K Number
K202137
Date Cleared
2021-03-04

(216 days)

Product Code
Regulation Number
878.4040
Panel
HO
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

Disposable Medical Masks are single use, three-layer, flat -folded masks with ear loops and nose clip. The Disposable Medical Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose which are welded to the facemask. There are two types of ear loops for the disposable medical masks, one shaped in thin rope is made of Nylon and spandex, and the other shaped in narrow band is made of spun-bond polypropylene and elastic body. The ear loops are not made with natural rubber latex. The nose clip in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polypropylene coating with iron wire. The medical masks will be provided in blue. The medical masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

This is a 510(k) premarket notification for a Disposable Medical Mask (Model: Y01, Y02) and as such, it does not involve the use of AI. The provided document details the substantial equivalence of the proposed medical mask to a predicate device based on non-clinical performance and biocompatibility testing. Therefore, the questions related to AI-specific acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and standalone algorithm performance are not applicable.

Here's the information extracted from the document regarding the acceptance criteria and the study results for the medical mask:

1. Table of acceptance criteria and the reported device performance

ItemReported Device Performance (3 lots)Acceptance Criteria (Level 1, as per ASTM F2100)Result
Fluid Resistance Performance ASTM F186232 out of 32 pass at 80 mmHg≥ 29 out of 32 pass at 80 mmHgPASS
Particulate Filtration Efficiency ASTM F229999.12%, 99.45%, 99.56%≥ 95%PASS
Bacterial Filtration Efficiency ASTM F210199.92%, 99.93%, 99.92%≥ 95%PASS
Differential Pressure (Delta P) EN 14683 Annex C3.0mmH2O/cm², 4.2mmH2O/cm², 3.7mmH2O/cm²

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.