K Number
K202954
Date Cleared
2021-07-02

(275 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Surgical Mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided sterile.

Device Description

The Disposable Surgical Mask is composed of mask body, nose clip and ear loop. The mask is composed of three layers: the inner and outer layers are made of Spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter, the nose clip is made of plastic materials and iron wire, ear loop is made of spandex.

The size of the disposable surgical mask is 17.5*9.5cm with tolerance +-5%, the length of the ear loop is 16cm with tolerance +-5%, and the length of the nose clip should no less than 8.0cm.

The outer layer of disposable surgical mask will be provided in blue, the inner layer of the disposable surgical mask will be provided in white, and it will be provided with sterile and is intended to be single use, disposable devices.

AI/ML Overview

The document describes the non-clinical testing performed to demonstrate that the Disposable Surgical Mask (Model MP9017) manufactured by XIAMEN PROBTAIN MEDICAL TECHNOLOGY CO., LTD meets its acceptance criteria and is substantially equivalent to a predicate device.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

ItemReported Device PerformanceAcceptance Criteria (Level 2)Result
Fluid Resistance (ASTM F1862)32 out of 32 pass at 120 mmHg29 out of 32 pass at 120 mmHgPASS
Particulate Filtration Efficiency (ASTM F2299)99.77%≥ 98%PASS
Bacterial Filtration Efficiency (ASTM F2101)99.7%≥ 98%PASS
Differential Pressure (Delta P) (EN 14683 Annex C)4.26 mmH2O/cm²< 6.0 mmH2O/cm²PASS
Flammability (16 CFR 1610)Class 1Class 1PASS
CytotoxicityUnder the conditions of the study, the device is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder the conditions of the study, the device is non-irritating.Non-IrritatingPASS
SensitizationUnder the conditions of the study, the device is non-sensitizing.Non-SensitizingPASS

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for each test in the acceptance criteria table (Table 2 and 3). However:

  • Fluid Resistance (ASTM F1862): The acceptance criteria and performance indicate a sample size of 32 test articles.
  • Other performance tests (PFE, BFE, Differential Pressure, Flammability): The specific sample sizes for these tests are not provided.
  • Biocompatibility tests (Cytotoxicity, Irritation, Sensitization): The specific sample sizes are not provided, but these tests are typically conducted in vitro or on animal models following ISO 10993 guidelines.

Data Provenance: The non-clinical tests were conducted by the manufacturer, XIAMEN PROBTAIN MEDICAL TECHNOLOGY CO., LTD, based in China (Xiamen, Fujian). The document does not specify if the data is retrospective or prospective, but as it's part of a 510(k) submission for a new device, these would typically be newly conducted prospective tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the acceptance criteria are based on established international standards (ASTM, EN, ISO, CFR) for medical device performance and biocompatibility, not on expert consensus for a ground truth.

4. Adjudication Method for the Test Set

This information is not applicable as the tests are objective laboratory measurements against defined standards, not subjective assessments requiring expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." Surgical masks are typically evaluated through non-clinical performance and biocompatibility testing against established standards, not through studies involving human readers or comparative effectiveness with AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. This device is a physical product (surgical mask), not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the test results is based on objective measurements performed according to established international standards and test methods. For example:

  • Fluid Resistance: Measured according to ASTM F1862.
  • Filtration Efficiency: Measured according to ASTM F2299 (PFE) and ASTM F2101 (BFE).
  • Differential Pressure: Measured according to EN 14683 Annex C.
  • Flammability: Measured according to 16 CFR 1610.
  • Biocompatibility: Evaluated according to ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization).

8. The Sample Size for the Training Set

This information is not applicable. This device is a physical surgical mask and does not involve an AI algorithm that requires a training set. The term "training set" is relevant for AI/machine learning models.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 2, 2021

Xiamen Probtain Medical Techology Co., Ltd % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 13th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K202954

Trade/Device Name: Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 19, 2021 Received: May 19, 2021

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202954

Device Name Disposable Surgical Mask

Indications for Use (Describe)

The Disposable Surgical Mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided sterile

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K202954

Summary prepared date: 2021-05-14

A. Applicant:

XIAMEN PROBTAIN MEDICAL TECHNOLOGY CO., LTD Address: 4th Floor,No.1 Building,No.6 Ji'an Road,Tong'an District,Xiamen, Fujian,361100, China Contact Person: Kangjianli Tel: +86-592-7557106 Fax: +86-592-7199255 Email: kangjianli@probtain.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device:

Trade Name: Disposable Surgical Mask Common Name: Disposable Surgical Mask Model(s): MP9017

Regulatory Information Classification Name: Disposable Surgical Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

C. Predicate device:

K100846 IMC Surgical Face Mask (non-sterile and sterile, yellow) International Medsurg Connection

D. Indications for use of the device:

{4}------------------------------------------------

The Disposable Surgical Mask are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided sterile.

E. Device Description:

The Disposable Surgical Mask is composed of mask body, nose clip and ear loop. The mask is composed of three layers: the inner and outer layers are made of Spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter, the nose clip is made of plastic materials and iron wire, ear loop is made of spandex.

The size of the disposable surgical mask is 17.5*9.5cm with tolerance +-5%, the length of the ear loop is 16cm with tolerance +-5%, and the length of the nose clip should no less than 8.0cm.

The outer layer of disposable surgical mask will be provided in blue, the inner layer of the disposable surgical mask will be provided in white, and it will be provided with sterile and is intended to be single use, disposable devices.

F. Comparison of technological characteristics with the predicate device.

Table 1 General Comparison

DeviceSubject DevicePredicate DeviceComparison
ManufacturerXIAMEN PROBTAIN MEDICALTECHNOLOGY CO., LTDINTERNATIONAL MEDSURGCONNECTION-
510K numberK202954K100846-
Model NameDisposable Surgical MaskIMC Surgical Face Mask(non-sterile and sterile, yellow)-
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21CFR878.4040)Same
Intended/Indications for useThe Disposable Surgical Masks areintended to be worn to protect boththe patient and healthcare personnelfrom transfer of microorganisms,body fluids and particulate material.These face masks are intended touse in infection control practices toreduce the potential exposure toblood and body fluids. This is adisposable device(s), providedsterile.This device is intended to be wornby operating room personnel duringsurgical procedures to protect boththe surgical patient and theoperating room personnel fromtransfer of microorganisms, bodyfluids, and particulate material.Similar
ModelMP9017--
Design featureEar Loops, Flat Pleated, 3 layersEar Loops, Tie-On, Flat Pleated, 3layersSimilar
MaterialOuter layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Inner layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Nose clipPlastic materials and iron wireNADifferent
Ear loopsSpandexNADifferent
ColorBlueYellowDifferent
Dimension(Length)17.5±5%cm17.8cm (7 inches)Different
(Width)9.5cm±5%cm8.9cm (3.5inches)Different
OTC useYesYesSame
SterilitySterileSterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100 levelLevel 2Level 2Same
Fluid ResistancePer ASTM F1862, an acceptable quality limit of 4.0% is met for a normal single sampling plan when ≥ 29 of 32 test articles show passing results.The surgical face mask were tested in accordance with ASTM F1862-07 test methods and meets its acceptance Criteria of Low =80, Moderate =120,High = 160.Similar
Filtration EfficiencyParticulateThe test results in average of 98.6% are in accordance with the requirements under ASTM F2299The surgical face mask were tested in accordance with ASTM F2299 test methods and meets its acceptance Criteria of ≥98%Similar
Filtration EfficiencyBacterialThe test results in average of 99.6% are in accordance with the requirement under ASTM F2101-19The surgical face mask were tested in accordance with ASTM F2101-07 test methods and meets its acceptance criteria of Low≥95%, Moderate =≥98%, High ≥98%.Similar
PressureDifferentialThe test results in average of 5.0mmH2O/cm² are in accordance with the requirement < 6.0mmH2O/cm² under EN 14683 Annex CThe surgical face mask were tested in accordance with MIL-M36945C 4.4.1.1.1 test methods and meets its acceptance criteria of < 4 mmH2O/cm²Similar
FlammabilityThe test results is Class I under test method of 16 CFR Part 1610The surgical face mask were tested in accordance with 16 CFR Part 1610 test methods and meets its acceptance criteria of Class ISimilar
In VitroCytotoxicity testUnder the conditions of the study, the device is non-cytotoxic.The materials of the surgical face mask were tested in accordance with ISO10993-5 test methods and were found to be acceptable for theSimilar
intended use.
Skin IrritationUnder the conditions of the study,the device is non-irritating.The materials of the surgical facemask were tested in accordance withISO10993-10 test methods and werefound to be acceptable for theintended use.Similar
Skin SensitizationUnder the conditions of the study,the device is non-sensitizingThe materials of the surgical facemask were tested in accordance withISO10993-10 test methods and werefound to be acceptable for theintended use.Similar

{5}------------------------------------------------

{6}------------------------------------------------

XIAMEN PROBTAIN MEDICAL TECHNOLOGY CO., LTD Address: 4th Floor,No.1 Building,No.6 Ji'an Road,Tong'an District,Xiamen, Fujian,361100, China

From the comparison we found the material of the current nose clip, the ear loop and the color were different from the predicate device. The biocompatibility tests were conducted to ensure their compliance to the ISO10993-5 and ISO10993-10. There is no new risk generated from the difference of the material.

Also, the size of the mask was different from the predicate device. The performance testing was conducted and although the test results are not identical to each other, they are similar and they both meet the requirement of Level 2 medical mask according to the ASTM F2100.

G. Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

  • A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
  • ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

  • ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks

  • ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity);

  • A EN 14683, Medical Face Masks-Requirements and Test Methods;
  • ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

  • A ASTM F2299, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
  • A 16 CFR 1610, Standard for the Flammability of clothing textiles;
ItemProposed deviceAcceptance Criteria (level 2)Result
FluidResistancePerformanceASTM32 out of 32 pass at 120 mmHg29 out of 32 pass at 120 mmHgPASS

Table 2 - Performance Testing

{7}------------------------------------------------

K202954

XIAMEN PROBTAIN MEDICAL TECHNOLOGY CO., LTD Address: 4th Floor,No.1 Building,No.6 Ji'an Road,Tong'an District,Xiamen,Fujian,361100,China

F1862
ParticulateFiltrationEfficiency ASTM F229999.77%≥ 98%PASS
BacterialFiltrationEfficiency ASTM F210199.7%≥ 98%PASS
DifferentialPressure(Delta P) EN 14683Annex C4.26mmH2O/cm²< 6.0mmH2O/cm²PASS
Flammability 16 CFR1610Class 1Class 1PASS

Table 3 Biocompatibility Comparison

ItemProposed deviceAcceptance CriteriaResult
CytotoxicityUnder the conditions of the study, thedevice is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder the conditions of the study, thedevice is non-irritating.Non-IrritatingPASS
SensitizationUnder the conditions of the study, thedevice is non-sensitizingNon-SensitizingPASS

H. Clinical Tests

No clinical study is included in this submission.

I. Conclusion

The conclusions drawn from the nonclinical tests that the subject is as safe, as effective, and performs as well as the legally marketed predicate device, IMC Surgical Face Mask (non-sterile, yellow) cleared under K100846.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.