K Number
K202331
Date Cleared
2021-02-02

(169 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Medical Face Masks are single use, disposable device, provided non-sterile.

Device Description

The Medical Face Mask is composed of three layers and is flat-pleated. The mask materials consist of an outer layer (spun-bond polypropylene), a middle layer (meltblown polypropylene), and an inner layer (spun-bond polypropylene). Each mask contains tie strings (spun-bond polypropylene) or ear loops (spandex elastic cord) to secure the mask over the users' mouth and face and includes a malleable nose piece (iron wire with white plastic covering) to provide a firm fit over the nose.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for a Medical Face Mask, not an AI-powered diagnostic device. Therefore, the questions related to AI device performance metrics, such as human reader improvement with AI assistance, standalone algorithm performance, AI training/test set details, and expert adjudication, are not applicable to this document.

However, I can extract the information relevant to the medical device's performance as described in the provided text.

Here's a breakdown of the requested information based on the document:

1. A table of acceptance criteria and the reported device performance

Performance CharacteristicsAcceptance CriteriaTest Result
Bacterial Filtration Efficiency Performance (%)≥9899.9%
Differential Pressure (Delta-P) (mm H2O/cm²)<6.02.11mmH2O/cm²
Particulate Filtration Efficiency at 0.1 micron Performance (%)≥9899.35%
Resistance to penetration by synthetic blood, Minimum pressure in mmHg for pass results12032/32 passed at 120 mmHg
Flammability ClassClass 1Class 1
CytotoxicityIf viability is reduced to < 70% of the blank, it has a cytotoxic potential.Non-cytotoxic
Irritation(0-0.4) negligible;(0.5-1.9) slight;(2-4.9) moderate;(5-8) severeNon-sensitizing
SensitizationMagnusson and Kingman grades of 1 or greater in the test group generally indicate sensitization.Non-irritating

2. Sample size used for the test set and the data provenance

The document does not specify general "sample sizes" for the test sets in the typical sense of a clinical trial. Instead, it references specific test methodologies with their own inherent sample requirements. For example:

  • Resistance to penetration by synthetic blood (ASTM F1862): The result is "32/32 passed," indicating a sample size of 32 masks were tested.
  • For other tests (BFE, PFE, Delta-P, Flammability, Biocompatibility), the specific sample sizes used for the tests are not explicitly stated, but the methodology standards (e.g., ASTM F2101, F2299, etc.) would define these.

Data Provenance: The document does not explicitly state the country of origin of the data for these performance tests. It mentions the manufacturer is Gemtier Medical (Shanghai) Inc. in China. The testing was non-clinical. The document implies these tests were conducted as part of the submission to the FDA. It does not mention if the data was retrospective or prospective, as these terms are typically reserved for clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The device is a medical face mask, and its performance acceptance is based on objective, standardized laboratory tests (e.g., filtration efficiency, fluid resistance, biocompatibility), not on expert interpretations of data like in an AI diagnostic setting.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable. This is a non-clinical device performance study based on standardized test methodologies, not a study involving human readers or subjective interpretations that would require adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document pertains to the regulatory submission for a medical face mask, not an AI-assisted diagnostic device. No human reader studies (MRMC) were conducted or are relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic device. The performance data presented is for the physical medical face mask.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the device's performance is established by adherence to recognized international and national standards for medical face masks and biological evaluation. These standards define the specific methodologies and acceptance criteria for measuring characteristics like bacterial filtration efficiency, particulate filtration efficiency, fluid resistance, flammability, and biocompatibility. For example:

  • ASTM F2100, ASTM F1862, ASTM F2299, ASTM F2101, MIL-M-36954C, 16 CFR Part 1610 (for physical properties)
  • ISO 10993-1, ISO 10993-5, ISO 10993-10 (for biological evaluation)

The performance results are objectively measured against these predefined thresholds, rather than based on subjective expert consensus, pathology, or outcomes data.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI system that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As this is not an AI system, there is no training set or associated ground truth establishment process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 2, 2021

Gemtier Medical (Shanghai) Inc % Julie Chen Technical Manager Shanghai Medical Business Consulting Co.,Ltd. No. 170 Huajiang Road, Jiading District Shanghai, 201803 China

Re: K202331

Trade/Device Name: Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 24, 2020 Received: January 6, 2021

Dear Julie Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K202331

Device Name Medical Face Mask

Indications for Use (Describe)

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Medical Face Masks are single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

{3}------------------------------------------------

510(K) Summary-K202331

I. SUBMITTER:

Gemtier Medical (Shanghai) Inc. No.18 Jianding road, Fengjing town, Jinshan district, Shanghai, China, 201502

Contact Person: Li Zhi Qing Title: RA Director Phone: (+86) 189 1703 5908 Email: lizhiqing@jintayiqi.com

Submission Correspondent: Julie Chen Title: RA Manager Shanghai Mind-link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District, Shanghai, China, 200040 Email: julie.chen@medicalbc.net Tel: +86 139 1804 5781

Summary prepared: 01/18/2021

II. DEVICE

Name of Device: Medical Face Mask Regulation Number: 21 CFR PART 878.4040 Common Name: Surgical Mask Classification Name: Surgical Mask Regulatory Class: II Product Code: FXX

III. PREDICATE DEVICE

Primary predicate device: Disposable Surgical Face Mask (K153496)

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The Medical Face Mask is composed of three layers and is flat-pleated. The mask materials consist of an outer layer (spun-bond polypropylene), a middle layer (meltblown polypropylene), and an inner layer (spun-bond polypropylene). Each mask contains tie strings (spun-bond polypropylene) or ear loops (spandex elastic cord) to secure the mask over the users' mouth and face and includes a malleable nose piece (iron wire with white plastic covering) to provide a firm fit over the nose.

REF No.Product SizeModel Description
MaskEar LoopTie-on
NS2R-D117.5×9.5cmXX
NS2R-0117cm×9.5cmXX
NS2R-C114.5cm×9.5cmXX
NS2R-I117.5×9.5cmXXX
NS2R-E117cm×9.5cmXXX
NS2R-H114.5cm×9cmXXX

V. Available Models

VI. INDICATIONS FOR USE

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Medical Face Masks are single use, disposable device, provided non-sterile

{5}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VII. PREDICATE DEVICE

The Medical Face Masks are compared with the predicate device (Disposable Surgical Face Mask (K153496)). The results are shown below in the Technological Characteristics Comparison Table:

Subject Device MedicalPrimary Predicate DeviceRemark
DEVICEFace Mask (K202331)Disposable Surgical FaceMask (K153496)
Intended UseThe Medical Face Masks areintended to be worn to protect boththe patient and healthcare personnelfrom transfer of microorganisms,body fluids and particulatematerial. These face masks areintended for use in infection controlpractices to reduce the potentialexposure to blood and body fluids.The Medical Face Masks are singleuse, disposable device, providednon-sterile.The Disposable Surgical FaceMasks are intended to be wornto protect both the patient andhealthcare personnel fromtransfer of microorganisms,body fluids and particulatematerial. These face masks areintended for use in infectioncontrol practices to reduce thepotential exposure to blood andbody fluids. This is a single use,disposable device(s), providednon-sterile.Same
Classification ProductCodeFXXFXXSame
Ear Loop Model andTie-onModelEar Loops, Tie-OnEar Loops, Tie-OnSame
Materials
Outer Facing LayerSpun-bond PolypropyleneSpun-bond polypropyleneSimilar
Middle LayerMelt-blown polypropyleneMelt blown polypropylenefilterSimilar
Inner Facing LayerSpun-bond PolypropyleneSpun-bond polypropyleneSimilar
Nose PieceMalleable iron wire withplastic coveringMalleable aluminum wireDifferent 1
Tie StringsSpun-bond PolypropyleneSpun-bond polypropyleneSimilar

{6}------------------------------------------------

Ear LoopsSpandex elastic cordPolyesterSimilar
Design Features
ColorBlueBlueSame
StyleFlat - PleatedFlat - PleatedSame
Multiple Layers3 Layers3 LayersSame
Single UseSingle useSingle useSame
Sterility
SterileNon-sterileNon-sterileSame
Dimensions
Length × Width175×95mm (±5%)170×95mm (±5%)145×95mm (±5%)17.5×9.5cm (±1cm)Similar
Technological Characteristics Product Barrier Specifications Per ASTM F2100 – Meets Level 2
Fluid Resistance ASTMF186232/32 passed at 120 mmHg32/32 passed at 120 mmHgSimilar
Particulate FiltrationEfficiency(PFE) ASTM F229999.35%98.46%Similar
Bacterial FiltrationEfficiency(BFE) ASTM F210199.9%98.7%Similar
Differential Pressure(Delta P)MIL-M-36954C2.11mmH2O/cm²4.2mmH2O/cm²Similar
Flammability16 CFR PART 1610Class 1 Non-FlammableClass 1 Non-FlammableSimilar
Biocompatibility
Cytotoxicity (ISO 10993-5)Non-cytotoxicNon-cytotoxicSame
Irritation (ISO 10993-10)Non-sensitizingNon-sensitizingSame
Sensitization (ISO10993-10)Non-irritatingNon-irritatingSame

Analysis 1

The material of nose piece of the subject device is different from the material of nose piece of the predicate device. To illustrate the difference does not raise any safety and effectiveness questions, we conducted non-clinical tests. The data drawn from the tests shows that the subject device is as safe as effective as predicate device.

{7}------------------------------------------------

VIII. PERFORMANCE DATA

Non-Clinical Performance Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the Medical Face Mask complies with the following standards:

  • . ASTM F2100 Standard Specification for Performance of Materials Used in Medical Face Masks
  • ASTM F1862 Standard Test Method for Resistance of Medical Face Masks to . Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
  • ASTM F2299 Standard Test Method for Determining the Initial Efficiency of . Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
  • ASTM F2101 Standard Method for Evaluating the Bacterial Filtration ● Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
  • MIL-M- 36954C Military Specification, Mask, Surgical, Disposable ●
  • 16 CFR Part 1610 Standard for the Flammability of Clothing ●
  • ISO10993-1 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process
  • . ISO10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity of medical devices
  • ISO10993-10 Biological Evaluation of Medical Devices Part 10: Tests for . Irritation and Skin Sensitization

Clinical Test Conclusion

No clinical study is included in this submission.

{8}------------------------------------------------

Table 1 Summary of Performance Test

PerformanceCharacteristicsMethodologyPurposeAcceptanceCriteriaTest Result
Bacterial FiltrationEfficiencyPerformance(%)ASTM F2101This performancetest aims to verifyBacterial FiltrationEfficiency (BFE)of the Medical facemask reaches level2 as required inASTM F 2100-2019≥9899.9%
Differential Pressure (Delta-P)(mm H2O/cm²)MIL-M-36954CThis performancetest aims to verifythe breathingresistance ofMedical face maskreaches level 2 asrequired in ASTMF 2100-2019<6.02.11mmH2O/cm²
Particulate FiltrationEfficiencyat 0.1micron Performance(%)ASTM F2299This performancetest aims to verifythe ParticulateFiltrationEfficiency ofMedical face maskreaches level 2 asrequired in ASTMF 2100-2019≥9899.35%
Resistance to penetration bysynthetic blood, Minimumpressure in mmHg for passresultsASTM F1862To verify the fluidresistance of theMedical face maskreaches level 2 asrequired in ASTMF 2100-2019,through performthe fluid resistancetesting.12032/32 passed at120 mmHg
Flammability Class16 CFR PART 1610To verify theflammability of theMedical face maskreaches level 2 asrequired in ASTMF 2100-2019,through conductthe flammabilitytestingClass 1Class 1
CytotoxicityISO 10993-5:2009The purpose of thetest is to determinethe biologicalreactivity ofmammalian cellculture (mousefibroblast L929If viability isreduced to < 70%of the blank, it hasa cytotoxicpotential.Non-cytotoxic
cells) in responseto the test article
IrritationISO 10993-10: 2010The teat wasdesigned toevaluate thepotential of a testarticle cause skinirritationUse only 24±2h,48±2h and 72±2hobservation forcalculation. Theprimary irritationindex for the testarticle wasevaluatedaccording to asfollow: (0-0.4)negligible;(0.5-1.9)slight;(2-4.9)moderate;(5-8)severeNon-sensitizing
SensitizationISO 10993-10: 2010The teat wasdesigned toevaluate thepotential of a testarticle cause skinsensitizationMagnusson andKingman grades of1 or greater in thetest groupgenerally indicatesensitization.Non-irritating

{9}------------------------------------------------

IX. CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device Disposable Surgical Face Mask (K153496).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.