K Number
K202565
Date Cleared
2020-12-02

(89 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When properly worn, the surgical face masks are intended to patient and healthcare workers from transfer of microorganisms, body fluids and airborne particles. This device is non-sterile and for single use only.

Device Description

The surgical face mask is a non-sterile, single use, three-layer, flat, pleated style with ear loops and a nose piece. The outer and inner layers are spunbond polypropylene, and the middle layer consists of melt-blown polypropylene filter. Each mask is secured to the face with ear loops with a nosepiece to conform the mask to the shape of the nose.

AI/ML Overview

This document describes the validation of a surgical face mask. As such, the concept of "AI assistance for human readers" or "ground truth established by experts with many years of experience" is not applicable in the context of device performance testing for such simple medical devices. The evaluations are based on standardized physical and biological tests against established criteria.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Fluid Resistance (ASTM F1862)29 out of 32 pass at 120 mm Hg31 out of 32 pass at 80 mm Hg30 out of 32 pass at 120 mm Hg
Particulate Filtration Efficiency (ASTM F2299)≥ 98%98.8%
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%> 99.9%
Differential Pressure (MIL- M-36954C)< 5.0 mm H20/cm²4.4 mmH20/cm²
Flammability (16 CFR 1610)Class 1Class 1
BiocompatibilityNon-Cytotoxic, Non-Sensitizing, Non-IrritatingNon-Cytotoxic, Non-Sensitizing, Non-Irritating

2. Sample Size Used for the Test Set and Data Provenance

  • Fluid Resistance (ASTM F1862): 32 samples were tested for fluid resistance.
  • Other Tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability, Biocompatibility): The document does not explicitly state the sample size for these specific tests, but performance data is provided.
  • Data Provenance: The tests were conducted on the "proposed device," which implies new testing of the specific product being submitted for clearance. The document does not specify the country of origin for the testing itself, but the manufacturer is Ningbo Green Textile Co., Ltd, which is likely in China. The data would be considered prospective for the purposes of this submission as it was generated specifically for the device's clearance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable for this type of device and testing. The "ground truth" is established by adherence to standardized testing protocols (e.g., ASTM, MIL, ISO, CFR standards) with measurable outcomes, not by expert consensus or interpretation of medical images.

4. Adjudication Method for the Test Set

Not applicable. Testing is against objective, measurable criteria defined by international and national standards.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a surgical face mask, not an AI-powered diagnostic tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a surgical face mask, not an AI algorithm.

7. The Type of Ground Truth Used

The ground truth is based on standardized physical and biological testing methods and their predefined acceptance criteria. These standards include:

  • ASTM F2100: Standard Specification for Performance of Materials Used in Medical Face Masks
  • ASTM F1862/F1862M-17: Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood
  • ASTM F2299/F2299M-03: Determining the Initial Efficiency of Materials Used in Medical Face Masks to the Penetration of Particulates Using Latex Spheres
  • ASTM F2101: Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials
  • MIL - M-36954C: Medical face masks - Requirements and test methods, Annex C
  • 16 CFR 1610: Flammability Test Method
  • ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for In Vitro Cytotoxicity
  • ISO 10993-10: Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization

8. The Sample Size for the Training Set

Not applicable. This device is a physical medical device (surgical face mask), not a machine learning model; therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. No training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 1, 2020

Base4 Group, Inc. % Thomas Knott Senior Regulatory Advisor Benjamin L. England and Associates, LLC 810 Landmark Dr, Suite 126 Glen Burnie, Maryland 21061

Re: K202069

Trade/Device Name: BASE4 Disposable Medical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 2, 2020 Received: November 2, 2020

Dear Thomas Knott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202069

Device Name Surgical Face Mask

Indications for Use (Describe)

When properly worn, the surgical face masks are intended to patient and healthcare workers from transfer of microorganisms, body fluids and airborne particles. This device is non-sterile and for single use only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Date of Summary Prepared: November 25, 2020 510(k) Number: K202069

Submitted By:Benjamin L. England And Associates, LLC
Address:810 Landmark DriveSuite 126Glen Burnie, MD 21061
Phone:(410) 220-2800
Fax:(443) 583-1464
E-mail:tcknott@englandlawgroup.com
Contact:Thomas C. Knott, Senior Regulatory Consultant
On Behalf of:BASE4 Group, Inc.
Address4393 Sunbelt DriveAddison, TX 75001
Phone:(972) 331-6619
Fax:
E-Mail:Jon@ebase4.com
Contact:Jon Filipski

1. Regulatory Information

Proprietary Name:BASE4 Disposable Medical Mask (NB Green2020-12; ModelNumber is 07787)
Classification Name:Surgical Face Mask, Apparel
Common Name:Surgical face mask
Classification:Class II
Product Code:FXX
PanelGeneral Surgery Devices
Regulation:21 CFR 878.4040

2. Predicate Device

510(k) Number:K200847
Device Name:Mask, Surgical Apparel

3. Indications for Use

When properly worn, the surgical face masks are intended to protect both patient and healthcare workers from transfer of microorganisms, body fluids and airborne particles. This device is non-sterile and for single use only.

{4}------------------------------------------------

The surgical face mask is a non-sterile, single use, three-layer, flat, pleated style with ear loops and a nose piece. The outer and inner layers are spunbond polypropylene, and the middle layer consists of melt-blown polypropylene filter. Each mask is secured to the face with ear loops with a nosepiece to conform the mask to the shape of the nose.

5. Summary of Comparison and Technological Characteristics

Table 1 - General Comparison

  1. Device Description
CLASSIFICATION INFORMATION
510(k) NumberK202069K200847N/A
ManufacturerNingbo Green Textile Co., LtdMexpo International, Inc.N/A
CommonNameSurgical Face MaskSurgical Face MaskSame
Product CodeFXXFXXSame
ClassificationClass IIClass IISame
Regulation21 CFR 878.404021 CFR 878.4040Same
Intended useWhen properly worn, thesurgical face masks areintended to protect both patientand healthcare workers fromtransfer of microorganisms,body fluids and airborneparticles. This device is non-sterile and for single use onlyWhen properly worn, the surgicalface masks are intended toprotect both patient andhealthcare workers from transferof microorganisms, body fluidsand airborne particles. Thisdevice is non-sterile and forsingle use onlySame
ModelThree ply, flat pleated, earloops3 Ply, Ear Loops, Flat-PleatedStyleSame
TensileStrength10 – 15 NN/AN/A
MATERIALS
Outer LayerSpunbond PolypropyleneSpunbond PolypropyleneSame
Middle LayerMelt Blown PolypropyleneFilterMelt Blown PolypropyleneFilterSame
Inner LayerSpunbond PolypropyleneSpunbond PolypropyleneSame
Nose PieceGalvanized wire coated by PVCSingle Galvanize Wire, CoatedBy PESimilar
Ear Loops70% polyester; 30% spandexnot made with natural rubberlatexUnknown
ColorBlue outer layerWhiteDiffers
Width$95 mm = 9.5 cm$$9.0cm \pm 0.5cm$Same
Length$175 mm = 17.5 cm$$17.5cm \pm 0.5cm$Same
OTC Use?YesYesSame
SterilityNon-SterileNon-SterileSame
Single Use?YesYesSame
ASTM F2100Level 2Level 2Same

{5}------------------------------------------------

ASTM F1862FluidResistancePerformance31 out of 32 pass at 80 mm Hg30 out of 32 pass at 120 mm Hg30 Out of 32 pass at 120 mm HgSame
ASTM F2299ParticulateFiltrationEfficiency98.8%99.9%Both PassSimilar
ASTM F2101BacterialFiltrationEfficiency> 99.9%> 99.9%Same
DifferentialPressure (DeltaP) MIL- M-36954C4.4 mmH20/cm²3.0 mmH20/cm²Both PassSimilar
Flammability16 CFR 1610Class 1Class 1Same
BiocompatibilityTestingNon-Cytotoxic, Non-Sensitizing, Non-IrritatingNon-Cytotoxic, Non-Sensitizing, Non-IrritatingSame

The outer layer of the subject mask is colored blue, and the predicate device is white. Otherwise, they are the same material, polypropylene. The outer layer does not contact the patient's skin and therefore presents minimal risk to the health of the user.

The nose piece for both devices is galvanized wire. The subject device is covered by PVC and the predicate device is covered by PE. However, neither material will contact the user's skin so there is no difference in risk to the health of the user.

The 510(k) Summary for the predicate device does not show the material for the ear loops. Therefore, no comparison can be made.

The Particulate Filtration Efficiency and Differential Pressures are similar for both masks and pass the specification required by ASTM F2100 for a level 2 mask.

6. Non-clinical Tests Performed on the Proposed Device

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification [510(k)] Submission issued on March 5, 2004.

  • . ASTM F2100: Standard Specification for Performance of Materials Used in Medical Face Masks
  • . ASTM F1862/F1862M-17: Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
  • . ASTM F2299/F2299M-03: Determining the Initial Efficiency of Materials Used in Medical Face Masks to the Penetration of Particulates Using Latex Spheres
  • . ASTM F2101: Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
  • . MIL - M-36954C: Medical face masks - Requirements and test methods, Annex C

{6}------------------------------------------------

  • 16 CFR 1610 CPSC CS-191-53 Flammability Test Method (16 CFR 1610) Standard for ● Flammability of Clothing Textiles.
  • ISO 10993-5: Biological evaluation of medical devices-Part 5: Tests for In Vitro ● Cytotoxicity
  • ISO 10993-10: Biological evaluation of medical devices Part 10:Tests for irritation and skin sensitization
Table 2 - Performance Testing
TestProposed DeviceAcceptance CriteriaResult
FluidResistancePerformanceASTM F186231 out of 32 pass at 80 mm Hg30 out of 32 pass at 120 mm Hg29 out of 32 pass at 120 mmHgPassLevel 2
ParticulateFiltrationEfficiencyASTM F229998.8%≥ 98%Pass
BacterialFiltrationEfficiencyASTM F2101> 99.9%≥ 98%Pass
DifferentialPressure (DeltaP) MIL- M-36954C4.4 mmH20/cm²< 5.0 mm H20/cm²Pass
Flammability16 CFR 1610Class 1Class 1Pass

Table 3 - Biocompatibility Testing

ItemProposed DeviceAcceptance CriteriaResult
ResultsNon-Cytotoxic, Non-Sensitizing, Non-IrritatingNon-Cytotoxic, Non-Sensitizing, Non-IrritatingPass

7. Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.

END OF 510(k) SUMMARY*

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).