K Number
K210032
Date Cleared
2021-09-01

(239 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fingertip Pulse Oximeter (Wireless) are intended for measuring function of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches) and for patients during nomotion condition.

Device Description

The main function of the Fingertip Pulse Oximeter (Wireless) is to measure the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate as non-invasive spot checking monitoring. The subject determines the functional oxygen saturation of arterial hemoglobin (SpO2) by measuring the absorption of red and infrared light passing through perfused tissue. Changes in absorption caused by the pulsation of blood in the vascular bed are used to determine SpO2 reading and pulse rate. The characteristic of the device is a stand-alone device with 1 AAA Alkaline battery. It includes an OLED display screen and a warning or indicator function for abnormal readings. The Fingertip Pulse Oximeter (Wireless) is also built in with a Bluetooth low energy (BLE) function to perform data transmission.

AI/ML Overview

The provided documentation is for a 510(k) premarket notification for the Leadtek Fingertip Pulse Oximeter (Wireless). It establishes substantial equivalence to a predicate device (K193350) and does not present detailed new clinical study data for the subject device beyond referencing the predicate. Therefore, some of the requested information cannot be fully provided from the given document as it refers to the predicate device's clinical testing.

Here's a breakdown of the requested information based on the provided text, focusing on the subject device's demonstrated performance and criteria where available, and referencing the predicate device's clinical study for SpO2 accuracy.


1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Predicate)Reported Device Performance (Subject Device)
SpO2 AccuracyAccuracy: 70% to 100% range ± 2% (as per predicate device, implying compliance with ISO 80601-2-61 requirements for pulse oximeters)Algorithm accuracy for SpO2 is considered unchanged from the predicate device. The document states: "The laboratory evaluations for the performance of the subject device was conducted to ensure the modification do not change the performance that root-mean-square (Arms) value is less than 2% within the range 70% - 100%..." and "Accuracy: 70% to 100% range ± 2%, less than 70% unspecified."
Pulse Rate AccuracyAccuracy: ± 1 bpm or ± 1%, whichever is greater (as per predicate device)Unchanged from predicate device: "Accuracy: ± 1 bpm or ± 1%, whichever is greater"
Electrical SafetyCompliance with IEC 60601-1:2012, IEC 60601-1-11:2015, and ISO 80601-2-61:2011Complied with IEC 60601-1:2012, IEC 60601-1-11:2015, and ISO 80601-2-61:2011 (Laboratory tests performed)
EMC TestingCompliance with IEC 60601-1-2:2014Complied with IEC 60601-1-2:2014 (Laboratory tests performed)
BiocompatibilityCompliance with ISO 10993-1:2009 (cytotoxicity, skin sensitization, skin irritation studies showing no potential risks)Skin contacting materials are the same as the predicate (K193350) and passed cytotoxicity, skin sensitization, and skin irritation studies per ISO 10993-1:2009.
Software Verification & ValidationCompliance with FDA Guidance for the Content of the Premarket Submission for Software Contained in Medical Devices (Moderate level of concern)Software verification and validation provided in compliance with FDA Guidance; deemed "moderate" level of concern. Tests support substantial equivalence.
Wireless CoexistenceCompliance with IEEE/ANSI C63.27:2009, AAM TIR69:2017, and FDA Guidance for Radio Frequency Wireless Technology in Medical Device.Coexistence testing performed according to IEEE/ANSI C63.27:2009, AAM TIR69:2017, and FDA Guidance. Demonstrated ability to coexist with interfering networks (2.4 GHz ISM band) and BLE.
Non-clinical PerformanceRoot-mean-square (Arms) value less than 2% within the 70%-100% SpO2 range.Confirmed that modifications did not change performance, with Arms value less than 2% within the 70%-100% range.

2. Sample Size Used for the Test Set and the Data Provenance (e.g., country of origin of the data, retrospective or prospective)

For the SpO2 accuracy, the document explicitly states that no new clinical testing was performed for the subject device because its pulse oximetry algorithm and measuring technology are the same as the predicate device (K193350). Therefore, the clinical data refers to the predicate device's clearance.

The document does not provide the specific sample size, country of origin, or whether the study was retrospective or prospective for the predicate device's clinical test. It only states that "A clinical study was conducted on the predicate device for K193350 clearance to demonstrate the safety and performance."

For non-clinical performance and wireless coexistence, these were laboratory-based tests and not human clinical studies.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts (e.g., radiologist with 10 years of experience)

This information is not provided in the document. As the clinical study was for the predicate device and the details are not elaborated here, the specifics of ground truth establishment and expert qualifications are absent. Typically, for pulse oximeter accuracy studies, arterial blood gas measurements (co-oximetry) performed by trained medical personnel serve as the reference standard (ground truth).


4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is not provided in the document.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a Fingertip Pulse Oximeter, which directly measures physiological parameters (SpO2 and pulse rate). It does not involve "human readers" interpreting images or data that would be enhanced by AI assistance in the way an MRMC study would assess. The "Wireless" aspect refers to data transmission, not AI-driven interpretation.


6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the core SpO2 and pulse rate measurement algorithm performs in a standalone manner. The device outputs direct measurements based on light absorption. The "Wireless" functionality is for data transmission, not for an AI algorithm interpreting the results for a human. The non-clinical performance testing validated the algorithm's accuracy (Arms value < 2%) in a standalone capacity.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

As the clinical study for SpO2 accuracy was referenced from the predicate device, and details are not in this submission, the specific type of ground truth is not explicitly stated here. However, for pulse oximeters, the accepted ground truth for SpO2 accuracy is generally arterial blood gas analysis (co-oximetry), where blood samples are directly analyzed for oxygen saturation.


8. The sample size for the training set

This information is not provided. Since the device uses the same algorithm as the predicate, and no new clinical study was performed for the subject device, details on a training set would pertain to the development of the original predicate algorithm.


9. How the ground truth for the training set was established

This information is not provided. Similar to point 8, this would refer to the predicate device's development.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

September 1, 2021

Leadtek Research Inc. Sharon Peng Regulatory Affairs Specialist 18F, No. 166, Jian-Yi Rd., Chung Ho Dist. New Taipei City, 23511 Taiwan

Re: K210032

Trade/Device Name: Leadtek Fingertip Pulse Oximeter (Wireless) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: July 26, 2021 Received: August 2, 2021

Dear Sharon Peng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210032

Device Name Fingertip Pulse Oximeter (Wireless)

Indications for Use (Describe)

The Fingertip Pulse Oximeter (Wireless) are intended for measuring function of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches) and for patients during nomotion condition.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(K) information is being submitted in accordance with the requirements of 21 CFR 807.92

1. General Information

Applicant:Leadtek Research Incorporation
18F, No.166, Jian-Yi Rd., Chung Ho Dist.
23511 New Taipei City, Taiwan
Phone: +886-2-8226-5800
Fax: +886-2-8226-5801
Contact Person:Sharon Peng
Regulatory Affairs Specialist
18F, No.166, Jian-Yi Rd., Chung Ho Dist.
23511 New Taipei City, Taiwan
Phone: +886-2-8226-5800
Fax: +886-2-8226-5801
Email: Sharon_peng@leadtek.com.tw
Date Prepared:December 21, 2020

2. Device Information

Proprietary Name:Fingertip Pulse Oximeter (Wireless)
Common Name:Pulse Oximeter
Classification Name:Oximeter
Regulatory Class:Class II
Regulation Number:21CFR 870.2700
Product Code:DQA
Review Panel:Anesthesiology

3. Predicate Device

Proprietary Name:Fingertip Pulse Oximeter
510(K) Number:K193350
Manufacturer:Leadtek Research Inc.

{4}------------------------------------------------

4. Intended Use

The Fingertip Pulse Oximeter (Wireless) are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition.

5. Device Description

The main function of the Fingertip Pulse Oximeter (Wireless) is to measure the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate as non-invasive spot checking monitoring. The subject determines the functional oxygen saturation of arterial hemoglobin (SpO2) by measuring the absorption of red and infrared light passing through perfused tissue. Changes in absorption caused by the pulsation of blood in the vascular bed are used to determine SpO2 reading and pulse rate. The characteristic of the device is a stand-alone device with 1 AAA Alkaline battery. It includes an OLED display screen and a warning or indicator function for abnormal readings.

The Fingertip Pulse Oximeter (Wireless) is also built in with a Bluetooth low energy (BLE) function to perform data transmission.

6. Substantial Equivalence Information:

The intended use and test principle of Fingertip Pulse Oximeter (Wireless) is the same that used in the predicate devices, Fingertip Pulse Oximeter cleared under K193350. The main function of Fingertip Pulse Oximeter (Wireless) is to measure the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate as non-invasive spot checking by measuring the absorption of red and infrared light passing through perfused tissue. No changes of intended use and technological characteristics support the substantial equivalence of the subject device to the predicate.

As compared to the predicate device, the main modification of the subject device is that Fingertip Pulse Oximeter (Wireless) uses Bluetooth Low Energy (BLE) technology to transfer the measurement results from the oximeter to a mobile device or PC tablet enabled Bluetooth function. The Bluetooth icon will light up on the oximeter display when a connection is established successfully. The data of the oximeter will be synchronized with the enable device since the successful connection. The verification and validation tests were found to support that wireless transmission function will not raise different questions of safety and effectiveness.

TEL: 886-2-8226-5800 FAX: 886-2-8226-5801

{5}------------------------------------------------

ItemSubject DevicePredicate Device
ProductNameFingertip Pulse Oximeter (Wireless)Fingertip Pulse Oximeter
Model No.8D01H and 8D01I8D01B and 8D01C
510(k) Information
RegulationNumber870.2700870.2700
ClassificationClass IIClass II
Product CodeDQADQA
ComparisonUnchanged as previous submission
Indication for Use
Intended UseFingertip Pulse Oximeter (Wireless) are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition.Fingertip Pulse Oximeter are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition.
Populationadults and adolescent patientsadults and adolescent patients
ApplicationsiteFingerFinger
Performancenormal conditionnormal condition
Stand-aloneor modulestand-alonestand-alone
Single use ornotmultiple usemultiple use
Useenvironmenthome and professional caring environmenthome and professional caring environment
ComparisonUnchanged as previous submission
Test Principle
PrincipleDetermine the functional oxygenDetermine the functional oxygen
ItemSubject DevicePredicate Device
ProductNameFingertip Pulse Oximeter (Wireless)Fingertip Pulse Oximeter
Model No.8D01H and 8D01I8D01B and 8D01C
saturation of arterial hemoglobin(SpO2) and pulse rate by measuring theabsorption of red and infrared lightpassing through perfused tissue.Changes in absorption caused by thepulsation of blood in the vascular bedare used to determine SpO2 readingand pulse rate.saturation of arterial hemoglobin (SpO2)and pulse rate by measuring theabsorption of red and infrared lightpassing through perfused tissue.Changes in absorption caused by thepulsation of blood in the vascular bedare used to determine SpO2 reading andpulse rate.
WavelengthDual wavelength LED(660 nanometers @ 0.8mW and 940nanometers @ 1.2mW; both as maxaverage)Dual wavelength LED(660 nanometers @ 0.8mW and 880nanometers @ 1.2mW; both as maxaverage)
ComparisonUnchanged as previous submission
Energy
TypeBatteryBattery
BatteryAAA Alkaline battery x 1AAA Alkaline battery x 1
ComparisonUnchanged as previous submission
Operation Features
On/OffAutomatic turn on and offAutomatic turn on and off
DisplayFull color OLEDFull color OLED
Input KeyA 5-directional key (8D01H) or a singlepush-down (8D01I) keyA 5-directional key (8D01B) or a singlepush-down (8D01C) key
Warning /Indicator8D01H: Audio and visual warning8D01I: Visual indicator8D01B: Audio and visual warning8D01C: Visual indicator
8D01H:Appear red color with beep soundswhen SpO2 and pulse rate out of thesetting range.8D01B:Appear red color with beep soundswhen SpO2 and pulse rate out of thesetting range.
Warning /IndicatorFunction- Low SpO2 warning:default 87%; setting range: 50% to95%- Low SpO2 warning:default 87%; setting range: 50% to95%
ItemSubject DevicePredicate Device
ProductNameFingertip Pulse Oximeter (Wireless)Fingertip Pulse Oximeter
Model No.8D01H and 8D01I8D01B and 8D01C
High SpO2 warning:default off; setting range: 80% to 100% Low HR warning:default off; setting range: 30 to 110 bpm High HR warning:default off; setting range: 75 to 250 bpm 8D01I:Appear red color when SpO2 is lower than 87%.High SpO2 warning:default off; setting range: 80% to 100% Low HR warning:default off; setting range: 30 to 110 bpm High HR warning:default off; setting range: 75 to 250 bpm 8D01C:Appear red color when SpO2 is lower than 87%.
DisplayRotationYesYes
WirelessConnectionBLE BT 4.0None
ComparisonThe differences of operation features is an additional Bluetooth module withdata transmission. With the VIEW TRACKER app, "8D01" Leadtek Fingertip PulseOximeter (Wireless) can transmit the real-time heart rate, SpO2, and IRwaveform to the compatible mobile device for the sole purpose of datarecording, daily review, and IR waveform display, not for continuous monitoring.The software is validated according to FDA's Guidance for Content of PremarketSubmissions for Management of Cybersecurity in Medical Devices. The wirelesstesting is conducted according to FDA's Guidance Radio Frequency WirelessTechnology in Medical Devices. The difference does not raise different questionsof safety and effectiveness.General Specification
Usage Life> 18 hrs typical operation under defaultsetting
OperatingTemp.5 ºC to 40 ºC ( 41 ºF to 104 ºF)5 ºC to 40 ºC ( 41 ºF to 104 ºF)
StorageTemp.-30ºC to 70 ºC (-22 ºF to 158 ºF)-30ºC to 70 ºC (-22 ºF to 158 ºF)
ItemSubject DevicePredicate Device
ProductNameFingertip Pulse Oximeter (Wireless)Fingertip Pulse Oximeter
Model No.8D01H and 8D01I8D01B and 8D01C
Humidity10% to 90%, non-condensing for bothoperating and storage10% to 90%, non-condensing for bothoperating and storage
AtmosphericPressure700 hPa - 1013 hPa for both operatingand storage700 hPa - 1013 hPa for both operatingand storage
WaterResistanceIP22IP22
ComparisonThe differences of operation features is the usage life for supporting wirelesscommunication and it will not be considered as a NSE between the subject andpredicate device. The Fingertip Pulse Oximeter (Wireless) is substantiallyequivalent to the predicate device (K193350) concerning the generalspecification.
Classification
Applied PartType BFType BF
SafetyIEC 60601-1IEC 60601-1
EMCIEC 60601-1-2IEC 60601-1-2
HarmonizedStandardISO 80601-2-61ISO 80601-2-61
Mode ofOperationSpot checkingSpot checking
ComparisonUnchanged as previous submission
Appearance
Weightweight without battery: 26g (0,92ounces)weight without battery: 26g (0,92ounces)
SizeL67.5 mm (2.63") x W38 mm (1.48") xH25 mm (0.98")L67.5 mm (2.63") x W38 mm (1.48") xH25 mm (0.98")
ComparisonUnchanged as previous submission
Pulse Oximetry and Heat Rate Specification
Range0% to 100%0% to 100%
Resolution1%1%
Accuracy70% to 100% range ± 2%, less than 70%70% to 100% range ± 2%, less than 70%
ItemSubject DevicePredicate Device
ProductNameFingertip Pulse Oximeter (Wireless)Fingertip Pulse Oximeter
Model No.8D01H and 8D01I8D01B and 8D01C
ComparisonUnchanged as previous submissionUnchanged as previous submission
Heat Rate Specification
Range30 to 250 bpm30 to 250 bpm
Resolution1 bpm1 bpm
Accuracy$\pm$ 1 bpm or $\pm$ 1%, whichever is greater$\pm$ 1 bpm or $\pm$ 1%, whichever is greater
ComparisonUnchanged as previous submissionUnchanged as previous submission

Table 6-1. A comparison table between subject and predicate device

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

7. Functional and Safety Testing:

The Fingertip Pulse Oximeter (Wireless) is designed and conducted in accordance with the related standards. The following tests were performed to verify and validate that Leadtek Fingertip Pulse Oximeter (Wireless) meets all requirements of related standards and demonstrates substantial equivalence to the predicates.

Electrical Safety and EMC Testing

The laboratory tests of electrical safety electromagnetic compatibility testing were conducted and show that the subject device complies with IEC 60601-1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in home healthcare environment, and ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

Biocompatibility Testing

The skin contacting materials used with the subject device are the same that used for the predicate device under K193350. These materials were evaluated for biocompatibility testing in accordance with ISO 10993-1:2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" as recognized by FDA. The studies of cytotoxicity, skin sensitization, and skin irritation were performed and show that the skin contacting materials do not cause the potential risks.

{10}------------------------------------------------

Software Verification and Validation

Software verification and validation were provided in compliance with FDA Guidance for the Content of the Premarket Submission for Software Contained in Medical Devices. These verifications and validations demonstrate that the subject device work functionally and the software for the device is considered as a "moderate" level of concern, as defined by the FDA guidance, and it is identical to the predicate devices. A failure or latent flaw in the software could not directly cause serious injury or death to the patient or operator. According to FDA Guidance document, the software validation documentation summarized the required for a Moderate level of concern device and the test results are found to support the substantial equivalence of the subject device.

Coexistence Testing

Coexistence testing was performed according to IEEE/ANSI C63.27:2009 American National Standard for Evaluation of Wireless Coexistence, AAM TIR69:2017 Risk Management radio-frequency wireless coexistence for medical devices and systems, and FDA Guidance for Radio Frequency Wireless Technology in Medical Device to demonstrate that the subject device is able to coexist with interfering networks in the 2.4 GHz ISM band and the communication links for the subject device can coexist with BLE as the interfering network operating at maximum throughput and minimum transmit power. The wireless coexistence and the security of wireless data is within the limits declared in accordance with the characteristics of the subject device.

Non-clinical Performance Testing

The laboratory evaluations for the performance of the subject device was conducted to ensure the modification do not change the performance that root-mean-square (Arms) value is less than 2% within the range 70% - 100% and do not raise different questions of safety and effectiveness.

Clinical Testing

There is no change to the subject device in pulse oximetry algorithm since the subject device use the same test principle and measuring technology as the predicate device. A clinical study was conducted on the predicate device for K193350 clearance to demonstrate the safety and performance and additional clinical testing of the subject device performance is not required necessary to support the substantial equivalence.

8. Conclusion

The Fingertip Pulse Oximeter (Wireless) has the same intended use and technology and similar characteristics as that of the predicate device cleared by FDA under K193350 and manufactured

{11}------------------------------------------------

by Leadtek Research Inc. Based on the information contained in this submission demonstrate that the subject device is substantially equivalent to the predicate device and any different characteristics do not raise different questions of safety and effectiveness.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).