K Number
K202905
Date Cleared
2021-03-26

(178 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Device Description

The proposed device, Disposable Face Mask is a three-layer, single-use, flat-pleated mask. The ear loops to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. The disposable surgical mask is available in four different specifications: 14.5cmx9.5cm and 17.5cmx9.5cm. The device is single use and provided non-sterile.

AI/ML Overview

The provided text describes the 510(k) submission for a Disposable Face Mask (K202905) and its equivalence to a predicate device (K160269). However, it does not contain information related to an AI/ML-enabled medical device or a clinical study involving human readers and AI assistance.

Therefore, I cannot provide the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies for an AI device. The document is strictly about the non-clinical testing and comparison of a traditional medical device (face mask) to a predicate device to demonstrate substantial equivalence.

The document covers:

  • Device: Disposable Face Mask
  • Purpose: To protect against transfer of microorganisms, body fluids, and particulate material.
  • Regulatory Class: Class II, Product Code FXX.
  • Acceptance Criteria/Performance (based on standards for face masks):
    • Fluid Resistance (ASTM F1862)
    • Particulate Filtration Efficiency (ASTM F2299)
    • Bacterial Filtration Efficiency (ASTM F2101)
    • Differential Pressure (EN 14683)
    • Flammability (16 CFR 1610)
    • Biocompatibility (ISO 10993-5 and ISO 10993-10)

Here's a breakdown of what can be extracted from the document, tailored as much as possible to your request, but highlighting the absence of AI/ML specifics:

1. Table of Acceptance Criteria and Reported Device Performance

CriterionAcceptance Criteria (Predicate)Reported Device Performance (Proposed Device K202905)Remark
Fluid Resistance (ASTM F1862)Level 1 Pass at 80mmHg; Level 2 Pass at 120mmHgLevel 1 Pass at 80mmHg; Level 2 Pass at 120mmHgIdentical
Particulate Efficiency (ASTM F2299)Pass at 99.6% (implied for Level 1 & 2 for predicate)$\ge 95%$ (Level 1), $\ge 98%$ (Level 2)Similar (Meets Level 2 requirements)
Bacterial Filtration Efficiency (ASTM F2101)Pass at >98% (implied for Level 1 & 2 for predicate)$\ge 95%$ (Level 1), $\ge 98%$ (Level 2)Similar (Meets Level 2 requirements)
Differential Pressure (EN 14683)Passed at 2.0 mmH2O/cm² (predicate's Level 1), Passed at 1.6 mmH2O/cm² (predicate's Level 2)$<$ 5.0mm H2O/cm² (Level 1), $<$ 6.0mm H2O/cm² (Level 2)Similar (Meets Level 2 requirements)
Flammability (16 CFR 1610)Class 1Class 1Identical
Biocompatibility (Cytotoxicity - ISO 10993-5)No CytotoxicityNo CytotoxicityIdentical
Biocompatibility (Sensitization - ISO 10993-10)No SensitizationNo SensitizationIdentical
Biocompatibility (Irritation - ISO 10993-10)No IrritationNo IrritationIdentical

Note: The 'acceptance criteria' are implicitly derived from the performance of the predicate device and the relevant ASTM/EN standards for medical face masks. The proposed device had to meet or exceed these to demonstrate substantial equivalence.

Regarding AI/ML specific information (which is not in the document):

  1. Sample size used for the test set and data provenance: Not applicable to this device type or study.
  2. Number of experts used to establish the ground truth and qualifications: Not applicable. Ground truth for a face mask's performance is established through standardized laboratory testing (e.g., measuring filtration efficiency, fluid resistance) against predefined physical criteria, not expert interpretation of AI outputs.
  3. Adjudication method: Not applicable.
  4. Multi-reader multi-case (MRMC) comparative effectiveness study: No, this was not done as it's not an AI-enabled device requiring human-in-the-loop assessment.
  5. Standalone (algorithm only) performance: Not applicable.
  6. Type of ground truth: (expert consensus, pathology, outcomes data, etc.): Ground truth for this device is based on physical and biological laboratory measurements according to established international standards (ASTM, EN, ISO).
  7. Sample size for the training set: Not applicable, as this is not an AI/ML device that requires a training set.
  8. How the ground truth for the training set was established: Not applicable.

In summary, the provided document details the regulatory clearance of a disposable face mask, focusing on non-clinical performance testing against established standards and comparison to a predicate device, not on the performance of an AI/ML system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 26, 2021

Jiangsu Excellence Medical Supplies Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai. 200120 China

Re: K202905

Trade/Device Name: Disposable Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 25, 2021 Received: February 26, 2021

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K202905

Device Name

Disposable Face Mask

Indications for Use (Describe)

The Disposable Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The

assigned 510(k) Number: K202905

    1. Date of Preparation: 03/26/2020
    1. Sponsor Identification

Jiangsu Excellence Medical Supplies Co., Ltd.

No.65 Huiling West Road, Zouqu Town, Zhonglou District, Changzhou City, Jiangsu Province, 203144, China

Establishment Registration Number: 3010138147

Contact Person: Jing Zhu Position: General Manager Tel: +86-519-83635058 Fax: +86-519-83631033 Email:ritazhu@hualiandressing.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd.

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net

    1. Identification of Proposed Device
      Common Name: Surgical Mask Trade Name: Disposable Face Mask Models: 14.5cm×9.5cm(L1), 17cm×9.5cm(L1), 17.5cm×9.5cm(L1) 14.5cm×9.5cm(L2), 17cm×9.5cm(L2), 17.5cm×9.5cm(L2)

Regulatory Information Classification Name: Mask, Surgical

{4}------------------------------------------------

Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040 Review Panel: General Hospital

Indication for use:

The Disposable Face Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Device Description:

The proposed device, Disposable Face Mask is a three-layer, single-use, flat-pleated mask. The ear loops to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. The disposable surgical mask is available in four different specifications: 14.5cmx9.5cm and 17.5cmx9.5cm. The device is single use and provided non-sterile.

    1. Identification of Predicate Device
      510(k) Number: K160269 Product Name: Surgical Face Masks (Ear loops and Tie-on) Models: EL 10000, EL 10010, TO 10000, TO 10010 EL 20000, EL 20010, TO 20000, TO 20010

EL 30000, EL 30010, TO 30000, TO 30010

  • Comparison of technological characteristics between the subject and predicate devices 6.
ITEMProposed Device K202905Predicate Device K160269Remark
Product CodeFXXFXXIdentical
Regulation No.21 CFR 878.404021 CFR 878.4040Identical
ClassIIIIIdentical
Indication for UseThe Disposable Face Mask isintended to be worn to protectboth the patient and healthcarepersonnel from transfer ofmicroorganisms, body fluids,and particulate material. Theseface masks are intended for usein infection control practices toreduce the potential exposure toblood and body fluids. This is asingle-use, disposable device,provided non-sterile.The surgical face masks are intended to beworn to protect both the patient andhealthcare personnel from transfer ofmicroorganisms, body fluids, andparticulate material. These face masks areintended for use in infection controlpractices to reduce the potential exposureto blood and body fluids. This is a single-use, disposable device,provided non-sterile.Same
Mask styleFlat pleatedFlat pleatedIdentical

Table 1 Comparison of Disposable Face Mask

{5}------------------------------------------------

Design featureEar strapEar strap/ Tie-onSame
Dimension145mm×95mm,175mm×90mm,Similar
170mm×95mm,180mm×90mm
175mm×95mm
ASTM F2100 LevelLevel 1Level 1Identical
Level 2Level 2Identical
Fluid resistanceLevel 1 Passat 80mmHgLevel 1 Passat 80mmHgIdentical
ASTM F1862Level 2 Passat 120mmHgLevel 2 Passat 120mmHgIdentical
Particulate efficiencylevel ASTM F2299≥95%≥98%Pass at 99.6%Pass at 99.6%Similar
Bacterial filtrationlevel ASTM F2101≥95%≥98%Pass at >98%Pass at >98%Similar
Differential pressureEN 14683$<$ 5.0mmH2O/ cm²$<$ 6.0mmH2O/ cm²Passed at 2.0mmH2O/cm²Passed at 1.6mmH2O/cm²Similar
Flammability16 CFR 1610Class 1Class 1Identical
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Identical
Patient Contacting Material
ear strapSpandex and PolyesterPolyester, polyurethane
nose clipHDPE and IronPolyethylene and steel wire
mask body25g/m² PP spunbond non-woven cloth and 25g/m² PP meltblown non-woven clothPolypropylene spun bond and Polypropylene melt-blownDifferent
Biocompatibility
CytotoxicityComply with ISO 10993-5 No CytotoxicityComply with ISO 10993-5 No Cytotoxicity
SensitizationComply with ISO 10993-10 No SensitizationComply with ISO 10993-10 No SensitizationIdentical
IrritationComply with ISO 10993-10 No IrritationComply with ISO 10993-10 No Irritation

Similar - Dimension

The dimension for the proposed device is different from predicate device. This difference does not affect intended use and will not raise any safety issues. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Similar - Particulate efficiency level

The test result for particulate efficiency for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Similar - Bacterial filtration level

The test result for bacteria efficiency for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

{6}------------------------------------------------

Similar - Differential pressure

The test result for different pressure for the proposed device is different from predicate device. However, the test result for the proposed device can meet the requirements of level 2 mask. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device.

Different- Patient Contacting Material

The patient contact material for the propose device is different from predicate device. However, biocompatibility test has been conducted on the propose device and the test result does not show any adverse effect. Therefore, this difference does not affect substantially equivalence between the proposed device and predicate device.

    1. Non-Clinical Test Summary
      Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
  • A ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)

  • ASTM F2299/F2299M-03(2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres

  • ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of A Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus

  • ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity

  • ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization

  • Clinical Test Conclusion 8.

No clinical study is included in this submission.

Substantially Equivalent (SE) Conclusion 9.

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K202905, the Disposable Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K160269.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.