K Number
K202628
Date Cleared
2021-03-19

(189 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Disposable Medical Face Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Disposable Medical Face Masks are manufactured with three layers. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric with blue color. The middle layer with filtration function is made of melt blown polypropylene (PP) non-woven fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric with white color. The Disposable Medical Face Masks are single use, disposable device, provided non-sterile.

AI/ML Overview

The provided text is a 510(k) Premarket Notification from the FDA for a Disposable Medical Face Mask. It outlines the device's characteristics, comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.

Based on the provided information, I can answer some of your questions, but many of the specific details you requested (especially regarding AI/ML models, expert ground truth, and human reader studies) are not applicable as this document pertains to a physical medical device (a face mask) and not a software-based diagnostic or AI-powered system.

Here's the breakdown:

Acceptance Criteria and Device Performance (for a Medical Face Mask)

The acceptance criteria are essentially the performance requirements for medical face masks as defined by the listed ASTM and EN standards. The "device performance" is whether the mask meets these criteria.

1. A table of acceptance criteria and the reported device performance

TestAcceptance Criteria (typically from ASTM F2100-19 for Medical Face Masks)Reported Device Performance (Proposed Device K202628)Predicate Device Performance (K173062)
Fluid Resistance (ASTM F1862)Pass at 120mmHg (Level 1, 2 or 3)Pass at 120mmHgPass at 120mmHg
Bacterial Filtration Efficiency (BFE) (ASTM F2101)≥98% (Level 2 or 3)≥98%Average 99.4%
Particulate Filtration Efficiency (PFE) (ASTM F2299)≥98% (Level 2 or 3 for 0.1µm particles)≥98%Average 99.74% at 0.1µm
Differential Pressure (Delta P) (EN 14683)<6.0 mmH2O/cm²<6.0 mmH2O/cm²Average 2.7 mmH2O/cm²
Flammability (16 CFR 1610)Class 1 (non-flammable)Class 1 non-flammableClass 1 non-flammable
Biocompatibility (ISO 10993 series)Non-cytotoxic, non-irritating, non-sensitizingConfirm to requirements of ISO 10993 series standards (non-cytotoxic, non-irritating, non-sensitizing)Confirm to requirements of ISO 10993 series standards (non-cytotoxic, non-irritating, non-sensitizing)

Study Proving Device Meets Acceptance Criteria:

The study refers to the non-clinical testing conducted.

2. Sample sized used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the number of mask samples tested for each criterion (e.g., how many masks were tested for BFE). This level of detail is typically in the test reports, which are not part of this summary.
  • Data Provenance: The tests were conducted by the manufacturer, Qinghai Zhong Dao Win-win Medical Protective Equipment Co., Ltd. The country of origin for the data is China, as this is where the manufacturer is based and presumably where the testing was performed. The data is prospective in the sense that the tests were performed on newly manufactured devices specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This is a physical device (mask), not an AI/ML model for image analysis or diagnosis. The "ground truth" is established by direct physical or biological testing against a standard, not by expert consensus on observational data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. As above, the "truth" is determined by standardized test procedures and measurements, not by expert adjudication of interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical device, not an algorithm. The performance metrics are "standalone" in that they are direct measurements of the mask's physical and biological filtration properties.

7. The type of ground truth used:

  • Analytical/Empirical Data derived from Standardized Test Methods: The "ground truth" for the mask's performance (e.g., BFE, PFE, fluid resistance) is the result obtained from specific, validated laboratory test methods (e.g., ASTM F2101 for BFE, ASTM F1862 for fluid resistance). These methods have defined procedures and measurement techniques to determine the mask's properties.

8. The sample size for the training set:

  • Not Applicable. This is not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established:

  • Not Applicable. No training set is used for this type of device.

In summary, the document describes a successful 510(k) submission for a disposable medical face mask, demonstrating its substantial equivalence to a legally marketed predicate device through a series of non-clinical, standardized performance tests. The concepts of AI/ML model evaluation, expert ground truth, and human reader studies are entirely outside the scope of this particular device's regulatory review.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 19, 2021

Qinghai Zhong Dao Win-win Medical Protective Equipment Co., % Shelley Li Director Shanghai Landlink Medical Information Technology Co., Ltd. Room 703, 705, Baohua International Plaza, West Guangzhong Road 555, Jingan Shanghai, 200071 China

Re: K202628

Trade/Device Name: Disposable Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 5, 2021 Received: February 16, 2021

Dear Shelley Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202628

Device Name Disposable Medical Face Mask

Indications for Use (Describe)

The Disposable medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This is a single use, disposable device(s), provided nonsterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) summary: K202628

I Submitter

Qinghai Zhong Dao Win-win Medical Protective Equipment Co., Ltd. No.36 Jingsan Road, Biological Park, Xining City 810000, Qinghai Province, China Establishment Registration Number: 3016998435 Contact person: Zhaochu Chen Position: QC Manager Tel.: +86-13381230018 Fax: ++086-010-67083032 E-mail: chenzhaochu@jingzhuzangyao.com

Preparation date: March 17, 2021

II Proposed Device

Trade Name of Device:Disposable Medical Face Mask
Regulation Name:Surgical Apparel
Regulation Number:21 CFR 878.4040
Regulatory Class:Class II
Product code:FXX
Review Panel:General Hospital

III Predicate Devices

510(k) Number:K173062
Trade name:Non Woven Face Mask (Models: VQN0185W (earloop) and VQN0185B (ties)
Regulation Name:Surgical Apparel
Classification:Class II
Product Code:FXX
ManufacturerV&Q Manufacturing Corporation

IV Device description

The Disposable Medical Face Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Disposable Medical Face Masks are manufactured with three layers. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric with blue color. The middle layer with filtration function is made of melt blown

{4}------------------------------------------------

polypropylene (PP) non-woven fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric with white color.

The Disposable Medical Face Masks are single use, disposable device, provided non-sterile.

V Indication for use

The Disposable medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This is a single use, disposable device(s), provided non-sterile.

ItemProposed device(K202628)Predicate device(K173062)Remark
Product nameDisposable Medical Face MaskNon Woven Face Mask(Models: VQN0185W(earloop) and VQN0185B(ties)/
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassClass IIClass IISame
Mask styleFlat-pleated, ear loop, 3layersFlat-pleated, ear loop, 3LayersSame
Indication for useThe Disposable medicalFace Masks are intendedto be worn to protect boththe patient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. This is a singleuse, disposable device(s),provided non-sterile.Non Woven Face Mask(Models: VQN0185W(earloop) and VQN0185B(ties) is intended forsingle use by operatingroom personnel and othergeneral healthcareworkers to protectboth patients andhealthcare workersagainst transfer ofmicroorganisms, bloodand body fluids, andparticulate material.Same
MaterialInnerSpun-bond polypropyleneSpun-bond polypropyleneSame
layer
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Outer layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Ear loopPolyamide fiber+PolyurethaneUrethane elastic fiber orspun-bond polypropyleneDifferent1
Nose pieceIron wire covered bypolyethyleneWhite aluminum stripcovered by PP coveringDifferent1
ColorBlueBlueSame
Length17.5cm17.5cmSame
Width9.5cm9.5cmSame
OTC useYesYesSame
sterileNon-sterileNon-sterileSame
Single for useYesNon-sterileSame
LatexNot made with naturalrubber latexNot made with naturalrubber latexSame
Fluid ResistancePerformanceASTM F1862Pass at 120mmHgpass at 120mmHgSimilar 2
ParticulateFiltrationEfficiencyASTM F2299≥98%Average 99.74% at 0.1µm
Bacterial FiltrationEfficiencyASTM F2101≥98%Average 99.4%
DifferentialPressureEN 14683<6.0mmH2O/cm²Average 2.7 mmH2O/cm²
Flammability16 CFR 1610Class I non flammableClass 1Non Flammable
BiocompatibilityConfirm to therequirements of ISOConfirm to therequirements of ISOSame
10993 series standardsnon-cytotoxicnon-irritatingnon-sensitizing10993 series standardsnon-cytotoxicnon-irritatingnon-sensitizing

VI Comparison of technological characteristics with the predicate devices

{5}------------------------------------------------

{6}------------------------------------------------

Analysis:

1 The difference in the materials and colors does not raise additional questions for safety and effectiveness of the device. The biocompatibility evaluation test of the subject devices have been performed on the final finished device which includes all construction materials and color additives. The test results shows pass the requirements.

2 The performance of proposed device have been performed the performance test according to method given in the ASTM F2100-19. The test results demonstrate that met the requirements in the standards. The minor difference between the proposed and predicate device does not affect the safety and effectiveness of the device.

VII Non-Clinical Testing

Non clinical tests were conducted to verify that the proposed device met all design specifications as the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For . In Vitro Cytotoxicity
  • ISO 10993-10: 2010 Biological Evaluation Of Medical Devices Part 10: Tests . For Irritation And Skin Sensitization.
  • ASTM F2100-19, Standard Specification for Performance of Materials Used In ● Disposable Surgical Mask.
  • ASTM F1862M-17, Standard Test Method For Resistance Of Disposable . Surgical Mask To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity)
  • EN 14683:2019 Medical Face Masks-Requirements and Test Methods ●
  • ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration . Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus;
  • ASTM F2299-03, Stand test method for determining the initial efficiency of . materials used in Disposable Surgical Mask to penetration by particulates using latex spheres;
  • 16 CFR 1610, Standard for the Flammability of clothing textiles; .

VIII Clinical Testing

{7}------------------------------------------------

No clinical study is included in this submission.

IX Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device identified.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.