K Number
K203038

Validate with FDA (Live)

Date Cleared
2021-03-26

(171 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Toro-L Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or cortical cancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle screw/rod systems) cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Toro-L Interbody Fusion System is intended for use in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracy disc herniation (with myelopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Toro-L Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scollosis.

Device Description

Integrity Implants' Toro-L Interbody Fusion System incorporates a bi-directional expandable interbody fusion device and a monolithic interbody fusion device intended for use in the thoracolumbar spine.

Once implanted, the Toro-L interbody fusion device is designed to restore intervertebral disc height, provide anterior column support, and maintain structural stability of the motion segment to facilitate intervertebral body fusion. The Toro-L interbody fusion device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone, and with supplemental fixation instrumentation that has been cleared for use in the lumbar spine. The Toro-L Interbody Fusion System includes manual surgical instruments for delivery of the implant device and for disc preparation.

The Toro-L Interbody Fusion System implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136 and Ti-6Al-4V per ASTM F2924) and are offered in a range of sizes and lordotic options to accommodate variations in patient anatomy.

The Toro-L Interbody Fusion System expandable implant has integrated ramps and a threaded actuator that allow the implant to be inserted in a non-expanded form, and then subsequently expanded to full width and desired height through continuous expansion. When the desired height is achieved, the locking system is engaged to secure the construct in its final expanded configuration. The implant is then post-packed with bone graft.

The monolithic implant is a non-expanding conventional spacer with the interior space of the device divided into two compartments for bone graft containment.

Both the expandable and monolithic configurations have protrusions on the superior and inferior surfaces of the endplates of the implant to grip the adjacent vertebral endplates to resist expulsion.

The subject implant and instruments are offered non-sterile and are intended to be steam sterilized by the user prior to use.

AI/ML Overview

This is a medical device 510(k) summary for the Toro-L Interbody Fusion System, which is a spinal implant. The document does not contain information related to AI/ML device performance, acceptance criteria for such, or studies involving human readers or ground truth as it pertains to AI/ML.

Therefore, I cannot fulfill your request for the specific information regarding AI/ML device performance, acceptance criteria, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details.

The provided text focuses on the mechanical and biological testing of the physical implant to demonstrate substantial equivalence to previously cleared devices.

Here's what the document does provide:

  • Device Name: Toro-L Interbody Fusion System
  • Regulatory Class: Class II (Product Codes MAX, PHM)
  • Intended Use: Intervertebral body fusion of the spine in skeletally mature patients, with
    autogenous and/or allogeneic bone graft and supplemental internal spinal fixation systems. Intended for use in the thoracic spine (T1-T12) and thoracolumbar junction (T12-L1), and lumbar spine (L1-S1) for symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation. Can also be used as an adjunct to fusion in patients with multilevel degenerative scoliosis.
  • Predicate Devices:
    • Amplify Surgical's DualX LLIF, K181397
    • Nuvasive's AP Expandable XLIF System, K173025
    • Globus Medical's Rise-L Spacers, K113447
    • Medtronic's Capstone-L Spinal System, K123978
    • Integrity Implants' FlareHawk7 Interbody Fusion System, K183184
  • Non-Clinical Tests Conducted (Bench Performance Testing):
    • Static Axial Compression in accordance with ASTM F2077
    • Dynamic Axial Compression in accordance with ASTM F2077
    • Static Compression Shear in accordance with ASTM F2077
    • Dynamic Compression Shear in accordance with ASTM F2077
    • Wear Debris and Particle Characterization in accordance with ASTM F1877
    • Subsidence in accordance with ASTM F2267
  • Biocompatibility Assessment: In accordance with ISO 10993-1, including cytotoxicity testing in accordance with ISO 10993-5.

The conclusions state that based on these bench tests and biocompatibility assessment, the device demonstrates substantial equivalence and that any differences do not impact safety and effectiveness. This entire document describes a physical medical device, not an AI/ML-driven one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 26, 2021

Integrity Implants Inc Lauren Kamer Senior Director of Regulatory and Clinical Affairs 354 Hiatt Drive Palm Beach Gardens, Florida 33418

Re: K203038

Trade/Device Name: Toro-L Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, PHM Dated: February 19, 2021 Received: February 24, 2021

Dear Lauren Kamer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

K203038

Page 1 of 1

510(k) Number (if known)

K203038

Device Name Toro-L Interbody Fusion System

Indications for Use (Describe)

The Toro-L Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or cortical cancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle screw/rod systems) cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Toro-L Interbody Fusion System is intended for use in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracy disc herniation (with myelopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Toro-L Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scollosis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Toro-L Interbody Fusion System

September 15, 2020

I.Company:Integrity Im lants Inc.354 Hiatt DrivePalm Beach Gardens, FL 33418Telephone: 561-529-3861
II.Contact:merLauren KaSenior Director of Regulatory and Clinical Affairs
III.Proprietary Trade Name:Toro-L Interbody Fusion System
IV.Common Name:Intervertebral Body Fusion Devicewith Bone Graft, Lumbar;Intervertebral Body Fusion Devicewith Bone Graft, Thoracic
V.Classification Name:Class:Product Code:Intervertebral Body Fusion Device (21 CFR 888.3080)IIMAX, PHM

Product Description VI.

Integrity Implants' Toro-L Interbody Fusion System incorporates a bi-directional expandable interbody fusion device and a monolithic interbody fusion device intended for use in the thoracolumbar spine.

Once implanted, the Toro-L interbody fusion device is designed to restore intervertebral disc height, provide anterior column support, and maintain structural stability of the motion segment to facilitate intervertebral body fusion. The Toro-L interbody fusion device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone, and with supplemental fixation

{4}------------------------------------------------

instrumentation that has been cleared for use in the lumbar spine. The Toro-L Interbody Fusion System includes manual surgical instruments for delivery of the implant device and for disc preparation.

VII. Indications for Use

The Toro-L Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or cortical cancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems (e.g., pedicle screw/rod systems) cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Toro-L Interbody Fusion System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Toro-L Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis.

VIII. Summary of Technological Characteristics

The Toro-L Interbody Fusion System implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136 and Ti-6Al-4V per ASTM F2924) and are offered in a range of sizes and lordotic options to accommodate variations in patient anatomy.

The Toro-L Interbody Fusion System expandable implant has integrated ramps and a threaded actuator that allow the implant to be inserted in a non-expanded form, and then subsequently expanded to full width and desired height through continuous expansion. When the desired height is achieved, the locking system is engaged to secure the construct in its final expanded configuration. The implant is then post-packed with bone graft.

{5}------------------------------------------------

The monolithic implant is a non-expanding conventional spacer with the interior space of the device divided into two compartments for bone graft containment.

Both the expandable and monolithic configurations have protrusions on the superior and inferior surfaces of the endplates of the implant to grip the adjacent vertebral endplates to resist expulsion.

The subject implant and instruments are offered non-sterile and are intended to be steam sterilized by the user prior to use.

IX. Identification of Legally Marketed Predicate Devices Used to Claim Substantial Equivalence

To demonstrate the substantial equivalence of the subject Toro-L Interbody Fusion System to legally marketed predicate devices, Amplify Surgical's DualX LLIF, K181397 (SE 09/27/2018) is used as the primary predicate device. Additionally, the following devices are being used as predicate devices for this submission:

  • . Nuvasive's AP Expandable XLIF System, K173025 (SE 02/08/2018)
  • Globus Medical's Rise-L Spacers, K113447 (SE 01/26/2012) ●
  • . Medtronic's Capstone-L Spinal System, K123978 (SE 04/09/2013)
  • Integrity Implants' FlareHawk7 Interbody Fusion System, K183184 (SE 04/03/2019)

X. Brief Discussion of the Non-Clinical and Clinical Tests Submitted

Integrity Implants has conducted bench performance testing in support of this premarket notification submission as follows:

  • Static Axial Compression in accordance with ASTM F2077 ●
  • . Dynamic Axial Compression in accordance with ASTM F2077
  • . Static Compression Shear in accordance with ASTM F2077
  • . Dynamic Compression Shear in accordance with ASTM F2077
  • Wear Debris and Particle Characterization in accordance with ASTM F1877
  • . Subsidence in accordance with ASTM F2267

Integrity Implants has conducted a biocompatibility assessment in accordance with ISO 10993-1, including cytotoxicity testing in accordance with ISO 10993-5.

{6}------------------------------------------------

XI. Conclusions Drawn for the Non-Clinical and Clinical Tests

Based on the bench performance testing, biocompatibility assessment/testing, and other supporting documentation provided in this premarket notification, the subject Toro-L Interbody Fusion System demonstrates substantial equivalence to legally marketed predicate devices, and any differences between the subject device and the predicate devices do not impact the safety and effectiveness of Toro-L Interbody Fusion System.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.