K Number
K202528
Manufacturer
Date Cleared
2020-10-26

(55 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MESA and DENALI Spinal Systems (including ARI Staples) and the EVEREST Spinal System are cleared for the following indications:

Posterior non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudoarthosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

Except for the ARI staples, the MESA, DENALI and EVEREST Spinal Systems are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior non-cervical fixation in pediatric patients. The MESA, DENALI and EVEREST Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Device Description

This traditional 510(k) premarket notification is to introduce the Cannulated Power Driver Attachment instrument for use with Stryker Power Systems.

The Cannulated Power Driver Attachment is an accessory to the Everest and Range (Mesa and Denali) pedicle screw systems intended to facilitate the insertion of pedicle screws using powered instrumentation. The Cannulated Power Driver Attachment, also referred to as the power adaptor or subject device, serves as a mechanical interface between the power driver and screwdriver instrument. When the adaptor is attached, the driver (corded and cordless) provides appropriate power to rotate screw drivers for the insertion of pedicle screws. No changes have been made to the indications for use of the associated thoracolumbar spinal implant systems Everest and Range (Mesa and Denali) Spinal Systems. The indications for use of each spinal system remain consistent with their most recent 510(k) clearance.

AI/ML Overview

The provided text is a 510(k) summary for the "Cannulated Power Driver Attachment". This document is for a medical device that assists in the surgical insertion of pedicle screws, not an AI/ML-driven device for diagnosis or prediction. Therefore, the questions related to AI/ML device performance, such as sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth types, are not applicable.

However, I can extract information related to the acceptance criteria and how the device met them.

1. A table of acceptance criteria and the reported device performance

The document states that the "Cannulated Power Driver Attachment Instrument has demonstrated substantial equivalence to the predicate device." It further states: "The performance of the Cannulated Power Driver Attachment has met all acceptance criteria."

While specific numerical acceptance criteria (e.g., torque values, cycles to failure) are not provided in this summary, the general acceptance criteria can be inferred from the context of "substantial equivalence" and "verification and validation activities."

Acceptance Criteria (Inferred)Reported Device Performance
Substantial Equivalence to Predicate DeviceDemonstrated substantial equivalence to the predicate device (K200666 and others).
Compatibility with Stryker Power DriversConnection to various Stryker Power Drivers does not represent a new worst case for powered pedicle screw insertion.
Functionality for Pedicle Screw InsertionAids in the rotation of the bone screw to facilitate insertion (when used with Stryker Power Drivers).
Mechanical Integrity/PerformanceMet all acceptance criteria through Verification and Validation activities. (Specific criteria, e.g., torque, fatigue, not detailed in summary).
Material and Design EquivalenceHas equivalent materials, fundamental scientific technologies, and design compared with the predicate device.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This device is a mechanical instrument. Testing would involve engineering evaluations (e.g., mechanical stress tests, fatigue tests) rather than data from human subjects or medical images. Therefore, the concepts of "test set sample size" and "data provenance" as typically applied to AI/ML systems are not relevant here. The "test set" would be the manufactured Cannulated Power Driver Attachment instruments subjected to various non-clinical performance evaluations. The number of units tested is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the device is a mechanical instrument. Ground truth in this context would be defined by engineering standards and specifications, not expert medical opinion on clinical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for a mechanical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable as this is a mechanical medical instrument, not an AI/ML diagnostic or predictive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable as this is a mechanical medical instrument, not an AI/ML algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for a mechanical device would be established by engineering specifications, validated test methods, and performance benchmarks against predicate devices. The summary indicates that "Verification and validation activities" were conducted to demonstrate that the device "met all acceptance criteria" and "does not represent a new worst case for powered pedicle screw insertion," implying a comparison to established performance characteristics of the predicate device.

8. The sample size for the training set

Not applicable as this is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable as this is not an AI/ML device.

{0}------------------------------------------------

October 26, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

K2M Inc. Megan Callanan Senior Regulatory Affairs Specialist 600 Hope Parkway SE Leesburg, Virginia 20175

Re: K202528

Trade/Device Name: Cannulated Power Driver Attachment Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: August 31, 2020 Received: September 1, 2020

Dear Ms. Callanan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K202528

Device Name Cannulated Power Driver Attachment

Indications for Use (Describe)

The MESA and DENALI Spinal Systems (including ARI Staples) and the EVEREST Spinal System are cleared for the following indications:

Posterior non-cervical fixation as an adjunct to following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);

spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kydosis); tumor; pseudoarthosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

Except for the ARI staples, the MESA, DENALI and EVEREST Spinal Systems are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior non-cervical fixation in pediatric patients. The MESA, DENALI and EVEREST Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and are closely spaced together. The word is black and is set against a white background.

510(k) Summary
510(k) Summary: Cannulated Power Driver Attachment
Submitter:K2M, Inc.600 Hope Pkwy SELeesburg, VA 20175
Contact Person :Name: Megan CallananPhone: (551)262-2429Email: megan.callanan1@stryker.com
Date Prepared:8/31/2020
Trade Name:Cannulated Power Driver Attachment
Common Name:Spinal Fixation Appliances, Instrument Accessory
Proposed Class:Class II
Classification Name:Thoracolumbosacral Pedicle Screw System
Regulation Number:888.3070
Product Code:NKB
Predicate Device:Primary Predicate: K200666Additional Predicates: K172724, K170496, K152632, K120434, K122845, K111478
Device Description:This traditional 510(k) premarket notification is to introduce the Cannulated PowerDriver Attachment instrument for use with Stryker Power Systems.The Cannulated Power Driver Attachment is an accessory to the Everest and Range(Mesa and Denali) pedicle screw systems intended to facilitate the insertion of pediclescrews using powered instrumentation. The Cannulated Power Driver Attachment,also referred to as the power adaptor or subject device, serves as a mechanicalinterface between the power driver and screwdriver instrument. When the adaptor isattached, the driver (corded and cordless) provides appropriate power to rotatescrew drivers for the insertion of pedicle screws. No changes have been made to theindications for use of the associated thoracolumbar spinal implant systems Everestand Range (Mesa and Denali) Spinal Systems. The indications for use of each spinalsystem remain consistent with their most recent 510(k) clearance.
Intended UseThe intended use for the Cannulated Power Driver Attachment is to facilitate theplacement of pedicle screws using the power technique (corded and cordless). TheCannulated Power Driver Attachment is intended for exclusive use with the StrykerCordless and Corded Power Drivers. When the power adaptors are attached, theStryker Power Drivers (corded and cordless) provide power to rotate screwdriversfor the insertion of pedicle screws.Pedicle screws from select Stryker Spine implant systems may be implanted in thenon-cervical spine using powered (corded and cordless) instrumentation The
systems included are the family of Everest and Range (Mesa and Denali) SpinalSystems.
Indications for Use:The MESA and DENALI Spinal Systems (including ARI Staples) and the EVERESTSpinal System are cleared for the following indications:
Posterior non-cervical fixation as an adjunct to fusion for the following indications:degenerative disc disease (defined as back pain of discogenic origin withdegeneration of the disc confirmed by history and radiographic studies);spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e.scoliosis, kyphosis and/or lordosis); tumor; pseudoarthosis; and/or failed previousfusion.
Except for hooks, when used as an anterolateral thoracic/lumbar system the RangeSpinal System may also be used for the same indications as an adjunct to fusion.
Except for the ARI staples, the MESA, DENALI and EVEREST Spinal Systems areindicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used forposterior non-cervical fixation in pediatric patients. The MESA, DENALI and EVERESTSpinal System for pediatric use is intended to be used with autograft and/or allograft.Pediatric pedicle screw fixation is limited to a posterior approach.
Summary of theTechnologicalThe Cannulated Power Driver Attachment has equivalent materials, fundamentalscientific technologies and design compared with the predicate device.
CharacteristicsThe screws are intended to be inserted manually or with Stryker's corded andcordless power drivers. The power adaptor accessory instrument assembly aids inthe rotation of the bone screw to facilitate insertion.
Non-ClinicalPerformanceEvaluationThe Cannulated Power Driver Attachment Instrument has demonstrated substantialequivalence to the predicate device. Verification and validation activitiesdemonstrated connection of the Cannulated Power Driver Attachment to variousStryker Power Drivers does not represent a new worst case for powered pediclescrew insertion. The performance of the Cannulated Power Driver Attachment hasmet all acceptance criteria.
ConclusionBased on the design features, the use of established well known materials, featurecomparisons, and indications for use the subject devices have demonstratedsubstantial equivalence to the identified predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is horizontally oriented and takes up most of the image.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.