpha (VA30) is based on the predicate Cios Alpha (VA30).

Technological differences between the Subject Device and the Predicate Device is provided in Table 3 below for all modifications.

Comparison of Modifications to Predicate Devices
Subject DeviceCios Alpha (VA30)ModificationsPrimary Predicate DeviceComparisonResults
1.Target PointerCios AlphaK181560Secondary Predicate DeviceCios FlowK203504(The following modificationsare derived from the secondarypredicate device Cios Flow)Target PointerSame: Target Pointer has thesame functionality as cleared inthe Secondary PredicateDevice Cios Flow K202504 .There are no technologicaldifferences in these featuresand the functionality of thesefeatures have not changed.
2.Interactive User Touch Control(Software components of VA30)A. Collimation ControlsB. Brightness Contrast ControlsC. Rotate and Flip ControlsD. Zoom and Pan ControlsE. Spot Adapt (Dose regulation withmovable circle)Interactive User Touch ControlCollimation ControlsBrightness Contrast ControlsRotate and Flip ControlsZoom and Pan ControlsSpot Adapt (Dose regulation withmovable circle)Same: 2.A-E has the samefunctionality as cleared in theSecondary Predicate device.The functionality of thesefeatures has not changed fromthe Secondary Predicatedevice Cios Flow K203504 .
3.New Dose Regulation IndicatorDose Regulation IndicatorSame:The Dose regulation is equal tothe Secondary Predicate CiosFlow K203504
4.New Sound during Radiation DelaySound during Radiation DelaySame:The Sound during RadiationDelay is equal to theSecondary Predicate CiosFlow K203504
5.New Product Software SecurityProduct Software SecuritySame:The updated Product SoftwareSecurity functionality is thesame as cleared in theSecondary Predicate deviceCios Flow K181560 .

Table 3: Summary of Comparison of Technological Characteristics

9. Nonclinical Performance Testing:

Non-clinical tests were conducted for the Cios Alpha (VA30) during product development.

The Cios Alpha (VA30) was certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance and Electromagnetic Compatibility:

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.

  • AAMI ANSI ES60601-1:2005/(R)2012 .
  • IEC 60601-1-2:2014 .
  • IEC 60601-1-3:2013 ●
  • IEC 60601-1-6:2010/A1:2013 ●
  • IEC 60825-1:2014 ●
  • IEC 62304:2015 ●
  • IEC 60601-2-28:2010 ●
  • IEC 60601-2-43:2010 .
  • IEC 60601-2-54:2009/A1:2015 ●
  • ISO 14971:2019 .
  • IEC 62366-1:2015/ Cor.1:2016 .

Table 4: FDA Guidance Documents

FDA Guidance Document and Effective Date
1.Guidance for Industry and FDA Staff – User Fees and Refunds for PremarketNotification Submissions 510(k)Document issued on October 2, 2017
2.Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept Policyfor 510(k)sDocument issued on September 13, 2019
3.Guidance for Industry and FDA Staff: The Special 510(k) ProgramDocument issued on September 13, 2019
4.Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a changeto an existing device.Document issued on October 25, 2017
5.Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program:Evaluating Substantial Equivalence in Premarket Notifications [510(k)]Document Issued on July 28, 2014
6.Guidance for Industry and FDA Staff: Guidance for the Submission Of 510(k)'s for SolidState X-ray Imaging DevicesDocument issued on September 1, 2016
7.Guidance for Industry and FDA Staff: Guidance for the Content of PremarketSubmission for Software in Medical DevicesDocument issued on May 11, 2005
8.Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use inMedical DevicesDocument issued on September 27, 2019
9.Guidance for Industry and FDA Staff: Applying Human Factors and UsabilityEngineering to Medical Devices.Document issued February 3, 2016
10.Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging DevicePremarket Notifications.Document issued on November 28, 2017
11.Guidance for Industry and FDA Staff: Content of Premarket Submissions forManagement of Cybersecurity in Medical devices.Document issued on October 2, 2014
12.Guidance for Industry and FDA Staff: Appropriate Use of Voluntary ConsensusStandards in Premarket Submission for Medical DevicesDocument issued on September 14, 2018
13.Guidance for Industry and FDA Staff: Medical Device Accessories - DescribingAccessories and Classification PathwaysDocument issued on December 20, 2017
14.Guidance for Industry and FDA Staff: Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket SubmissionsDocument issued on December 20, 2019

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

The modifications described in this Premarket Notification are supported with verification and validation testing.

Verification and Validation:

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Nonclinical tests were conducted on Cios Alpha (VA30) during product development.

The Risk analysis was completed, and risk control implemented to mitigate identified The testing results support that all the software specifications have met the hazards. acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.

The Cios Alpha software (VA30) was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual and in clinical use tests with customer report and feedback form. Customer employees are adequately trained in the use of this equipment.

Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. Compliance with IEC 80001-1-2010 is the responsibility of the hospital. Provided in the Software Section is the required cybersecurity information.

Summary:

Performance tests were conducted to test the functionality of Cios Alpha (VA30) System. These tests have been performed to assess the functionality of the subject device. Results of all conducted testing and clinical assessment were found acceptable and do not raise any new issues of safety or effectiveness.

10. General Safety and Effectiveness Concerns:

Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.

11. Conclusion as to Substantial Equivalence:

{9}------------------------------------------------

Special 510(k) Submission: Cios Alpha (VA30)

IEMEN

The predicate devices were cleared based on non-clinical supportive information and clinical images and data. Similar non-clinical test results demonstrate that the Cios Alpha (VA30) System acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data and software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device that is currently marketed for the same intended use.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.