(268 days)
The Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
The Medical Face Masks are single use, three-layer, flat-pleated style with ear loops and nose piece. The Medical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Galvanized iron wire+ Polyethylene(PE). The medical face masks will be provided in blue. The medical face masks are sold non-sterile and are intended to be single use, disposable devices.
The provided document is a 510(k) Premarket Notification for a Medical Face Mask, not an AI/ML medical device. Therefore, the information requested about acceptance criteria and study proving an AI/ML device meets them (including sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth establishment, and training set details) is not applicable and cannot be extracted from this document.
The document describes the non-clinical performance testing of a physical medical device (face mask) to demonstrate its substantial equivalence to a predicate device, focusing on material properties, fluid resistance, filtration efficiency, differential pressure, flammability, and biocompatibility.
Here's the relevant information that can be extracted, pertaining to the physical medical face mask, structured to highlight the difference from an AI/ML device study:
1. A table of acceptance criteria and the reported device performance:
| Item | Acceptance Criteria | Reported Device Performance (Result) | Conclusion |
|---|---|---|---|
| Fluid Resistance Performance (ASTM F1862) | ≥ 29 Out of 32 pass at 120 mmHg (16.0 kPa) | 32 Out of 32 pass at 120 mmHg(16.0 kPa) | Pass |
| Particulate Filtration Efficiency (ASTM F2299) | ≥ 98% | 99.7% | Pass |
| Bacterial Filtration Efficiency (ASTM F2101) | ≥ 98% | 98.7% | Pass |
| Differential Pressure (Delta P) MILM-36954C | < 6.0 mmH20/cm² | 1.98 mmH20/cm² | Pass |
| Flammability (16 CFR 1610) | Class 1 | Class 1 | Pass |
(Note: The "Reported Device Performance" and "Acceptance Criteria" columns from the document were swapped in this table for logical flow, as the "Result" column in the document appears to be the actual measured performance, and the "Acceptance Criteria" column lists the threshold required for passing.)
2. Sample sizes used for the test set and the data provenance:
- Sample Size: For Fluid Resistance Performance, the sample size was 32 (32 out of 32 tested). Sample sizes for other tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability, Biocompatibility) are not specified in this document, though these are typically standardized tests with defined sample sizes.
- Data Provenance: Not explicitly stated as "data provenance" in the context of an AI/ML study. However, the manufacturer is CHONGQING CHAOKE INDUSTRY DEVELOPMENT CO.,LTD. in Chongqing, China. The tests are non-clinical (laboratory-based) and are likely performed in a controlled laboratory environment conforming to the specified ASTM and other standards. The document does not specify if the testing was retrospective or prospective, but rather that it was performed for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not Applicable. This is a physical medical device (face mask) undergoing laboratory performance testing against established standards, not an AI/ML algorithm whose performance is measured against expert human-labeled ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. See point 3. Testing involves standard laboratory procedures and measurements, not human adjudication of AI output.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. See point 3. This device is not an AI/ML system and does not involve human readers or image interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. See point 3.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Ground truth in this context refers to established physical/chemical standards and measurement methods (e.g., ASTM F1862 for fluid resistance, ASTM F2299 for particulate filtration, etc.). The device's performance is measured against these technical specifications, not against expert human interpretations or clinical outcomes in the way an AI diagnostic tool would be.
8. The sample size for the training set:
- Not Applicable. This is not an AI/ML device; there is no training set.
9. How the ground truth for the training set was established:
- Not Applicable. This is not an AI/ML device; there is no training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
June 11, 2021
Chongqing Chaoke Industry Development Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K202714
Trade/Device Name: Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX, Dated: April 30, 2021 Received: May 14, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K202714
Device Name Medical Face Mask
Indications for Use (Describe)
The Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary (K202714)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
1.0 submitter's Information
Name: CHONGQING CHAOKE INDUSTRY DEVELOPMENT CO.,LTD. Address: C31, Yangtze River Industry Park, Tea Garden New Zone, Nan'an District, Chongqing, China 400067 Tel: 86-23-62809222 Contact: Liu Xincheng Date of Preparation: Apr.30,2020
Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device Information
Trade name: Medical Face Mask Common name: Surgical Face Mask Classification name: Surgical Face Mask Model: CK-MFM-002
3.0 Classification
Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Panel: Surgical Apparel
4.0 Predicate Device Information
Manufacturer: WUHAN DYMEX HEALTHCARE CO., LTD. SURGICAL FACE MASK Device: 510(k) number: K182515
{4}------------------------------------------------
5.0 Device Description
The Medical Face Masks are single use, three-layer, flat-pleated style with ear loops and nose piece. The Medical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of Galvanized iron wire+ Polyethylene(PE). The medical face masks will be provided in blue. The medical face masks are sold non-sterile and are intended to be single use, disposable devices.
6.0 Indication for Use Statement
The Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
7.0 Comparison to the Predicate Device
| Item | Subject DeviceK202714 | Predicate DeviceK182515 | Remark |
|---|---|---|---|
| Product Name | Medical Face Mask | Surgical Face Mask | -- |
| Product Code | FXX | FXX | Same |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Intended Use&Indications for use | The Medical Face Mask isintended to be worn toprotect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial. These facemasks are intended for usein infection controlpractices to reduce thepotential exposure to blood | The Surgical FaceMasks are intended tobe worn to protect boththe patient andhealthcare personnelfrom transfer ofmicroorganisms,bodyfluids and particulatematerial. These facemasks are intended foruse in infection control | Same |
Table 1 General Comparison
{5}------------------------------------------------
| and body fluids. This is asingle use, disposabledevice(s),providednon-sterile. | practices to reduce thepotential exposure toblood and body fluids.This is a single use,disposable device(s),provided non-sterile. | |||
|---|---|---|---|---|
| Design features | Ear Loops, 3 layers | Ear Loops, 3 layers | Same | |
| Mask Styles | Flat pleated | Flat pleated | Same | |
| Material | Outerfacinglayer | Spun-bond polypropylene | Spun-bondpolypropylene | Same |
| Middlelayer | Melt blown polypropylenefilter | Melt blownpolypropylene filter | Same | |
| InnerFacinglayer | Spun-bond polypropylene | Spun-bondpolypropylene | Same | |
| Nosepiece | Galvanized iron wire+Polyethylene(PE) | Malleable polyethylenewire | Different* | |
| Earloops | not made with naturalrubber latex | not made with naturalrubber latex | Same | |
| Color | Blue | Blue | Same | |
| Dimension(Length) | 175 mm +/- 8mm | 17.5 cm +/- 1cm | Same | |
| Dimension (Width) | 95 mm +/- 4mm | 9.5 cm +/- 1cm | Same | |
| OTC use | Yes | Yes | Same | |
| Sterility | Non-Sterile | Non-Sterile | Same | |
| Single Use | Yes | Yes | Same | |
| Sterile | No | No | Same | |
| ASTM F2100 Level | Level 2 | Level 2 | Same |
*The difference in the materials does not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials.
8.0 Non-Clinical Test Conclusion
The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Medical face masks - Premarket Notification [510(k)] Submission issued on March 5, 2004.
Table 2 - Performance Testing
| Item | Subject DeviceK202714 | Acceptance Criteria | Result |
|---|---|---|---|
| ------ | --------------------------- | --------------------- | -------- |
{6}------------------------------------------------
| FluidResistancePerformanceASTM F1862 | 32 Out of 32 passat 120 mmHg(16.0 kPa) | 29 Out of 32 pass at120 mmHg (16.0 kPa) | Pass |
|---|---|---|---|
| ParticulateFiltrationEfficiencyASTM F2299 | 99.7% | ≥ 98% | Pass |
| BacterialFiltrationEfficiencyASTM F2101 | 98.7% | ≥ 98% | Pass |
| DifferentialPressure(Delta P) MILM-36954C | 1.98 mmH20/cm² | < 6.0 mmH20/cm² | Pass |
| Flammability16 CFR 1610 | Class 1 | Class 1 | Pass |
Table 3 - Biocompatibility Testing
| Item | Subject DeviceK202714 | Result |
|---|---|---|
| Cytotoxicity | Under the conditions of the study, the device is noncytotoxic. | Pass |
| Irritation | Under the conditions of the study, the device is nonirritating. | Pass |
| Sensitization | Under the conditions of the study, the device is nonsensitizing | Pass |
9.0 Clinical Test Conclusion
No clinical study is included in this submission.
10.0 Conclusion
The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K182515.
§ 878.4040 Surgical apparel.
(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.