AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The disposable surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device provided non-sterile.

Device Description

The 3-ply surgical mask is non-sterile, disposable, single-use surgical mask. It is manufactured with three layers, the inner and outer layers are made of non-woven, spun-bond polypropylene, while the middle filter layer is made of a non-woven, melt-blown polypropylene.

AI/ML Overview

The provided text describes the acceptance criteria and performance study for a Surgical Mask - Level 3 & Procedure Mask - Level 3. This is a medical device (PPE), not an AI/software device, so the requested information regarding AI-specific studies (experts establishing ground truth, MRMC studies, standalone algorithm performance, training set details) is not applicable to this submission.

Here is the information directly relevant to the acceptance criteria and study for this physical device based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Device Name: Surgical Mask - Level 3 & Procedure Mask - Level 3 (K202607)

ItemAcceptance CriteriaReported Device Performance (Subject Device)Predicate Device PerformanceResult
Level 3 Fluid Resistance (ASTM F1862)29 Out of 32 pass at 160 mmHg32 Out of 32 pass at 160 mmHg32 Out of 32 pass at 160 mmHgPass
Particulate Filtration Efficiency (ASTM F2299)≥ 98%99.67-99.80%≥99%Pass
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%99.6 - 99.9%≥99%Pass
Differential Pressure (ASTM F2100)< 6.0 mmH₂0/cm²4.8-5.0 mmH₂0/cm²Avg. 3.6 mmH₂0/cm²Pass
Class 1 Flammability (16 CFR 1610)< 3.5 second burn timeDid not ignite or Ignited but ExtinguishedClass 1Pass
Cytotoxicity (ISO 10993-5)Non-cytotoxicUnder the conditions of the study the device is non-cytotoxicNon-cytotoxicPass
Skin Sensitization (ISO 10993-10)Non-sensitizingUnder the conditions of the study the device is non-sensitizingNon-sensitizingPass
Skin Irritation (ISO 10993-10)Non-irritatingUnder the conditions of the study the device is non-irritatingNon-irritatingPass

2. Sample Size Used for the Test Set and Data Provenance

The sample size is explicitly stated for the Level 3 Fluid Resistance test: 32 samples (32 Out of 32 pass). For the other tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability, Biocompatibility), specific sample sizes are not explicitly provided in the summary, but the results are reported as meeting the criteria, implying sufficient samples were tested according to the respective standards.

Data Provenance: The data comes from non-clinical benchtop performance testing and biocompatibility testing. The country of origin of the data is not specified, but the submission is to the U.S. FDA, implying testing conducted to U.S. and international standards (ASTM, ISO). The testing is implicitly prospective, as it's conducted to demonstrate meeting the criteria for regulatory submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not Applicable. This is a physical medical device (surgical mask), not an AI/software device that requires human experts to establish ground truth for image interpretation or diagnosis. The "ground truth" for this device is established through standardized laboratory testing protocols of its physical properties and biological compatibility.

4. Adjudication Method for the Test Set

Not Applicable. As this is a physical device tested against established engineering and biological standards, "adjudication" in the sense of expert review for ambiguous cases is not relevant. The results are quantitative measurements against defined thresholds.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. MRMC studies are typically performed for AI or diagnostic devices where human reader performance is a key metric. This is a physical protective device. The study performed was non-clinical benchtop testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not Applicable. This is a physical medical device. There is no algorithm to assess.

7. The Type of Ground Truth Used

The "ground truth" for this device is based on objective measurements from standardized laboratory tests according to recognized standards (e.g., ASTM F1862, ASTM F2299, ASTM F2101, ASTM F2100, 16 CFR 1610, ISO 10993-5, ISO 10993-10). These standards define the test methods and the performance criteria for surgical masks.

8. The Sample Size for the Training Set

Not Applicable. This is a physical medical device. There is no "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

Not Applicable. As there is no training set, there is no ground truth to establish for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

April 23, 2021

String King Lacrosse, LLC % Thomas Frasca President 19100 S Vermont Ave. Gardena, CA, 90248 USA

Re: K202607

Trade/Device Name: Surgical Mask - Level 3 & Procedure Mask - Level 3 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 9, 2020 Received: September 9, 2020

Dear Thomas Frasca:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Ryan Ortega, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202607

Device Name Surgical Mask - Level 3 & Procedure Mask - Level 3

Indications for Use (Describe)

The disposable surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device provided non-sterile.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY - K202607

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

  1. SUBMITTER String King Lacrosse, LLC 19100 S Vermont Ave. Gardena, CA. 90248 USA Tel: +1.508.654.1988 Fax: N/A Contact Person: Thomas Frasca

April 22, 2021

II. DEVICE
Name of Device:Surgical Mask – Level 3 & Procedure Mask – Level 3
Classification Name:Surgical Mask
Regulation:21 CFR § 878.4040
Regulatory Class:Class II
Product Classification Code:FXX

PREDICATE DEVICE |||.

Date Prepared:

Predicate Manufacturer:DemeTECH Corporation
Predicate Trade Name:DemeMASK
Predicate 510(k):K201479

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The 3-ply surgical mask is non-sterile, disposable, single-use surgical mask. It is manufactured with three layers, the inner and outer layers are made of non-woven, spun-bond polypropylene, while the middle filter layer is made of a non-woven, melt-blown polypropylene.

V. INDICATIONS FOR USE

The disposable surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device provided non-sterile.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE

The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence in Table 1 below.

{4}------------------------------------------------

DeviceProposed Device -Surgical Mask - Level 3& Procedure Mask -Level 3 (K202607)Predicate Device -DemeMASK(K201479)Result
ManufacturerStringKingDemeTECH CorporationN/A
510K NumberK202607K201479N/A
Product CommonNameSURGICAL FACE MASKSURGICAL FACE MASKIdentical
Product CodeFXXFXXIdentical
ClassificationClass II (21 CFR 878.4040)Class II (21 CFR 878.4040)Identical
Intended UseThe disposable surgical facemasks are intended to beworn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, body fluidsand particulate material.These face masks areintended for use in infectioncontrol practices to reducethe potential exposure toblood and body fluids. Thisis a single-use, disposabledevice provided non-sterile.The Disposable SurgicalFace Masks are intended tobe worn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, body fluidsand particulate material.These face masks areintended for use in infectioncontrol practices to reducethe potential exposure toblood and body fluids. Thisis a single use, disposabledevice provided non-sterile.Identical
Model3 Ply, Ear Loops, Flat-PleatedStyle3 Ply, Ear Loops or Tie-OnNo tie-onoption forsubject device
Materials
Outer FacingLayerSpun-Bond Polypropylene(same material as inner layer)Spunbond PolypropyleneSame
Middle LayerMelt-blown PolypropyleneMelt Blown PolypropyleneFilterSame
Inner Facing LayerSpun-Bond PolypropyleneSpunbond PolypropyleneSame
Nose PieceLDPE cover and galvanized wirecoreSingle Galvanize Wire, CoatedBy PESame
Ear LoopsNylon, Spandex, Polyesternot made with natural rubber latexSame
ColorBlueWhiteDifferent, performance supported by biocompatibility and performance test data
Dimension (Width)9.5 cm ± 0.5 cm9.5 cm ± 1.0 cmSimilar
Dimension (Length)17.5 cm ± 1 cm17.5 cm ± 1.0 cmSame
OTC UseYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle UseSingle UseSame
ASTM F2100 LevelLevel 3Level 3Same, see Table 2
Biocompatibility ISO 10993-5 and ISO 10993-10Non-cytotoxic, nonsensitizing, and nonirritatingNon-cytotoxic, nonsensitizing, and nonirritatingSame, see Table 3

Table 1 – Comparison of Technological Characteristics

{5}------------------------------------------------

{6}------------------------------------------------

VII. NON-CLINICAL DATA

The subject device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004. A summary of the benchtop performance testing results is provided below in Table 2 along with the acceptance criteria and the predicate device's reported results.

ItemProposed Device -Surgical Mask - Level 3 &Procedure Mask - Level 3Predicate Device -DemeMASK (K201479)Acceptance CriteriaSubjectDeviceResult
Level 3 FluidResistancePerformanceASTM F186232 Out of 32 pass at 160 mmHg32 Out of 32 pass at160 mm Hg29 Out of 32 pass at 160 mmHgPass
ParticulateFiltrationEfficiencyASTM F229999.67-99.80%≥99%≥ 98%Pass
BacterialFiltrationEfficiencyASTM F210199.6 - 99.9%≥99%≥ 98%Pass
DifferentialPressureASTM F21004.8-5.0 mmH₂0/cm²Avg. 3.6 mmH₂0/cm²< 6.0 mmH₂0/cm²Pass
Class 1Flammability16 CFR 1610Did not ignite or Ignited butExtinguishedClass 1< 3.5 second burn timePass

Table 2 - Benchtop Performance Testing

Sterilization & Shelf-life Testing

Not Applicable (This is a non-sterile device and shelf-life is not applicable to this device because of low likelihood of time-dependent product degradation.)

Biocompatibility Testing

Biocompatibility testing was performed in accordance with ISO 10993-1:2018. Specifically, the following testing endpoints were evaluated.

{7}------------------------------------------------

StandardsProposed Device -Surgical Mask - Level 3 &Procedure Mask - Level 3Predicate Device -DemeMASK (K201479)Result
Cytotoxicity -ISO 10993-5Under the conditions of thestudy the device is non-cytotoxicUnder the conditions of thestudy the device is non-cytotoxicPass
SkinSensitization -ISO 10993-10Under the conditions of thestudy the device is non-sensitizingUnder the conditions of thestudy the device is non-sensitizingPass
Skin Irritation -ISO 10993-10Under the conditions of thestudy the device is non-irritatingUnder the conditions of thestudy the device is non-irritatingPass

Table 3 - Biocompatibility Testing

Software Verification and Validation Testing

Not Applicable (Passive Device)

Electrical safety and electromagnetic compatibility (EMC)

Not Applicable (Passive Device)

Human Clinical Performance Testing

Clinical testing was not required to demonstrate the safety and effectiveness of the device.

VIII. CONCLUSIONS

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well or better than the legally marketed predicate device (K201479), manufactured by DemeTECH Corporation.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.