K Number
K203527
Date Cleared
2022-05-09

(523 days)

Product Code
Regulation Number
862.3650
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use.

The Immunalysis Tapentadol Urine HEIA™ is a homogeneous enzyme immunoassay with a cutoff of 200 ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of tapentadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against tapentadol. This in vitro diagnostic device is for prescription use only.

The Immunalysis Tapentadol Urine HEIA™ provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation using a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

Device Description

The Immunalysis Tapentadol Urine HEIA™ is a homogeneous enzyme immunoassay intended for use in laboratories for the qualitative and semi-quantitative analysis of tapentadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against tapentadol. This in vitro diagnostic device is for prescription use only.

The Immunalysis Tapentadol Urine HEIA™ provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation using a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

AI/ML Overview

The document describes the performance characteristics of the Immunalysis Tapentadol Urine HEIATM, a homogeneous enzyme immunoassay intended for qualitative and semi-quantitative analysis of tapentadol in human urine.

Here's an breakdown of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of "acceptance criteria" alongside the "reported device performance" in a single, consolidated table. However, the performance characteristics section (G) details various studies with implicit acceptance criteria and their corresponding results. I've compiled this information into a table format based on the presented data:

Study/Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
Precision - Qualitative (Concentrations Relative to Cutoff)Samples at -100%, -75%, -50%, -25% of cutoff should be Negative. Samples at +25%, +50%, +75%, +100% of cutoff should be Positive. Sample at Cutoff (200 ng/mL) expected to show a mix of positive/negative.Lot 1: -100% to -25%: 80 Negative Cutoff: 41 Neg / 39 Pos +25% to +100%: 80 Positive Lot 2: -100% to -25%: 80 Negative Cutoff: 35 Neg / 45 Pos +25% to +100%: 80 Positive Lot 3: -100% to -25%: 80 Negative Cutoff: 42 Neg / 38 Pos +25% to +100%: 80 Positive
Precision - Semi-Quantitative (%CV)%CV should be within acceptable limits (typically <10-20% for immunoassays depending on concentration).%CV ranged from 3.6 to 9.7 for all lots tested.
Specificity and Cross-Reactivity (Qualitative)Structurally and functionally similar compounds (except known metabolites) should not cross-react (i.e., yield negative results or equivalent concentration <200 ng/mL).All tested compounds (e.g., Chlorpromazine, Clomipramine, Tramadol) showed < 0.2% cross-reactivity and were negative, except for N-desmethyl tapentadol (15.7% cross-reactivity) and tapentadol glucuronide (0.5% cross-reactivity), which are expected metabolites and showed positive results at equivalent concentrations.
Specificity and Cross-Reactivity (Semi-Quantitative % Cross-Reactivity)% of cross-reactivity for structurally and functionally similar compounds (except known metabolites) should be low.% of cross-reactivity was < 0.2% for all compounds tested except for N-desmethyl tapentadol (15.7%) and tapentadol glucuronide (0.5%).
Interference – Structurally Unrelated CompoundsNo interference observed at specified concentrations.No interference observed for all 90+ listed compounds at the tested concentrations (mostly 100,000 ng/mL, some 50,000 ng/mL).
Interference – Endogenous Compounds and Urine PreservativesNo interference observed at specified concentrations.No interference observed for listed endogenous compounds (e.g., Acetone, Bilirubin, Glucose) and urine preservatives (Sodium Azide, Sodium Fluoride) at specified concentrations. Boric acid showed no interference at ±50% of cutoff.
Interference – pHNo positive or negative interference over the tested pH range.No positive or negative interference observed at urine pH values ranging from 3.0 to 11.0.
Interference – Specific GravityNo positive or negative interference over the tested specific gravity range.No positive or negative interference observed at urine specific gravity values ranging from 1.000 to 1.030.
Linearity/RecoveryThe assay should demonstrate linearity and acceptable recovery over the specified range.Confirmed linear range 100-1100 ng/mL. Assay drug recovery percentage ranged from 95.8 to 110.4%.
Calibration DurationMaintain performance for the recommended calibration frequency.Test results met acceptance criteria up to 14 days, supporting a recommended 14-day calibration frequency.
Tapentadol Stability in UrineTapentadol samples should remain stable for a specified duration under defined storage conditions.Urine samples stable for up to 14 days stored at 2°C - 8°C.
Method Comparison (Qualitative & Semi-Quantitative)High agreement (PPA, NPA) with LC-MS/MS.PPA: 100% (95/95) and NPA: 100% (65/65) for both qualitative and semi-quantitative modes.

2. Sample size used for the test set and the data provenance

  • Precision Study:
    • Sample Size: 9 panel members (concentrations from 0 to 400 ng/mL, including drug-free negative and spiked concentrations relative to the 200 ng/mL cutoff). Each panel member was tested in 80 replicates (20 days, 2 runs/day, 2 replicates/run), across 3 lots of reagent.
    • Data Provenance: Drug-free negative urine was used as the base sample. Spiked concentrations were confirmed by mass spectrometry (LC-MS/MS). The document does not explicitly state the country of origin of the data, nor whether it was retrospective or prospective, but precision studies are typically prospective laboratory experiments.
  • Specificity and Cross-Reactivity:
    • Sample Size: Varies per compound tested (implied multiple replicates to determine exact cross-reactivity percentage). Compounds were spiked into drug-free urine.
    • Data Provenance: Laboratory study using spiked samples.
  • Interference (Structurally Unrelated Compounds, Endogenous Compounds, Urine Preservatives, pH, Specific Gravity):
    • Sample Size: Varies per compound/condition. Potential interferents were spiked into drug-free urine containing tapentadol at ±25% of the cutoff (and ±50% for boric acid).
    • Data Provenance: Laboratory study using spiked samples.
  • Linearity/Recovery:
    • Sample Size: A drug-free urine pool and a pool spiked with high concentration of tapentadol. Serially diluted to achieve 12 concentrations (0 to 1100 ng/mL). Each pool was tested in triplicate.
    • Data Provenance: Laboratory study using spiked samples.
  • Calibration Duration:
    • Sample Size: Drug-free negative urine spiked with tapentadol at ±25% of the cutoff. Tested at multiple time points up to 14 days.
    • Data Provenance: Laboratory study using spiked samples.
  • Tapentadol Stability in Urine:
    • Sample Size: Urine samples from 4 participants who reported taking tapentadol.
    • Data Provenance: Clinical samples, stored and tested over time.
  • Method Comparison:
    • Sample Size: 160 de-identified remnant unaltered clinical urine samples.
    • Data Provenance: Retrospective; samples obtained from clinical testing laboratories. Country of origin not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This device is an in vitro diagnostic (IVD) assay for substance detection, not an AI/imaging device requiring expert interpretation for ground truth.

  • For the precision, specificity, linearity, and interference studies, the ground truth was established by known concentrations of analytes/compounds prepared in the laboratory, often confirmed by mass spectrometry (LC-MS/MS), which is considered a gold standard analytical method. No human experts were involved in establishing this type of ground truth.
  • For the method comparison study, the Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) acted as the reference method (ground truth). This is an analytical chemistry technique, not dependent on expert visual interpretation. The document references "Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry" as the instrument used.

Therefore, the concept of "number of experts" and "qualifications of those experts" does not apply in the same way it would for imaging-based AI diagnostics.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The ground truth for this chemical assay is primarily based on quantitative analytical methods (LC-MS/MS for confirmation/comparison) and precisely prepared spiked samples with known concentrations. There is no human interpretation or subjective assessment that would require an adjudication method among experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic tool that involves human readers interpreting results. Therefore, an MRMC study or a study on human reader improvement with AI assistance was not conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a standalone immunoassay (a chemical test) that provides a result (qualitative or semi-quantitative). Its performance is evaluated directly against analytical ground truth (LC-MS/MS). The concept of "algorithm only" or "human-in-the-loop" isn't directly applicable in the same way it is for AI-driven software where human input might influence the output. The assay generates a quantifiable signal which is then interpreted against a cutoff, which is an automated process on a clinical chemistry analyzer.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The primary type of ground truth used was:

  • Known concentrations/preparations: For precision, specificity, interference, linearity, calibration duration.
  • Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS): Considered a gold standard confirmatory method for drug testing, used as the reference method (ground truth) for the method comparison study.

8. The sample size for the training set

This document describes a traditional in vitro diagnostic immunoassay, not an AI/machine learning model. Therefore, there is no "training set" in the context of machine learning. The assay's performance characteristics are inherent to its chemical/biological design and reagents, not learned from data.

9. How the ground truth for the training set was established

Not applicable, as there is no "training set" for a traditional immunoassay.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 9, 2022

Immunalysis Corporation Elina Arroyo Manager Regulatory Affairs 829 Towne Center Drive Pomona, California 91767

Re: K203527

Trade/Device Name: Immunalysis Tapentadol Urine HEIA™ Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG Dated: February 10, 2022 Received: February 11, 2022

Dear Elina Arroyo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203527

Device Name Immunalysis Tapentadol Urine HEIA™

Indications for Use (Describe)

For in vitro diagnostic use.

The Immunalysis Tapentadol Urine HEIA™ is a homogeneous enzyme immunoassay with a cutoff of 200 ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of tapentadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against tapentadol. This in vitro diagnostic device is for prescription use only.

The Immunalysis Tapentadol Urine HEIA™ provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation using a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

510(K) NUMBER: K203527

A. GENERAL INFORMATION

Applicant Name:Immunalysis Corporation
829 Towne Center Drive
Pomona, CA 91767
Establishment # 2020952
Company Contact:Elina Arroyo
Associate Director, Regulatory Affairs
1 312-208-1514
elina.arroyo@abbott.com
Date Prepared:February 10, 2022

B. DEVICE IDENTIFICATION

Trade or Proprietary Names: Immunalysis Tapentadol Urine HEIA™

Common Name: Tapentadol Urine Enzyme Immunoassay

C. REGULATORY INFORMATION

Device Classification Name: Enzyme Immunoassay, Opiates
Product Codes:DJG
Regulatory Class:Class II
Classification Regulation:21 CFR 862.3650, Opiate Test System
Panel:Toxicology (91)
Predicate Device:Immunalysis Tramadol Enzyme Immunoassay [K141803]

{4}------------------------------------------------

D. DEVICE DESCRIPTION

The Immunalysis Tapentadol Urine HEIA™ is a homogeneous enzyme immunoassay intended for use in laboratories for the qualitative and semi-quantitative analysis of tapentadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against tapentadol. This in vitro diagnostic device is for prescription use only.

The Immunalysis Tapentadol Urine HEIA™ provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation using a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

Tapentadol is a centrally acting analgesic approved by the FDA in 2008 to treat moderate to severe pain in adults. It is a u-opioid receptor agonist and is classified as a Schedule II controlled substance. The therapeutic doses of tapentadol are generally between 50 and 600 mg/ day (the range specified by the package insert)} with a median of 200 mg/day. About 99% of ingested tapentadol undergoes glucuronide conjugation (55%) and sulfate conjugation (15%) before excretion into the urine233. Wu et a14. described urine concentrations of tapentadol and its metabolites in 736 samples. The parent drug and all three metabolites were detected together in 79.9% of positive specimens (n = 586); the parent drug was detected together with glucuronide conjugate in 7.1% of positive samples. Parent drug was detected without metabolites in 49 samples. The median

1 Janssen Pharmaceuticals, Inc. https://www.nucynta.com/hcp/ir/individualized-dosing/

1 Terlinden R, Ossig J, Fliegert F, Lange C, Goehler K. Absorption, metabolism, and excretion of 14C-labeled tapentadol HCI in healthy male subjects. Eur J Drug Metab Pharmacokinetics 2007; 32:163-169.

3 DePriest AZ, Puet BL, Holt AC, Roberts A, Cone EJ. Metabolism and disposition opioids: A Review. Forensic Sci Rev. 2015 Jul;27(2):115-45. Review.

4 Wu F, Slawson MH, Johnson-Davis KL. Metabolic patterns of fentanyl, meperidine, methylphenidate, tapentadol and tramadol observed in urine, serum or plasma. J Anal Toxicol. 2017;41(4):289-299

{5}------------------------------------------------

concentration observed for tapentadol was 3,104 ng/mL, and the concentrations of tapentadol exceeding 5,000 ng/mL were found in 37.5% of positive samples.

E. INTENDED USE

For in vitro diagnostic use.

The Immunalysis Tapentadol Urine HEIA™ is a homogeneous enzyme immunoassay with a cutoff of 200 ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of tapentadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against tapentadol. This in vitro diagnostic device is for prescription use only.

The Immunalysis Tapentadol Urine HEIA™ provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation using a confirmatory method such as Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

F. COMPARISON WITH PREDICATE

Immunalysis Tramadol Enzyme Immunoassay was selected as the predicate because both Tapentadol and Tramadol are in the opioid family and share identical characteristics except for the analyte being detected. The following Table 5-1 includes a summary of the technological characteristics.

{6}------------------------------------------------

AttributeCandidate DeviceImmunalysis Tapentadol UrineHEIATMPredicate DeviceImmunalysis Tramadol EnzymeImmunoassay [K141803]
Similarities
Intended UseIdenticalFor the qualitative and semi-quantitative analysis of an opioid inhuman urine with automatedclinical chemistry analyzers.
Test PrincipleIdenticalHomogeneous enzymeimmunoassay
UserEnvironmentIdenticalFor use in laboratories
Sample MatrixIdenticalHuman urine
MassSpectrometryConfirmationIdenticalRequired for preliminary positiveanalytical results
Reagent StorageIdentical2-8°C until expiration date
Assay MaterialsIdenticalTwo analytical reagents:antibody/substrate reagent andenzyme labeled conjugate reagent
Assay CutoffLevelIdentical200 ng/mL
InstrumentationIdenticalAutomated Clinical chemistryanalyzer
Differences
AntibodyRecombinant FAB antibody toTapentadolGoat Polyclonal Antibody toTramadol
CalibratorTapentadolTramadol

Table 5-1 Comparison to Predicate Device

G. PERFORMANCE CHARACTERISTICS

The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis Tapentadol Urine HEIA™ to the predicate device. Assay performance was established using the Beckman Coulter AU480 chemistry analyzer.

1. Precision

Precision study was performed over 20 days, two runs per day in duplicates (20 x 2 x 2 replicates per panel member) for a total of 80 replicates (N=80) on 3 lots of reagent. Nine panel members were made using Drug free negative urine as the base sample and 8 panel members were spiked to concentrations of assay cutoff and ±25%, ±50%, ±75%, ±100%

{7}------------------------------------------------

of the cutoff (200 ng/mL). The concentrations used for spiking were confirmed by mass spectrometry (LC-MS/MS). The data demonstrates all sample concentration ranging from -100% to -25% of the cutoff were negative and all sample concentrations ranging from +25% to +100% of cutoff were positive for both qualitative and semi-quantitative interpretations. The %CV of the semi-quantitative results ranged from 3.6 to 9.7 for all lots tested. The study established the repeatability of the assay. Test results in qualitative and semi-quantitative modes for a representative lot are presented in Tables 2 to 7.

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%8080 Negative
50-75%8080 Negative
100-50%8080 Negative
150-25%8080 Negative
200Cutoff8041 Neg / 39 Pos
250+25%8080 Positive
300+50%8080 Positive
350+75%8080 Positive
400+100%8080 Positive

Table 2. Precision – Qualitative Lot#1

Table 3. Precision - Qualitative Lot#2

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%8080 Negative
50-75%8080 Negative
100-50%8080 Negative
150-25%8080 Negative
200Cutoff8035 Neg / 45 Pos
250+25%8080 Positive
300+50%8080 Positive
350+75%8080 Positive
400+100%8080 Positive

{8}------------------------------------------------

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%8080 Negative
50-75%8080 Negative
100-50%8080 Negative
150-25%8080 Negative
200Cutoff8042 Neg / 38 Pos
250+25%8080 Positive
300+50%8080 Positive
350+75%8080 Positive
400+100%8080 Positive

Table 4. Precision – Qualitative Lot#3

Table 5. Precision - Semi-Quantitative Lot#1

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%80-380 Negative
50-75%804880 Negative
100-50%809980 Negative
150-25%8016180 Negative
200Cutoff802152 Neg / 78 Pos
250+25%8026880 Positive
300+50%8030480 Positive
350+75%8037280 Positive
400+100%8043180 Positive

Table 6. Precision - Semi-Quantitative Lot#2

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%80-480 Negative
50-75%805080 Negative
100-50%8010080 Negative
150-25%8016280 Negative
200Cutoff802138 Neg / 72 Pos
250+25%8026380 Positive
300+50%8029880 Positive
350+75%8035980 Positive
400+100%8041780 Positive

{9}------------------------------------------------

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%80-380 Negative
50-75%804980 Negative
100-50%8010280 Negative
150-25%8016280 Negative
200Cutoff802157 Neg / 73 Pos
250+25%8026680 Positive
300+50%8030380 Positive
350+75%8036980 Positive
400+100%8042280 Positive

Table 7. Precision - Semi-Quantitative Lot#3

2. Specificity and Cross-Reactivity

Structurally and functionally similar compounds were spiked into drug free urine at levels that will yield a result that is equivalent to the cutoff. The data demonstrates all sample compounds tested were negative by both the qualitative and semi-quantitative interpretations except for N-desmethyl tapentadol glucuronide. The % of cross- reactivity of the semi-quantitative results was < 0.2 % for all compounds tested except for N-desmethyl tapentadol with 15.7% and tapentadol glucuronide with 0.5% of cross- reactivity. Cross-reactivity test results in qualitative mode are presented in Table 8. Cross-reactivity test results in semi-quantitative mode are presented in Table 9.

CompoundCompound Conc.(ng/mL)TapentadolEquivalentConc.(ng/mL)ResultCross-Reactivity(%)
Chlorpromazine100,000< 200NEG<0.2
Clomipramine100,000<200NEG<0.2
Cyclobenzaprine100,000<200NEG<0.2
Doxepin100,000<200NEG<0.2
Imipramine100,000<200NEG<0.2
O-desmethyl tramadol100,000<200NEG<0.2

Table 8. Cross-Reactivity - Qualitative

{10}------------------------------------------------

CompoundCompound Conc.(ng/mL)TapentadolEquivalentConc.(ng/mL)ResultCross-Reactivity(%)
O-desmethyl venlafaxine100,000<200NEG<0.2
N-desmethyl tapentadol1,275200POS15.7
N-desmethyl tramadol100,000<200NEG<0.2
N-desmethyl venlafaxine100,000<200NEG<0.2
Tapentadol glucuronide43,000200POS0.5
Tramadol100,000<200NEG<0.2
Trimipramine100,000<200NEG<0.2
Venlafaxine100,000<200NEG<0.2

Table 9. Cross-Reactivity - Semi-Quantitative

CompoundCompoundConc.(ng/mL)TapentadolEquivalentConc.(ng/mL)MeanValue(ng/mL)ResultCross-Reactivity (%)
Chlorpromazine100,000< 20047.3NEG<0.2
Clomipramine100,000<20047.6NEG<0.2
Cyclobenzaprine100,000<2001.0NEG<0.2
Doxepin100,000<2000.9NEG<0.2
Imipramine100,000<20069.7NEG<0.2
O-desmethyl tramadol100,000<200101.0NEG<0.2
O-desmethyl venlafaxine100,000<2001.8NEG<0.2
N-desmethyl tapentadol1,275200214.3POS15.7
N-desmethyl tramadol100,000<2006.0NEG<0.2
N-desmethyl venlafaxine100,000<2001.9NEG<0.2
Tapentadol glucuronide43,000200211.7POS0.5
Tramadol100,000<2002.5NEG<0.2
Trimipramine100,000<20010.8NEG<0.2
Venlafaxine100,000<200-1.6NEG<0.2

3. Interference – Structurally Unrelated Compounds

Structurally unrelated compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free urine containing tapentadol at

{11}------------------------------------------------

±25% of the cutoff. The levels of structurally unrelated compounds that did not interfere in the assay are presented in Table 10.

CompoundConc. Tested (ng/mL)
4-Bromo-2,5,Dimethoxyphenethylamine100,000
6-Acetylcodeine100,000
6-Acetylmorphine100,000
Alprazolam100,000
7-Aminoclonazepam100,000
7-Aminoflunitrazepam100,000
7-Aminonitrazepam100,000
Amitriptyline100,000
d-Amphetamine100,000
Amobarbital100,000
Atomoxetine100,000
Benzoylecgonine100,000
Benzylpiperazine100,000
Bromazepam100,000
Brompheniramine100,000
Buprenorphine100,000
Bupropion100,000
Butabarbital100,000
Butalbital100,000
Cannabidiol100,000
Cannabinol100,000
Carbamazepine100,000
Cetirizine100,000
Chlordiazepoxide100,000
1-(3-Chlorophenylpiperazine) mCPP100,000
Chlorpheniramine100,000
Cimetidine100,000
Citalopram100,000
Clobazam100,000
Clonazepam100,000
Clozapine100,000
Cocaine100,000
Codeine100,000
Cotinine100,000
CompoundConc. Tested (ng/mL)
Dehydronorketamine50,000
Demoxepam100,000
Desakylflurazepam100,000
Desipramine100,000
Dextromethorphan100,000
Dihydrohydroxcarbamazepine100,000
Diazepam100,000
Digoxin100,000
Dihydrocodeine100,000
Doxylamine100,000
Duloxetine100,000
Ecgonine100,000
Ecgonine Methyl Ester100,000
EDDP100,000
EMDP100,000
1S,2R (+)-Ephedrine100,000
Ethylmorphine100,000
Ethyl-β-D-Glucuronide100,000
Fentanyl100,000
Fenfluramine100,000
Flunitrazepam100,000
Fluoxetine100,000
Flurazepam100,000
Haloperidol100,000
Heroin100,000
Hexobarbital100,000
Hydrocodone100,000
Hydromorphone100,000
Ketamine100,000
Lamotrigine100,000
Levorphanol100,000
Lidocaine100,000
Lorazepam100,000
Lorazepam Glucuronide50,000
Lormetazepam100,000
LSD100,000
Maprotiline100,000
MDA100,000
CompoundConc. Tested (ng/mL)
MDEA100,000
MDMA100,000
Meperidine100,000
Meprobamate100,000
d-Methamphetamine100,000
Methaquolone100,000
Methoxetamine100,000
Methylone100,000
Methylphenidate100,000
Midazolam100,000
Morphine100,000
Morphine-3-Glucuronide100,000
Morphine-6-Glucuronide100,000
Nalorphine100,000
Naloxone100,000
Naltrexone100,000
Naproxen100,000
Nitrazepam100,000
Norbuprenorphine100,000
Norcodeine100,000
Nordiazepam100,000
Norketamine100,000
Normorphine100,000
Noroxycodone100,000
Norpropoxyphene100,000
Norpseudoephedrine100,000
Nortriptyline100,000
Olanzapine100,000
Oxazepam100,000
Oxazepam glucuronide50,000
Oxycodone100,000
Oxymorphone100,000
Delta-9-THC100,000
11-hydroxy-delta-9-THC100,000
11-nor-9 carboxy THC100,000
PCP100,000
Pentazocine100,000
Pentobarbital100,000
CompoundConc. Tested (ng/mL)
Phenobarbital100,000
Phentermine100,000
R(-)-Phenylephrine100,000
Phenylpropanolamine (PPA)100,000
Phenytoin100,000
PMA100,000
PMMA100,000
Prazepam100,000
Propoxyphene100,000
Propranolol100,000
Protriptyline100,000
R,R (-)-Pseudoephedrine100,000
S,S (+)-Pseudoephedrine100,000
Risperidone100,000
Ritalinic Acid100,000
Salicylic Acid100,000
Secobarbital100,000
Sertraline100,000
Sufentanil50,000
Temazepam100,000
Theophylline100,000
Thioridazine100,000
Trazadone100,000
Triazolam100,000
3-Trifluoromethylphenyl-piperazine100,000
Tyramine100,000
Verapamil100,000
Zolpidem100,000
Carisoprodol100,000
1R,2S (-)-Ephedrine100,000
Acetaminophen500,000
Acetylsalicyclic Acid500,000
α-hydroxyalprazolam100,000
Barbital100,000
Caffeine500,000
Cyclopentobarbital100,000
Diphenhydramine300,000
Ibuprofen500,000
CompoundConc. Tested (ng/mL)
LAAM100,000
Labetalol100,000
Loratadine100,000
Mephenytoin100,000
Methadone500,000
Methylphenylsuccinimide (mCPP)100,000
Mirtazapine100,000
n-desmethylcitalopram100,000
Nor-LAAM100,000
Noroxymorphone100,000
Normesuximide100,000
PEMA100,000
Phenazepam100,000
Procaine100,000

Table 10. Non-Interfering Structurally Unrelated Compounds

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

4. Interference – Endogenous Compounds and Urine Preservatives

Endogenous compounds and urine preservatives were evaluated in qualitative and semiquantitative modes by spiking the potential interferent into drug free urine containing tapentadol at ±25% of the cutoff. Due to the interference of boric acid observed at ±25% of the cutoff, potential interference was also evaluated at ±50% of the cutoff. Endogenous compounds and urine preservative tested that did not interfere in the assay are presented in Tables 11 and 12. Boric acid interference test results at ±50% of the cutoff in qualitative and semi-quantitative modes are presented in Table 13.

{16}------------------------------------------------

CompoundConcentration Tested
Acetone1.0 g/dL
Ascorbic Acid1.5 g/dL
Bilirubin0.002 g/dL
Creatinine0.5 g/dL
Ethanol1.0 g/dL
Galactose0.01 g/dL
y-Globulin0.5 g/dL
Glucose2.0 g/dL
Hemoglobin0.3 g/dL
Human Serum Albumin0.5 g/dL
Oxalic Acid0.1 g/dL
Riboflavin0.0075 g/dL
Sodium Chloride6.0 g/dL
Urea6.0 g/dL

Table 11. Non-interfering Endogenous Compounds

Table 12. Non-interfering Urine Preservative

CompoundConcentration Tested
Sodium Azide1% w/v
Sodium Fluoride1% w/v

Table 13. Interference at ±50% of the Cutoff

CompoundConcentrationTested-50% Cutoff (50 ng/mL)+50% Cutoff (150 ng/mL)
QualitativeResultSemi-QuantitativeResultQualitativeResultSemi-QuantitativeResult
Boric Acid1% w/vNegativeNegativeNegativeNegative

5. Interference – pH

To evaluate potential interference from the effect of urine pH on the assay's ability to detect tapentadol, device performance in the qualitative and semi-quantitative modes was tested using a range of urine pH values (3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0, 10.0 and 11.0). All test samples were prepared in drug free urine containing tapentadol at ±25% of the cutoff. No positive or negative interference was observed at urine pH values ranging from 3.0 to 11.0 for each test mode.

{17}------------------------------------------------

6. Interference - Specific Gravity

To evaluate potential interference from the specific gravity of urine on the assay's ability to detect tapentadol, device performance in the qualitative and semi-quantitative modes was tested using a range of physiologically relevant urine specific gravity values (1.000, 1.002, 1.005, 1.010, 1.015, 1.020, 1.025 and 1.030). All test samples were prepared in drug free urine containing tapentadol at ±25% of the cutoff. No positive or negative interference was observed at urine specific gravity values ranging from 1.000 to 1.030 for each test mode.

7. Linearity/Recovery

A linearity study in the semi-quantitative mode was conducted by spiking a drug free urine pool with a high concentration of tapentadol above the highest calibrator. Additional pools were made by serially diluting the high concentration specimen with drug free urine to achieve concentrations ranging from 1100 to 100 ng/mL. The 0 ng/mL specimen was made from drug free urine. Each pool was tested in triplicate to calculate the mean concentration values that were used to calculate drug recovery of tapentadol. The study confirmed the linear range to be 100-1100 ng/mL. The assay drug recovery percentage ranged from 95.8 to 110.4 %. Linearity test results in semi-quantitative mode are presented in Tables 14.

Expected ConcentrationMean ConcentrationRecovery (%)
(ng/mL)(ng/mL)
03.0N/A
100110.4110.4
200211.6105.8
300298.299.4
400417.3104.3
500531.2106.2
600614.1102.4
700721.7103.1
800827.3103.4
900907.8100.9
1000980.198.0
11001054.395.8

Table 14. Linearity/Recovery - Tapentadol

{18}------------------------------------------------

8. Calibration Duration

Drug free negative urine spiked with tapentadol at ±25% of the cutoff were tested in qualitative and semi-quantitative mode at time points up to 14 days. At the initial time point, a two-point calibration curve was established in qualitative mode and a multi-point calibration curve was established in semi-quantitative mode. This calibration was used through the duration of this study. The test results met acceptance criteria at each time point. The recommended frequency of calibration is 14 days.

9. Tapentadol Stability in Urine

Urine samples were collected from 4 participants who reported taking tapentadol in the last 24 hours. These samples were tested within 4 hours of sample collection. Test results indicated that urine samples containing tapentadol are stable for up to 14 days stored at 2°C - 8°C.

10. Method Comparison

A method comparison study was performed using 160 deidentified remnant unaltered clinical urine samples obtained from clinical testing laboratories. The urine samples were analyzed for tapentadol using the Immunalysis tapentadol Urine HEIA™ in both qualitative and semi-quantitative modes and Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS). The instruments used were the Beckman Coulter AU480 chemistry analyzer and an Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry. The method comparison study demonstrates that the Immunalysis Tapentadol Urine HEIA™ is an accurate test method in comparison to LC-MS/MS with a positive percent agreement (PPA) and negative percent agreement (NPA) of 100% and 100%, respectively. The results are presented from Tables 15 and 16.

{19}------------------------------------------------

LC-MS/MS Tapentadol Concentration
ImmunalysisTapentadol UrineHEIA™ Result< 100ng/mL(less than-50%cutoff)100 - 199ng/mL(between -50% cutoffand cutoff)200 - 300ng/mL(betweencutoff and+50% cutoff)> 300 ng/mL(greaterthan +50%cutoff)Agreement (%)
Qual.Positive001481100% (95/95)
Qual.Negative461900100% (65/65)
Semi-Quant.Positive001481100% (95/95)
Semi-Quant.Negative461900100% (65/65)

Table 15. Method Comparison Results by Concentration Range

Table 16. Method Comparison Results Compared to LC-MS/MS

Immunalysis Tapentadol Urine HEIA™LC-MS/MS
(+)(-)
Qualitative(+)950
(-)065
Semi-Quantitative(+)950
(-)065

H. CONCLUSION

The information provided in this pre-market notification demonstrates that the Immunalysis Tapentadol Urine HEIA™ is substantially equivalent to the legally marketed predicate device.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).