K Number
K202847

Validate with FDA (Live)

Device Name
AccuBrain
Date Cleared
2021-01-15

(112 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AccuBrain is a fully automated post-processing software that provides automatic labeling, visualization and volumetric quantification of hippocampus from a set of T1W MRIs and returns an analysis report.

Device Description

AccuBrain is a fully automated post-processing software that provides automatic labeling, visualization and volumetric quantification of hippocampus from a set of T1W MRIs and returns an analysis report. The resulting output is a morphometric report in PDF format. The software is suitable for use in both clinical trial research and routine patient care as a support tool for clinicians in assessment of structural MRIs.

AccuBrain provides morphometric measurements based on 3D T1 MRI series. The output of the software includes volumes of hippocampus.

The AccuBrain processing architecture includes an automated internal pipeline that performs artifact correction, segmentation, hippocampus quantification, volume calculation and report generation.

Additionally, automated safety measures include automated quality control functions, such as DICOM check, age check and image resolution check and image quality check.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document for the AccuBrain device:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Mean Dice coefficient for right hippocampus0.89 (std: 0.03)
Mean Dice coefficient for left hippocampus0.89 (std: 0.03)
Mean Dice coefficient for total hippocampus0.89 (std: 0.03)
Mean intrascanner Coefficient of Variation (CV) for left hippocampus3.20%
Mean intrascanner Coefficient of Variation (CV) for right hippocampus1.23%
Mean percentage absolute volume differences (DIFF) for left hippocampus4.52%
Mean percentage absolute volume differences (DIFF) for right hippocampus1.74%

Study Details

2. Sample Size Used for the Test Set and the Data Provenance:

  • Sample Size: 135 subjects.
  • Data Provenance: The data was provided by the EADC-ADNI HarP (European Alzheimer's Disease Consortium - Alzheimer's Disease Neuroimaging Initiative Harmonized Protocol for Hippocampal Segmentation). The document does not specify the country of origin, but ADNI is primarily a North American initiative with international collaborations, and EADC is European. The study states "The subjects upon whom the device was tested include healthy subjects, Alzheimer's disease patients and Mild Cognitive Impairment patients," implying retrospective use of pre-existing data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

  • The document states that the ground truth was "manual segmentation results." It does not specify the number of experts or their qualifications for the test set.

4. Adjudication Method for the Test Set:

  • The document refers to "manual segmentation results" as the ground truth but does not describe an explicit adjudication method (e.g., 2+1, 3+1). The use of the term "manual segmentation results" in the singular might suggest a single expert, or a consensus that is not explicitly detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • A MRMC comparative effectiveness study was not explicitly described. The study focused on the standalone performance of AccuBrain against manual segmentation (ground truth) and compared its performance metrics (Dice coefficients, CV, DIFF) to previously established performance ranges or specific values of predicate devices. There is no mention of human readers improving with AI vs without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Yes, a standalone study was done. The performance metrics (Dice coefficient, CV, DIFF) reported for AccuBrain are based on the algorithm's output compared to manual segmentation or repeated scans, without human intervention in the segmentation process.

7. Type of Ground Truth Used:

  • Expert Consensus / Expert-driven segmentation: The ground truth for the test set was "manual segmentation results," which implies segmentation performed by human experts. The reference to the EADC-ADNI HarP further indicates that these manual segmentations likely followed a standardized protocol for consistency and accuracy.

8. Sample Size for the Training Set:

  • The training set, referred to as the "atlas pool," consisted of 300 brain MRIs.

9. How the Ground Truth for the Training Set Was Established:

  • The ground truth for the training set (atlas pool) was established through "prior encoded radiologist-specified anatomy information for hippocampus." This indicates that human experts (radiologists) manually segmented or annotated the hippocampi in the 300 ATLAS MRIs, which then served as the ground truth for training the atlas-based segmentation method.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

BrainNow Medical Technology Limited % You Yijie General Manager Qimmiq Medical Consulting Service Co., Ltd. RM.1711, Building K. NO.101 Science Ave International Creative Valley Guangzhou, Guangdong 510663 CHINA

January 15, 2021

Re: K202847

Trade/Device Name: AccuBrain Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: September 11, 2020 Received: December 11, 2020

Dear You Yijie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202847

Device Name AccuBrain

Indications for Use (Describe)

AccuBrain is a fully automated post-processing software that provides automatic labeling, visualization and volumetric quantification of hippocampus from a set of MRIs and returns an analysis report.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

"510(k) Summary" as required by 21 CFR Part 807.92.

1. Submitter's Information

Establishment Registration Information

Name: BrainNow Medical Technology Limited Address: Unit 201, 2/F, Lakeside 2, No. 10 Science Park West Avenue, Hong Kong Science Park, Shatin, N.T., Hong Kong ZIP/Postal Code: 999077

Contact Person of applicant

Contact Person: Junbing Huang Telephone Number: 86-13416183887 Fax Number: 852-36221760 Email: 417731983@qq.com

Contact Person of the Submission:

Name: You Yijie Address: RM.1711, Building K, NO.101 Science Ave International Creative Valley Development Zone, Guangzhou China TEL: +86 020-8224 5821 FAX: +86 020-8224 5821 Email: Jet.you@qimmig-med.com

Date to prepare: 9/11/2020

2. Device Information

Device Name: AccuBrain Common Name: Neuroimage Analysis Software Model: AccuBrain_Intl Software version: V1.0.0.200703 Classification Name: System, Image Processing, Radiological Regulation Number: 21 CFR 892.2050 Regulation Description: Picture archiving and communications system Product Code: LLZ Classification Panel: Radiology Regulation Class: -

3. Predicate Device Information

ItemPrimary Predicate(A)Predicate or Reference Device (B)
510(k) submitter/holderCorTechs Labs, IncBrainreader Aps
510(K) NumberK170981K140828
Device nameNeuroQuantNeuroReader MedicalImage Processing Software
Common nameMedical ImageProcessing SoftwareNeuroreader

{4}------------------------------------------------

Regulation DescriptionPicture archiving andcommunications systemPicture archiving andcommunication system
Review panelRadiologyRadiology
Product codeLLZLLZ, LNH
Regulation ClassClass IIClass II
Regulation Number21 CFR 892.205021 CFR 892.2050

4. Device Description

AccuBrain is a fully automated post-processing software that provides automatic labeling, visualization and volumetric quantification of hippocampus from a set of T1W MRIs and returns an analysis report. The resulting output is a morphometric report in PDF format. The software is suitable for use in both clinical trial research and routine patient care as a support tool for clinicians in assessment of structural MRIs.

AccuBrain provides morphometric measurements based on 3D T1 MRI series. The output of the software includes volumes of hippocampus.

The AccuBrain processing architecture includes an automated internal pipeline that performs artifact correction, segmentation, hippocampus quantification, volume calculation and report generation.

Additionally, automated safety measures include automated quality control functions, such as DICOM check, age check and image resolution check and image quality check.

5. Principle of operation

AccuBrain automatically segmented the subject's hippocampi using the uploaded T1W MRIs in a multi-atlas-based segmentation manner. The quantification and visualization of hippocampal segmentation results were output in the form of a morphometric analysis report.

The hippocampal segmentation procedure is described as follows. 1) Pre-processing to increase the image quality, including noise reduction, bias field correction, and intensity normalization to normalize intensity level of MRIs from different scanners. For noise reduction method, we used non-local mean filtering method [1]. Bias correction method used in AccuBrain is N4 bias correction [2]. Intensity Normalization method used in AccuBrain is histogram matching [3]. 2) Atlas selection. The atlas pool, consisting of 300 brain MRIs together with their segmentation labels, were previously obtained from different individuals using different scanners and have highly variable appearances. Each atlas contains both brain MRI and the prior encoded radiologist-specified anatomy information for hippocampus. The detailed of demographic information about the atlas data was described in Table 1. During processing, AccuBrain selects a number of brain images from the atlas pool based on similarity with the subject images. The similarity measures used in this step is Normalized Cross correlation (NCC). AccuBrain will select 10 images from the atlas pool with highest NCC scores. 3) Image segmentation. The non-rigid image registration is performed to match the selected image with the subject image. The resulting transformation field will be applied to transform the predefined atlas label to the subject image. As 10 template images are selected, 10 segmentation results are obtained and will be merged using STAPLE label fusion method l41 to fuse the final segmentation labels of the subject image.

{5}------------------------------------------------

Demographic CategoriesFrequency (subjectnumber)Percentage(%)
Gender
Female15150.3
Male14949.7
Disease Status
AD12341
NC11538.3
MCI6220.7
Age
51-60258.3
61-709632
71-8011839.3
81-906120.3
Magnetic Field Strength
1.5T6822.7
3T23277.3
Manufacturer
GE13143.7
Philips4414.7
Siemens12541.7
In-plane resolution
1x121471.3
0.9375x0.93758428
0.8594x0.859420.7
FOV (mm²)
22020.7
23013545
2408428
2567926.3
Slice Thickness(mm)
121672
1.28428

Table 1. Demographic information of the atlas data

6. Indications for Use

AccuBrain is a fully automated post-processing software that provides automatic labeling, visualization and volumetric quantification of hippocampus from a set of T1W MRIs and returns an analysis report.

7. Comparison of Predicate Devices

Summary Comparison Table for the subject device and predicate devices (K170981 and K140828):

ComparisonElementsSubject DevicePredicateDevice (A)Predicate orReferenceDevice (B)Discussion ofdifference
Device NameAccuBrainNeuroQuantNeuroReader/
510(k) No/K170981K140828/

{6}------------------------------------------------

Regulation No21 CFR892.205021 CFR892.205021 CFR892.2050Sameaccording to the ANSIAAMI IEC62304:2006/A1:2016(SoftwareDocumentation,section 004) and theAccuracy andReproducibility of theSubject Device wasverified (section 006).The difference doesnot affect thedetermination ofsubstantialequivalence.
RegulationDescription"Picturearchiving andcommunicationssystem""Picturearchiving andcommunicationssystem""Picturearchiving andcommunicationssystem"SameProcessingarchitectureAutomatedinternal pipelinethat performs:Automatedinternal pipelinethat performs:Information notpubliclyavailableSE-within thepredicate
ClassificationnameSystem, ImageProcessing,RadiologicalSystem, ImageProcessing,RadiologicalSystem, ImageProcessing,RadiologicalSame-artifactcorrection-artifactcorrectionBoth are performed:
ClassificationClass IIClass IIClass IISame-segmentation-segmentation-artifact correction
Product codeLLZLLZLLZSame-hippocampusquantification-lesionquantification-segmentation
Indications foruseAccuBrain is afully automatedpost-processingsoftware thatprovidesautomaticlabeling,visualizationand volumetricquantification ofhippocampusfrom a set ofT1W MRIs andreturns ananalysis report.Automaticlabeling,visualization andvolumetricquantification ofsegmentablebrain structuresand lesions froma set of MRimages.Volumetric datamay becompared toreferencepercentile dataAutomaticlabeling,visualizationand volumetricquantification ofsegmentablebrain structuresfrom a set ofMR images.This software isintended toautomate thecurrent manualprocess ofidentifying,labeling andquantifying thevolume ofsegmentablebrain structuresidentified on MRimages.SE-within thepredicateBoth are indicated toautomatic labeling,visualization andvolumetricquantification ofhippocampus from aset of T1W MRIs Isand returns ananalysis report.-volumecalculation-volumecalculation-hippocampusquantification
Design andincorporatedtechnology• Automatedmeasurement ofhippocampusvolumes• Automaticatlas-basedsegmentationandquantification ofhippocampususing an atlaspool consistingof difference• Automatedmeasurement ofbrain tissuevolumes andstructures andlesions• Automaticsegmentationandquantification ofbrain structuresusing a dynamicprobabilisticInformation notpubliclyavailableSEBoth are indicated toautomatemeasurement ofhippocampusvolumes throughautomatic atlas-basedsegmentation andquantification ofhippocampus base onan atlas poolconsisting ofdifference template-reportgeneration-reportgeneration-volume calculation
templateimages withhighly variableappearancetogether withtheir priorencodedradiologist-specifiedanatomyinformationimages with highlyvariable appearancetogether with theirprior encodedradiologist-specifiedanatomy informationneuroanatomicalatlas, with ageand genderspecificity,based on theMR imageintensityInformation notpublicly availableimages with highlyvariable appearancetogether with theirprior encodedradiologist-specifiedanatomy information.The difference doesnot affect thedetermination ofsubstantialequivalence.-report generation
Physicalcharacteristics• Web-basedapplication• Operates onoff-the -shelfhardware(multiplevendors)• Softwarepackage• Operates onoff-the-shelfhardware(multiplevendors)Information notpubliclyavailableSEThe Cybersecurity ofSubject Device wasverified(CybersecurityInformationDocument, section005).And the SubjectDevice was verifiedaccording to theANSI AAMI IEC62304:2006/A1:2016(section 004) andthe Accuracy andReproducibility ofthe Subject Devicewas verified (section006).The difference doesnot affect thedetermination ofsubstantialequivalence.The difference doesnot affect thedetermination ofsubstantialequivalence.
OperatingsystemSupportsWindowsSupports Linux,Mac OS X andWindows.Information notpubliclyavailableSE--within thepredicate.The Operatingsystem of SubjectDevice is fewer thanPredicate Device (A),the risk of the SubjectDevice is fewer thanPredicate Device (A).And the SubjectDevice was verifiedData source• MRI scanner:3D T1 MRIscans acquiredwith specifiedprotocols• MRI scanner:3D T1 MRIscans acquiredwith specifiedprotocolsInformation notpubliclyavailableSame
• AccuBrainsupportsDICOM formatas input• NeuroQuantsupports DICOMformat as input
Output-Providesvolumetric• ProvidesvolumetricmeasurementsInformation notpubliclyavailableSE-within thepredicate

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

measurementsof hippocampus-Includessegmentedcolor overlaysand an analysisreport (justvolumetric ofthehippocampus)of brainstructures andlesions• Includessegmented coloroverlays andmorphometricreports• Automaticallycomparesresults toreferencepercentile dataand to priorscans whenavailable• SupportsDICOM formatas output ofresults that canbe displayed onDICOMworkstationsand PictureArchive andCommunicationsSystemsBoth providevolumetricmeasurements ofhippocampus-Includes segmentedcolor overlays and ananalysis reportincluding thevolumetric of thehippocampus.The difference doesnot affect thedetermination ofsubstantialequivalence.
AccuracyThe mean DICEof AccuBrainresults andmanualsegmentationresults is 0.89,0.89 and 0.89for right, left andtotalhippocampus,respectively.For majorsubcortical brainstructures Dice'scoefficients arein the range of80%-90%.NeuroReadercan segmentthehippocampuswith a Dicesimilarity indexof 0.87 for boththe right and lefthippocampus.SEThe accuracy andreproducibility ofhippocampussegmentation ofAccuBrain with T1WMRI images arecomparable withPredicate Device(A) and PredicateDevice (B). Theaccuracy andreproducibility ofSubject Device wasverified (Accuracyand ReproducibilityTest Report, section006).
The difference doesnot affect thedetermination ofsubstantialequivalence.
Safety• Automatedquality controlfunctions- DICOM check- Age check- Imageresolution check- Image qualitycheckDiagnosticdecisionsshould be madeby trainedclinicians.• Automatedquality controlfunctions- Tissue contrastcheck- Scan protocolverification- Atlas alignmentcheck• Results mustbe reviewed bya trainedphysicianInformation notpubliclyavailableSEBoth are conductedthe - DICOM check,Age check, Imageresolution check,Image quality check.The Cybersecurity ofSubject Device wasverified(CybersecurityInformationDocument, section005).And the SubjectDevice was verifiedaccording to the IEC62304(SoftwareDocumentation,section 004) and theAccuracy andReproducibility ofthe Subject Devicewas verified(Accuracy andReproducibility testreport, section 006).The difference doesnot affect thedetermination ofsubstantialequivalence.

{10}------------------------------------------------

Subject device and predicate devices are softwares for automatically identifying and quantifying volumes of brain structures. Subject and predicate devices take 3D MR images of the brain as input and generate electronic report with similar quantitative information.

AccurBrain and NeuroQuant achieve the intended use based on similar principle and processing architecture, since the quantification systems implement brain segmentation and quantification using atlas-based segmentation scheme. Both hippocampi are segmented and the volumes are calculated.

{11}------------------------------------------------

AccurBrain and NeuroQuant are DICOM compatible and operate on off-the-shelf hardware. Meanwhile, both devices are used by medical professional, such as radiologists, neurologists and neuroradiologists, as well as by clinical researchers, as a support tool in assessment of structural MRIs.

The output volumes which both devices provide include volumes of left hippocampus, right hippocampus and whole hippocampi.

8. Performance Testing

To demonstrate the performance of AccuBrain (model: AccuBrain_Intl), the measured volumes and volume differences of hippocampus are validated for accuracy and reproducibility. The subjects upon whom the device was tested include healthy subjects, Alzheimer's disease patients and Mild Cognitive Impairment patients. AccuBrain segmentation accuracy with 3D T1 MRI scans was evaluated using Dice coefficient metric. With 135 data provided by the EADC-ADNI HarP, the mean Dice coefficient by comparing AccuBrain results and manual segmentation results was 0.89 (std: 0.03), 0.89 (std: 0.03) and 0.89 (std: 0.03) for right, left and total hippocampal volumes, respectively. Segmentation reproducibility of repeated 3D T1 MRI scans for the same subjects was evaluated using Coefficient of Variation (CV). The mean intrascanner CV values were 3.20% and 1.23%, the mean percentage absolute volume differences DIFF values were 4.52% and 1.74% for left and right hippocampus, respectively. Compared with the performances of the predicate devices, the results presented above shows that the subject device is safe and effective and performs as well as the predicate devices. The AccuBrain (Model: AccuBrain_Intl) was designed, verified, and validated according to the company's Design Control process and has been subjected to extensive safety and performance testing as shown in the test results provided in this submission. Verification and Validation testing data demonstrate that the device meets all of its specifications.

The Accuracy and Reproducibility of AccuBrain (Model: AccuBrain Intl) was verified please see section 006 for the Accuracy and Reproducibility Test Report.

The software of AccuBrain (Model: AccuBrain Intl) was verified according to the ANSI AAMI IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes [Including Amendment 1 (2016)]. Please see section 004 for the Software Documentation.

The Cybersecurity of AccuBrain (Model: AccuBrain Intl) was verified. Please see section 005 for the Cybersecurity Information Document.

During the verification and validation activity the following quidance documents were used:

General Principles of Software Validation: Guidance for Industry and FDA Staff Postmarket Management of Cybersecurity in Medical Devices: Guidance for Industry and Food and Drug Administration Staff

Content of Premarket Submissions for Management of Cybersecurity in Medical Devices: Draft Guidance for Industry and Food and Drug Administration Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: Guidance for Industry and FDA Staff

9. Conclusions

The performance testing presented above shows that the device is as safe, as effective and performs as well as the predicate devices(A) and predicate devices(B),

{12}------------------------------------------------

and as well as gold standard-computer-aided expert manual segmentation. By virtue of the physical characteristics and intended user, AccuBrain(AccuBrain_Intl) is substantially equivalent to its predicate devices (A) (K170981) and predicate devices (B)(K140828).

10. Bibliography

[1] Coll, Bartomeu & Morel, Jean-Michel. (2005). A non-local algorithm for image denoising. Proceedings of the IEEE Computer Society Conference on Computer Vision and Pattern Recognition. 2. 60- 65 vol. 2.

[2] Tustison NJ, Avants BB, Cook PA, et al. N4ITK: improved N3 bias correction. IEEE Trans Med Imaging. 2010;29(6):1310-1320.

[3] Laszlo G. Nyul, Jayaram K. Udupa, and Xuan Zhang, "New Variants of a Method of MRI Scale Standardization", IEEE Transactions on Medical Imaging, 19(2):143-150, 2000.

[4] Warfield SK, et al. Simultaneous truth and performance level estimation (STAPLE): an algorithm for the validation of image segmentation. Medical Imaging, IEEE Transactions on. 2004;23:903-921.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).