K Number
K202439
Date Cleared
2021-03-08

(195 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These mask are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use , disposable device, provided nonsterile.

Device Description

The proposed devices are single use, three-layer, flat masks with ear-loops and nose piece. The Disposable Medical Face Masks are manufactured with three layers, the outer layer is made of PPSB non-woven fabric, which the chemical composition is the mixture of Polypropylene and color master batch. Inner layer is made of Polyethylene, polypropylene and mixture of fiber finishes. And the middle layer is made of melt blown polypropylene filter. The model of proposed device ,YX001, is held in place over the user's mouth and nose bu two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural rubber latex. The nose piece contained in the proposed device is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Malleable aluminum wire. The proposed devices are sold non-sterile and are intended to be single use, disposable devices. The colorants used for mask are Copper phthalocyanine and Titanium dioxide.

AI/ML Overview

This document does not describe an AI medical device study. It is a 510(k) summary for a Disposable Medical Face Mask (K202439), demonstrating substantial equivalence to a predicate device (K153496).

Therefore, I cannot provide information on:

  • Acceptance criteria and device performance for an AI/ML model
  • Sample size for a test set or data provenance in the context of AI
  • Number of experts or their qualifications for ground truth in AI
  • Adjudication methods for AI test sets
  • MRMC studies or effect sizes for AI assistance
  • Standalone AI performance
  • Training set details for AI
  • How ground truth for a training set was established for AI

The document focuses on non-clinical performance characteristics of the medical face mask, such as:

  • Resistance to penetration by synthetic blood
  • Sub-micron particulate filtration efficiency (PFE)
  • Bacterial Filtration Efficiency (BFE)
  • Differential Pressure (breathability)
  • Flame spread
  • Biocompatibility (Cytotoxicity, Irritation, Sensitization)

The acceptance criteria and performance data are entirely related to these physical and biological characteristics of a face mask, not an AI/ML algorithm.

Here is the relevant information from the document as it pertains to the face mask, not an AI device:

1. Table of Acceptance Criteria and Reported Device Performance (for the face mask):

ItemsAcceptance CriteriaSubject Device Test Result (K20249)Predicate Device (K153496)Purpose/Test Standard
Resistance to penetration by synthetic blood (ASTM F1862)Pass at 120 mmHgPass at 120 mmHgPass at 120 mmHg, Pass at 160 mmHgEvaluate protection against fluid penetration.
Sub-micron particulate filtration efficiency (PFE) (ASTM F2299)> 98%99.9%> 98%Evaluate non-viable particle filtration efficiency using monodispersed polystyrene latex spheres.
Bacterial Filtration Efficiency (BFE) (ASTM F2101)> 98%99.9%> 98%Determine filtration efficiency by comparing bacterial counts upstream and downstream of the test article.
Differential Pressure (Delta P) (EN 14683:2019)< 6.0 mmH2O/cm²4.7 mmH2O/cm²< 6.0 mmH2O/cm²Determine breathability by measuring differential air pressure across the mask at a constant flow rate.
Flame spread (16 CFR 1610)Class 1Class 1Class 1Evaluate flammability of clothing textiles by measuring ease of ignition and speed of flame spread.
Cytotoxicity (ISO 10993-5: 2009)Non-cytotoxicNon-cytotoxicNon-cytotoxicBiological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
Irritation (ISO 10993-10: 2010)Non-irritatingNon-irritatingNon-irritatingBiological Evaluation Of Medical Devices Part 10: Tests For Irritation And Skin Sensitization.
Sensitization (ISO 10993-10: 2010)Non-sensitizingNon-sensitizingNon-sensitizingBiological Evaluation Of Medical Devices Part 10: Tests For Irritation And Skin Sensitization.

2. Sample Size Used for the Test Set and the Data Provenance:
This document does not specify general "test sets" in the AI/ML sense. For the performance tests (e.g., BFE, PFE), specific sample sizes for each individual test are not explicitly stated in this summary, but these are standard laboratory tests typically performed on a defined number of mask samples as per the respective ASTM/ISO standards.
The data provenance is from Hunan EEXI Technology & Service Co.,Ltd. in China, as they are the submitter (K202439) and presumably conducted these tests on their own product. The tests are non-clinical, essentially lab-based performance evaluations of the physical product, not data-driven AI evaluations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
Not applicable. Ground truth as typically understood in AI/ML (e.g., expert labels on medical images) is not relevant here. The "ground truth" for a face mask's physical properties are the specific, measurable outcomes of the standardized ASTM/ISO tests. These are determined by lab measurements, not expert consensus in the diagnostic sense.

4. Adjudication Method for the Test Set:
Not applicable. There is no adjudication process described for these performance tests as they are objective quantitative or qualitative (pass/fail) lab measurements based on established standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
No, this is not an AI device. An MRMC study is not applicable for a physical product like a face mask.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Not applicable. This is not an AI device.

7. The Type of Ground Truth Used:
The "ground truth" for the face mask's performance is established by objective laboratory measurements performed according to recognized international standards (ASTM, ISO, etc.). For example:

  • Fluid resistance: Measured penetration at specific pressure.
  • Filtration efficiency: Measured percentage of particles/bacteria filtered.
  • Differential pressure: Measured pressure drop.
  • Biocompatibility: Lab test results indicating cellular response (cytotoxicity), skin reactions (irritation), or immune response (sensitization).

8. The Sample Size for the Training Set:
Not applicable. This is not an AI device. There is no "training set."

9. How the Ground Truth for the Training Set Was Established:
Not applicable. As there is no training set for an AI model, there is no ground truth established for one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 8, 2021

Hunan EEXI Technology & Service Co.,Ltd. % Joyce Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518100 China

Re: K202439

Trade/Device Name: Disposable Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 18, 2021 Received: February 8, 2021

Dear Joyce Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202439

Device Name Disposable Medical Face Mask

Indications for Use (Describe)

The Disposable Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These mask are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use , disposable device, provided nonsterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k) Number: K202439

Date of Summary prepared: March 4, 2021

1. Submission Sponsor

Applicant NameHunan EEXI Technology & Service Co.,Ltd.
AddressNo. 6 North of Pingtou Road, Liuyang Hi-tech Industrial Development Zone, Hunan China.
Contact personZhang Xianliang
Phone+0086-15874115252

2. Submission correspondent

NameShenzhen Joyantech Consulting Co., Ltd
Address1713A, 17th Floor, Block A, Zhongguan Times
Square, Nanshan District, Shenzhen
Post Code518000
Phone No.+86-755-86069197
Contact PersonJoyce Yang
Emailjoyce@cefda.com

3. Device Identification

Type of 510(k) submission:Traditional
Trade Name:Disposable Medical Face Mask
Common name:Surgical face mask
Model:YX001
Classification name:Mask, Surgical
Product Code:FXX
Device Class:II
Regulation Number:878.4040

{4}------------------------------------------------

4. Legally Marketed Predicate Device

Trade Name:Disposable Medical Face Mask
Regulation number:21 CFR 878.4040
Regulation class:II
Regulation name:Surgical Apparel
510(k) Number:K153496
Product Code:FXX
Manufacturer:Xiantao Rayxin Medical Products Co.,Ltd.

5. Device Description

The proposed devices are single use, three-layer, flat masks with ear-loops and nose piece. The Disposable Medical Face Masks are manufactured with three layers, the outer layer is made of PPSB non-woven fabric, which the chemical composition is the mixture of Polypropylene and color master batch. Inner layer is made of Polyethylene, polypropylene and mixture of fiber finishes. And the middle layer is made of melt blown polypropylene filter.

The model of proposed device ,YX001, is held in place over the user's mouth and nose bu two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural rubber latex.

The nose piece contained in the proposed device is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Malleable aluminum wire.

The proposed devices are sold non-sterile and are intended to be single use, disposable devices. The colorants used for mask are Copper phthalocyanine and Titanium dioxide.

6. Intended Use/ Indications for Use

The Disposable Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These mask are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use ,disposable device, provided non-sterile.

potential exposure to blood and body fluids.

7. Technological characteristics comparison

The Disposable Medical Face Mask is compared with the predicate device Disposable Medical Face Mask(K153496). The product characteristics are shown as follow:

{5}------------------------------------------------

ComparisonitemSubject Device(K202439)Predicate Device(K153496)Comments
ProductCodeFXXFXXSame
RegulationNumber21 CFR § 878.404021 CFR § 878.4040Same
ClassificationClass IIClass IISame
OTC useYesYesSame
Intendeduse &Indications for UseThe Disposable MedicalFace Masks are intended tobe worn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, body fluidsand particulate material.These mask are intended foruse in infection controlpractices to reduce thepotential exposure to bloodand body fluids. This is asingle use ,disposabledevice, provided non-sterile.The Disposable SurgicalFace Masks are intendedto be worn to protect boththe patient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. These face maskare intended for use ininfection controlpractices to reduce thepotential exposure toblood and body fluids.This is a singleuse ,disposable device,provided non-sterile.Same
DesignfeatureEar-loopEar-loop, Tie-onSimilarIssue 1
UsageSingle useSingle useSame
ColorBlueBlueSame
Size(175±10)mm×(95±10)mm(17.5±1)cm×(9.5±1)cmSame
SterileNon-sterileNon-sterileSame
Outer layer: Spun-bondpolypropyleneOuter layer: Spun-bondpolypropyleneSame
Middle layer: Melt blownpolypropylene filterMiddle layer: Melt blownpolypropylene filterSame
ComparisonitemSubject Device(K202439)Predicate Device(K153496)Comments
MaterialInner layer: Spun-bondpolypropyleneInner layer:Spun-bondpolypropyleneSame
Nose piece: PE+ Steel wireNose piece: Malleablealuminum wireDifferentIssue 2
Ear-loops: Elastic fiberEar-loops: PolyesterSimilarIssue 3
ASTM F2100 LevelLevel 2Level 2Same
FluidResistancePerformanceASTM F1862-13Meet the ASTM F2100Requirements for Level 2ClassificationMeet the ASTM F2100Requirements for Level 2ClassificationSame
ParticulateFiltrationEfficiencyASTM F2299Meet the ASTM F2100Requirements for Level 2ClassificationMeet the ASTM F2100Requirements for Level 2ClassificationSame
BacterialFiltrationEfficiencyASTM F2101Meet the ASTM F2100Requirements for Level 2ClassificationMeet the ASTM F2100Requirements for Level 2ClassificationSame
DifferentialPressure(Delta P)MIL-M-36954CMeet the ASTM F2100Requirements for Level 2ClassificationMeet the ASTM F2100Requirements for Level 2ClassificationSame
Flammability16CFR 1610Class 1Class 1Same
CytotoxicityComply with ISO 10993-5Non cytotoxicComply with ISO 10993-5Non cytotoxicSame
IrritationComply with ISO 10993-10Non irritatingComply with ISO 10993-10Non irritatingSame
SensitizationComply with ISO 10993-10Non sensitizingComply with ISO 10993-10Non sensitizingSame

Table 1: General Comparison

{6}------------------------------------------------

lssue 1: Design differences do not introduce different questions of safety and effectiveness.

lssue 2 : The nose piece of the proposed device is made by PE+Steel wire,which of the predicate device is made by Malleable aluminum wire. The Nose piece is between the inner and outer layers of the mask, which does not contact with the human body directly when used.

{7}------------------------------------------------

Moreover, the whole product has been tested for biocompatibility, and the test results confirm that they have good biocompatibility, these differences do not introduce different questions of safety and effectiveness.

Issue 3: The Ear-loops of the proposed device are made by elastic fiber, which of the predicate device is made by polyester. The major chemical composition of the elastic fiber is seqmented polyurethane-urea, which is similar to polyester. In addition, the proposed devices have been tested for biocompatibility, and the test results confirm that they have good biocompatibility, these differences do not introduce different questions of safety and effectiveness.

8. Summary of Non-clinical Testing

Non clinical tests were conducted and conformed to the following standards and the requirements state in the Guidance for Industry and FDA Staff: Surgical Masks -Premarket Notification [510(k)] Submission issued on March 05, 2004.

Standards:

  • ASTM F2100-19 Standard Specification For Performance of Materials used in Medical Face Masks.
  • ASTM F1862-13 Standard Test Method For Resistance of Medical Face Masks to Penetration by Synthetic Blood.
  • ASTM F2299-03 Stand Test Method For Determining The Initial Efficiency Of Materials Used In Medical Face Masks To Penetration By Particulates Using Latex Spheres.
  • 16 CFR 1610 Standard For The Flammability Of Clothing Textiles.
  • ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
  • ANSI/AAMI/ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for Development, validation, and routine control of a sterilization process for Medical devices
  • ISO 10993-5: 2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
  • ISO 10993-10: 2010 Biological Evaluation Of Medical Devices Part 10: Tests ● For Irritation And Skin Sensitization.

{8}------------------------------------------------

ItemsPurposeAcceptancecriteriaSubjectdevice testresult(K202439)Predicatedevice(K153496)
Resistance topenetration bysyntheticblood(ASTMF1862)This procedure wasperformed to evaluatesurgical facemasks andother types of protectiveclothing materials designedto protect against fluidpenetration.Pass at120mmHgPass at120 mmHgPass at120 mmHgPass at160 mmHg
Sub-micronparticulatefiltrationefficiency( ASTMF2299)This procedure wasperformed to evaluate thenon-viable particle filtrationefficiency (PFE) of the testarticle. Monodispersedpolystyrene latex spheres(PSL) were nebulized(atomized), dried, andpassed through the testarticle.> 98%99.9%> 98%
BacterialFiltrationEfficiency(ASTM F2101)The BFE test is performedto determine the filtrationefficiency of test articles bycomparing the bacterialcontrol counts up stream ofthe test article to thebacterial countsdownstream.> 98%99.9%> 98%
DifferentialPressure(EN14683:2019)The Delta P test isperformed to determine thebreathability of test articlesby measuring thedifferential air pressure oneither side of the test articleusing a manometer, at aconstant flow rate.< 6.0mmH2O/cm²4.7mmH2O/cm²< 6.0mmH2O/cm²
ItemsPurposeAcceptancecriteriaSubjectdevice testresult(K202439)Predicatedevice(K153496)
Flame spread(16CFR 1610)This procedure wasperformed to evaluate theflammability of plain surfaceclothing textiles bymeasuring the ease ofignition and the speed offlame spread.Class 1Class 1Class 1

Table 2: Performance Characteristic Comparison

{9}------------------------------------------------

Table 3: Biocompatibility Comparison

ItemsSubject device(K202439)Predicate device(K153496)
CytotoxicityUnder the conditions of thestudy, the test article was foundto be non-cytotoxicUnder the conditions of thestudy, the test article was foundto be non-cytotoxic
IrritationUnder the conditions of thestudy, the test article was foundto be non-irritationUnder the conditions of thestudy, the test article was foundto be non-irritation
SensitizationUnder the conditions of thestudy, the test article was foundto be non-sensitizingUnder the conditions of thestudy, the test article was foundto be non-sensitizing

9. Brief discussion of clinical tests

No clinical tests were performed.

10. Conclusions

The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(K) submission K202439, the Disposable Medical Face Masks are as safe and effective, and performs as well as or better than the legally marketed predicate device cleared under K153496.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.