K Number
K203640

Validate with FDA (Live)

Device Name
CoolSeal Reveal
Date Cleared
2021-05-06

(143 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.

The CoolSeal™ Reveal is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, and parotidectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.

The CoolSeal™ Reveal has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Reveal for these procedures.

Device Description

The CoolSeal™ Reveal device is a sterile, single-use, hand-held bipolar vessel sealing device designed for use with the CoolSeal™ Generator. The CoolSeal™ Reveal is a hemostat-style design with a 5 mm shaft diameter and 10 cm shaft length. The jaw is a curved Maryland-style jaw that rotates. The Reveal creates seals by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between its jaws. A blade within the instrument is surgeon-actuated to divide tissue. The double-action jaws have been designed to dissect tissue, which includes separating tissue planes and widening openings as necessary for the surgical procedure.

AI/ML Overview

This is a 510(k) premarket notification for the CoolSeal Reveal electrosurgical cutting and coagulation device. The document does not describe the acceptance criteria and study for AI/ML performance. Instead, it focuses on the substantial equivalence of the CoolSeal Reveal device to predicate devices (LigaSure Exact Dissector, Nano-Coated and JustRight Sealer) based on various performance data types.

Therefore, the following information regarding AI/ML performance acceptance criteria and study specifics cannot be extracted from the provided text:

  • A table of acceptance criteria and the reported device performance
  • Sample size used for the test set and the data provenance
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts
  • Adjudication method for the test set
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
  • The type of ground truth used
  • The sample size for the training set
  • How the ground truth for the training set was established

The document details the following performance data to support substantial equivalence:

  • Biocompatibility Testing: Conducted in accordance with ISO 10993-1.
  • Electrical Safety and Electromagnetic Compatibility (EMC): Complies with ANSI/AAMI/EC 60601-2-2 and ANSI/AAMI/IEC 60601-1-2 standards.
  • Mechanical and Functional Testing: Carried out to verify expected performance.
  • Ex-vivo and In-vivo Vessel Burst Pressure:
    • Ex-vivo: Performed on excised fresh porcine blood vessels for both the subject device and the primary predicate.
    • In-vivo: Performed on lymphatics in live animals for both the subject device and the primary predicate.
  • In-vivo Thermal Spread Comparison: Open laparotomy performed in porcine and ovine models, using both the subject device and the primary predicate device to seal vessels and tissue bundles. Samples were excised for three-dimensional histological assessments (length, width, and depth) to quantify thermal spread.
  • Chronic Animal Study: A study of 23 days to assess the safety and performance of bipolar vessel sealing with the subject device. All animals survived without complications, and vessel sealing effects maintained chronic hemostasis and healed as anticipated.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

May 6, 2021

Bolder Surgical, LLC Nicholas Wong Regulatory Affairs Manager 331 S. 104th Street, Suite 200 Louisville, Colorado 80027

Re: K203640

Trade/Device Name: CoolSeal Reveal Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 9, 2021 Received: April 12, 2021

Dear Nicholas Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203640

Device Name CoolSeal Reveal

Indications for Use (Describe)

The CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.

The CoolSeaT™ Reveal is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, and parotidectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermallysensitive structures such as nerves and parathyroid glands.

The CoolSeal™ Reveal has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Reveal for these procedures.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue letters. The word "SURGICAL" is in smaller, thinner, blue letters below the word "BOLDER". The letter "O" in "BOLDER" has a black diamond shape in the upper left quadrant.

510(k) Summary

SUBMITTER

Bolder Surgical, LLC. 331 S. 104th Street, Suite 200 Louisville, CO 80027

Phone: 720-287-7130 Fax: 720-287-7135

Contact Person: Nick Wong Regulatory Affairs Manager

Date Prepared: May 6th, 2021

DEVICE

Trade Name / Model #:CoolSeal™ Reveal / CSL-RV105-10
Common Name:Bipolar Vessel Sealing System
Classification Name:Electrosurgical cutting and coagulation device and accessories
Regulatory Class:II
Produce Code:GEI

PREDICATE DEVICES

Trade Name:LigaSure Exact Dissector, Nano-Coated
510(k):K173281
Common or Usual Name:Bipolar Vessel Sealing System
Classification Name:Electrosurgical cutting and coagulation device and accessories
Regulatory Class:II
Produce Code:GEI
Trade Name:JustRight™ Sealer
510(k)K160602
Common or Usual Name:Bipolar Vessel Sealing System
Classification Name:Electrosurgical cutting and coagulation device and accessories
Regulatory Class:II
Produce Code:GEI

DEVICE DESCRIPTION

CoolSeal™ Reveal:

The CoolSeal™ Reveal device is a sterile, single-use, hand-held bipolar vessel sealing device designed for use with the CoolSeal™ Generator. The CoolSeal™ Reveal is a hemostat-style

{4}------------------------------------------------

design with a 5 mm shaft diameter and 10 cm shaft length. The jaw is a curved Maryland-style jaw that rotates. The Reveal creates seals by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between its iaws. A blade within the instrument is surgeon-actuated to divide tissue. The double-action jaws have been designed to dissect tissue, which includes separating tissue planes and widening openings as necessary for the surgical procedure.

INDICATIONS FOR USE

CoolSeal™ Reveal:

The CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.

The CoolSeal™ Reveal is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, and parotidectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.

The CoolSeal™ Reveal has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Reveal for these procedures.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue font, with a circle in place of the letter "O". A small black diamond is in the upper left quadrant of the circle. Below "BOLDER" is the word "SURGICAL" in a smaller, thinner, black font.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

Tables 1 and 2 below present the comparison of technological characteristics between the subject device and the two identified predicate devices.

DescriptionCoolSeal™ Reveal (Subject Device)LigaSure Exact (Primary Predicate)
Indications for UseThe CoolSeal™ Reveal is a bipolar electrosurgicalinstrument intended for use in open surgical procedures inadults and pediatrics where ligation and division ofvessels, tissue bundles, and lymphatics is desired. TheCoolSeal™ Reveal can be used on vessels (arteries andveins) up to and including 6 mm in diameter. It isindicated for use in general surgery and in such surgicalspecialties as urologic, thoracic, plastic, andreconstructive. Procedures may include but are not limitedto, bowel resections, gall bladder procedures, Nissenfundoplication, and adhesiolysis.The CoolSeal™ Reveal is also indicated for open ENTprocedures in adults (thyroidectomy, radical neckdissection, and parotidectomy) for ligation and division ofvessels, lymphatics and tissue bundles 2-3 mm away fromunintended thermally-sensitive structures such as nervesand parathyroid glands.The CoolSeal™ Reveal has not been shown to be effectivefor tubal sterilization or tubal coagulation for sterilizationprocedures. Do not use the CoolSeal™ Reveal for theseprocedures.The LigaSure Sealer/Divider is a bipolar electrosurgical instrumentintended for use in open surgical procedures where ligation anddivision of vessels, tissue bundles, and lymphatics is desired. TheLigaSure Sealer/ Divider can be used on vessels (arteries and veins)up to and including 7 mm. It is indicated for use in general surgeryand in such surgicalspecialties as urologic, thoracic, plastic, and reconstructive.Procedures may include, but are not limited to, bowel resections, gallbladder procedures, Nissen fundoplication, and adhesiolysis.The instrument is also indicated for open ENT procedures in adults(thyroidectomy, radical neck dissection, parotidectomy, andtonsillectomy) for ligation and division of vessels, lymphatics andtissue bundles 2-3 mm away from unintended thermally-sensitivestructures such as nerves and parathyroid glands.The LigaSure system has not been shown to be effective for tubalsterilization or tubal coagulation for sterilizationprocedures. Do not use the LigaSure system for these procedures.
Where used(environment)Operating RoomOperating Room
Intended UserSurgeonSurgeon
DescriptionCoolSeal™ Reveal (Subject Device)LigaSure Exact (Primary Predicate)
Anatomical SitesVessels, tissue bundles, and lymphaticsVessels, tissue bundles, and lymphatics
Anatomical SizeUp to and including 6 mm diameterUp to and including 7 mm diameter
PatientPopulationAdult and Pediatric in General Surgery.Adult in ENT surgery.Not specified (Adults assumed)
Power SourceBipolar energy platformBipolar energy platform
PrimaryFunctionsGrasp, Dissect, Seal, DivideGrasp, Dissect, Seal, Divide
Mechanism ofGraspingHand actuated lever allows user to open or closeHand actuated lever allows user to open or close
Mechanism ofDissection(Separation)Bilateral jaw allows user to separate planes of tissue
Mechanism ofAction (Sealing)Seal is created by application of RF energy to structuresinterposed between the jaws of the instrument.Seal is created by application of RF energy to structures interposedbetween the jaws of the instrument.
Seal ActivationButton on the sealer instrument is activated by user fingerdepression.Button on the sealer instrument is hand activated by closing thehandles until the button is depressed.
Knife ActivationCutting trigger – non-energizedCutting trigger – non-energized
Automaticsealing cycleYesYes
Rated Voltage(Vpeak)190 Vpeak288 Vpeak
Shaft Diameter5 mmN/A
Shaft Length10 cmN/A
Jaw Rotation$≤$ 335°N/A
Jaw Dimensions– Jaw Lengthand Side WidthJaw Length: 12 mmProximal Side Width: 4.3 mmDistal Side Width: 2.0 mmJaw Length: 20.6 mmProximal Side Width: 4.2 mmDistal Side Width: 2.3 mm
Jaw Dimensions- Tip Width(proximal anddistal)Proximal: 5.2 mmDistal: 3.2 mmProximal: 6.7 mmDistal: 3.6 mm
Seal Plate WidthSeal Plate Width: 1.8 mm throughoutSeal Plate Width:
DescriptionCoolSeal™ Reveal (Subject Device)LigaSure Exact (Primary Predicate)
Proximal: 4.0 mmDistal: 1.75 mm
Maximum sizestructure (vessel,tissue bundles,lymphatic) forsealingUp to 6 mmUp to 7 mm
Tissue ContactMaterialsStainless steel, titanium, polymers, silicone, polymeradhesives and insulating coatingsStainless steel, titanium, polymers, silicone, polymer adhesives andinsulating coatings
SurgicalApproachOpenOpen
SterilizationEthylene OxideEthylene Oxide
SterilityAssurance Level10-610-6

Table 1. CoolSeal™ Reveal vs. LigaSure™ Exact

{6}------------------------------------------------

{7}------------------------------------------------

Table 2. CoolSeal™ Reveal vs. JustRight™ Sealer Comparison Table

DescriptionCoolSeal™ Reveal (Subject Device)JustRight Sealer (Pediatric Predicate)
Indications for UseThe CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.The CoolSeal™ Reveal is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, and parotidectomy) for ligation and divisionThe JustRight Surgical® Vessel Sealing System is intended for use in open and laparoscopic general surgical procedures to seal blood vessels and vascular bundles up to and including 5 mm in diameter for use in adult and pediatric populations, wherever vessel ligation is required.The device is contraindicated for use in ENT procedures.
DescriptionCoolSeal™ Reveal (Subject Device)JustRight Sealer (Pediatric Predicate)
of vessels, lymphatics and tissue bundles 2-3 mm awayfrom unintended thermally-sensitive structures such asnerves and parathyroid glands.The CoolSeal™ Reveal has not been shown to beeffective for tubal sterilization or tubal coagulation forsterilization procedures. Do not use the CoolSeal™Reveal for these procedures.
Where used(environment)Operating RoomOperating Room
Intended UserSurgeonSurgeon
AnatomicalSitesVessels, tissue bundles, and lymphaticsVessels, and tissue bundles
AnatomicalSizeUp to and including 6 mm diameterUp to and including 5 mm diameter
PatientPopulationAdult and Pediatric in General Surgery.Adult in ENT surgery.Adult and Pediatric in General Surgery.
Power SourceBipolar energy platformBipolar energy platform
PrimaryFunctionsGrasp, Dissect, Seal, DivideGrasp, Dissect, Seal
Mechanism ofGraspingHand actuated lever allows user to open or closeHand actuated lever allows user to open or close
Mechanism ofDissection(Separation)Bilateral jaw allows user to separate planes of tissueBilateral jaw allows user to separate planes of tissue
Mechanism ofAction(Sealing)Seal is created by application of RF energy to structuresinterposed between the jaws of the instrument.Seal is created by application of RF energy to structures interposedbetween the jaws of the instrument.
Seal ActivationButton on the sealer instrument is activated by userfinger depression.Button on the sealer instrument is activated by user finger depression.
KnifeActivationCutting trigger – non-energizedNo cutting functionality.
DescriptionCoolSeal™ Reveal (Subject Device)JustRight Sealer (Pediatric Predicate)
Automaticsealing cycleYesYes
Rated Voltage(Vpeak)190 Vpeak190 Vpeak
MaximumOutput Power30 Watts25 Watts
Shaft Diameter5 mm3 mm
Shaft Length10 cm20 cm
Jaw Rotation≤ 335°≤ 340°
JawDimensions –Jaw Length andSide WidthJaw Length: 12 mmProximal Side Width: 4.3 mmDistal Side Width: 2.0 mmJaw Length: 10 mmProximal Side Width: 3.3 mmDistal Side Width: 1.4 mm
JawDimensions –Tip Width(proximal anddistal)Proximal: 5.2 mmDistal: 3.2 mmProximal: 3.3 mmDistal: 2.2 mm
Seal PlateWidthSeal Plate Width: 1.8 mm throughoutSeal Plate Width:Proximal: 3.3 mmDistal: 1.4 mm
ThermalCoating onJawsPresentPresent
Maximum sizestructure(vessel, tissuebundles,lymphatic) forsealingUp to 6 mmUp to 5 mm
Tissue ContactMaterialsStainless steel, titanium, polymers, silicone, polymeradhesives, and insulating coatingsStainless steel, titanium, polymers, silicone, polymer adhesives, andinsulating coatings
DescriptionCoolSeal™ Reveal (Subject Device)JustRight Sealer (Pediatric Predicate)
SurgicalApproachOpenOpen and Laparoscopic
SterilizationEthylene Oxide
SterilityAssuranceLevel$10^{-6}$$10^{-6}$

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue letters, with a black diamond shape inside the "O". Below "BOLDER" is the word "SURGICAL" in smaller, thinner, blue letters. The logo is simple and modern.

{9}------------------------------------------------

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue letters, with a black diamond shape inside the "O". Below "BOLDER" is the word "SURGICAL" in smaller, thinner, black letters. The logo is simple and modern, and the use of blue and black gives it a professional look.

PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing

The biocompatibility evaluation for the CoolSeal™ Reveal was conducted in accordance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," as recognized by FDA.

Electrical Safety and Electromagnetic Compatibility (EMC)

The CoolSeal™ Reveal complies with relevant clauses of the ANSI/AAMI/EC 60601-2-2 standards for electrical safety and ANSI/AAMI/IEC 60601-1-2 standard for EMC.

Mechanical and Functional Testing

Mechanical, electrical, and functional testing was carried out to verify that the proposed device performed as expected.

Ex-vivo and In-vivo Vessel Burst Pressure

Ex-vivo burst pressure testing of excised fresh porcine blood vessels was conducted on the subject device and primary predicate to demonstrate effective bipolar electrosurgical vessel sealing performance. Additionally, in-vivo burst pressure testing of lymphatics was conducted on the subject device and primary predicate to demonstrate effective bipolar electrosurgical vessel sealing performance.

In-vivo Thermal Spread Comparison

Open laparotomy was performed in porcine and oving the subject device and primary predicate device. Each device was used to seal vessels and tissue bundles. Samples were excised for three-dimensional histological assessments (e.g., length, width, and

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue font, with a stylized "O" that includes a black triangle. Below "BOLDER" is the word "SURGICAL" in a smaller, thinner, blue font.

depth) to quantify thermal spread in seals created by both demonstrated that the subject device is as safe and effective as the predicate device.

Chronic Animal Study

A chronic study was conducted to assess the safety and performance of bipolar vessel sealing with the subject device of 23 days. All animals survived 23 days post-op without any complications. All vessel sealing effects on tissue maintained chronic hemostasis and healed as anticipated.

CONCLUSIONS

Based on a review of performance data, comparison of the device classification, intended use, operating principles, technological characteristics, sterility, and biocompatibility, the subject device is safe, as effective, and performs as well as the legally marketed predicate devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.