(56 days)
Not Found
No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and describes the device as a tool for manual planning and measurement, requiring clinical judgment.
No.
The explanation clearly states it "is indicated for assisting healthcare professionals in preoperative planning of orthopedic surgery" and for "evaluating digital images while performing various pre-operative treatment planning, evaluation of images and post-operative treatment planning." It does not directly treat or diagnose a disease or condition.
No
The device is intended for preoperative planning of orthopedic surgery, which falls under treatment planning, not diagnosis. It assists in overlaying templates and performing measurements but explicitly states that clinical judgments and experience are required, indicating it does not provide a diagnosis.
Yes
The device description explicitly states it is a "web-based platform module system" and a "software application," with no mention of accompanying hardware components being part of the device itself. The testing described focuses solely on software performance and usability.
Based on the provided information, the OrthoNext ™ Platform system is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs are used to examine specimens derived from the human body. The OrthoNext ™ Platform system operates on radiological images (x-rays), which are not biological specimens.
- The intended use is for preoperative planning of orthopedic surgery. This involves analyzing images and assisting in surgical strategy, not performing tests on bodily fluids or tissues to diagnose or monitor a condition.
- The device description focuses on image manipulation, template overlay, and measurement tools. These are functions related to image analysis and surgical planning, not in vitro testing.
The device falls under the category of medical image processing software used for surgical planning, which is distinct from IVDs.
N/A
Intended Use / Indications for Use
The OrthoNext ™ Platform system is indicated for assisting healthcare professionals in preoperative planning of orthopedic surgery. The device allows for overlaying of Orthofix Product templates on radiological images, and includes tools for performing measurements on the image and for positioning the template. Clinical judgments and experience are required to properly use the software.
The OrthoNext ™ Platform system is not to be used for mammography.
Product codes
LLZ
Device Description
The OrthoNextTM Platform is a web-based platform module system, to allow surgeons to evaluate digital images while performing various pre-operative treatment planning, evaluation of images and post-operative treatment planning. This software application enables surgeons to import radiological images, display various 2D views of the images, overlays the positioning of the Orthofix devices template and simulate the treatment options, generate parameters and/or measurements to be verified or adjusted by the surgeons based on their clinical judgment.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
x-ray images
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
healthcare professionals / Hospital
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
The testing for each release consisted of Unit, System/Integration and Acceptance test levels. Testing included also security, negative testing, error message handling, stress testing, platform testing, workflow testing, functional testing, multi-user/external access testing, data integrity testing, compatibility testing, load testing, regression testing, and hazard mitigation testing. In case a test was failed any necessary corrections were made, the relevant test was executed and repeated again until all passed.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical software testing on operative treatment planning of orthopedic surgery using OrthoNext ™ Platform system produces results comparable to planning using acetate overlays but with the additional advantages of digital planning and simulations including ease of use, library, case documentation, access to a wider arrange of tools, and secure accessibility. Any potential hazards have been evaluated and controlled through Risk Management activities. The review of clinical literatures on similar devices support the clinical performance of the Subject device with no additional clinical data. Usability testing have been performed by simulating a clinical environment requiring the test participants to perform treatment
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
October 27, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The logos are placed side-by-side.
Orthofix Srl % Ms. Cheryl Wagoner Consultant Wagoner Consulting LLC P O Box 15729 WILMINGTON NC 28408
Re: K202519
Trade/Device Name: OrthoNext™ Platform System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 25, 2020 Received: September 1, 2020
Dear Ms. Wagoner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
1
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K202519
Device Name OrthoNext TM Platform system
Indications for Use (Describe)
The OrthoNext ™ Platform system is indicated for assisting healthcare professionals in preoperative planning of orthopedic surgery. The device allows for overlaying of Orthofix Product templates on radiological images, and includes tools for performing measurements on the image and for positioning the template. Clinical judgments and experience are required to properly use the software.
The OrthoNext ™ Platform system is not to be used for mammography.
Type of Use (Select one or both, as applicable) |
---|
Prescription Use (Part 21 CFR 801 Subpart D) |
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape above the word "ORTHOFIX" in black, with a registered trademark symbol next to it. The abstract shape is a curved, geometric design.
510(k) Summary (21 CFR 807.92) K202519 Submitter information Submitter Name Orthofix Srl Via delle Nazioni, 9 | 37012 Bussolengo (VR) - Italy Address Telephone + 39 045 6719.000 Fax + 39 045 6719.380
| Contact Person | Gianluca Ricadona
Sr. Quality & Regulatory Affairs Manager |
|--------------------|---------------------------------------------------------------|
| Address | Via delle Nazioni, 9
37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719 000 |
| Fax | + 39 045 6719 380 |
| Email address | GianlucaRicadona@orthofix.it |
| Date of submission | October 22, 2020 |
Trade Name, Common Name, Classification
Trade Name | OrthoNext ™ Platform system |
---|---|
Device Classification name | Image Processing System Radiological |
Product code | LLZ |
Panel Code | Radiology |
Class | Class II |
Classification Regulation Number | 21 CFR § 892.2050 |
Predicate devices
Primary Predicate Device | 510(k) Number | Manufacturer |
---|---|---|
TraumaCad Mobile 1.0 | K142923 | ORTHOCRAT,LTD. |
291Hillside Avenue | ||
Somerset, MA 02726 | ||
Additional Predicate Device | ||
TraumaCAD Version 2.0 | K073714 | ORTHOCRAT,LTD. |
291Hillside Avenue | ||
Somerset, MA 02726 |
| Device description | The OrthoNextTM Platform is a web-based platform module system, to
allow surgeons to evaluate digital images while performing various pre-
operative treatment planning, evaluation of images and post-operative
treatment planning.
This software application enables surgeons to import radiological
images, display various 2D views of the images, overlays the
positioning of the Orthofix devices template and simulate the treatment
options, generate parameters and/or measurements to be verified or
adjusted by the surgeons based on their clinical judgment. |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The OrthoNextTM Platform system is indicated for assisting healthcare
professionals in preoperative planning of orthopedic surgery. The
device allows for overlaying of Orthofix Product templates on
radiological images, and includes tools for performing measurements
on the image and for positioning the template. Clinical judgments and
experience are required to properly use the software.
The OrthoNextTM Platform system is not to be used for mammography. |
4
| Technological
Characteristics and
Intended Use | | The OrthoNext™ Platform operating principles and technological characteristic, including the intended use and users are the same as, or similar to, the chosen predicate devices.
Summary of the equivalence in technological characteristics and Intended Use:
✓ Intended use: identical.
✓ Operating principles, technological characteristics and conditions of use are substantially equivalent to predicates:
the OrthoNext ™ Platform system is a web-based software executed on a common web browser (Chrome, Internet Explorer, Safari), intended to run on a PC, MAC by Windows and Mac OS package , accessible in a secure environment by a license activation code and password provided by the manufacturer.
Principle of operation includes:
• Importation medical images format (x-ray images)
• Processing tools
• Measurements and parameters analysis tools
• Surgical planning tools
• Enable SW Modules (operative treatment planning) for overlaying template for simulation. |
|------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance
Analysis | | Subject device has similar configuration, and operating principle as the predicate device. Non-clinical software testing on operative treatment planning of orthopedic surgery using OrthoNext ™ Platform system produces results comparable to planning using acetate overlays but with the additional advantages of digital planning and simulations including ease of use, library, case documentation, access to a wider arrange of tools, and secure accessibility. Any potential hazards have been evaluated and controlled through Risk Management activities.
The review of clinical literatures on similar devices support the clinical performance of the Subject device with no additional clinical data. Usability testing have been performed by simulating a clinical environment requiring the test participants to perform treatment |
Basis for Substantial Equivalence
| | SUBJECT DEVICE | PRIMARY
PREDICATE DEVICE |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| System
Features | OrthoNext ™ Platform | TraumaCad Mobile 1.0 |
| Product Code | LLZ | Identical |
| Indications for Use | The OrthoNext ™ Platform system is
indicated for assisting healthcare
professionals in preoperative planning of
orthopedic surgery. The device allows for
overlaying of Orthofix Product templates
on radiological images, and includes tools
for performing measurements on the
image and for positioning the template.
Clinical judgments and experience are
required to properly use the software. The
OrthoNext ™ Platform system is not to be
used for mammography | Identical |
| Intended Enviroment | Hospital | Identical |
5
SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | |
---|---|---|
Specialties Sites | Easy-to-use solutions for various orthopedic subspecialties for skeletal appendicular trauma and deformity analysis | Identical |
Configuration | Web- based | Identical |
Image Input | Can receive digital images | Identical |
Run on server | YES | Identical |
Digital Device Template | YES | Identical |
Interactive Template positioning | YES | Identical |
Automatic scaling | YES | Identical |
Template support from the manufacturer | YES | Identical |
Permits template rotation | YES | Identical |
Treatment operative planning | YES | Identical |
Patient contacting | NO | Identical |
Control of Life-Saving Devices | NO | Identical |
HCP intervention for interpretation and manipulation of images | YES | Identical |
Conclusion | The successful non-clinical testing demonstrates the safety and effectiveness of the OrthoNext ™ Platform system when used for the defined indications for use and demonstrates that the subject device, for which this Traditional 510(k) is submitted, performs as well as or better than the legally marketed predicate devices. OrthoNext ™ Platform contains a subset of the same features and algorithms as those that are in the predicate devices. The testing for each release consisted of Unit, System/Integration and Acceptance test levels. Testing included also security, negative testing, error message handling, stress testing, platform testing, workflow testing, functional testing, multi-user/external access testing, data integrity testing, compatibility testing, load testing, regression testing, and hazard mitigation testing. In case a test was failed any necessary corrections were made, the relevant test was executed and repeated again until all passed. |
---|---|
------------ | ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |