K Number
K203503
Date Cleared
2022-09-02

(641 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Camber Sacroiliac (SI) Fixation System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis. When the Camber SI Fixation System is implanted, it must be used with a SICONUS SI Joint Fixation System screw implanted across the same sacroiliac joint.

Device Description

The Camber SI Fixation System is a fusion device consisting primarily of an open architecture 3D generated titanium body to permit bone growth (fusion) throughout the implant. All internal surfaces have a roughened texture. The upper and lower faces have specifically designed surface approximately 0.5 mm thick to provide a trabecular support structure. In addition, the device utilizes a set of two sharpened anchor plates that translate from within the device in slightly angular opposing lateral directions which provide an anchoring system to fixate the implant between ilium and sacrum. The Camber SI Fixation System device has one footprint: 23x26 mm. Implant heights range from 9 to 13mm in 2 mm increments with 8° angulation.

AI/ML Overview

This document is a 510(k) summary for a medical device (Camber Sacroiliac (SI) Fixation System), not a study analyzing AI performance. Therefore, most of the requested information about acceptance criteria, study design, and AI performance metrics is not applicable.

However, I can extract the relevant information regarding the performance testing of the device itself and how it meets acceptance criteria based on predicate devices.

1. A table of acceptance criteria and the reported device performance

Test PerformedAcceptance Criteria (Defined by Predicate Device Performance)Reported Device Performance (Camber SI Fixation System)
Static Vertical ShearMet performance of predicate device(s)Met
Static Vertical Shear StiffnessMet performance of predicate device(s)Met
Dynamic Vertical Shear EnduranceMet performance of predicate device(s)Met
Implant DislodgementMet performance of predicate device(s)Met
Anchor Collapse ForceMet performance of predicate device(s)Met

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the given document. The document states "Testing performed indicate that the Camber SI Fixation System is as mechanically sound as the cleared devices," but does not detail the specific sample sizes for mechanical testing or cadaveric/usability testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as this document describes mechanical and cadaveric testing for a physical implant, not an AI or diagnostic device that requires expert-established ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable for mechanical or cadaveric testing of a physical implant.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable as this document describes a physical medical device (sacroiliac joint fixation system), not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable as this document describes a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mechanical testing, the "ground truth" or benchmark was established by the performance of the predicate devices. For cadaveric and usability testing, the "ground truth" would be related to anatomical feasibility, surgical ease of use, and stabilization, which is typically assessed by surgeons or technical experts during the testing process. The document does not specify the exact methods or criteria for "ground truth" in these non-mechanical tests beyond indicating they were performed to establish substantial equivalence.

8. The sample size for the training set

This information is not applicable as this document describes a physical medical device and its mechanical testing, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable as this document describes a physical medical device and its mechanical testing, not a machine learning model.

{0}------------------------------------------------

September 2, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Camber Spine Technologies, LLC Christine Scifert Official Correspondent 501 Allendale Road King of Prussia, Pennsylvania 19604

Re: K203503

Trade/Device Name: Camber Sacroiliac (SI) Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: August 24, 2022 Received: August 26, 2022

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anne D. Talley -S

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

for

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203503

Device Name Camber Sacroiliac (SI) Fixation System

Indications for Use (Describe)

The Camber Sacroiliac (SI) Fixation System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroilitis. When the Camber SI Fixation System is implanted, it must be used with a SICONUS SI Joint Fixation System screw implanted across the same sacroiliac joint.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)X
Over The Counter Use (21 CFR 801 Subpart C)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Device Trade Name:Camber Sacroiliac (SI) Fixation System
Manufacturer:Camber Spine Technologies501 Allendale RoadKing of Prussia, PA 19406
Company Contact:Noel HetrickNhetrick@cambermedtech.com
Official Correspondent:Christine ScifertPartner, MRC GlobalChristine.scifert@askmrcglobal.com901-831-8053
Date Prepared:September 2, 2022
Classifications:21 CFR §888.3040, Smooth or Threaded Metallic Bone FixationFastener
Class:II
Product Code:OUR
Primary Predicate:Tenon Medical Catamaran Sacroiliac Joint Fixation System(K180818)
Reference Predicate(s):Camber Spine ENZA-A Titanium ALIF device (K173432)

Indications For Use:

The Camber Sacroiliac (SI) Fixation System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis. When the Camber SI Fixation System is implanted, it must be used with a SICONUS SI Joint Fixation System screw implanted across the same sacroiliac joint.

Device Description:

The Camber SI Fixation System is a fusion device consisting primarily of an open architecture 3D generated titanium body to permit bone growth (fusion) throughout the implant. All internal surfaces have a roughened texture. The upper and lower faces have specifically designed surface approximately 0.5 mm thick to provide a trabecular support structure. In addition, the device utilizes a set of two sharpened anchor plates that translate from within the device in slightly angular opposing lateral directions which provide an anchoring system to fixate the implant between ilium and sacrum. The Camber SI Fixation System device has one footprint: 23x26 mm. Implant heights

{4}------------------------------------------------

range from 9 to 13mm in 2 mm increments with 8° angulation.

Predicate Device:

The subject device, the Camber Sacroiliac (SI) Fixation System, is substantially equivalent to the Tenon Catamaran SI Joint Fixation System (K180818) in terms of Indications for use. The reference predicate, ENZA-A Titanium ALIF device (K173432) has an identical anchoring system as the Camber SI Fixation System.

Performance Testing Summary:

Testing performed indicate that the Camber SI Fixation System is as mechanically sound as the cleared devices. Testing included static vertical shear, static vertical shear stiffness, dynamic vertical shear endurance, implant dislodgement and anchor collapse force. The results demonstrate that the acceptance criteria defined by predicate device performance were met. Cadaveric and usability testing was done to establish substantial equivalence.

Substantial Equivalence:

The subject Camber Sacroiliac (SI) Fixation System was demonstrated to be substantially equivalent to primary predicate, Tenon Catamaran Sacroiliac Joint Fixation System (K180818) and the reference predicate, the ENZA-A Titanium ALIF device (K173234) with regards to indications for use, design inputs, materials, performance and/or manufacturing.

Conclusion:

Camber Spine Technologies has provided sufficient information to demonstrate the Camber Sacroiliac (SI) Fixation System is substantially equivalent to primary predicate, Tenon Catamaran Sacroiliac Joint Fixation System (K180818) and the reference predicate, the ENZA-A Titanium ALIF device (K173234) with regards to indications for use, design inputs, materials, performance and/or manufacturing. .

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.