K Number
K210018
Date Cleared
2021-04-11

(97 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The disposable medical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The proposed device is a three-layer, flat pleated mask. Each mask is composed of a mask body, a nose piece and two ear loops. The mask body is manufactured with three layers, the inner layer and the outer layer are made of spunbond polypropylene nonwoven fabric, and the middle layer is made of meltblown polypropylene nonwoven fabric. The model of proposed device, ear-loop, is held in place over the user's mouth and nose by two elastic ear loops welded to the mask body. The elastic ear loops are made of spandex and nylon, not made from natural rubber latex. The nose piece is in the layers of face mask to allow the user to fit the face mask around his nose, which is a iron wire with polypropylene covering. The proposed device is provided non-sterile and is intended to be a single use, disposable device.

AI/ML Overview

The provided document describes the performance testing and acceptance criteria for a "Disposable Medical Mask" (K210018).

Here's the breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

TestAcceptance Criteria (per ASTM F2100-19 Level 3)Reported Device Performance (ASTM F2100-19 Level 3)Average Result
Fluid Resistance (ASTM F1862)Pass at 160 mmHgPass at 160 mmHg (96/96)N/A
Bacterial filtration efficiency (BFE) (ASTM F2101)≥ 98%Pass (96/96)99.9%
Particulate filtration efficiency (PFE) (ASTM F2299)≥ 98%Pass (96/96)99.8%
Differential pressure (Delta-P) (EN 14683)< 6.0 mmH2O/cm²Pass (96/96)4.5 mmH2O/cm²
Flammability (16 CFR 1610)Class 1Pass (96/96)N/A
Biocompatibility (ISO 10993-5 and ISO 10993-10)Device is non-cytotoxic, non-irritating, non-sensitizingDevice is non-cytotoxic, non-irritating, non-sensitizingN/A

2. Sample size used for the test set and the data provenance

  • Sample Size for Bench Testing: For each performance test (Fluid Resistance, BFE, PFE, Differential Pressure, Flammability), the test set consisted of 96 samples. These samples were composed of "32 Samples each from 3 non-consecutive lots."
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether the tests were retrospective or prospective. Given that it's a premarket notification for a device manufactured in China, it's highly probable the testing was conducted by or on behalf of the manufacturer in China, prospectively for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the document describes performance testing for a medical mask, which involves objective laboratory measurements against established standards (e.g., ASTM F2100-19, EN 14683, ISO 10993). Expert consensus or human interpretation is not used to establish "ground truth" for these types of physical and biological performance tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human readers/interpreters where there might be subjective assessments requiring conflict resolution. The tests described are objective, quantitative laboratory tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a disposable medical mask, not an AI-assisted diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical medical mask, not an algorithm or AI system.

7. The type of ground truth used

The "ground truth" for the performance tests is based on established quantitative standards and methodologies defined by organizations like ASTM, EN, and ISO. For example:

  • Fluid Resistance: Defined by ASTM F1862.
  • Bacterial Filtration Efficiency: Defined by ASTM F2101.
  • Particulate Filtration Efficiency: Defined by ASTM F2299.
  • Differential Pressure: Defined by EN 14683.
  • Flammability: Defined by 16 CFR 1610.
  • Biocompatibility: Defined by ISO 10993-5 and ISO 10993-10.

8. The sample size for the training set

This information is not applicable. There is no "training set" for a physical medical mask's performance testing. Training sets are relevant for machine learning or AI models.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

April 11, 2021

Diasia Biomedical Technology Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co., Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China

Re: K210018

Trade/Device Name: Disposable Medical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 16, 2021 Received: March 23, 2021

Dear Grace Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for

combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega -S

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210018

Device Name Disposal Medical Mask

Indications for Use (Describe)

The disposable medical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K210018

1. Contact Details


1.1 Applicant information

Applicant NameDiasia Biomedical Technology Co., Ltd.
Address2F/3F301/4F (Building A) & 2F (Building B), FuxinlinIndustrial Park, Hangcheng Industrial Zone, Taoyuan,Xixiang, Bao'an District, Shenzhen, Guangdong, China
Contact personJason Luo
Phone No.+86 150 1299 9080
E-mailluojinxin@diasia.com.cn
Date Prepared2021-04-05

1.2 Submission Correspondent

Image: logoShenzhen Joyantech Consulting Co., Ltd1713A, 17th Floor, Block A, Zhongguan Times Square,Nanshan District, Shenzhen, Guangdong Province, China
卓远天成
Phone No.+86-755-86069197
Contact personGrace Liu; Field Fu;
Contact person's e-mailgrace@cefda.com; field@cefda.com
Websitehttp://www.cefda.com

2. Device Information

Trade nameDisposable Medical Mask
Common nameSurgical Face Mask
ModelTP175x94
ClassificationII
Classification nameMask, Surgical
Product codeFXX
Regulation No.21 CFR 878.4040

3. Legally Marketed Predicate Device

Trade NameDemeMASK Surgical Mask
510(k) NumberK201479
Product CodeFXX
ManufacturerDemeTECH Corporation

4. Device Description

{4}------------------------------------------------

The proposed device is a three-layer, flat pleated mask. Each mask is composed of a mask body, a nose piece and two ear loops. The mask body is manufactured with three layers, the inner layer and the outer layer are made of spunbond polypropylene nonwoven fabric, and the middle layer is made of meltblown polypropylene nonwoven fabric.

The model of proposed device, ear-loop, is held in place over the user's mouth and nose by two elastic ear loops welded to the mask body. The elastic ear loops are made of spandex and nylon, not made from natural rubber latex.

The nose piece is in the layers of face mask to allow the user to fit the face mask around his nose, which is a iron wire with polypropylene covering.

The proposed device is provided non-sterile and is intended to be a single use, disposable device.

5. Intended Use/Indication for Use

The disposable medical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Table 1 Substantial Equivalence Comparison
ItemProposed Device(K210018)Predicate Device(K201479)Comment
Product nameDisposable Medical MaskDemeMASK Surgical MaskNone
ManufacturerDiasia BiomedicalTechnology Co., Ltd.DemeTECH CorporationNone
Product CodeFXXFXXSame
Regulation Number21 CFR § 878.404021 CFR § 878.4040Same
ClassificationClass IIClass IISame
OTC useYesYesSame
ASTM Level (ASTMF2100-19)Level 3Level 3Same
Indications for useThe disposable medicalmasks are intended to beworn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. These masks areintended for use in infectioncontrol practices to reducethe potential exposure toThe Disposal Surgical FaceMasks are intended to beworn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, body fluidsand particulate material.These face masks areintended for use in infectioncontrol practices to reducethe potential exposure toSame
blood and body fluids. This isa single use, disposabledevice(s), providednon-sterile.blood and body fluids. This isa single use, disposabledevice provided non-sterile.
Design featureEar-loopEar-loopSame
Mask styleFlat PleatedFlat PleatedSame
UseSingle Use, DisposableSingle Use, DisposableSame
Specifications andDimensionsLength: 17.5cm±0.5cmWidth: 9.4cm ±0.5cmLength: 17.5cm±1cmWidth: 9.5cm±1cmSimilar
SterilityNon-SterileNon-SterileSame
Materials
Outer layerSpunbond polypropyleneSpunbond polypropyleneSame
Middle layerMeltblown polypropylenefilterMeltblown polypropylenefilterSame
Inner layerSpunbond polypropyleneSpunbond polypropyleneSame
Nose pieceIron wire with polypropylenecoveringGalvanized wire coated withpolyethyleneDifferent(Issue 1)
Ear loopSpandex and Nylon - Notmade from natural rubberlatexSpandex and Nylon - Notmade from natural rubberlatexSame
Performance
Fluid ResistanceASTM F1862Pass at 160 mmHgPass at 160 mmHgSame
Bacterial FiltrationEfficiencyASTM F2101Pass at ≥99%Pass at ≥99%Same
Particulate FiltrationEfficiencyASTM F2299Pass at ≥99%Pass at ≥99%Same
Differential Pressure(Delta-P)Pass at <6.0 mmH2O/cm²Pass at <6.0 mmH2O/cm²Same
Flammability16 CFR 1610Class 1Class 1Same
BiocompatibilityISO 10993-5 andISO 10993-10Under the conditions of thestudies employed, thedevice is non-cytotoxic,non-sensitizingUnder the conditions of thestudies employed, thedevice is non-cytotoxic,non-sensitizingSame

6. Substantial Equivalence Comparison

Table 1 Substantial Equivalence Comparison

{5}------------------------------------------------

lssue 1: The differences in the materials do not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials.

{6}------------------------------------------------

7. Non-clinical Testing

7.1 Biocompatibility testing

The Medical Surgical Mask has been subjected to biocompatibility studies to demonstrate the safety of device. The biocompatibility studies are in accordance with ISO10993:

  • A In Vitro Cytotoxicity (ISO 10993-5): the device was non-cytotoxic;
  • ♪ Skin Irritation (ISO 10993-10): the device was non-irritating;
  • Skin Sensitization (ISO 10993-10): the device was non-sensitizing.

There is no additional safety risk for the proposed device when compared with the predicate device.

7.2 Performance testing - Bench

The performance testing was determined according to "ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks" to demonstrate the effectiveness of device.

TestPurposeAcceptance Criteriaper ASTM F2100-19Level 3 (AQL=4.0%)Subject Device Test ResultsASTM F2100-19Level 3Average
Fluid Resistance(ASTM F1862)Determine syntheticblood penetrationresistance.Pass at 160 mmHgPass at 160 mmHg(96/96)32 Samples each from3 non-consecutive lotsN/A
Bacterial filtrationefficiency (BFE)(ASTM F2101)Determine thebacterial filtrationefficiency.$\ge$ 98%Pass (96/96)32 Samples each from3 non-consecutive lots99.9%
Particulatefiltrationefficiency (PFE)(ASTM F2299)Determinesubmicronparticulate filtrationefficiency at 0.1micron.$\ge$ 98%Pass (96/96)32 Samples each from3 non-consecutive lots99.8%
Differentialpressure(Delta-P)(EN 14683)Determine breathingresistance ordifferential pressure.< 6.0 mmH2O/cm²Pass (96/96)32 Samples each from3 non-consecutive lots4.5mmH2O/cm²
Flammability(16 CFR 1610)Determineflammability or flamespread.Class 1Pass (96/96)32 Samples each from3 non-consecutive lotsN/A

Table 2 Summary of Performance Testing

8. Clinical Testing

{7}------------------------------------------------

No clinical study is included in this submission.

9. Conclusions

The proposed device has the same indication for use and similar technological characteristic as the predicate device. Non-clinical testing demonstrates that the proposed device performs as safe and effective as the predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.