(179 days)
The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile and sterile.
The proposed device, Disposable Surgical Mask is a three-layer, flat-folded mask with ear strap and nose clip. The ear straps are held in place over the users' mouth and nose by two elastic ear straps welded to the facemask. The color of the surgical mask is blue. The disposable surgical mask is available in three different specifications: 17.59.5cm, 14.59.5cm and 14.5*8cm. The device is single use and provided non-sterile and sterile.
The provided document is a 510(k) Pre-market Notification for a Disposable Surgical Mask, primarily focusing on proving substantial equivalence to a predicate device based on non-clinical performance standards. It explicitly states, "No clinical study is included in this submission." Therefore, it does not contain the information requested regarding acceptance criteria and a study proving a device meets acceptance criteria using clinical or AI-driven performance metrics.
The document outlines acceptance criteria for non-clinical tests (e.g., fluid resistance, particulate efficiency, bacterial filtration, differential pressure, flammability, biocompatibility) to demonstrate the proposed device performs as well as or better than the predicate surgical mask. However, these are physical and biological performance metrics for the mask itself, not measures of a complex device's performance in a clinical setting or where AI is involved.
Here's a breakdown of why I cannot fulfill your request based on the provided text, and what information is available related to performance:
Information NOT available in the provided document:
- A table of acceptance criteria and the reported device performance related to a clinical or AI-driven study: The document presents a comparison table (page 5) of the proposed and predicate device's specifications and physical performance parameters, not clinical performance against acceptance criteria for a device involving AI or complex decision-making.
- For example, it states "Fluid Resistance: Pass at 160 mmHg" as a characteristic, which is a physical test result, not a clinical outcome or AI performance metric.
- Sample size used for the test set and data provenance: Not applicable as no clinical or AI test set is mentioned.
- Number of experts used to establish ground truth and qualifications: Not applicable as no ground truth for a clinical or AI study is established.
- Adjudication method for the test set: Not applicable.
- If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done: "No clinical study is included in this submission." (page 7)
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
- The sample size for the training set: Not applicable, as no AI/machine learning component is described.
- How the ground truth for the training set was established: Not applicable.
Information related to "acceptance criteria" and "performance" that IS in the document, but for physical/biological mask properties:
The document lists performance measures for the physical characteristics of the surgical mask:
1. A table of acceptance criteria and the reported device performance (for physical/biological properties):
| ITEM | Proposed Device Performance (Reported as "Pass at...") | Predicate Device Performance (Reported as "Passed at...") | Comparison |
|---|---|---|---|
| Fluid Resistance | Pass at 160 mmHg | Pass at 160 mmHg | Same |
| Particulate efficiency | Pass at >98.01% | Passed at 99.7% | Similar |
| Bacterial filtration | Pass at >98.12% | Passed at >99% | Similar |
| Differential Pressure | Pass at <4.6 mmH2O/cm2 | Passed at 2.5 mmH2O/cm2 | Similar |
| Flammability | Class 1 | Class 1 | Same |
| Biocompatibility: Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Biocompatibility: Sensitization | No Sensitization | No Sensitization | Same |
| Biocompatibility: Irritation | No Irritation | No Irritation | Same |
2. Sample sizes for these physical/biological tests: Not explicitly stated in the document. The standards (e.g., ASTM, ISO) typically define the sample sizes for these types of tests.
3. Data provenance for these non-clinical tests: The tests were "conducted to verify that the proposed device met all design specifications" (page 6). This implies they were prospective tests performed on samples of the manufactured surgical masks. The country of origin for the manufacturer is China.
4. Experts involved in establishing ground truth for these non-clinical tests: Not applicable in the context of expert review of data/images. These are typically lab-based, standardized tests performed by qualified technicians following established protocols.
5. Adjudication method for these non-clinical tests: Not applicable.
6. Effect size of human readers improving with AI vs. without AI assistance: Not applicable as this is a surgical mask submission, not an AI device.
7. Standalone performance (algorithm only): Not applicable.
8. Type of ground truth for these non-clinical tests: For these physical and biological tests, the "ground truth" is defined by the objective results of the standardized test methods outlined (e.g., ASTM F1862/F1862M-17 for fluid resistance, ASTM F2101-2017 for BFE). These are laboratory measurements against accepted thresholds.
9. Sample size for the training set (for these non-clinical tests): Not applicable, as there's no machine learning training set involved.
10. How ground truth for training set was established (for these non-clinical tests): Not applicable.
In summary, the provided document is for a medical device (surgical mask) that is cleared based on its equivalence in physical and biological performance to a predicate device, as demonstrated through established non-clinical laboratory tests. It does not describe an AI-driven device or a clinical study that would involve the kind of acceptance criteria, test sets, and expert involvement you've asked about for an AI/clinical diagnostic device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2021
Shanghai Jianzhong Medical Packaging Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai, 200120 China
Re: K202341
Trade/Device Name: Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 13, 2021 Received: January 14, 2021
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For CAPT Elizabeth Claverie-Williams, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K202341
Device Name Disposable Surgical Mask
Indications for Use (Describe)
The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided nonsterile and sterile.
Models: 17.59.5cm-3ply (sterile), 14.59.5cm-3ply (sterile), 14.58cm-3ply (sterile) 17.59.5cm-3ply (non-sterile), 14.59.5cm-3ply (non-sterile), 14.58cm-3ply (non-sterile)
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K202341
- Date of Preparation: 02/12/2021 1.
-
- Sponsor Identification
Shanghai Jianzhong Medical Packaging Co., Ltd. Bldg 16, 789# Puxing Road, Minhang District, Shanghai 201114, P.R. China
Establishment Registration Number: 3005030944.
Contact Person: Nina Huang Position: Marketing Manager Tel: +86-21-54315666-5201 Fax: +86-21-54315801 Email: nina.huang@mpackchina.com
-
- Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
- Designated Submission Correspondent
Mid-Link Consulting Co., Ltd.
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
4. Identification of Proposed Device
Trade Name: Disposable Surgical Mask
Common Name: Surgical mask
Models: 17.59.5cm-3ply (sterile), 14.59.5cm-3ply (sterile), 14.5*8cm-3ply (sterile)
17.59.5cm-3ply (non-sterile), 14.59.5cm-3ply (non-sterile), 14.5*8cm-3ply (non-sterile)
Regulatory Information
Classification Name: Mask, Surgical; Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040; Review Panel: General Hospital;
Indication for use:
The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile and sterile.
Models: 17.59.5cm-3ply (sterile), 14.59.5cm-3ply (sterile), 14.58cm-3ply (sterile) 17.59.5cm-3ply (non-sterile), 14.59.5cm-3ply (non-sterile), 14.58cm-3ply (non-sterile)
Device Description:
The proposed device, Disposable Surgical Mask is a three-layer, flat-folded mask with ear strap and nose clip. The ear straps are held in place over the users' mouth and nose by two elastic ear straps welded to the facemask. The color of the surgical mask is blue. The disposable surgical mask is available in three different specifications: 17.59.5cm, 14.59.5cm and 14.5*8cm. The device is single use and provided non-sterile and sterile.
- న. Identification of Predicate Device
510(k) Number: K160269 Product Name: Surgical Face Masks (Ear loops and Tie-on) Models: EL 10000, EL 10010, TO 10000, TO 10010 EL 20000, EL 20010, TO 20000, TO 20010 EL 30000, EL 30010, TO 30000, TO 30010
{5}------------------------------------------------
| 6. Comparison of Technological Characteristics | ||||
|---|---|---|---|---|
| ITEM | Proposed DeviceK202341 | Predicate DeviceK160269 | Comparison | |
| Product Code | FXX | FXX | Same | |
| RegulationNo. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | |
| Class | II | II | Same | |
| Indication forUse | The Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile and sterile.Models: 17.59.5cm-3ply (sterile),14.59.5cm-3ply (sterile),14.58cm-3ply (sterile),17.59.5cm-3ply (non-sterile),14.59.5cm-3ply (non-sterile),14.58cm-3ply (non-sterile) | The surgical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.Models: EL 10000, EL 10010, TO 10000, TO 10010, EL 20000, EL 20010, TO 20000, TO 20010, EL 30000, EL 30010, TO 30000, TO 30010 | Similar | |
| Configuration | ear strapnose clipmask body | ear strap/ Tie-onnose clipmask body | Similar | |
| Dimension | 17.59.5 cm14.59.5 cm14.5*8 cm | 17.59 cm189 cm | Different | |
| Level | Level III | Level III | Same | |
| FluidResistance | Pass at 160 mmHg | Pass at 160 mmHg | Same | |
| Particulateefficiencylevel | Pass at >98.01% | Passed at 99.7% | Similar | |
| Bacterialfiltration level | Pass at >98.12% | Passed at >99% | Similar | |
| Differentialpressure | Pass at <4.6 mmH2O/cm2 | Passed at 2.5 mmH2O/cm2 | Similar | |
| Flammability | Class 1 | Class 1 | Same | |
| Patient Contacting Material | ||||
| Ear strap | Polyurethane | Polyester, polyurethane | Similar | |
| Nose clip | Polyvinylchloride coated iron wire | Polyethylene coated steel wire | Different | |
| Maskbody | Outer material | Polypropylene | Polypropylene | Same |
| Middle material | Polypropylene melt-blown | Polypropylene spun bond andPolypropylene melt-blown | Similar | |
| Inner material | Polypropylene | Polypropylene | Same | |
| Pigment | Blue PCTG Pigment | Unknown | Different | |
| Biocompatibility | ||||
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same | |
| Sensitization | No Sensitization | No Sensitization | Same | |
| Irritation | No Irritation | No Irritation | Same | |
| Sterilization | ||||
| Method | EO sterilized or non-sterile | Non-sterile | Different | |
| SAL | 10-6 | / | Different |
6. Comparison of Technological Characteristics
{6}------------------------------------------------
7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
-
ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals
-
ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization
-
ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)
-
ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
-
ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
-
ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- A ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres
- A ASTM F2101-2017 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- A BS EN 14683 Medical face masks - Requirements and test methods, Annex C: Method for
{7}------------------------------------------------
determination of breathability (differential pressure)
-
- Clinical Test Conclusion
No clinical study is included in this submission.
- Clinical Test Conclusion
-
- Conclusion
Based on the technological comparison and nonclinical tests described above, the subject devices, Disposable Surgical Masks, are as safe, as effective, and perform as well or better than the legally marketed predicate device cleared under K160269.
- Conclusion
§ 878.4040 Surgical apparel.
(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.