K Number
K203426
Date Cleared
2021-02-18

(90 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Face Mask (non-sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

Device Description

The proposed device(s) are blue color, and flat pleated type mask, utilizing ear loops way for wearing, and they all has Nose clips design for fitting the face mask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of nylon and spandex. The nose piece contained in the proposed device(s) is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of polypropylene. The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Surgical Face Mask (non-sterile)" as presented in the FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance:

ItemAcceptance Criteria (Level 2)Reported Device Performance (Proposed Device)Result
Fluid Resistance (ASTM F1862)29 out of 32 pass at 120 mmHg32 out of 32 pass at 120 mmHgPASS
Particulate Filtration Efficiency (ASTM F2299)≥ 98%99.16%PASS
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%99.74%PASS
Differential Pressure (Delta P) (EN 14683 Annex C)< 6.0 mmH₂O/cm²3.7 mmH₂O/cm²PASS
Flammability (16 CFR 1610)Class 1Class 1PASS
Cytotoxicity (ISO 10993-5)Non-CytotoxicUnder the conditions of the study, the device is non-cytotoxic.PASS
Irritation (ISO 10993-10)Non-IrritatingUnder the conditions of the study, the device is non-irritating.PASS
Sensitization (ISO 10993-10)Non-SensitizingUnder the conditions of the study, the device is non-sensitizing.PASS

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes for each specific test (e.g., number of masks tested for BFE). However, for Fluid Resistance (ASTM F1862), the criteria and result are given as "32 out of 32 passes," suggesting a sample size of 32 tests for this particular performance characteristic.

  • Data Provenance: The tests appear to be conducted by or on behalf of Nantong Taiweishi Medical Technology Co., Ltd. in China, as indicated by the company's address on the document. The document presents these tests as non-clinical laboratory evaluations. There is no information provided about whether these studies were retrospective or prospective, but given they are laboratory performance tests for device clearance, they would typically be considered prospective studies on manufactured samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This type of information is generally not applicable to the performance testing of a physical product like a surgical face mask. The "ground truth" here is established by the standardized test methods themselves (e.g., ASTM F2100, F1862, F2299, F2101, EN 14683, ISO 10993, 16 CFR 1610). The results are quantitative measurements against predefined criteria, not subjective expert interpretations. Therefore, no experts were used to establish ground truth in the traditional sense of clinical or image-based studies.

4. Adjudication Method for the Test Set:

Not applicable. As noted above, these are objective physical and biological performance tests against established standards, not studies requiring expert adjudication of subjective findings.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human reader performance (with or without AI assistance) is being evaluated against some ground truth. This document pertains to the performance characteristics of a physical medical device (surgical face mask).

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

No, this question is not applicable. This device is a physical product, not an algorithm or AI system.

7. The Type of Ground Truth Used:

The ground truth is established by standardized test methods and predefined acceptance criteria from recognized international and national standards such as ASTM, ISO, EN, and CFR. These standards define the methodology for measuring specific performance characteristics and the thresholds for acceptable performance (e.g., ≥ 98% BFE, Class 1 Flammability).

8. The Sample Size for the Training Set:

Not applicable. This device is a physical product, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As above, this information is not relevant for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

February 18, 2021

Nantong Taiweishi Medical Technology Co., Ltd. % Mandy Wu Consultant Shanghai Sungo Management Consulting Company Limited 13th Floor, 1500# Central Avenue Shangghai, Shanghai 200122 China

Re: K203426

Trade/Device Name: Surgical Face Mask (non-sterile) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 20, 2020 Received: November 20, 2020

Dear Mandy Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203426

Device Name Surgical Face Mask (non-sterile)

Indications for Use (Describe)

The Surgical Face Mask (non-sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

_ | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary
for K203426

{4}------------------------------------------------

Nantong Taiweishi Medical Technology Co., Ltd. Taihangshan Road, New District, Economic Development Zone, Rudong County, Nantong City, Jiangsu Province, China ,226400

510(K) Summary

K203426

Summary prepared date: 2021-02-18

A. Applicant:

Name: Nantong Taiweishi Medical Technology Co., Ltd. Address: Taihangshan Road, New District, Economic Development Zone, Rudong County, Nantong City, Jiangsu Province, China ,226400 Contact Person: ZhangXuewen

Tel: +8618761789767

Submission Correspondent: Primary contact: Ms. Mandy Wu Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: min.wu@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device:

Trade Name: Surgical Face Mask (non-sterile ) Common Name: Surgical Face Mask Model(s): NA

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

C. Predicate device:

K182514 Surgical Face Mask

D. Indications use of the device:

The Surgical Face Mask (non-sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids.

{5}------------------------------------------------

Nantong Taiweishi Medical Technology Co., Ltd. Taihangshan Road, New District, Economic Development Zone, Rudong County, Nantong City, Jiangsu Province, China ,226400

This a single use, disposable device(s), provided non-sterile.

E. Device Description:

The proposed device(s) are blue color, and flat pleated type mask, utilizing ear loops way for wearing, and they all has Nose clips design for fitting the face mask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of nylon and spandex. The nose piece contained in the proposed device(s) is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of polypropylene. The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

F. Comparison with predicate device

DeviceProposed DevicePredicate DeviceComparison
ManufacturerNantong Taiweishi MedicalTechnology Co., Ltd.Xiantao Zhibo Non-wovenProducts Co., Ltd-
510K numberK203426K182514-
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21CFR878.4040)Same
Indications for useThe Surgical Face Mask(non-sterile) is intended to beworn to protect both the patient andhealthcare personnel from transferof microorganisms, body fluids andparticulate material. These facemasks are intended for use ininfection control practices to reducethe potential exposure to blood andbody fluids. This a single use,disposable device(s), providednon-sterile.The Disposable Surgical FaceMasks are intended to be worn toprotect both the patient andhealthcare personnel from transferof microorganisms, body fluids andparticulate material. These facemasks are intended for use ininfection control practices to reducethe potential exposure to blood andbody fluids. This is a single use,disposable device(s), providednon-sterile.Same
MaterialOuter layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Inner layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Nose clipPolyethyleneMalleable aluminum wireDifferent
Ear loopsNylon and SpandexPolyesterDifferent
ColorBlueBlueSame
Dimension (Length)$175mm+/-5%$$17.5cm+/-1cm$Similar
Dimension (Width)$95mm+/-5%$$9.5cm+/-1cm$Similar

Table 1 General Comparison

{6}------------------------------------------------

Nantong Taiweishi Medical Technology Co., Ltd.

Taihangshan Road, New District, Economic Development Zone, Rudong County, Nantong City, Jiangsu Province, China 226400

China ,226400
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100 levelLevel 2Level 2Same
BiocompatibilityISO10993ISO10993Same
FluidResistancePerformance ASTMF186232 out of 32 passes at 120 mmHg32 out of 32 passes at 120 mmHgSimilar
ParticulateFiltration EfficiencyASTM F229999.16%99.88%Similar
Bacterial FiltrationEfficiencyASTMF210199.74%99.6%Similar
DifferentialPressure (Delta P)EN 14683 Annex C3.7mm H2O/cm²3.0mm H2O/cm²Similar
Flammability 16CFR 1610Class 1Class 1Similar

From the comparison we found the material of the current nose clip and the ear loop were different from the predicate device. The biocompatibility tests were conducted to both components to ensure their compliance to the ISO10993-5 and ISO10993-10. There is no new risk generated from the difference of the material.

For the performance testing, the test results are not identical to each other, but they are similar and they both meet the requirement of Level 2 medical mask according to the ASTM F 2100.

G. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

  • A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests for in Vitro cytotoxicity
  • A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization
  • A ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks
  • A ASTM F1862, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity);
  • EN 14683. Medical Face Masks-Requirements and Test Methods;

{7}------------------------------------------------

Nantong Taiweishi Medical Technology Co., Ltd.

Taihangshan Road, New District, Economic Development Zone, Rudong County, Nantong City, Jiangsu Province,

China ,226400

  • A ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus Aureus;
  • ASTM F2299, Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres;

  • A 16 CFR 1610, Standard for the Flammability of Clothing Textiles;
ItemProposed devicePredicate DeviceAcceptanceCriteria (level 2)Result
Fluid ResistancePerformance ASTMF186232 out of 32 pass at120 mmHg32 out of 32 pass at120 mmHg29 out of 32 passat 120 mmHgPASS
Particulate FiltrationEfficiency ASTMF229999.16%99.88%≥ 98%PASS
Bacterial FiltrationEfficiency ASTMF210199.74%99.6%≥ 98%PASS
Differential Pressure(Delta P) EN 14683Annex C3.7 mmH2O/cm²3.0 mmH2O/cm²< 6.0mmH2O/cm²PASS
Flammability 16 CFR1610ClassClass 1Class 1PASS

Table 2 - Performance Testing

Table 3 Biocompatibility Comparison

ItemProposed deviceAcceptance CriteriaResult
CytotoxicityUnder the conditions of the study, thedevice is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder the conditions of the study, thedevice is non-irritating.Non-IrritatingPASS
SensitizationUnder the conditions of the study, thedevice is non-sensitizingNon-SensitizingPASS

H. Clinical Test Conclusion

No clinical study is included in this submission.

I. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K182514 Xiantao Zhibo Non-woven Products Co., Ltd. Surgical Face Mask.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.