(127 days)
Not Found
No
The device description and performance studies focus on mechanical and material properties, with no mention of AI or ML.
Yes
The device is used to provide intraosseous access for resuscitation and "lifesaving fluid delivery," indicating its role in treating patients.
No
Explanation: The device is described as providing "intraosseous access" for "resuscitation and lifesaving fluid delivery," indicating its purpose is for treatment and delivery of substances, not for diagnosing a condition or disease.
No
The device description clearly outlines multiple hardware components including needles, drivers (powered and manual), extension sets, and securement dressings. There is no mention of any software component.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use describes a system for providing intraosseous access for fluid delivery. This is a procedure performed on the patient's body to deliver substances, not a test performed on a sample taken from the body to diagnose or monitor a condition.
- Device Description: The device components (needles, drivers, extension sets, dressings) are all designed for accessing the intraosseous space and delivering fluids. There is no mention of reagents, test strips, or any components used for analyzing biological samples.
- Lack of IVD Characteristics: The description does not mention any analysis of biological samples, diagnostic purposes, or measurement of analytes.
Therefore, the BD Intraosseous Infusion System is a medical device used for a therapeutic procedure (fluid delivery via intraosseous access), not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The BD Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Product codes
MHC
Device Description
The BD Intraosseous Infusion System provides clinicians and emergency personnel with access to the intraosseous space for resuscitation and lifesaving fluid delivery for up to 24 hours. The BD Intraosseous Infusion System consists of the following: a single use hypodermic needle (with needle safety cap), a powered or manual driver to assist with needle insertion, an extension set, and; an adhesive-backed securement dressing. For insertions using the powered driver, the hypodermic needle includes a needle hub that mates with a stylet connected to a drive adapter hub. The drive adapter hub includes a magnetic insert that attaches to the powered driver prior to needle insertion. The BD Intraosseous Infusion System is an easy-grip, hand-held, battery-powered device with a rechargeable lithium battery used to assist in the insertion of the subject device needle through the bone cortex. The assembly of the hypodermic needle and stylet with connected drive adapter hub is referred to as the needle set. For insertions using the manual driver, the needle and the needle hub mate with a stylet in the same way as the needle set that is used with the powered driver, except the stylet is integrated into the handle of the manual driver instead of a drive adaptor hub (i.e. the manual driver needle assembly does not include a drive adapter hub). The stylet was designed to include a passive safety feature to protect the placer from sharps injury. After the needle is inserted, the stylet is separated from the needle and needle hub. Upon separation of the stylet from the needle hub, the passive safety feature is released onto the stylet tip and can be safely discarded into a sharps container. Following needle insertion, the securement dressing can be applied to secure the needle hub to the skin. An extension set is available for access to the needle hub to support fluid exchange. The subject device BD Intraosseous Infusion System will be offered in needle set (for use with the powered driver) and manual driver needle kit configurations. Each kit configuration will include a securement dressing and an extension set.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients
Indicated Patient Age Range
Adult and pediatric patients
Intended User / Care Setting
Clinicians and emergency personnel; emergent, urgent, or medically necessary cases
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance tests were conducted to determine the substantial equivalence of the BD Intraosseous Infusion System to the predicate Piper GO-IO® Intraosseous Infusion System. These included: Needle Outer Diameter (OD), Needle Length, Needle Lubricity, Needle Quality, Surface Finish, and Cleanliness, Needle to Hub Assembly Tensile, Stylet to Drive Adapter Hub Tensile, Needle and Stylet Disassembly Force, Safety Activation, Stylet Safety Override (force to failure), Manual Driver Hub to Stylet Tensile, Needle Resistance to Corrosion, Needle Hub Luer, Needle Hub Cleanliness, Needle Point, Needle Resistance to Breakage, Needle Stiffness, Gravity Flow Rate, Liquid Leak Needle Hub, Limits for Acidity or Alkalinity (Needle), Limits for Extractable Metals (Needle), Depth Markings, Insertion Force, Needle Bone Retention - Needle Point OD, Packaging Integrity and Seal Strength, Device Usability/Simulated Use, Sharps Injury Prevention Feature (Simulated Clinical Use).
Sterilization, Packaging, and Shelf-Life studies were conducted including: Sterilization Validation/Adoption, Packaging/Shelf-Life Validations, Sterilant Residuals, Bacterial Endotoxin.
A biocompatibility evaluation was conducted per ISO 10993-1:2009. The biological tests included: Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemocompatibility.
The results of the user validation, performance (verification and validation testing) and biological tests conducted on the BD Intraosseous Infusion System met all predetermined acceptance criteria and demonstrated that the different technological characteristics of the subject device do not raise different questions of safety and effectiveness.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.5570 Hypodermic single lumen needle.
(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 4, 2021
Bard Access Systems, Inc. (BAS) [Wholly-owned subsidiary of BD] Connor Dahl Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, Utah 84116
Re: K203193
Trade/Device Name: BD Intraosseous Infusion System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MHC Dated: January 29, 2021 Received: February 1, 2021
Dear Connor Dahl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Rumi Young Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K203193
Device Name BD Intraosseous Infusion System
The BD Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
---|---|
--------------------------------------------------------------------------------------- | -------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.## DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or soonsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary for BD Intraosseous Infusion System
21 CFR 807.92(a)
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part(l)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is presented in the following table:
General Provisions | Submitter Name: | Bard Access Systems, Inc. (BAS) [Wholly-owned subsidiary of BD] |
---|---|---|
Submitter Address: | 605 North 5600 West | |
Salt Lake City, UT 84116 | ||
Contact Person: | Connor Dahl | |
Regulatory Affairs Specialist | ||
Telephone Number: | 801.522.5834 | |
Fax Number: | 801.522.5425 | |
Date of Preparation: | 1/26/2021 | |
Subject Device | Trade Name(s): | BD Intraosseous Infusion System |
Common Name: | Interosseous Infusion System | |
Classification Name: | Hypodermic single lumen needle | |
Class: | 2 | |
Regulation Number: | 21 CFR 880.5570 | |
Product Code: | MHC | |
Classification Panel | General Hospital | |
Predicate Device | Predicate Trade Name: | Piper GO-IO® Intraosseous Infusion System |
Classification Name: | Intraosseous Infusion System | |
Class: | 2 | |
Product Code: | FMI | |
Regulation Number: | 21 CFR 880.5570 | |
Premarket Notification #: | K191976 | |
Manufacturer: | Piper Access, LLC | |
Classification Panel: | General Hospital |
4
| Device Description | The BD Intraosseous Infusion System provides clinicians and emergency personnel with access to the intraosseous
space for resuscitation and lifesaving fluid delivery for up to 24 hours. The BD Intraosseous Infusion System consists
of the following:
a single use hypodermic needle (with needle safety cap), a powered or manual driver to assist with needle insertion, an extension set, and; an adhesive-backed securement dressing. For insertions using the powered driver, the hypodermic needle includes a needle hub that mates with a stylet connected to a drive adapter hub. The drive adapter hub includes a magnetic insert that attaches to the powered driver prior to needle insertion. The BD Intraosseous Infusion System is an easy-grip, hand-held, battery-powered device with a rechargeable lithium battery used to assist in the insertion of the subject device needle through the bone cortex. The assembly of the hypodermic needle and stylet with connected drive adapter hub is referred to as the needle set.
For insertions using the manual driver, the needle and the needle hub mate with a stylet in the same way as the needle set that is used with the powered driver, except the stylet is integrated into the handle of the manual driver instead of a drive adaptor hub (i.e. the manual driver needle assembly does not include a drive adapter hub).
The stylet was designed to include a passive safety feature to protect the placer from sharps injury. After the needle is inserted, the stylet is separated from the needle and needle hub. Upon separation of the stylet from the needle hub, the passive safety feature is released onto the stylet tip and can be safely discarded into a sharps container. Following needle insertion, the securement dressing can be applied to secure the needle hub to the skin. An extension set is available for access to the needle hub to support fluid exchange.
The subject device BD Intraosseous Infusion System will be offered in needle set (for use with the powered driver) and manual driver needle kit configurations. Each kit configuration will include a securement dressing and an extension set. |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The BD Intraosseous Infusion System is intended to provide clinicians and emergency personnel with access to the intraosseous space. |
| Indications for Use | The BD Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours. |
| Technological
Characteristics | The technological characteristics of the subject BD Intraosseous Infusion System are substantially equivalent with respect to the basic design and function as compared to the predicate Piper GO-IO® Intraosseous Infusion System. The technological characteristics between the subject and predicate devices are the same, with the exception of the addition of the 35 mm and 55 mm needle lengths under review in this submission. The technological differences listed below were evaluated using industry consensus standards, validation, and as defined in the risk assessment.
Therefore, the differences in technological characteristics between the subject and predicate devices do not raise new or different questions of safety or effectiveness. |
5
The following table provides a comparison between the subject and predicate devices. | ||
---|---|---|
Attribute | Subject Device - BD Intraosseous | |
Infusion System | Predicate Device - Piper GO-IO® | |
Intraosseous Infusion System | ||
Owner | Bard Access Systems, Inc. | Piper Access, LLC |
Classification | Same as predicate | FMI - 21 CFR 880.5570 |
510(k) Status | Subject of this Premarket Notification | K191976 - Concurrence date November |
13, 2019 | ||
Intended Use | Same as predicate | Intended to provide clinicians and |
emergency personnel with access to the | ||
intraosseous space. | ||
Indications for | ||
Use | The BD Intraosseous Infusion System | |
provides intraosseous access in the proximal | ||
tibia, distal tibia and humeral head (proximal | ||
humerus) of adult and pediatric patients, and | ||
the distal femur in pediatric patients when | ||
intravenous access is difficult or impossible | ||
to obtain in emergent, urgent, or medically | ||
necessary cases for up to 24 hours. | The Piper GO-IO® Intraosseous Infusion | |
System provides intraosseous access in the | ||
proximal tibia, distal tibia and humeral head | ||
(proximal humerus) of adult and pediatric | ||
patients, and the distal femur in pediatric | ||
patients when intravenous access is difficult | ||
or impossible to obtain in emergent, urgent, | ||
or medically necessary cases for up to 24 | ||
hours. | ||
Commercial | ||
Name | BD Intraosseous Infusion System | Piper GO-IO® Intraosseous Infusion |
System | ||
Target Patient | ||
Population | Same as predicate | Adults and Pediatrics |
Anatomical | ||
Insertion Site | Same as predicate | Adults: Proximal tibia, distal tibia, proximal |
humerus | ||
Pediatrics: Proximal tibia, distal tibia, | ||
proximal humerus, distal femur | ||
Primary IO | ||
System | ||
Components | Same as predicate | |
The previously cleared Securement | ||
Dressing and Power Driver are not | ||
under review in this submission | ● Hypodermic Needle w/Stylet | |
● Needle Safety Cap | ||
● Securement Dressing | ||
● Powered Driver |
6
Manual Driver | |||
---|---|---|---|
Needle: Dwell Time | Same as predicate | 24 hours or less | |
Needle: Use | Same as predicate | Single Use | |
Needle Lengths | Needle lengths with corresponding tissue depth recommendations: | ||
• 15mm (0-10 mm tissue depth) | |||
• 25mm (0-20 mm tissue depth) | |||
• 35mm (10-30 mm tissue depth) | |||
• 45mm (20-40 mm tissue depth) | |||
• 55mm (30-50 mm tissue depth) | Needle lengths with corresponding patient weight recommendations: | ||
• 15mm (3-39kg) | |||
• 25mm (>3kg) | |||
• 45mm (>40kg) | |||
Needle: Outer Diameter | Same as predicate | 15 gauge | |
Needle: Materials | Same as predicate | 304 Stainless Steel | |
Needle: Tip Design | Same as predicate | Faceted Tip | |
Needle: Depth Markers | Same as predicate | Depth markers every 1 cm | |
Needle: Hub Material | Same as predicate | Medical grade polycarbonate | |
Needle: Hub Connection | Same as predicate | Standard Luer Lock | |
Stylet: Materials | Same as predicate | Stainless Steel | |
Stylet: Sharps Injury Prevention Feature | Same as predicate | Includes a stylet tip safety feature | |
Drive Adapter Hub: Materials | Same as predicate | Polycarbonate and stainless steel |
7
| Safety &
Performance Tests | Inclusion of a
Needle Protective
Cover | Same as predicate | Yes, includes a needle cover made of
polypropylene |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Needle Set
Sterilization
Method & SAL | Same as predicate | Ethylene Oxide, 10-6 |
| | Manual Driver
Attachment | Same as predicate | Manual driver handle with integrated stylet
mates with internal lumen of needle and
needle hub attaches to manual driver |
| | Manual Driver
Component
Materials | Same as predicate | Handle: ABS
Stylet: Stainless Steel |
| | Manual Driver
Sterilization
Method and SAL | Same as predicate | Ethylene Oxide, 10-6 |
| | Means to Insert
Needle | Same as predicate | Manual or Powered Driver |
| | General Method
of Insertion | Same as predicate | Push needle through soft tissue until it
contacts bone. Confirm depth markings.
Insert needle set through bone until change
in pressure is felt or to desired depth.
Remove stylet. Connect IV extension set. |
| | The following performance tests were conducted in determining substantial equivalence of the BD Intraosseous
Infusion System to the predicate Piper GO-IO® Intraosseous Infusion System: | | |
| | Needle Set Kit and Manual Driver Kit
Performance Tests | | Standard Followed |
| | Needle Outer Diameter (OD) | | ISO 9626: 2016 and Internal Protocol/Standard |
| | Needle Length | | Internal Protocol/Standard |
| | Needle Lubricity | | ISO 7864: 2016 |
| | Needle Quality, Surface Finish, and Cleanliness | | ISO 9626: 2016 |
8
Needle to Hub Assembly Tensile | Internal Protocol/Standard |
---|---|
Stylet to Drive Adapter Hub Tensile | Internal Protocol/Standard |
Needle and Stylet Disassembly Force | Internal Protocol/Standard |
Safety Activation | FDA Guidance for Sharps Injury Prevention |
Features & ISO 23908: 2011 | |
Stylet Safety Override (force to failure) | ISO 23908: 2011 |
Manual Driver Hub to Stylet Tensile | Internal Protocol/Standard |
Needle Resistance to Corrosion | ISO 9626: 2016 |
Needle Hub Luer | ISO 594-1: 1986 and ISO 594-2: 1998 |
Needle Hub Cleanliness | ISO 7864: 2016 |
Needle Point | ISO 7864: 2016 |
Needle Resistance to Breakage | ISO 9626: 2016 |
Needle Stiffness | ISO 9626: 2016 and Internal Protocol/Standard |
Gravity Flow Rate | Internal Protocol/Standard |
Liquid Leak Needle Hub | Internal Protocol/Standard |
Limits for Acidity or Alkalinity (Needle) | ISO 9626: 2016 / ISO 7864: 2016 |
Limits for Extractable Metals (Needle) | ISO 7864: 2016 |
Depth Markings | Internal Protocol/Standard |
Insertion Force | Internal Protocol/Standard |
Needle Bone Retention - Needle Point OD | Internal Protocol/Standard |
Packaging Integrity and Seal Strength | ISO 11607-1:2006 |
ASTM F88/F88M: 2015 | |
ASTM F1886/F1886M: 2016 | |
ASTM F1929: 2015 | |
Device Usability/Simulated Use | Internal Protocol/Standard |
9
| Sharps Injury Prevention Feature (Simulated
Clinical Use) | FDA Guidance for Sharps Injury Prevention
Features & ISO 23908: 2011 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Sterilization, Packaging, and Shelf-Life | Standard Followed |
| Sterilization Validation/Adoption | ISO 11135:2014 |
| Packaging/Shelf-Life Validations | ISO 11607-1 AMD 1: 2014
ASTM F88/F88M: 2015
ASTM F1886/F1886M: 2016
ASTM F1929: 2015 |
| Sterilant Residuals | ISO 10993-7: 2008 |
| Bacterial Endotoxin | USP
USP |
| A biocompatibility evaluation was conducted on the subject device per ISO 10993-1:2009,
Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk
Management Process. According to the evaluation, the biological tests in the table below were
conducted. | |
| Biological Endpoint | Standard Followed |
| Cytotoxicity | ISO 10993-05: 2009 |
| Sensitization | ISO 10993-10: 2010 |
| Irritation/Intracutaneous Reactivity | |
| Acute Systemic Toxicity | ISO 10993-11: 2006 |
| Material Mediated Pyrogenicity | |
| Hemocompatibility | ISO 10993-4: 2017 |
10
| Technological
Comparison to
Predicate Device | The subject device, BD Intraosseous Infusion System, has the same intended use and the same fundamental scientific technology as the predicate device, Piper GO-IO® Intraosseous Infusion System. The main difference between the subject device is the addition of the 35 mm and 55 mm needle lengths as compared to the predicate device needle lengths. This technological difference was assessed by the performance of verification testing to applicable test standards and the performance of additional user validation to address the acceptability and risks associated with the new subject device needle lengths. |
---|---|
The results of the user validation, performance (verification and validation testing) and biological tests conducted on the BD Intraosseous Infusion System met all predetermined acceptance criteria and demonstrated that the different technological characteristics of the subject device do not raise different questions of safety and effectiveness. Based on the intended use, technological characteristics, performance and biological test results, the BD Intraosseous Infusion System can be considered substantially equivalent to the cited predicate device. | |
Summary of | |
Substantial | |
Equivalence | Based on the risk management activities and testing, the subject BD Intraosseous Infusion System has been demonstrated to be substantially equivalent to the cited predicate device. |