K Number
K202552
Date Cleared
2020-11-02

(60 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Midas Rex™ attachments and dissecting tools for Mazor are indicated for the incision/cutting, drilling, burring, and removal of hard tissue and bone in open and minimally invasive spine procedures.

Computer-assisted surgery and its associated applications are intended as an aid for locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition for which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a long bone or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.

Device Description

Midas Rex™ attachments and surgical dissecting tools for Mazor are intended for the incision/cutting, drilling, burring, and removal of hard tissue and bone. The subject instruments are designed to be utilized through a cannula, for use with the Mazor X Stealth Edition™ system in open and minimally invasive spine procedures.

The subject devices perform the intended function as part of the existing Medtronic Midas Rex TM surgical drill systems which consist of the subject attachments and dissecting tools, electric and pneumatic drill handpieces and system accessories.

AI/ML Overview

The provided text describes the acceptance criteria and a study for the "Midas Rex Attachments and Dissecting Tools" which are intended for surgical navigation.

Here's the detailed information based on your request:

1. Table of Acceptance Criteria and the Reported Device Performance

Acceptance CriteriaReported Device Performance
Mean Positional Error (mm): ≤ 2.0 mmMean Positional Error (mm): 1.12 mm
Mean Trajectory Error (degrees): ≤ 2.0 degreesMean Trajectory Error (degrees): 0.37 degrees

2. Sample size used for the test set and the data provenance

The document indicates that "Navigational Accuracy Analysis" was performed, which "Provides confirmation that the subject devices satisfy the necessary navigational accuracy requirements." However, the specific sample size used for this test set is not explicitly mentioned in the provided text.

Regarding data provenance, the study appears to be an internal performance verification study conducted by Medtronic Powered Surgical Solutions (MPSS) to demonstrate the device's adherence to specified accuracy metrics. The data would therefore be prospective in the context of this regulatory submission, generated specifically for the purpose of demonstrating substantial equivalence. The document doesn't specify the country of origin of the data beyond the manufacturer's location in Fort Worth, Texas.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number of experts used or their qualifications for establishing ground truth for the navigational accuracy analysis. The "Navigational Accuracy Analysis" is likely a technical measurement against a known reference, rather than an expert-adjudicated ground truth like in image interpretation studies.

4. Adjudication method for the test set

No adjudication method (e.g., 2+1, 3+1, none) is mentioned or implied for the navigational accuracy analysis. This type of test typically involves direct measurement against a known physical standard or simulated environment, rather than expert consensus on subjective interpretations.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No multi-reader multi-case (MRMC) comparative effectiveness study was done, as this device is a surgical tool with navigation capabilities, not an AI-assisted diagnostic or interpretive system involving human readers. The device itself provides assistance to the surgeon in locating anatomical structures.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance evaluation was conducted for the device's navigational accuracy. The "Navigational Accuracy Analysis" measured the device's performance (Mean Positional Error and Mean Trajectory Error) against predefined acceptance criteria, independent of a human operator actively engaged in a clinical task. This assessment is purely about the device's technical specifications.

7. The type of ground truth used

For the "Navigational Accuracy Analysis," the ground truth is based on technical specifications and precision measurements. The comparison is against a known, presumably highly accurate, reference or measurement standard that determines what constitutes "0 mm error" and "0 degrees error." It is not based on expert consensus, pathology, or outcomes data in the typical sense of a diagnostic device.

8. The sample size for the training set

The provided document describes a medical device (surgical tools with navigation) and its substantial equivalence submission, not a machine learning or AI model that requires a training set. Therefore, there is no mention of a "training set sample size."

9. How the ground truth for the training set was established

As mentioned above, this device is not an AI/ML model and therefore does not have a "training set" or a corresponding "ground truth for the training set" in the context of typical AI/ML development.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 2, 2020

Medtronic Powered Surgical Solutions (MPSS) Jainam Shah Sr Regulatory Affairs Specialist 4620 North Beach Street Fort Worth, Texas 76137

Re: K202552

Trade/Device Name: Midas Rex Attachments and Dissecting Tools Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 2, 2020 Received: September 3, 2020

Dear Jainam Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202552

Device Name Midas RexTM attachments and dissecting tools

Indications for Use (Describe)

The Midas Rex™ attachments and dissecting tools for Mazor are indicated for the incision/outting, drilling, burring, and removal of hard tissue and bone in open and minimally invasive spine procedures.

Computer-assisted surgery and its associated applications are intended as an aid for locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition for which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a long bone or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

September 02, 2020

  • I. Company: Medtronic Powered Surgical Solutions 4620 North Beach Street Fort Worth, Texas 76137 Telephone Number: (720) 890-3200
    • Jainam Shah Contact: Sr. Regulatory Affairs Specialist Telephone Number: (720) 890-2595 Fax: (720) 890-3500 Email: jainam.shah@medtronic.com

Jenna Groves (Alternate) Regulatory Affairs Manager Telephone number: (817) 788-6686 Email: jenna.a.groves@medtronic.com

  • II. Proprietary Trade Name: Midas Rex™ Attachments and Dissecting tools
  • III. Common Name: Orthopedic Stereotaxic Instrument
  • IV. Classification Name: Stereotaxic Instrument (21 CFR 882.4560)
  • V. Classification: Class II

VI. Product Code: OLO

VII. Predicate Devices

The legally marketed predicate devices are identified below:

Predicate510(k) Clearance
Primary PredicateMR8TM Drill System and MidasRexTM MR8TM ClearViewTM ToolsK183515S.E. 12-May-2019
Additional Predicate:Stealth Midas MR8TM SystemK183644S.E. 22-May-2019

{4}------------------------------------------------

VIII. Product Description:

Midas Rex™ attachments and surgical dissecting tools for Mazor are intended for the incision/cutting, drilling, burring, and removal of hard tissue and bone. The subject instruments are designed to be utilized through a cannula, for use with the Mazor X Stealth Edition™ system in open and minimally invasive spine procedures.

The subject devices perform the intended function as part of the existing Medtronic Midas Rex TM surgical drill systems which consist of the subject attachments and dissecting tools, electric and pneumatic drill handpieces and system accessories.

IX. Indications for Use:

The Midas Rex™ attachments and dissecting tools for Mazor are indicated for the incision/cutting, drilling, burring, and removal of hard tissue and bone in open and minimally invasive spine procedures.

Computer-assisted surgery and its associated applications are intended as an aid for locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition for which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a long bone or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.

X. Comparison of the Technological Characteristics:

The subject Midas Rex™ attachments and dissecting tools utilize the same fundamental scientific technology and have the same mode of operation and functionality as the predicate attachments and surgical dissecting tools (K183515), i.e. they are powered by the existing electric/pneumatic handpieces for the intended use. The subject devices have equivalent materials and sterilization/reprocessing methods as the primary predicate (K183515). The changes in design features between the subject and the predicates do not raise any new risks or any concerns about the safety and effectiveness. The subject navigated devices have an integrated optical tracker to enable real-time navigation in a manner identical to the additional predicate Stealth Midas™ (K183644). The subject navigated devices also meet the established system-level navigation accuracy requirements.

Table 1: Comparison of technological characteristics
FeatureSubject Devices(Midas RexTMattachments andsurgical dissecting tools)Primary PredicateMidas MR8TM Drill Systemand Midas RexTM MR8TMClearViewTM Tools(K183515)Additional PredicateStealth MidasTMMR8TM(K183644)
Regulation21 CFR 882.4560(Stereotaxic Instrument)21 CFR 882.4360 (Motor,Drill, Electric)21CFR 882.4370 (Motor,Drill, Pneumatic)21 CFR 882.4560(Stereotaxic Instrument)
FeatureSubject Devices (Midas Rex TM attachments and surgical dissecting tools)Primary Predicate Midas MR8 TM Drill System and Midas Rex TM MR8 TM ClearView TM Tools (K183515)Additional Predicate Stealth Midas TM MR8 TM (K183644)
21 CFR 882.4310 (Drills, Burs, Trephines & Accessories
Product CodeOLOHBC, HBB, HBEOLO, HAW
Intended UseIncision/cutting, drilling, burring, and removal of hard tissue and bone.Incision / cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials.Incision / cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials.
Indications for UseThe Midas Rex TM attachments and dissecting tools for Mazor are indicated for the incision/cutting, drilling, burring, and removal of hard tissue and bone in open and minimally invasive spine procedures.Computer-assisted surgery and its associated applications are intended as an aid for locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition for which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a long bone or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.The Medtronic MR8 Drill System is indicated for the incision/ cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials in Neurosurgical (Cranial and Craniofacial including craniotomy); Ear, Nose and Throat (ENT), Maxillofacial, Orthopedic, Arthroscopic, Spinal, Sternotomy, and General Surgical Procedures.Additionally, the MR8 Drill System is indicated for the incision/cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures:• Lumbar Microdiscectomy• Lumbar Stenosis Decompression• Posterior Lumbar Interbody Fusion (PLIF)• Transforaminal Lumbar Interbody Fusion (TLIF)• Anterior Lumbar Interbody Fusion (ALIF)• Direct Lateral Interbody Fusion (DLIF)The Midas Rex MR8 ClearView Tools are used only in conjunction with the MR8 Drill System to performThe Stealth-Midas MR8 TM System is indicated for the drilling, burring and removal of hard tissue and bone in spinal and cranial surgical procedures.Computer-assisted surgery and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, or vertebra, can be identified relative to a CT- or MR-based model, fluoroscopic images, or digitized landmarks of the anatomy.
FeatureSubject Devices(Midas Rex™attachments andsurgical dissecting tools)Primary PredicateMidas MR8™ Drill Systemand Midas Rex™ MR8™ClearView™ Tools(K183515)Additional PredicateStealth Midas™MR8™(K183644)
Midas Rex MR8 Drill Systemand associated User's Guidesfor the Indications of Use
GeneralSystemComponents• Attachments - Guidedand Navigated w/optical tracker• Surgical DissectingTools• System Accessories• Electric Handpieces• Pneumatic Handpieces• Attachments• Surgical Dissecting Tools• System Accessories• Electric Handpiecesw/ optical tracker• Pneumatic Handpiecesw/ optical tracker• Attachments• Surgical DissectingTools• System Accessories
MaterialsAttachments: StainlessSteelTracker: Aluminum,StainlessSteelSurgical DissectingTools: Tool (M2) SteelAttachments: StainlessSteel, Aluminum, Ceramic,Phenolic, Epoxy, ChromeCoated Brass, Torlon 4301Surgical Dissecting Tools:Stainless Steel, Tool Steel,Alloy Steel, Carbide, TDCCoating, Diamond Coatingin Nickel Substrate,Titanium NitrideAttachments - StainlessSteel, Aluminum,Ceramic, Phenolic,Epoxy, Chrome CoatedBrass, PolymericSurgical DissectingTools - Stainless Steel,Tool Steel, Alloy Steel,Carbide, TDC Coating,Diamond Coating inNickel Substrate,Titanium NitrideTracker: Aluminum
SurgicalDissectingTool TipStyleTwist DrillRound/Acorn, MatchHead, Ball, Cylinder,Oval, Tapered/SideCutting, Metal Cutting, TwistDrill, Hole Maker/Saw,Reverse TaperRound/Acorn, MatchHead, Ball, Cylinder,Oval, Tapered/SideCutting, Metal Cutting,Twist Drill, HoleMaker/Saw, ReverseTaper
SurgicalDissectingTool OverallLength23cm31cm3 - 42 cm3-42 cm
SurgicalDissectingTool HeadDiameter3.0 mm0.5 - 25 mm0.5 - 25 mm
AttachmentConfigurationStraightStraight, Angled,Variable, Double-Lock,Footed, Telescoping,Perforator, Jacobs Chuck, J-Latch, Contra Angle, MetalCuttingStraight, Angled,Variable, Double-Lock,Footed, Telescoping,Perforator, JacobsChuck, J-Latch, ContraAngle, Metal Cutting
FeatureSubject Devices(Midas RexTMattachmentsandsurgical dissecting tools)Primary PredicateMidas MR8TM Drill Systemand Midas RexTM MR8TMClearViewTM Tools(K183515)Additional PredicateStealth MidasTMMR8TM(K183644)
AttachmentLength23 cm31cm2 – 40 cm2 – 40 cm
Drill SystemOperatingPrinciple• Pneumatic (poweredby Pneumatic Pressure)• Electric (poweredby IPC)• Pneumatic (powered byPneumatic Pressure)• Electric (powered by IPC)• Pneumatic (poweredby PneumaticPressure)• Electric (powered byIPC)
Packaging -SterileSurgicalDissectingToolsIndividually packaged ina clear plastic cappedtube placed within apoly-polypouchIndividually packaged in aclear plastic capped tubeplaced within a poly-polypouchThe Surgical DissectingTools are individuallypackaged in a clearplastic capped tubeplaced within a poly-polypouch.
SterilizationMethod –Single UseToolsGamma with minimumradiation dose of 25 kGyGamma with minimumradiation dose of 25 kGyGamma with minimumradiation dose of 25 kGy
SterilizationMethod -ReusableAttachmentSteamSteamSteam
Shelf Life -SurgicalDissectingTools5 Years5 Years5 Years
NavigationOpticalTrackingPrincipleOptical (Infra-red)N/AOptical (Infra-red)
System levelNavigationAccuracyAcceptanceCriteriaMean Positional Error(mm) ≤ 2.0 mmMean Trajectory Error(degrees): ≤ 2.0 degreesN/AMean Positional Error(mm) ≤ 2.0 mmMean Trajectory Error(degrees): ≤ 2.0 degrees
System levelNavigationAccuracyValuesMean Positional Error(mm):1.12 mmMean Trajectory Error(degrees):0.37 degreesN/AMean Positional Error(mm):0.94 mmMean Trajectory Error(degrees):0.85 degrees

Table 1· Comnarison of technological characteristics

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

XI. Discussion of the Performance Testing:

Testing conducted to demonstrate equivalency of the subject device to the predicate devices is summarized as follows:

Performance TestingActivityDescription
Functional andReliability DesignVerificationDemonstrates the ability of the Midas Rex™ attachment anddissecting tools to perform throughout a single procedureand throughout expected product life thus, verifying thedesign robustness.
NavigationalAccuracy AnalysisProvides confirmation that the subject devices satisfy thenecessary navigational accuracy requirements.
CAD ModelAnalysisVerifies the correctness of CAD model representations ofthe subject devices for the Mazor X Stealth Edition™system.
Human Factors/Usability TestingDemonstrates that the usability and human factorsconsiderations were adequately addressed and validated, andthe product design performs according to its intended use.

The subject devices met the appropriate requirements and passed all the performance testing activities listed above. Additionally, biological endpoint testing, conducted per recommendations from ISO 10993-1:2018, indicates that the subject devices are noncytotoxic, non-sensitizing, non- irritating, non-toxic, non-pyrogenic, non-hemolytic and pose a negligible risk of adverse biological effects to patients when used as intended.

XII. Conclusions

The subject devices have shown through comparison to be substantially equivalent to the identified predicate devices.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).